Azurion R1.x and R2.x systems; Model Numbers: (1) 722063, (2) 722064, (3) 722067, (4) 722068, (5) 722078, (6) 722079, (7) 722221, (8) 722222, (9) 722223, (10) 722224, (11) 722225, (12) 722226, (13) 722227, (14) 722228, (15) 722280, (16) 722281, (17) 722282; Software Version: All software versions;
Class I - DangerousWhat Should You Do?
- Check if you have this product: Model Numbers: (1) 722063, (2) 722064, (3) 722067, (4) 722068, (5) 722078, (6) 722079, (7) 722221, (8) 722222, (9) 722223, (10) 722224, (11) 722225, (12) 722226, (13) 722227, (14) 722228, (15) 722280, (16) 722281, (17) 722282; UDI-DIs: (1) 00884838085275, (2) 00884838085282, (3) 00884838085350, (4) 00884838085367, (5) 00884838085251, (6) 00884838085268, (7) 00884838099203, (8) 00884838099210, (9) 00884838099241, (10) 00884838099258, (11) 00884838099265, (12) 00884838099272, (13) 00884838099227, (14) 00884838099234, (15) Not commercialized in the US, (16) Not commercialized in the US, (17) Not commercialized in the US;
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
- Reason for Recall:
- A software issue was identified in the internal communication process between the system software and the X-ray generator firmware. Loss of imaging (X-ray) functionality caused by the software issue in either of the two situations identified could result in a delay of therapy. The potential delay may result in serious adverse health outcomes, including the possibility of death, especially when the system is used with patients undergoing complex and/or urgent interventions for potentially life-threatening conditions (e.g., acute ischemic stroke, ST-segment elevation myocardial ischemia, life-threatening bleedings).
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
Azurion R1.x and R2.x systems; Model Numbers: (1) 722063, (2) 722064, (3) 722067, (4) 722068, (5) 722078, (6) 722079, (7) 722221, (8) 722222, (9) 722223, (10) 722224, (11) 722225, (12) 722226, (13) 722227, (14) 722228, (15) 722280, (16) 722281, (17) 722282; Software Version: All software versions;
Product Codes/Lot Numbers:
Model Numbers: (1) 722063, (2) 722064, (3) 722067, (4) 722068, (5) 722078, (6) 722079, (7) 722221, (8) 722222, (9) 722223, (10) 722224, (11) 722225, (12) 722226, (13) 722227, (14) 722228, (15) 722280, (16) 722281, (17) 722282; UDI-DIs: (1) 00884838085275, (2) 00884838085282, (3) 00884838085350, (4) 00884838085367, (5) 00884838085251, (6) 00884838085268, (7) 00884838099203, (8) 00884838099210, (9) 00884838099241, (10) 00884838099258, (11) 00884838099265, (12) 00884838099272, (13) 00884838099227, (14) 00884838099234, (15) Not commercialized in the US, (16) Not commercialized in the US, (17) Not commercialized in the US;
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1969-2025
Related Recalls
Philips Azurion 3M15, Model Numbers: 722064, 722222 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5 and R2.2.6 with the Roadmap Pro Software key
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
During the device phase of Roadmap Pro, a user may experience a dark image that is not clinically usable. The dark image may result after the user performs a pedal tap (shortly touching the footswitch pedal without generating an Xray image). The pedal tap may cause a dark image with the next fluoroscopy run.
Philips Azurion 3M12, Model Numbers: 722063, 722221 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5 and R2.2.6 with the Roadmap Pro Software key
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
During the device phase of Roadmap Pro, a user may experience a dark image that is not clinically usable. The dark image may result after the user performs a pedal tap (shortly touching the footswitch pedal without generating an Xray image). The pedal tap may cause a dark image with the next fluoroscopy run.
Philips Azurion 7B12/12, Model Numbers: 722067, 722225 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5 and R2.2.6 with the Roadmap Pro Software key
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
During the device phase of Roadmap Pro, a user may experience a dark image that is not clinically usable. The dark image may result after the user performs a pedal tap (shortly touching the footswitch pedal without generating an Xray image). The pedal tap may cause a dark image with the next fluoroscopy run.