Dexcom One Continuous Glucose Monitoring System, SKUs(Part Number): STK-DO-006 (MT27409-1), STK-DO-013 (MT27409-1), STK-DO-103 (MT27409-2), STK-DO-109 (MT27409-2)

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    All Lots/ UDI: STK-DO-006 -00386270003188 STK-DO-013 -00386270002327 STK-DO-103-00386270002969 STK-DO-109 -00386270002792
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Dexcom, Inc.
Reason for Recall:
Defective foam or an assembly error may cause the receiver speaker to lose contact with the printed circuit board, leading to missed audible alerts for low or high blood glucose values. A missed audible alert for low or high blood glucose values could lead to untreated hypo or hyperglycemia which can cause seizures, vomiting, loss of consciousness, or death.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
ongoing

Product Information

Full Description:

Dexcom One Continuous Glucose Monitoring System, SKUs(Part Number): STK-DO-006 (MT27409-1), STK-DO-013 (MT27409-1), STK-DO-103 (MT27409-2), STK-DO-109 (MT27409-2)

Product Codes/Lot Numbers:

All Lots/ UDI: STK-DO-006 -00386270003188 STK-DO-013 -00386270002327 STK-DO-103-00386270002969 STK-DO-109 -00386270002792

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-1946-2025

Related Recalls

A software defect in version v1.15.0 of the G6 Android app can cause the app to terminate unexpectedly, which may result in the user not receiving estimated glucose values, alarms, alerts or notifications. This could result in the missed detection of a hyperglycemic or hypoglycemic event, protentional resulting in severe hyperglycemia, diabetic ketoacidosis (DKA), or hyperosmolar hyperglycemic state (HHS).

Aug 28, 2025 Diagnostic Equipment View Details →

The affected devices are the Dexcom G7 Continuous Glucose Monitoring System (CGM System) using iOS, watchOS and Android App software (versions 2.8.0 or earlier) and Dexcom ONE+ CGM System using iOS and Android App software (versions 1.4.0 or earlier), henceforth referred to as the App. The affected App does not provide an expected "Sensor Failed" alert when the transmitter sends a "transmitterFailed" error message to the App when the CGM experiences a hardware or firmware failure. Instead, the App ends the CGM sensor session, stops reporting glucose values and displays the Start Sensor screen or No active sensor message without alerting the user, which can lead to missed detection of a hyperglycemic or hypoglycemic event and a delay in treatment.

Jul 24, 2025 Diagnostic Equipment Nationwide View Details →

The affected devices are the Dexcom G7 Continuous Glucose Monitoring System (CGM System) using iOS, watchOS and Android App software (versions 2.8.0 or earlier) and Dexcom ONE+ CGM System using iOS and Android App software (versions 1.4.0 or earlier), henceforth referred to as the App. The affected App does not provide an expected "Sensor Failed" alert when the transmitter sends a "transmitterFailed" error message to the App when the CGM experiences a hardware or firmware failure. Instead, the App ends the CGM sensor session, stops reporting glucose values and displays the Start Sensor screen or No active sensor message without alerting the user, which can lead to missed detection of a hyperglycemic or hypoglycemic event and a delay in treatment.

Jul 24, 2025 Diagnostic Equipment Nationwide View Details →