MIS Rod 2, a component of the NewPort System. Model No: 60-0545 (45mm); 60-0550 (50mm); 60-0555 (55mm); 60-0560 (60mm); 60-0575 (75mm); 60-0580 (80mm). The NewPort MIS Rod 2 is a 5.5 diameter titanium spinal rod used to immobilize adjacent NewPort pedicle screws

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Lot No. Z21781A (45mm NewPort MIS Rod, 2) Z21956A (50mm NewPort MIS Rod, 2) Z21782A (55mm NewPort MIS Rod, 2) Z21957A (60mm NewPort MIS Rod, 2) Z21784A (75mm NewPort MIS Rod, 2) Z21958A (80mm NewPort MIS Rod, 2)
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Integra LifeSciences Corp.
Reason for Recall:
Integra Lifesciences is recalling NewPort System MIS Rod, 2 because it may be up to 30 degrees offset from specification requirements causing misalignment with the inserter. This could potentially result in the lordosis of the rod not aligning with the handle when the inserter is used in accordance with the current NewPort MIS System Surgical Technique.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

MIS Rod 2, a component of the NewPort System. Model No: 60-0545 (45mm); 60-0550 (50mm); 60-0555 (55mm); 60-0560 (60mm); 60-0575 (75mm); 60-0580 (80mm). The NewPort MIS Rod 2 is a 5.5 diameter titanium spinal rod used to immobilize adjacent NewPort pedicle screws

Product Codes/Lot Numbers:

Lot No. Z21781A (45mm NewPort MIS Rod, 2) Z21956A (50mm NewPort MIS Rod, 2) Z21782A (55mm NewPort MIS Rod, 2) Z21957A (60mm NewPort MIS Rod, 2) Z21784A (75mm NewPort MIS Rod, 2) Z21958A (80mm NewPort MIS Rod, 2)

Distribution:

Distributed in: US, CA, FL, IL, MD, MO, TN, TX

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-1961-2014

Related Recalls

Potential that the induction seal is not completely sealed to the device tube packaging.