Drager PT 4000 Phototherapy System
Class I - DangerousWhat Should You Do?
- Check if you have this product: Catalog number 2M21700 and serial numbers ASCF-0037 TO ASCF-0061; ASCJ-0037 TO ASCJ-0152; ASCL-0001 TO ASCL-0025; ASCM-0001 TO ASCM-0025; and ASDA-002, 0005, 0006, 0007, 0008, 0009, 0012, 0019, 0025.
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Draeger Medical Systems, Inc.
- Reason for Recall:
- A routine internal quality check found that the Drager PT 4000 Phototherapy System is fitted with a 16 Amp rated fuse instead of the specified 1.6 Amp rated fuse. This may mean that the internal fuse would not trip promptly in the event of an internal short-circuit. The possible consequence is the wall outlet circuit breaker would trip first on the device may severely overheat. There have not b
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Drager PT 4000 Phototherapy System
Product Codes/Lot Numbers:
Catalog number 2M21700 and serial numbers ASCF-0037 TO ASCF-0061; ASCJ-0037 TO ASCJ-0152; ASCL-0001 TO ASCL-0025; ASCM-0001 TO ASCM-0025; and ASDA-002, 0005, 0006, 0007, 0008, 0009, 0012, 0019, 0025.
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1950-2012
Related Recalls
Edifier speakers (shipped in or after April 2023) may not annunciate when configured in analog mode which can lead to delayed treatment to the patient.
Infinity Acute Care System (IACS) Monitoring System, Infinity M500 with Power, Model No. MS20407. Rev 20, 21, and 23
Draeger Medical Systems
Device is not in full compliance to Type CF requirements of IEC 60601-1 and IEC 60601-2-34 standards. Therefore, there is a potential risk of electric shock. Health consequences may include electrical shock, burns, and cardiac arrhythmia.
Draeger Softbed Resuscitaire and/or Birthing Room Warmer with Scale [Model(s): RW82VHA-1, RW82VHA-1C, RW82VHA-1D, RW82-1, RW82-1C, RW82-1D, WBR82-1 and WBR82-1C]
Draeger Medical Systems
The Resuscitaire Infant Radiant Warmer with the optional scale could display inaccurate weight values.