Atec Insignia Anterior Cervical Plate System, REF 136-0364 Insignia, ACP, 3-Level, 64 mm, Rx Only, Non-Sterile, UDI: (01)00190376268637
Class I - DangerousWhat Should You Do?
- Check if you have this product: 8609329R
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Alphatec Spine, Inc.
- Reason for Recall:
- There is a potential for the screw blocking mechanism on the anterior cervical plate system to disassociate intraoperatively/postoperatively or unlock postoperatively.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Atec Insignia Anterior Cervical Plate System, REF 136-0364 Insignia, ACP, 3-Level, 64 mm, Rx Only, Non-Sterile, UDI: (01)00190376268637
Product Codes/Lot Numbers:
8609329R
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1823-2021
Related Recalls
Due a design issue where the navigated array connection geometry is incorrect.
Complaints have been received regarding post-operative implant collapse.
Due to reports of intraoperative graft bolt implantation breakages.