Abacus 3.1, Pharmacy Calculator Product Usage: Provide pharmacy calculations for automated compounding.
Class I - DangerousWhat Should You Do?
- Check if you have this product: Product Code: 8300-0167, 8300-0168, 8300-0169
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Baxter Corporation Englewood
- Reason for Recall:
- Baxter Corporation is initiating a field correction following an upgrade to ABACUS Software v3.1, when Clinimix 4.25110 is used in an order, the Nutritional Summary and any labels containing energy information may display 1,000 times less protein than is actually present in the solution.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Abacus 3.1, Pharmacy Calculator Product Usage: Provide pharmacy calculations for automated compounding.
Product Codes/Lot Numbers:
Product Code: 8300-0167, 8300-0168, 8300-0169
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1847-2014
Related Recalls
0.2 Micron Filter, 50 mm Product Usage: The 0.2 Micron Filter is a bacteria and particulate filter for aqueous solutions. It attaches to a standard Baxter Pharmacy Pump Tube Set.
Baxter Corporation Englewood
Baxter Healthcare Corporation is issuing a voluntary product recall for all unexpired lots of the 50mm 0.2 Micron Filter (Product Code: H93835) due to the potential for the filter membrane layer to be missing and for particulate matter to be present.
Baxter, RAPIDFILL Syringe Strip, 10 ml. For use with RAPIDFILL Automated Syringe Filler (ASF)
Baxter Corporation Englewood
All lots of RAPIDFILL Syringe Strips lack evidence supporting the packaging claim for 12-month shelf life on released product.
ABACUS SE (Single-Workstation Edition) and ABACUS ME (Multi-Workstation Edition) Product Usage: The ABACUS Software is a Windows-based order entry software application for comprehensive total parenteral nutrition (TPN) calculations and label printing.
Baxter Corporation Englewood
Baxter Corporation is conducting a field action for the ABACUS SE and ME models due to the possibility that the compounder will load the incorrect formula upon scanning the bag label when two or more different formula files have been created with the same order number.