Thermo Scientific General Purpose Dynamic Bath Precision 5 & 10L Dual GP Bath Model Number: TSGP15D

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Serial Numbers: 300246816 300246817 300247411 300247412
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Thermo Fisher Scientific (Asheville) LLC
Reason for Recall:
Product may have been assembled incorrectly,the assembly error can cause the bath to malfunction and raise the temperature of the fluid in the reservoir above the temperature set point selected by the user.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

Thermo Scientific General Purpose Dynamic Bath Precision 5 & 10L Dual GP Bath Model Number: TSGP15D

Product Codes/Lot Numbers:

Serial Numbers: 300246816 300246817 300247411 300247412

Distribution:

Distributed in: CA, GA, KY, PA

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-1871-2019

Related Recalls