VITROS 5600 Integrated System, Catalog Number 6802413, Unique Device Identifier No. 10758750002740; and VITROS 5600 Integrated System (refurbished), Catalog Number 6802915, Unique Device Identifier No. 10758750007110; IVD.
Class I - DangerousWhat Should You Do?
- Check if you have this product: Software Version 3.2.2 & below; VITROS 5600 Serial Numbers J56000024 - J56002387; VITROS 5600 (Refurbished) US Serial Numbers: 56000186, 56000308, 56000315, 56000327, 56000339, 56000458, 56000559, 56000718, 56000727, 56000741, 56000747, 56000750, 56000786, 56000792, 56000824, 56001062, 56001235, 56001635, 56001641; Foreign Serial Numbers: 56000147, 56000200, 56000214, 56000245,56000283, 56000330, 56000331, 56000362, 56000377, 56000399, 56000413, 56000416, 56000428, 56000437, 56000438, 56000439, 56000459, 56000463, 56000464, 56000466, 56000515, 56000528, 56000537, 56000555, 56000560, 56000632, 56000642, 56000669, 56000680, 56000749, 56000867, 56001133, 56001135, 56001246, 56001461, 56001482, 56001487, 56001859. Unique Device Identifier No. 10758750002740
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Ortho-Clinical Diagnostics
- Reason for Recall:
- Ortho Clinical Diagnostics confirmed a software timing anomaly in which the VITROS 3600 and 5600 Systems may unexpectedly aspirate from or dispense into an unintended sample that may lead to erroneous or contaminated results.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
VITROS 5600 Integrated System, Catalog Number 6802413, Unique Device Identifier No. 10758750002740; and VITROS 5600 Integrated System (refurbished), Catalog Number 6802915, Unique Device Identifier No. 10758750007110; IVD.
Product Codes/Lot Numbers:
Software Version 3.2.2 & below; VITROS 5600 Serial Numbers J56000024 - J56002387; VITROS 5600 (Refurbished) US Serial Numbers: 56000186, 56000308, 56000315, 56000327, 56000339, 56000458, 56000559, 56000718, 56000727, 56000741, 56000747, 56000750, 56000786, 56000792, 56000824, 56001062, 56001235, 56001635, 56001641; Foreign Serial Numbers: 56000147, 56000200, 56000214, 56000245,56000283, 56000330, 56000331, 56000362, 56000377, 56000399, 56000413, 56000416, 56000428, 56000437, 56000438, 56000439, 56000459, 56000463, 56000464, 56000466, 56000515, 56000528, 56000537, 56000555, 56000560, 56000632, 56000642, 56000669, 56000680, 56000749, 56000867, 56001133, 56001135, 56001246, 56001461, 56001482, 56001487, 56001859. Unique Device Identifier No. 10758750002740
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1914-2016
Related Recalls
VITROS Chemistry Products LAC Slides. Catalog Number: 8150112. 1 Sales Unit = contains 5 cartridges of 18 slides, for a total of 90 tests.
Ortho-Clinical Diagnostics
VITROS Chemistry Products LAC Slides on VITROS XT 3400 and VITROS XT 7600 Systems are experiencing an increased occurrence of calibration failure related to lots from coating 0130 and above with Condition Code TH4-63J during calibration, causing the delay of lac test results and potentially impact patient management. The issue was identified by QuidelOrtho through an increase in customer complaints compared to the historical baseline with average failure rate of 1% for coating 0130 in 2025 while average of failure rate for other coating was 0.06% in 2025 and average of failure rate for all coatings is 0.03% for 2024.
Chemistry Products LAC Slides. Catalog Number: 8433880. 1 Sales Unit = contains 5 cartridges of 60 slides, for a total of 300 tests.
Ortho-Clinical Diagnostics
VITROS Chemistry Products LAC Slides on VITROS XT 3400 and VITROS XT 7600 Systems are experiencing an increased occurrence of calibration failure related to lots from coating 0130 and above with Condition Code TH4-63J during calibration, causing the delay of lac test results and potentially impact patient management. The issue was identified by QuidelOrtho through an increase in customer complaints compared to the historical baseline with average failure rate of 1% for coating 0130 in 2025 while average of failure rate for other coating was 0.06% in 2025 and average of failure rate for all coatings is 0.03% for 2024.
VITROS Chemistry Products Ca Slides; Catalog Number: 145 0261;
Ortho-Clinical Diagnostics
Generations 67 and above of VITROS Ca Slides are experiencing an increased occurrence of Condition Code TH4-63J during calibration on VITROS XT 3400 and VITROS XT 7600 Systems, causing calibration failures that delay calcium test results and potentially impact patient management. The imaging reflectometer algorithm detects excessive "spikes" (imperfections) in slide images and generates the TH4-63J condition code when the threshold of 600 spikes is exceeded. Updated analysis revealed a 19-fold increase in failure rates from 0.0043% (2024 baseline) to 0.082% for Generation 67+ slides in 2025, with 100% of calibrator failures occurring at the lowest concentration level. This can result in delayed diagnosis and treatment of conditions such as hypercalcemic crisis, parathyroid disorders, and other calcium-related emergencies requiring urgent intervention.