The Better-Bladder (BB), part of extracorporeal membrane oxygenation (ECMO) cardiopulmonary bypass circuits. Model: BB14

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Lot Number: 5290-201001 UDI: (01)00851997007004(10)5290-201001(17)20231021(11)20201021
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Circulatory Technology Inc.
Reason for Recall:
The device is used as part of extracorporeal membrane oxygenation (ECMO) cardiopulmonary bypass circuits and may collapse during use which can increase resistance to flow in the venous line and cause a drop in circuit blood flow.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
ongoing

Product Information

Full Description:

The Better-Bladder (BB), part of extracorporeal membrane oxygenation (ECMO) cardiopulmonary bypass circuits. Model: BB14

Product Codes/Lot Numbers:

Lot Number: 5290-201001 UDI: (01)00851997007004(10)5290-201001(17)20231021(11)20201021

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-1862-2021

Related Recalls

The device is used as part of extracorporeal membrane oxygenation (ECMO) cardiopulmonary bypass circuits and may collapse during use which can increase resistance to flow in the venous line and cause a drop in circuit blood flow.

Sep 22, 2017 Infusion Pumps View Details →