Pressure Injectable Arrowg+ard Blue Plus Multi-Lumen Central Venous Catheterization Kit. Permits venous access to central circulation by way of subclavian, jugular and femoral veins. The ARROWg+ard technology is intended to provide protection against catheter-related bloodstream infections (CRBSIs). It is not intended to be used as a treatment for existing infections nor is it indicated for long-term use (>30 days). Clinical effectiveness of the ARROW g+ard Blue Plus catheter in preventing CRBSIs compared to the original ARROWg+ard Blue catheter has not been studied
Class I - DangerousWhat Should You Do?
- Check if you have this product: Product Codes ASK-45703-PRWJ1, ASK-45703-PUAB1, and ASK-45703-PUPM Lot # 13F16L0105, 23F16K0180 and 13F16L0199
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Arrow International Inc
- Reason for Recall:
- Arrow International is recalling due to possible voids in the catheter hub, which could lead to an interlumen crossover.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Pressure Injectable Arrowg+ard Blue Plus Multi-Lumen Central Venous Catheterization Kit. Permits venous access to central circulation by way of subclavian, jugular and femoral veins. The ARROWg+ard technology is intended to provide protection against catheter-related bloodstream infections (CRBSIs). It is not intended to be used as a treatment for existing infections nor is it indicated for long-term use (>30 days). Clinical effectiveness of the ARROW g+ard Blue Plus catheter in preventing CRBSIs compared to the original ARROWg+ard Blue catheter has not been studied
Product Codes/Lot Numbers:
Product Codes ASK-45703-PRWJ1, ASK-45703-PUAB1, and ASK-45703-PUPM Lot # 13F16L0105, 23F16K0180 and 13F16L0199
Distribution:
Distributed in: US, NJ, IL, PA
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1879-2017
Related Recalls
Reports have been received regarding open seals on the packaging.
Arrow FiberOptix Intra-Aortic Balloon Catheter Kit: a) REF IAB-05830-LWS; b) REF IAB-05840-LWS; c) REF IAB-05850-LWS
ARROW INTERNATIONAL
Teleflex received reports indicating an infrequent condition that, when not identified and corrected promptly, could result in serious health consequences. The issue may manifest as: -failure of the intra-aortic balloon to completely inflate over its full length -damaged or broken central lumen in the segment contained within the balloon -helium loss or blood in the helium pathway.
Arrow UltraFlex Intra-Aortic Balloon Catheter Kit, a) REF IAB-06830-U; b) REF IAB-06840-U; c) REF IAB-06850-U
ARROW INTERNATIONAL
Teleflex received reports indicating an infrequent condition that, when not identified and corrected promptly, could result in serious health consequences. The issue may manifest as: -failure of the intra-aortic balloon to completely inflate over its full length -damaged or broken central lumen in the segment contained within the balloon -helium loss or blood in the helium pathway.