NxStage Dialysate Sack (SAK) with Concentrate for Use with PureFlow SL REF: SAK-301 Lactate 45, Potassium 1, Sodium 140, Calcium 3, Magnesium 1, Chloride 100, Glucose (mg/dL) 100 Batch Size (liters) 60 The SAK is one of the disposables used with NxStage PureFlow SL Dialysate Preparation System. The SAK is a disposable bag pre-filled with dialysate concentrate that is then filled with purified water prepared by NxStage PureFlow SL Dialysate Preparation System to prepare a batch of dialysate for use with the NxStage System One. SAKs are packaged in a case containing two SAKs. SAK product SKUs vary in the dialysate concentrate and volume of the final dialysate batch.
Class I - DangerousWhat Should You Do?
- Check if you have this product: Lot codes: 40279015 40179163 40179162 31279127 3067903 3057931 3057930 3047933 3047928
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- NxStage Medical, Inc.
- Reason for Recall:
- SAK Dialysate Concentrate contain aluminum levels which exceed internal product specification
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
NxStage Dialysate Sack (SAK) with Concentrate for Use with PureFlow SL REF: SAK-301 Lactate 45, Potassium 1, Sodium 140, Calcium 3, Magnesium 1, Chloride 100, Glucose (mg/dL) 100 Batch Size (liters) 60 The SAK is one of the disposables used with NxStage PureFlow SL Dialysate Preparation System. The SAK is a disposable bag pre-filled with dialysate concentrate that is then filled with purified water prepared by NxStage PureFlow SL Dialysate Preparation System to prepare a batch of dialysate for use with the NxStage System One. SAKs are packaged in a case containing two SAKs. SAK product SKUs vary in the dialysate concentrate and volume of the final dialysate batch.
Product Codes/Lot Numbers:
Lot codes: 40279015 40179163 40179162 31279127 3067903 3057931 3057930 3047933 3047928
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1854-2014
Related Recalls
NxStage PureFlow-B Solution. Catalog numbers: RFP-400, RFP-401, RFP-402, RFP-403, RFP-404, RFP-406, RFP-453, RFP-454, RFP-456.
NxStage Medical
Potential for the perimeter seal on the small chamber of the two-chamber bag to fail causing the electrolyte fluid to splash into the face of the user when attempting to open the peel seam to mix the two chambers prior to treatment.
The smaller chamber of the two chamber bag is bursting when nurses are attempting to mix the two chambers prior to use. The electrolyte fluid in the small chamber of the NxStage PureFlow B Solution has a pH of less than 2.0. The solution can be harmful if it comes in contact with the eyes or skin, and/or as a result of a slip or fall injury due to spilled fluid.
There is a potential patient health risk while performing chronic hemodialysis treatments in the home setting with for patients with low blood volume (2.4 Liters or less in a patient with body weight of 30kg or less) when prescribing larger dialysate volumes.