NxStage Dialysate Sack (SAK) with Concentrate for Use with PureFlow SL REF: SAK-407 Lactate 40, Potassium 1, Sodium 140, Calcium 3, Magnesium 1, Chloride 105, Glucose (mg/dL) 100 Batch Size (liters) 50
Class I - DangerousWhat Should You Do?
- Check if you have this product: Lot code: 40179074
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- NxStage Medical, Inc.
- Reason for Recall:
- SAK Dialysate Concentrate contain aluminum levels which exceed internal product specification
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
NxStage Dialysate Sack (SAK) with Concentrate for Use with PureFlow SL REF: SAK-407 Lactate 40, Potassium 1, Sodium 140, Calcium 3, Magnesium 1, Chloride 105, Glucose (mg/dL) 100 Batch Size (liters) 50
Product Codes/Lot Numbers:
Lot code: 40179074
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1861-2014
Related Recalls
NxStage PureFlow-B Solution. Catalog numbers: RFP-400, RFP-401, RFP-402, RFP-403, RFP-404, RFP-406, RFP-453, RFP-454, RFP-456.
NxStage Medical
Potential for the perimeter seal on the small chamber of the two-chamber bag to fail causing the electrolyte fluid to splash into the face of the user when attempting to open the peel seam to mix the two chambers prior to treatment.
The smaller chamber of the two chamber bag is bursting when nurses are attempting to mix the two chambers prior to use. The electrolyte fluid in the small chamber of the NxStage PureFlow B Solution has a pH of less than 2.0. The solution can be harmful if it comes in contact with the eyes or skin, and/or as a result of a slip or fall injury due to spilled fluid.
There is a potential patient health risk while performing chronic hemodialysis treatments in the home setting with for patients with low blood volume (2.4 Liters or less in a patient with body weight of 30kg or less) when prescribing larger dialysate volumes.