REF 00-4309-071-36 Trabecular Metal Reverse Shoulder System Instrumentation, Glenosphere Helmet, 36 mm Diameter Nonsterile, Qty -1 Instrumentation used in shoulder replacement surgery. The glenosphere helmet is used to insert the glenosphere onto the base plate prior to impaction with the glenospere impactor.
Class I - DangerousWhat Should You Do?
- Check if you have this product: 60729528 60720059 60782350 60803204 60811712 60889765 60909408 60943022 60948237 60989841 60985366 61019823 61051119 61034821 61140638 61178017 61207698 61237374 61304005 61338570 61356591 6381187 61400935 61431685 61463330 61487660 61521267 61544005 61585968 61627734 61704564 61735029 61772406 61794587 61817713 61858701 61850103 61877862 61950512 61998686 61972672 62010251 62018071 62041751 62078963 62096387 62121179 62136094 62141873 62158509 62180443 62203390 62215390 62249372 62271566 62298060
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
ā ļø Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Zimmer, Inc.
- Reason for Recall:
- There is potential for one or both tabsof theTrabecular Metal Reverse Glenosphere Helmet to fracture during use and become disassociated from the device. There have been 47 reported complaints of tab fracture.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
REF 00-4309-071-36 Trabecular Metal Reverse Shoulder System Instrumentation, Glenosphere Helmet, 36 mm Diameter Nonsterile, Qty -1 Instrumentation used in shoulder replacement surgery. The glenosphere helmet is used to insert the glenosphere onto the base plate prior to impaction with the glenospere impactor.
Product Codes/Lot Numbers:
60729528 60720059 60782350 60803204 60811712 60889765 60909408 60943022 60948237 60989841 60985366 61019823 61051119 61034821 61140638 61178017 61207698 61237374 61304005 61338570 61356591 6381187 61400935 61431685 61463330 61487660 61521267 61544005 61585968 61627734 61704564 61735029 61772406 61794587 61817713 61858701 61850103 61877862 61950512 61998686 61972672 62010251 62018071 62041751 62078963 62096387 62121179 62136094 62141873 62158509 62180443 62203390 62215390 62249372 62271566 62298060
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1953-2013
Related Recalls
Zimmer Biomet is conducting a medical device recall for two (2) lots of the Dual Mobility Vivacit-EĀæ Bearing due to a commingle. The outer package is labeled as a Size F 44 mm, however, the implant inside the package is a Size G 46 mm, and vice versa.
Zimmer Biomet is conducting a medical device recall for two (2) lots of the Dual Mobility Vivacit-EĀæ Bearing due to a commingle. The outer package is labeled as a Size F 44 mm, however, the implant inside the package is a Size G 46 mm, and vice versa.
The associated product labeling does not adequately provide compatibility information for use of the 22.2 mm Diameter Ceramic Femoral Heads with cobalt-chrome femoral stems or stainless-steel femoral stems.