NxStage Dialysate Sack (SAK) with Concentrate for Use with PureFlow SL REF: SAK-305 Lactate 45, Potassium 1, Sodium 140, Calcium 3, Magnesium 1 Chloride 100, Glucose (mg/dL) 100 Batch Size (liters) 40

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Lot codes: 40179026 3057901 3047930
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
NxStage Medical, Inc.
Reason for Recall:
SAK Dialysate Concentrate contain aluminum levels which exceed internal product specification
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

NxStage Dialysate Sack (SAK) with Concentrate for Use with PureFlow SL REF: SAK-305 Lactate 45, Potassium 1, Sodium 140, Calcium 3, Magnesium 1 Chloride 100, Glucose (mg/dL) 100 Batch Size (liters) 40

Product Codes/Lot Numbers:

Lot codes: 40179026 3057901 3047930

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-1858-2014

Related Recalls

The smaller chamber of the two chamber bag is bursting when nurses are attempting to mix the two chambers prior to use. The electrolyte fluid in the small chamber of the NxStage PureFlow B Solution has a pH of less than 2.0. The solution can be harmful if it comes in contact with the eyes or skin, and/or as a result of a slip or fall injury due to spilled fluid.

May 14, 2024 Other Medical Devices Nationwide View Details →

There is a potential patient health risk while performing chronic hemodialysis treatments in the home setting with for patients with low blood volume (2.4 Liters or less in a patient with body weight of 30kg or less) when prescribing larger dialysate volumes.

Aug 27, 2019 Other Medical Devices View Details →