Cardinal BI Stat Biological Test Pack with Instant Readout Integrator, SteriTec Product Code PL 160/A, Distributor Product Code T40311PE SteriTec Steam Biological Test Packs with Instant Readout Integrator Card: The Steam Biological Test Pack with Instant Readout Integrator is a steam sterilization monitor designed specifically for biological testing of pre-vacuum steam sterilizers operating at 132¿C (270¿F) for 3 minutes or longer. The Instant Readout Integrator Card gives the operator a prediction of the biological test outcome.

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Affected Biological Indicator Lot No.:130401
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Steritec Products Mfc. Co., Inc.
Reason for Recall:
Mesa Labs, based out of Bozeman, MT, informed SteriTec Products, Inc. of a product recall involving the EZTest Steam and Smart-Read Biological Indicators (BI) because the readout time is not in specification with the label claims.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

Cardinal BI Stat Biological Test Pack with Instant Readout Integrator, SteriTec Product Code PL 160/A, Distributor Product Code T40311PE SteriTec Steam Biological Test Packs with Instant Readout Integrator Card: The Steam Biological Test Pack with Instant Readout Integrator is a steam sterilization monitor designed specifically for biological testing of pre-vacuum steam sterilizers operating at 132¿C (270¿F) for 3 minutes or longer. The Instant Readout Integrator Card gives the operator a prediction of the biological test outcome.

Product Codes/Lot Numbers:

Affected Biological Indicator Lot No.:130401

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-1829-2013

Related Recalls

Mesa Labs, based out of Bozeman, MT, informed SteriTec Products, Inc. of a product recall involving the EZTest Steam and Smart-Read Biological Indicators (BI) because the readout time is not in specification with the label claims.

Jun 10, 2013 Diagnostic Equipment Nationwide View Details →