ICD COBALT VR MRI DF4, Model Number DVPB3D4; Implantable Cardioverter Defibrillators
Class I - DangerousWhat Should You Do?
- Check if you have this product: GTIN 00763000178512, Lot Serial Numbers: RSF600012S, RSF600254S, RSF600591S, RSF600655S, RSF600822S, RSF601247S, RSF602653S, RSF603908S, RSF601102S, RSF607914S, RSF605967S, RSF600633S, RSF600649S, RSF600867S, RSF600868S, RSF600869S, RSF600872S, RSF600873S, RSF600874S, RSF600875S, RSF600876S, RSF600878S, RSF600880S, RSF600883S, RSF602930S, RSF602991S, RSF602998S, RSF603002S, RSF603003S, RSF603013S, RSF603018S, RSF603019S, RSF603022S, RSF603026S, RSF603027S, RSF603030S, RSF603032S, RSF603033S, RSF603034S, RSF603035S, RSF603036S, RSF603736S, RSF603737S, RSF603741S, RSF603742S, RSF603744S, RSF603746S, RSF603747S, RSF603749S, RSF603754S, RSF603755S, RSF603756S, RSF603758S, RSF603759S, RSF603762S, RSF603763S, RSF603764S, RSF603799S, RSF603802S, RSF603804S, RSF603806S, RSF603810S, RSF603813S, RSF603818S, RSF603821S, RSF603824S, RSF602665S, RSF602703S, RSF603249S, RSF603648S, RSF603939S, RSF607644S, RSF601171S, RSF601174S, RSF602584S, RSF602743S, RSF600887S, RSF602678S, RSF608130S, RSF600767S, RSF608097S, RSF604855S, RSF606263S, RSF601206S, RSF601486S, RSF601567S, RSF601571S, RSF600924S, RSF601224S, RSF601226S, RSF602817S, RSF603405S, RSF603521S, RSF603678S, RSF602595S, RSF604129S, RSF600687S, RSF600210S, RSF602849S, RSF607243S, RSF607244S, RSF607291S, RSF604110S, RSF604146S, RSF605150S, RSF607995S, RSF600352S, RSF601546S, RSF607225S, RSF606605S, RSF600269S, RSF600940S, RSF601190S, RSF603550S, RSF604130S, RSF608014S, RSF600303S, RSF602434S, RSF602435S, RSF602449S, RSF602461S, RSF602466S, RSF607173S, RSF607174S, RSF600954S, RSF600955S, RSF602586S, RSF602587S, RSF602597S, RSF602804S, RSF603465S, RSF600982S, RSF603213S, RSF605644S, RSF606195S, RSF601058S, RSF603647S, RSF603724S, RSF604582S, RSF600219S, RSF601397S, RSF601539S, RSF602709S, RSF602757S, RSF603794S, RSF603973S, RSF604054S, RSF600443S, RSF600706S, RSF600721S, RSF600729S, RSF601434S, RSF603726S, RSF606532S, RSF603464S, RSF608114S, RSF601093S, RSF603450S, RSF607337S, RSF607359S, RSF600793S, RSF600805S, RSF600808S, RSF600864S, RSF601132S, RSF601165S, RSF601176S, RSF601540S, RSF601547S, RSF601587S, RSF602604S, RSF602874S, RSF603248S, RSF603399S, RSF603415S, RSF603428S, RSF603909S, RSF605018S, RSF605037S, RSF605025S, RSF601119S, RSF601462S, RSF607023S, RSF605918S, RSF606278S, RSF606884S, RSF600539S, RSF600523S, RSF602564S, RSF602565S, RSF602570S, RSF602571S, RSF603242S, RSF606921S, RSF601482S, RSF602165S, RSF602753S, RSF602822S, RSF602877S, RSF603585S, RSF605489S, RSF605974S, RSF607360S, RSF602220S, RSF602786S, RSF602789S, RSF604164S, RSF606324S, RSF606934S, RSF601167S, RSF601169S, RSF601172S, RSF602790S, RSF603216S, RSF603470S, RSF603549S, RSF604227S, RSF605701S, RSF605928S, RSF607222S, RSF607740S, RSF607741S, RSF607911S, RSF600260S, RSF601106S, RSF602684S, RSF603240S, RSF603468S, RSF603685S, RSF604089S, RSF604112S, RSF604121S, RSF604263S, RSF604389S, RSF604470S, RSF604700S, RSF604988S, RSF605889S, RSF600835S, RSF603897S, RSF603963S, RSF603966S, RSF605021S, RSF600263S, RSF600534S, RSF601142S, RSF601572S, RSF602208S, RSF602373S, RSF602632S, RSF602706S, RSF602848S, RSF603416S, RSF603527S, RSF603528S, RSF603662S, RSF603683S, RSF603892S, RSF604013S, RSF605570S, RSF607887S, RSF601305S, RSF600610S, RSF603260S, RSF603490S, RSF601446S, RSF605645S, RSF601534S, RSF602463S, RSF602698S, RSF603191S, RSF603510S, RSF603689S, RSF603916S, RSF603986S, RSF604266S, RSF605205S, RSF605206S, RSF605505S, RSF605670S, RSF600312S, RSF600889S, RSF604284S, RSF604471S, RSF604502S, RSF605697S, RSF605876S, RSF606313S, RSF606942S, RSF607052S, RSF607058S, RSF607714S, RSF607720S, RSF607721S, RSF603734S, RSF603870S, RSF603911S, RSF604010S, RSF604156S, RSF604289S, RSF604302S, RSF604425S, RSF604748S, RSF605054S, RSF607050S, RSF607733S, RSF607739S, RSF607763S, RSF608003S, RSF603960S, RSF605443S, RSF603543S, RSF607240S, RSF607378S, RSF607397S, RSF607406S, RSF600153S, RSF600328S, RSF600705S, RSF602707S, RSF603595S, RSF603773S, RSF600977S, RSF601471S, RSF601588S, RSF602343S, RSF602889S, RSF600162S, RSF600557S, RSF605875S, RSF606245S, RSF607055S, RSF607056S, RSF607716S, RSF601028S, RSF601033S, RSF601091S, RSF601189S, RSF601384S, RSF604197S, RSF604660S, RSF604701S, RSF604751S, RSF604905S, RSF607935S, RSF600717S, RSF603024S, RSF608070S, RSF603406S, RSF604073S, RSF600315S, RSF600420S, RSF606967S, RSF606367S, RSF606885S, RSF607025S, RSF607066S, RSF608102S, RSF608104S, RSF605699S, RSF605909S, RSF606985S, RSF607380S, RSF607697S, RSF607900S, RSF608009S, RSF600414S, RSF600671S, RSF601013S, RSF602591S, RSF602702S, RSF603130S, RSF603261S, RSF603449S, RSF603452S, RSF603492S, RSF603695S, RSF603823S, RSF603957S, RSF604087S, RSF604232S, RSF604268S, RSF604387S, RSF605559S, RSF606193S, RSF606194S, RSF606309S, RSF606534S, RSF600508S, RSF604838S, RSF606282S, RSF606513S, RSF603775S, RSF606311S, RSF607877S, RSF600708S, RSF600894S, RSF602727S, RSF602831S, RSF604604S, RSF606233S, RSF606894S, RSF607205S, RSF600815S, RSF604219S, RSF604448S, RSF607869S, RSF605419S, RSF607870S, RSF607030S, RSF607035S, RSF607934S, RSF605610S, RSF605964S, RSF606907S, RSF607073S, RSF607916S, RSF600472S, RSF600475S, RSF600696S, RSF600783S, RSF600802S, RSF600847S, RSF600911S, RSF601053S, RSF601192S, RSF601531S, RSF601560S, RSF602607S, RSF603100S, RSF603456S, RSF607366S, RSF602326S, RSF608093S, RSF603933S, RSF600241S, RSF605661S, RSF605908S, RSF600503S, RSF600886S, RSF601086S, RSF601214S, RSF601260S, RSF601279S, RSF601394S, RSF602547S, RSF602559S, RSF602654S, RSF602795S, RSF603785S, RSF601573S, RSF600801S, RSF601345S, RSF601467S, RSF603596S, RSF603649S, RSF604082S, RSF600701S, RSF605865S, RSF605866S, RSF605869S, RSF606217S, RSF606950S, RSF607044S, RSF607062S, RSF607067S, RSF601395S, RSF601469S, RSF601586S, RSF602589S, RSF602683S, RSF603214S, RSF603858S, RSF607904S, RSF606523S, RSF607928S, RSF600195S, RSF601250S, RSF601475S, RSF602578S, RSF603553S, RSF603559S, RSF603729S, RSF603832S, RSF603837S, RSF603875S, RSF603902S, RSF603982S, RSF604598S, RSF605436S, RSF600433S, RSF600615S, RSF600674S, RSF600762S, RSF600785S, RSF602744S, RSF602892S, RSF603713S, RSF603727S, RSF604237S, RSF605856S, RSF600566S, RSF601215S, RSF603602S, RSF603946S, RSF605030S, RSF603042S, RSF603587S, RSF601323S, RSF607949S, RSF607950S, RSF600552S, RSF600623S, RSF600711S, RSF600782S, RSF601088S, RSF602504S, RSF602770S, RSF605695S, RSF606528S, RSF603655S, RSF602205S, RSF600736S, RSF600865S, RSF600866S, RSF601003S, RSF601098S, RSF602177S, RSF602895S, RSF603227S, RSF603511S, RSF603544S, RSF604135S, RSF604823S, RSF605990S, RSF601100S, RSF601147S, RSF601185S, RSF601353S, RSF601354S, RSF602888S, RSF603489S, RSF604401S, RSF604968S, RSF605991S, RSF606221S, RSF607867S, RSF600360S, RSF601533S, RSF600561S, RSF600676S, RSF601549S, RSF603999S, RSF602623S, RSF600410S, RSF600937S, RSF602816S, RSF602827S, RSF603378S, RSF605945S, RSF608117S, RSF600282S, RSF600324S, RSF604759S, RSF600751S, RSF604115S, RSF607662S, RSF602592S, RSF602983S, RSF602985S, RSF603497S, RSF603519S, RSF603958S, RSF603987S, RSF603994S, RSF604186S, RSF604195S, RSF606189S, RSF606498S, RSF606535S, RSF606947S, RSF607330S, RSF602905S, RSF603104S, RSF604977S, RSF605086S, RSF607890S, RSF608062S, RSF602626S, RSF604260S, RSF602713S, RSF608041S, RSF608042S, RSF608043S, RSF608044S, RSF608045S, RSF605968S, RSF600749S, RSF600985S, RSF601212S, RSF603199S, RSF603402S, RSF603651S, RSF603652S, RSF603815S, RSF603817S, RSF603886S, RSF604217S, RSF605708S, RSF606256S, RSF600476S, RSF600477S, RSF601331S, RSF602334S, RSF602359S, RSF602629S, RSF602777S, RSF603221S, RSF603600S, RSF603735S, RSF603829S, RSF604413S, RSF604906S, RSF607227S, RSF607386S, RSF600492S, RSF600806S, RSF600975S, RSF601021S, RSF601025S, RSF601031S, RSF601265S, RSF601570S, RSF602494S, RSF602921S, RSF604154S, RSF604730S, RSF604883S, RSF604886S, RSF600154S, RSF600267S, RSF600843S, RSF601555S, RSF601574S, RSF602637S, RSF602700S, RSF602711S, RSF603054S, RSF603201S, RSF603369S, RSF603789S, RSF603918S, RSF603932S, RSF603955S, RSF603991S, RSF604098S, RSF604188S, RSF604766S, RSF600666S, RSF600949S, RSF602672S, RSF602988S, RSF603864S, RSF602619S, RSF607247S, RSF601449S, RSF603223S, RSF602423S, RSF606943S, RSF600395S, RSF600427S, RSF600504S, RSF601231S, RSF601278S, RSF601542S, RSF602802S, RSF603945S, RSF600309S, RSF600314S, RSF600560S, RSF600747S, RSF602327S, RSF602329S, RSF602411S, RSF605810S, RSF605849S, RSF605851S, RSF605852S, RSF605853S, RSF606197S, RSF606198S, RSF606200S, RSF606201S, RSF607098S, RSF607099S, RSF607101S, RSF607114S, RSF607116S, RSF607951S, RSF607952S, RSF601109S, RSF602897S, RSF603060S, RSF603062S, RSF603063S, RSF603065S, RSF603066S, RSF603069S, RSF603070S, RSF603071S, RSF603107S, RSF603639S, RSF603676S, RSF603801S, RSF603803S, RSF603807S, RSF603808S, RSF603809S, RSF603811S, RSF603812S, RSF605366S, RSF607097S, RSF607392S, RSF607393S, RSF607953S, RSF607954S, RSF607955S, RSF607956S, RSF607957S, RSF607958S, RSF607959S, RSF607960S, RSF607961S, RSF607962S, RSF607963S, RSF607964S, RSF605650S, RSF600398S, RSF600399S, RSF600401S, RSF600407S, RSF600743S, RSF600836S, RSF600838S, RSF600839S, RSF601261S, RSF601306S, RSF601314S, RSF601532S, RSF603106S, RSF603800S, RSF605678S, RSF607074S, RSF602425S, RSF602741S, RSF605956S, RSF603576S, RSF600511S, RSF600512S, RSF602372S, RSF602751S, RSF602752S, RSF602808S, RSF602932S, RSF602969S, RSF602974S, RSF603132S, RSF603362S, RSF603538S, RSF603697S, RSF603876S, RSF604012S, RSF604020S, RSF604173S, RSF604205S, RSF604344S, RSF604481S, RSF604482S, RSF604485S, RSF604500S, RSF602512S, RSF601491S, RSF606519S, RSF607271S, RSF602427S, RSF602859S, RSF602958S, RSF603675S, RSF600274S, RSF602553S, RSF607212S, RSF603430S, RSF603632S, RSF603653S, RSF608036S, RSF600142S, RSF600670S, RSF603625S, RSF601240S, RSF600532S, RSF600739S, RSF600741S, RSF604032S, RSF600967S, RSF605007S, RSF605008S, RSF606286S, RSF607865S, RSF608022S, RSF600209S, RSF600346S, RSF601075S, RSF601096S, RSF602605S, RSF602628S, RSF602780S, RSF603266S, RSF603522S, RSF603529S, RSF603491S, RSF607387S, RSF600668S, RSF602575S, RSF603512S, RSF607942S, RSF607943S, RSF607944S, RSF607945S, RSF607946S, RSF605684S, RSF605950S, RSF605962S, RSF606173S, RSF606228S, RSF606229S, RSF606230S, RSF606264S, RSF606265S, RSF606503S, RSF606504S, RSF606521S, RSF606963S, RSF607077S, RSF607266S, RSF607283S, RSF607377S, RSF600466S, RSF600605S, RSF600812S, RSF600951S, RSF601232S, RSF601235S, RSF601238S, RSF601242S, RSF601246S, RSF601262S, RSF601559S, RSF602462S, RSF602522S, RSF602524S, RSF602526S, RSF602527S, RSF602745S, RSF602748S, RSF602749S, RSF602809S, RSF603349S, RSF603494S, RSF603619S, RSF603738S, RSF603879S, RSF603935S, RSF603938S, RSF603971S, RSF603976S, RSF604055S, RSF604153S, RSF604170S, RSF604181S, RSF604255S, RSF604440S, RSF604488S, RSF604496S, RSF604668S, RSF604669S, RSF604670S, RSF604723S, RSF604821S, RSF604826S, RSF604904S, RSF605020S, RSF605976S, RSF600413S, RSF602638S, RSF602720S, RSF602841S, RSF602869S, RSF603429S, RSF603445S, RSF603940S, RSF600558S, RSF600927S, RSF601191S, RSF601213S, RSF602873S, RSF600431S, RSF600432S, RSF601120S, RSF602844S, RSF603226S, RSF604822S, RSF600598S, RSF607735S, RSF601333S, RSF602868S, RSF600374S, RSF601241S, RSF603784S, RSF600319S, RSF602511S, RSF600377S, RSF600926S, RSF601343S, RSF602862S, RSF604216S, RSF605676S, RSF606882S, RSF607027S, RSF607772S, RSF607780S, RSF607791S, RSF607927S, RSF608001S, RSF600555S, RSF600640S, RSF600658S, RSF600685S, RSF600686S, RSF600795S, RSF601228S, RSF601264S, RSF603099S, RSF603352S, RSF603397S, RSF603424S, RSF603541S, RSF603623S, RSF603708S, RSF603856S, RSF604176S, RSF604231S, RSF604310S, RSF604952S, RSF604990S, RSF605035S, RSF607771S, RSF600393S, RSF601014S, RSF603943S, RSF601162S, RSF602364S, RSF602928S, RSF600699S, RSF601070S, RSF601139S, RSF601154S, RSF601402S, RSF602784S, RSF604245S, RSF604258S, RSF604052S, RSF605870S, RSF601448S, RSF602609S, RSF604200S, RSF601272S, RSF602961S, RSF604661S, RSF605660S, RSF605710S, RSF605925S, RSF605926S, RSF606169S, RSF600392S, RSF600424S, RSF600521S, RSF600619S, RSF600626S, RSF600637S, RSF600638S, RSF600667S, RSF600684S, RSF600773S, RSF600989S, RSF601253S, RSF601386S, RSF601558S, RSF602518S, RSF602573S, RSF602613S, RSF602733S, RSF602806S, RSF603050S, RSF603384S, RSF603407S, RSF603461S, RSF603463S, RSF603567S, RSF603637S, RSF603924S, RSF604165S, RSF604207S, RSF604285S, RSF604323S, RSF604758S, RSF601406S, RSF602716S, RSF603263S, RSF601281S, RSF602696S, RSF603869S, RSF605015S, RSF605681S, RSF605758S, RSF606175S, RSF600438S, RSF600635S, RSF600902S, RSF600993S, RSF601036S, RSF601230S, RSF601403S, RSF601429S, RSF603488S, RSF604316S, RSF604327S, RSF605046S, RSF605528S, RSF600663S, RSF600665S, RSF601087S, RSF600179S, RSF600198S, RSF600243S, RSF600332S, RSF600341S, RSF600813S, RSF600994S, RSF601039S, RSF601322S, RSF602468S, RSF602785S, RSF603179S, RSF605338S, RSF605649S, RSF605920S, RSF606053S, RSF606317S, RSF606482S, RSF607034S, RSF607038S, RSF607040S, RSF607041S, RSF607776S, RSF607778S, RSF607800S, RSF600340S, RSF600525S, RSF600925S, RSF601445S, RSF601553S, RSF602585S, RSF602612S, RSF602670S, RSF603374S, RSF603377S, RSF603389S, RSF603612S, RSF603615S, RSF603626S, RSF603719S, RSF603774S, RSF603903S, RSF604053S, RSF604238S, RSF604947S, RSF604951S, RSF604978S, RSF604999S, RSF607765S, RSF608049S, RSF608050S, RSF600354S, RSF600355S, RSF600919S, RSF601063S, RSF601177S, RSF601374S, RSF602833S, RSF603692S, RSF603972S, RSF604224S, RSF604225S, RSF604898S, RSF601286S, RSF600653S, RSF601027S, RSF602354S, RSF603383S, RSF603388S, RSF603448S, RSF603525S, RSF604787S, RSF605034S, RSF605970S, RSF606059S, RSF600295S, RSF600368S, RSF603951S, RSF604229S, RSF603867S, RSF600594S, RSF600791S, RSF601263S, RSF602692S, RSF604635S, RSF607065S, RSF606226S, RSF606270S, RSF607236S, RSF607334S, RSF607372S, RSF604250S, RSF604278S, RSF604279S, RSF607022S, RSF600161S, RSF600956S, RSF601234S, RSF605641S, RSF606181S, RSF607348S, RSF607350S, RSF601376S, RSF603469S, RSF604027S, RSF602439S, RSF605639S, RSF601104S, RSF602721S, RSF606987S, RSF600498S, RSF600770S, RSF607898S, RSF603423S, RSF603479S, RSF603548S, RSF605947S, RSF600754S, RSF600809S, RSF600942S, RSF601133S, RSF601200S, RSF603043S, RSF603849S, RSF604460S, RSF601320S, RSF603148S, RSF603218S, RSF603690S, RSF604455S, RSF602938S, RSF607880S, RSF600936S, RSF601136S, RSF601194S, RSF602181S, RSF602852S, RSF603192S, RSF603363S, RSF603591S, RSF603705S, RSF606009S, RSF606010S, RSF606019S, RSF607316S, RSF603556S, RSF605087S, RSF605088S, RSF605089S, RSF605340S, RSF605526S, RSF607317S, RSF607320S, RSF608046S, RSF600568S, RSF603090S, RSF604230S, RSF605617S, RSF600217S, RSF601082S, RSF601083S, RSF601143S, RSF602320S, RSF603193S, RSF603599S, RSF600535S, RSF603120S, RSF601000S, RSF601584S, RSF607972S, RSF607973S, RSF607976S, RSF607977S, RSF607978S, RSF607970S, RSF607971S, RSF607974S, RSF607975S, RSF607979S, RSF605655S, RSF605878S, RSF606234S, RSF606489S, RSF606491S, RSF607195S, RSF607196S, RSF607197S, RSF607198S, RSF607199S, RSF607200S, RSF607275S, RSF600147S, RSF600148S, RSF600190S, RSF600342S, RSF600347S, RSF600351S, RSF600365S, RSF600517S, RSF600590S, RSF600682S, RSF600683S, RSF600846S, RSF600860S, RSF600984S, RSF600992S, RSF601029S, RSF601166S, RSF601188S, RSF601295S, RSF601303S, RSF601310S, RSF602085S, RSF602159S, RSF602164S, RSF602179S, RSF602767S, RSF602815S, RSF602857S, RSF602870S, RSF602882S, RSF603116S, RSF603187S, RSF603663S, RSF603835S, RSF603873S, RSF604261S, RSF604262S, RSF604400S, RSF604837S, RSF605012S, RSF605262S, RSF605898S, RSF605899S, RSF605900S, RSF605901S, RSF606209S, RSF606210S, RSF606322S, RSF600311S, RSF600344S, RSF600559S, RSF600588S, RSF600600S, RSF600695S, RSF600702S, RSF600719S, RSF600800S, RSF600819S, RSF600831S, RSF600972S, RSF601182S, RSF601259S, RSF601396S, RSF601401S, RSF601489S, RSF602339S, RSF602593S, RSF602621S, RSF602627S, RSF602645S, RSF602647S, RSF602773S, RSF602774S, RSF602776S, RSF602940S, RSF603359S, RSF603375S, RSF603408S, RSF603506S, RSF603561S, RSF603745S, RSF603777S, RSF603859S, RSF603878S, RSF603905S, RSF603919S, RSF604167S, RSF604177S, RSF604208S, RSF604233S, RSF604426S, RSF604450S, RSF604477S, RSF604657S, RSF604796S, RSF605014S, RSF605038S, RSF605048S, RSF605507S, RSF605533S, RSF605534S, RSF605535S, RSF605562S, RSF605564S, RSF605518S, RSF600516S, RSF603536S, RSF604325S, RSF606968S, RSF607886S, RSF600227S, RSF600272S, RSF600457S, RSF600837S, RSF600939S, RSF601071S, RSF601112S, RSF603606S, RSF607336S, RSF600434S, RSF601390S, RSF604269S, RSF600513S, RSF601332S, RSF603173S, RSF600536S, RSF600829S, RSF601138S, RSF602403S, RSF603129S, RSF605693S, RSF605921S, RSF605922S, RSF606208S, RSF606315S, RSF607364S, RSF603681S, RSF604469S, RSF604689S, RSF604792S, RSF605027S, RSF605215S, RSF605508S, RSF605566S, RSF602923S, RSF605683S, RSF602519S, RSF603728S, RSF604265S, RSF602537S, RSF600259S, RSF600369S, RSF603779S, RSF605961S, RSF605992S, RSF600164S, RSF606237S, RSF602975S, RSF603825S, RSF603828S, RSF604138S, RSF602695S, RSF600248S, RSF600766S, RSF600172S, RSF600173S, RSF601312S, RSF603552S, RSF607717S, RSF600182S, RSF604198S, RSF603573S, RSF603644S, RSF604024S, RSF604118S, RSF605002S, RSF600562S, RSF600816S, RSF605859S, RSF606269S, RSF606331S, RSF606516S, RSF600163S, RSF600888S, RSF602430S, RSF602830S, RSF603126S, RSF603515S, RSF603547S, RSF603558S, RSF603577S, RSF603766S, RSF603881S, RSF603887S, RSF603954S, RSF604685S, RSF604696S, RSF604763S, RSF605010S, RSF605671S, RSF607045S, RSF607279S, RSF607895S, RSF602163S, RSF602793S, RSF603820S, RSF604252S, RSF604281S, RSF608055S, RSF601308S, RSF604249S, RSF600439S, RSF600714S, RSF600715S, RSF600734S, RSF600778S, RSF600856S, RSF601342S, RSF602722S, RSF603023S, RSF603581S, RSF603871S, RSF604096S, RSF604162S, RSF603588S, RSF604839S, RSF604998S, RSF607939S, RSF603364S, RSF603365S, RSF603974S, RSF604847S, RSF604901S, RSF607201S, RSF607851S, RSF608040S, RSF600885S, RSF600907S, RSF601576S, RSF602730S, RSF603234S, RSF603555S, RSF604185S, RSF604967S, RSF605239S, RSF603628S, RSF603629S, RSF600425S, RSF605647S, RSF605648S, RSF605737S, RSF605756S, RSF605913S, RSF605914S, RSF605917S, RSF606054S, RSF606055S, RSF606058S, RSF606178S, RSF606179S, RSF606204S, RSF606205S, RSF606247S, RSF606248S, RSF606249S, RSF606250S, RSF606319S, RSF606481S, RSF606485S, RSF606591S, RSF606878S, RSF606879S, RSF606883S, RSF607026S, RSF607748S, RSF607749S, RSF607752S, RSF607754S, RSF607755S, RSF607756S, RSF607760S, RSF607932S, RSF607933S, RSF608047S, RSF600330S, RSF600363S, RSF600423S, RSF600461S, RSF600491S, RSF600520S, RSF600522S, RSF600726S, RSF600727S, RSF600730S, RSF600746S, RSF600768S, RSF600798S, RSF600857S, RSF601221S, RSF601284S, RSF602513S, RSF602538S, RSF602542S, RSF603031S, RSF603144S, RSF603351S, RSF603353S, RSF603381S, RSF603495S, RSF603571S, RSF603608S, RSF603616S, RSF603967S, RSF604069S, RSF604166S, RSF604247S, RSF604318S, RSF604443S, RSF604475S, RSF604646S, RSF604672S, RSF604690S, RSF604790S, RSF604960S, RSF604963S, RSF605000S, RSF605192S, RSF605193S, RSF605219S, RSF605487S, RSF605490S, RSF605504S, RSF605506S, RSF605538S, RSF605542S, RSF605551S, RSF605553S, RSF605554S, RSF608000S, RSF607638S, RSF600300S, RSF600929S, RSF603427S, RSF603460S, RSF607893S, RSF606212S, RSF607042S, RSF607043S, RSF607297S, RSF607394S, RSF607703S, RSF600197S, RSF600912S, RSF600945S, RSF602643S, RSF602677S, RSF602851S, RSF602856S, RSF603235S, RSF603391S, RSF603560S, RSF604008S, RSF604104S, RSF604212S, RSF606185S, RSF606186S, RSF606243S, RSF606524S, RSF607202S, RSF600629S, RSF600646S, RSF601095S, RSF601111S, RSF601454S, RSF602890S, RSF603185S, RSF603583S, RSF603992S, RSF605227S, RSF605503S, RSF607728S, RSF600184S, RSF600735S, RSF601124S, RSF607937S, RSF607938S, RSF607936S, RSF605960S, RSF606253S, RSF606512S, RSF606522S, RSF607078S, RSF607228S, RSF607344S, RSF600194S, RSF600871S, RSF601458S, RSF601536S, RSF602440S, RSF602688S, RSF602824S, RSF602864S, RSF603145S, RSF603434S, RSF603907S, RSF604490S, RSF604667S, RSF600609S, RSF600614S, RSF603481S, RSF600744S, RSF604257S, RSF606254S, RSF606280S, RSF606281S, RSF606958S, RSF607230S, RSF607391S, RSF601005S, RSF602636S, RSF602686S, RSF602964S, RSF602966S, RSF602971S, RSF602978S, RSF603882S, RSF603884S, RSF603894S, RSF604199S, RSF604744S, RSF607871S, RSF601134S, RSF601090S, RSF603952S, RSF605229S, RSF603704S, RSF607874S, RSF608011S, RSF605589S, RSF605672S, RSF605673S, RSF605871S, RSF605872S, RSF606213S, RSF606215S, RSF606246S, RSF606495S, RSF606944S, RSF607060S, RSF607363S, RSF600921S, RSF601016S, RSF601051S, RSF601123S, RSF602554S, RSF602582S, RSF602614S, RSF602639S, RSF603119S, RSF603453S, RSF603487S, RSF603715S, RSF603862S, RSF604011S, RSF604253S, RSF604259S, RSF604299S, RSF604419S, RSF604833S, RSF605198S, RSF605407S, RSF605532S, RSF607061S, RSF605682S, RSF600265S, RSF600405S, RSF602556S, RSF602576S, RSF602811S, RSF603380S, RSF603393S, RSF603419S, RSF603421S, RSF603508S, RSF603621S, RSF603739S, RSF603877S, RSF603906S, RSF604276S, RSF604414S, RSF604693S, RSF605039S, RSF605235S, RSF605493S, RSF600143S, RSF600644S, RSF601220S, RSF601565S, RSF600202S, RSF600381S, RSF600480S, RSF607249S, RSF607379S, RSF607894S, RSF608090S, RSF603127S, RSF604678S, RSF603636S, RSF603997S, RSF605942S, RSF600264S, RSF601335S, RSF603442S, RSF603843S, RSF603617S, RSF604682S, RSF602900S, RSF604864S, RSF606483S, RSF607079S, RSF607751S, RSF600678S, RSF604452S, RSF606954S, RSF602566S, RSF606274S, RSF600624S, RSF602625S, RSF600280S, RSF600913S, RSF601099S, RSF602682S, RSF603688S, RSF602371S, RSF602800S, RSF603385S, RSF603387S, RSF603590S, RSF603679S, RSF603860S, RSF604828S, RSF604972S, RSF605001S, RSF605026S, RSF605218S, RSF605220S, RSF606170S, RSF601435S, RSF607881S, RSF600859S, RSF606222S, RSF606956S, RSF600595S, RSF602724S, RSF602397S, RSF605675S, RSF603141S, RSF606492S, RSF604189S, RSF604881S, RSF605197S, RSF607303S, RSF600288S, RSF600764S, RSF601037S, RSF603707S, RSF603725S, RSF605919S, RSF607788S, RSF607789S, RSF607790S, RSF600601S, RSF600604S, RSF600606S, RSF600851S, RSF601216S, RSF604144S, RSF604145S, RSF605009S, RSF605234S, RSF605409S, RSF605552S, RSF605986S, RSF604458S, RSF604068S, RSF604076S, RSF604077S, RSF604078S, RSF605106S, RSF605107S, RSF600799S, RSF601381S, RSF602567S, RSF603039S, RSF603040S, RSF603390S, RSF603394S, RSF603431S, RSF603610S, RSF603872S, RSF603874S, RSF604126S, RSF604495S, RSF604778S, RSF604965S, RSF605216S, RSF605486S, RSF605753S, RSF605854S, RSF608026S, RSF608035S, RSF608056S, RSF605709S, RSF605963S, RSF606171S, RSF606174S, RSF606176S, RSF606255S, RSF607075S, RSF600361S, RSF600496S, RSF600950S, RSF603401S, RSF603446S, RSF603753S, RSF604097S, RSF604161S, RSF604286S, RSF604836S, RSF608092S, RSF605949S, RSF605954S, RSF606182S, RSF606183S, RSF606242S, RSF608002S, RSF600660S, RSF601243S, RSF601252S, RSF601452S, RSF601562S, RSF601583S, RSF602187S, RSF602399S, RSF602828S, RSF602832S, RSF602840S, RSF602879S, RSF605969S, RSF600738S, RSF608012S, RSF607245S, RSF607246S, RSF607021S, RSF607918S, RSF600487S, RSF606520S, RSF600448S, RSF605703S, RSF602659S, RSF600464S, RSF602525S, RSF602747S, RSF602803S, RSF600497S, RSF600796S, RSF601294S, RSF602457S, RSF602697S, RSF603594S, RSF604137S, RSF605051S, RSF601315S, RSF601316S, RSF601341S, RSF603029S, RSF603056S, RSF604707S, RSF602352S, RSF602853S, RSF603791S, RSF600997S, RSF600175S, RSF600178S, RSF600200S, RSF600388S, RSF601157S, RSF601159S, RSF601160S, RSF601161S, RSF601476S, RSF602756S, RSF603414S, RSF603706S, RSF604184S, RSF601441S, RSF601457S, RSF601363S, RSF601364S, RSF601368S, RSF600218S, RSF601126S, RSF603998S, RSF604445S, RSF607396S, RSF602875S, RSF605952S, RSF600308S, RSF600495S, RSF600553S, RSF600578S, RSF600579S, RSF600881S, RSF600916S, RSF600995S, RSF601219S, RSF601273S, RSF602594S, RSF602760S, RSF603896S, RSF604275S, RSF604454S, RSF601351S, RSF601367S, RSF600187S, RSF605147S, RSF600904S, RSF604768S, RSF605500S, RSF607413S, RSF603956S, RSF607090S, RSF607221S, RSF607407S, RSF607899S, RSF605545S, RSF608005S, RSF605642S, RSF606172S, RSF600642S, RSF601399S, RSF603044S, RSF603474S, RSF603578S, RSF605055S, RSF606511S, RSF601257S, RSF602791S, RSF605971S, RSF607070S, RSF607285S, RSF607358S, RSF600485S, RSF600784S, RSF601020S, RSF601163S, RSF601361S, RSF601460S, RSF601472S, RSF602269S, RSF602801S, RSF602929S, RSF603635S, RSF603953S, RSF607354S, RSF604641S, RSF605951S, RSF601418S, RSF603462S, RSF601195S, RSF604256S, RSF602521S, RSF600479S, RSF600551S, RSF601180S, RSF601199S, RSF601590S, RSF602189S, RSF602198S, RSF602199S, RSF603118S, RSF603554S, RSF603582S, RSF603671S, RSF604072S, RSF604123S, RSF604242S, RSF604014S, RSF603579S, RSF600893S, RSF603117S, RSF604066S, RSF600239S, RSF600541S, RSF603010S, RSF603095S, RSF603110S, RSF604322S, RSF604438S, RSF606332S, RSF600486S, RSF600519S, RSF601015S, RSF602959S, RSF603087S, RSF603959S, RSF605694S, RSF606238S, RSF601072S, RSF603370S, RSF605484S, RSF605238S, RSF603140S, RSF601197S, RSF601383S, RSF603094S, RSF603231S, RSF603233S, RSF603498S, RSF603667S, RSF607410S, RSF607411S, RSF602982S, RSF603250S, RSF603251S, RSF605686S, RSF605989S, RSF607866S, RSF600249S, RSF600285S, RSF601032S, RSF601034S, RSF601040S, RSF601041S, RSF601042S, RSF601046S, RSF601317S, RSF601412S, RSF602406S, RSF602633S, RSF602673S, RSF602675S, RSF602754S, RSF602814S, RSF602871S, RSF603121S, RSF603123S, RSF603124S, RSF603125S, RSF603128S, RSF603168S, RSF603220S, RSF603721S, RSF603795S, RSF603840S, RSF604459S, RSF600168S, RSF601385S, RSF600266S, RSF600718S, RSF600882S, RSF600803S, RSF600245S, RSF600326S, RSF600416S, RSF600584S, RSF600634S, RSF600978S, RSF601233S, RSF602487S, RSF602545S, RSF602768S, RSF603028S, RSF603531S, RSF600268S, RSF602821S, RSF604017S, RSF602726S, RSF602805S, RSF603673S, RSF603698S, RSF605272S, RSF605627S, RSF605653S, RSF606341S, RSF600316S, RSF600378S, RSF600415S, RSF600447S, RSF600556S, RSF600580S, RSF600582S, RSF600750S, RSF601304S, RSF601440S, RSF602766S, RSF602798S, RSF602907S, RSF603485S, RSF603524S, RSF603674S, RSF603768S, RSF604421S, RSF604764S, RSF605043S, RSF605491S, RSF600366S, RSF601023S, RSF601271S, RSF601371S, RSF602942S, RSF604132S, RSF604183S, RSF607290S, RSF601237S, RSF601348S, RSF603350S, RSF603780S, RSF605369S, RSF605413S, RSF602772S, RSF600772S, RSF602520S, RSF602640S, RSF602867S, RSF602919S, RSF603597S, RSF603888S, RSF603904S, RSF603989S, RSF603483S, RSF600533S, RSF604709S, RSF604713S, RSF604714S, RSF607746S, RSF600465S, RSF606890S, RSF606995S, RSF607203S, RSF600186S, RSF600199S, RSF600201S, RSF600230S, RSF600287S, RSF600293S, RSF600468S, RSF600469S, RSF601156S, RSF601578S, RSF602986S, RSF603021S, RSF603361S, RSF605217S, RSF600231S, RSF600286S, RSF600294S, RSF600317S, RSF600470S, RSF600810S, RSF600981S, RSF601004S, RSF601302S, RSF601485S, RSF602325S, RSF603459S, RSF606257S, RSF607054S, RSF600449S, RSF601248S, RSF602536S, RSF602539S, RSF602563S, RSF602642S, RSF602655S, RSF603420S, RSF603477S, RSF603503S, RSF603624S, RSF604241S, RSF604428S, RSF604706S, RSF604774S, RSF604775S, RSF604843S, RSF607210S, RSF600923S, RSF600930S, RSF601114S, RSF601115S, RSF601276S, RSF602365S, RSF602825S, RSF603410S, RSF603422S, RSF603786S, RSF603787S, RSF603790S, RSF603796S, RSF604650S, RSF604835S, RSF605029S, RSF605053S, RSF605226S, RSF607272S, RSF605883S, RSF600278S, RSF600301S, RSF600349S, RSF600362S, RSF600625S, RSF600792S, RSF600834S, RSF601010S, RSF601044S, RSF601130S, RSF601378S, RSF601455S, RSF601466S, RSF602161S, RSF602581S, RSF602583S, RSF602762S, RSF602878S, RSF603355S, RSF603836S, RSF604270S, RSF604695S, RSF605569S, RSF608085S, RSF605575S, RSF605880S, RSF606184S, RSF606926S, RSF600196S, RSF600277S, RSF601300S, RSF602548S, RSF602679S, RSF602731S, RSF602787S, RSF602866S, RSF603194S, RSF603486S, RSF603642S, RSF603797S, RSF603822S, RSF603845S, RSF603899S, RSF604074S, RSF604127S, RSF604141S, RSF604501S, RSF604769S, RSF605195S, RSF603085S, RSF603086S, RSF605033S, RSF600305S, RSF600542S, RSF600583S, RSF600586S, RSF600620S, RSF600690S, RSF601026S, RSF601218S, RSF601225S, RSF600509S, RSF600547S, RSF600565S, RSF600758S, RSF600760S, RSF600765S, RSF600776S, RSF600877S, RSF600879S, RSF600918S, RSF605646S, RSF605892S, RSF605906S, RSF605907S, RSF606190S, RSF606219S, RSF606240S, RSF606539S, RSF606988S, RSF607310S, RSF607339S, RSF607340S, RSF607355S, RSF607408S, RSF607409S, RSF608008S, RSF600236S, RSF600983S, RSF601293S, RSF601463S, RSF602624S, RSF602649S, RSF602650S, RSF602656S, RSF602687S, RSF602850S, RSF602972S, RSF602973S, RSF603025S, RSF603245S, RSF603247S, RSF603666S, RSF603730S, RSF604021S, RSF604474S, RSF604493S, RSF604644S, RSF604773S, RSF604941S, RSF604974S, RSF605222S, RSF605439S, RSF605441S, RSF605494S, RSF605498S, RSF605557S, RSF604626S, RSF604715S, RSF604716S, RSF605095S, RSF605099S, RSF605105S, RSF605108S, RSF605109S, RSF605110S, RSF605111S, RSF607115S, RSF600987S, RSF601043S, RSF601045S, RSF601047S, RSF601050S, RSF601052S, RSF601054S, RSF601055S, RSF601056S, RSF601057S, RSF601059S, RSF601066S, RSF601067S, RSF601068S, RSF601069S, RSF601074S, RSF602941S, RSF602953S, RSF602954S, RSF602957S, RSF602960S, RSF602962S, RSF604001S, RSF604004S, RSF604007S, RSF606321S, RSF608096S, RSF600574S, RSF600700S, RSF603722S, RSF606064S, RSF606065S, RSF603936S, RSF603975S, RSF605953S, RSF605957S, RSF606999S, RSF600693S, RSF600823S, RSF601389S, RSF601552S, RSF602488S, RSF602788S, RSF600226S, RSF601018S, RSF601110S, RSF601270S, RSF606517S, RSF600244S, RSF604067S, RSF607288S, RSF607346S, RSF601210S, RSF601292S, RSF601478S, RSF602704S, RSF603830S, RSF604226S, RSF605058S, RSF603038S, RSF606490S, RSF607047S, RSF607223S, RSF607269S, RSF607295S, RSF607381S, RSF603798S, RSF602740S, RSF603433S, RSF603609S, RSF604171S, RSF603059S, RSF604057S, RSF605668S, RSF607693S, RSF601236S, RSF601465S, RSF601223S, RSF603834S, RSF605983S, RSF608038S, RSF603944S, RSF604203S, RSF600654S, RSF601268S, RSF604680S, RSF600691S, RSF603514S, RSF605887S, RSF606241S, RSF606991S, RSF607241S, RSF607273S, RSF607305S, RSF607398S, RSF607999S, RSF600952S, RSF600953S, RSF602729S, RSF602742S, RSF603133S, RSF603395S, RSF605236S, RSF605438S, RSF608126S, RSF605704S, RSF605965S, RSF606232S, RSF606537S, RSF602503S, RSF602735S, RSF604094S, RSF604288S, RSF604964S, RSF602265S, RSF600167S, RSF600174S, RSF600242S, RSF600292S, RSF600587S, RSF600592S, RSF600613S, RSF603207S, RSF603478S, RSF603564S, RSF602955S, RSF606923S, RSF602569S, RSF603930S, RSF604009S, RSF604061S, RSF606008S, RSF606480S, RSF606881S, RSF607036S, RSF600529S, RSF600618S, RSF602769S, RSF603358S, RSF603922S, RSF604131S, RSF604449S, RSF604651S, RSF604652S, RSF605541S, RSF605689S, RSF600575S, RSF603254S, RSF603241S, RSF603670S, RSF604085S, RSF604752S, RSF600499S, RSF601084S, RSF601325S, RSF604155S, RSF604274S, RSF603096S, RSF603101S, RSF603366S, RSF603928S, RSF604673S, RSF604679S, RSF600379S, RSF606268S, RSF600181S, RSF603530S, RSF602492S, RSF602661S, RSF603669S, RSF600193S, RSF602846S, RSF605346S, RSF601329S, RSF602550S, RSF603640S, RSF606180S, RSF606206S, RSF606880S, RSF607031S, RSF607250S, RSF600526S, RSF600673S, RSF600970S, RSF602568S, RSF602606S, RSF602622S, RSF603584S, RSF603770S, RSF603900S, RSF603979S, RSF604046S, RSF604676S, RSF604686S, RSF604741S, RSF605023S, RSF607929S, RSF604113S, RSF603819S, RSF601481S, RSF602765S, RSF605665S, RSF605228S, RSF600372S, RSF602167S, RSF602718S, RSF602737S, RSF602823S, RSF606966S, RSF601362S, RSF602517S, RSF604026S, RSF601298S, RSF606887S, RSF600752S, RSF604149S, RSF600647S, RSF605094S, RSF603451S, RSF603714S, RSF603438S, RSF603948S, RSF605044S, RSF605045S, RSF604063S, RSF603993S, RSF605667S, RSF605905S, RSF606187S, RSF606220S, RSF606258S, RSF606279S, RSF606326S, RSF606515S, RSF606989S, RSF607209S, RSF607248S, RSF607265S, RSF607293S, RSF607347S, RSF607369S, RSF607415S, RSF607416S, RSF608007S, RSF601201S, RSF601203S, RSF601204S, RSF601319S, RSF602377S, RSF602415S, RSF602467S, RSF602681S, RSF602736S, RSF602829S, RSF603001S, RSF603020S, RSF603115S, RSF603224S, RSF603502S, RSF603694S, RSF603857S, RSF603988S, RSF604075S, RSF604239S, RSF604384S, RSF604442S, RSF604944S, RSF605124S, RSF605194S, RSF605221S, RSF605237S, RSF605258S, RSF605546S, RSF605558S, RSF607353S, RSF600335S, RSF600336S, RSF600339S, RSF601340S, RSF602712S, RSF600304S, RSF600306S, RSF600307S, RSF600284S, RSF601006S, RSF603509S, RSF603732S, RSF606541S, RSF606538S, RSF606244S, RSF601297S, RSF608075S, RSF608111S, RSF600412S, RSF600418S, RSF600460S, RSF600462S, RSF601183S, RSF601347S, RSF601380S, RSF601470S, RSF601556S, RSF602669S, RSF602705S, RSF602796S, RSF605858S, RSF608025S, RSF608027S, RSF604982S, RSF601344S, RSF601451S, RSF605657S, RSF605868S, RSF605981S, RSF607341S, RSF600960S, RSF604193S, RSF604330S, RSF605004S, RSF606553S, RSF604140S, RSF602558S, RSF602561S, RSF600367S, RSF602424S, RSF605685S, RSF600188S, RSF600189S, RSF600276S, RSF600291S, RSF600482S, RSF600506S, RSF600514S, RSF600763S, RSF600944S, RSF601024S, RSF601146S, RSF601431S, RSF602410S, RSF602549S, RSF602693S, RSF602758S, RSF603137S, RSF603138S, RSF603143S, RSF603466S, RSF604264S, RS
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
- Reason for Recall:
- There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
ICD COBALT VR MRI DF4, Model Number DVPB3D4; Implantable Cardioverter Defibrillators
Product Codes/Lot Numbers:
GTIN 00763000178512, Lot Serial Numbers: RSF600012S, RSF600254S, RSF600591S, RSF600655S, RSF600822S, RSF601247S, RSF602653S, RSF603908S, RSF601102S, RSF607914S, RSF605967S, RSF600633S, RSF600649S, RSF600867S, RSF600868S, RSF600869S, RSF600872S, RSF600873S, RSF600874S, RSF600875S, RSF600876S, RSF600878S, RSF600880S, RSF600883S, RSF602930S, RSF602991S, RSF602998S, RSF603002S, RSF603003S, RSF603013S, RSF603018S, RSF603019S, RSF603022S, RSF603026S, RSF603027S, RSF603030S, RSF603032S, RSF603033S, RSF603034S, RSF603035S, RSF603036S, RSF603736S, RSF603737S, RSF603741S, RSF603742S, RSF603744S, RSF603746S, RSF603747S, RSF603749S, RSF603754S, RSF603755S, RSF603756S, RSF603758S, RSF603759S, RSF603762S, RSF603763S, RSF603764S, RSF603799S, RSF603802S, RSF603804S, RSF603806S, RSF603810S, RSF603813S, RSF603818S, RSF603821S, RSF603824S, RSF602665S, RSF602703S, RSF603249S, RSF603648S, RSF603939S, RSF607644S, RSF601171S, RSF601174S, RSF602584S, RSF602743S, RSF600887S, RSF602678S, RSF608130S, RSF600767S, RSF608097S, RSF604855S, RSF606263S, RSF601206S, RSF601486S, RSF601567S, RSF601571S, RSF600924S, RSF601224S, RSF601226S, RSF602817S, RSF603405S, RSF603521S, RSF603678S, RSF602595S, RSF604129S, RSF600687S, RSF600210S, RSF602849S, RSF607243S, RSF607244S, RSF607291S, RSF604110S, RSF604146S, RSF605150S, RSF607995S, RSF600352S, RSF601546S, RSF607225S, RSF606605S, RSF600269S, RSF600940S, RSF601190S, RSF603550S, RSF604130S, RSF608014S, RSF600303S, RSF602434S, RSF602435S, RSF602449S, RSF602461S, RSF602466S, RSF607173S, RSF607174S, RSF600954S, RSF600955S, RSF602586S, RSF602587S, RSF602597S, RSF602804S, RSF603465S, RSF600982S, RSF603213S, RSF605644S, RSF606195S, RSF601058S, RSF603647S, RSF603724S, RSF604582S, RSF600219S, RSF601397S, RSF601539S, RSF602709S, RSF602757S, RSF603794S, RSF603973S, RSF604054S, RSF600443S, RSF600706S, RSF600721S, RSF600729S, RSF601434S, RSF603726S, RSF606532S, RSF603464S, RSF608114S, RSF601093S, RSF603450S, RSF607337S, RSF607359S, RSF600793S, RSF600805S, RSF600808S, RSF600864S, RSF601132S, RSF601165S, RSF601176S, RSF601540S, RSF601547S, RSF601587S, RSF602604S, RSF602874S, RSF603248S, RSF603399S, RSF603415S, RSF603428S, RSF603909S, RSF605018S, RSF605037S, RSF605025S, RSF601119S, RSF601462S, RSF607023S, RSF605918S, RSF606278S, RSF606884S, RSF600539S, RSF600523S, RSF602564S, RSF602565S, RSF602570S, RSF602571S, RSF603242S, RSF606921S, RSF601482S, RSF602165S, RSF602753S, RSF602822S, RSF602877S, RSF603585S, RSF605489S, RSF605974S, RSF607360S, RSF602220S, RSF602786S, RSF602789S, RSF604164S, RSF606324S, RSF606934S, RSF601167S, RSF601169S, RSF601172S, RSF602790S, RSF603216S, RSF603470S, RSF603549S, RSF604227S, RSF605701S, RSF605928S, RSF607222S, RSF607740S, RSF607741S, RSF607911S, RSF600260S, RSF601106S, RSF602684S, RSF603240S, RSF603468S, RSF603685S, RSF604089S, RSF604112S, RSF604121S, RSF604263S, RSF604389S, RSF604470S, RSF604700S, RSF604988S, RSF605889S, RSF600835S, RSF603897S, RSF603963S, RSF603966S, RSF605021S, RSF600263S, RSF600534S, RSF601142S, RSF601572S, RSF602208S, RSF602373S, RSF602632S, RSF602706S, RSF602848S, RSF603416S, RSF603527S, RSF603528S, RSF603662S, RSF603683S, RSF603892S, RSF604013S, RSF605570S, RSF607887S, RSF601305S, RSF600610S, RSF603260S, RSF603490S, RSF601446S, RSF605645S, RSF601534S, RSF602463S, RSF602698S, RSF603191S, RSF603510S, RSF603689S, RSF603916S, RSF603986S, RSF604266S, RSF605205S, RSF605206S, RSF605505S, RSF605670S, RSF600312S, RSF600889S, RSF604284S, RSF604471S, RSF604502S, RSF605697S, RSF605876S, RSF606313S, RSF606942S, RSF607052S, RSF607058S, RSF607714S, RSF607720S, RSF607721S, RSF603734S, RSF603870S, RSF603911S, RSF604010S, RSF604156S, RSF604289S, RSF604302S, RSF604425S, RSF604748S, RSF605054S, RSF607050S, RSF607733S, RSF607739S, RSF607763S, RSF608003S, RSF603960S, RSF605443S, RSF603543S, RSF607240S, RSF607378S, RSF607397S, RSF607406S, RSF600153S, RSF600328S, RSF600705S, RSF602707S, RSF603595S, RSF603773S, RSF600977S, RSF601471S, RSF601588S, RSF602343S, RSF602889S, RSF600162S, RSF600557S, RSF605875S, RSF606245S, RSF607055S, RSF607056S, RSF607716S, RSF601028S, RSF601033S, RSF601091S, RSF601189S, RSF601384S, RSF604197S, RSF604660S, RSF604701S, RSF604751S, RSF604905S, RSF607935S, RSF600717S, RSF603024S, RSF608070S, RSF603406S, RSF604073S, RSF600315S, RSF600420S, RSF606967S, RSF606367S, RSF606885S, RSF607025S, RSF607066S, RSF608102S, RSF608104S, RSF605699S, RSF605909S, RSF606985S, RSF607380S, RSF607697S, RSF607900S, RSF608009S, RSF600414S, RSF600671S, RSF601013S, RSF602591S, RSF602702S, RSF603130S, RSF603261S, RSF603449S, RSF603452S, RSF603492S, RSF603695S, RSF603823S, RSF603957S, RSF604087S, RSF604232S, RSF604268S, RSF604387S, RSF605559S, RSF606193S, RSF606194S, RSF606309S, RSF606534S, RSF600508S, RSF604838S, RSF606282S, RSF606513S, RSF603775S, RSF606311S, RSF607877S, RSF600708S, RSF600894S, RSF602727S, RSF602831S, RSF604604S, RSF606233S, RSF606894S, RSF607205S, RSF600815S, RSF604219S, RSF604448S, RSF607869S, RSF605419S, RSF607870S, RSF607030S, RSF607035S, RSF607934S, RSF605610S, RSF605964S, RSF606907S, RSF607073S, RSF607916S, RSF600472S, RSF600475S, RSF600696S, RSF600783S, RSF600802S, RSF600847S, RSF600911S, RSF601053S, RSF601192S, RSF601531S, RSF601560S, RSF602607S, RSF603100S, RSF603456S, RSF607366S, RSF602326S, RSF608093S, RSF603933S, RSF600241S, RSF605661S, RSF605908S, RSF600503S, RSF600886S, RSF601086S, RSF601214S, RSF601260S, RSF601279S, RSF601394S, RSF602547S, RSF602559S, RSF602654S, RSF602795S, RSF603785S, RSF601573S, RSF600801S, RSF601345S, RSF601467S, RSF603596S, RSF603649S, RSF604082S, RSF600701S, RSF605865S, RSF605866S, RSF605869S, RSF606217S, RSF606950S, RSF607044S, RSF607062S, RSF607067S, RSF601395S, RSF601469S, RSF601586S, RSF602589S, RSF602683S, RSF603214S, RSF603858S, RSF607904S, RSF606523S, RSF607928S, RSF600195S, RSF601250S, RSF601475S, RSF602578S, RSF603553S, RSF603559S, RSF603729S, RSF603832S, RSF603837S, RSF603875S, RSF603902S, RSF603982S, RSF604598S, RSF605436S, RSF600433S, RSF600615S, RSF600674S, RSF600762S, RSF600785S, RSF602744S, RSF602892S, RSF603713S, RSF603727S, RSF604237S, RSF605856S, RSF600566S, RSF601215S, RSF603602S, RSF603946S, RSF605030S, RSF603042S, RSF603587S, RSF601323S, RSF607949S, RSF607950S, RSF600552S, RSF600623S, RSF600711S, RSF600782S, RSF601088S, RSF602504S, RSF602770S, RSF605695S, RSF606528S, RSF603655S, RSF602205S, RSF600736S, RSF600865S, RSF600866S, RSF601003S, RSF601098S, RSF602177S, RSF602895S, RSF603227S, RSF603511S, RSF603544S, RSF604135S, RSF604823S, RSF605990S, RSF601100S, RSF601147S, RSF601185S, RSF601353S, RSF601354S, RSF602888S, RSF603489S, RSF604401S, RSF604968S, RSF605991S, RSF606221S, RSF607867S, RSF600360S, RSF601533S, RSF600561S, RSF600676S, RSF601549S, RSF603999S, RSF602623S, RSF600410S, RSF600937S, RSF602816S, RSF602827S, RSF603378S, RSF605945S, RSF608117S, RSF600282S, RSF600324S, RSF604759S, RSF600751S, RSF604115S, RSF607662S, RSF602592S, RSF602983S, RSF602985S, RSF603497S, RSF603519S, RSF603958S, RSF603987S, RSF603994S, RSF604186S, RSF604195S, RSF606189S, RSF606498S, RSF606535S, RSF606947S, RSF607330S, RSF602905S, RSF603104S, RSF604977S, RSF605086S, RSF607890S, RSF608062S, RSF602626S, RSF604260S, RSF602713S, RSF608041S, RSF608042S, RSF608043S, RSF608044S, RSF608045S, RSF605968S, RSF600749S, RSF600985S, RSF601212S, RSF603199S, RSF603402S, RSF603651S, RSF603652S, RSF603815S, RSF603817S, RSF603886S, RSF604217S, RSF605708S, RSF606256S, RSF600476S, RSF600477S, RSF601331S, RSF602334S, RSF602359S, RSF602629S, RSF602777S, RSF603221S, RSF603600S, RSF603735S, RSF603829S, RSF604413S, RSF604906S, RSF607227S, RSF607386S, RSF600492S, RSF600806S, RSF600975S, RSF601021S, RSF601025S, RSF601031S, RSF601265S, RSF601570S, RSF602494S, RSF602921S, RSF604154S, RSF604730S, RSF604883S, RSF604886S, RSF600154S, RSF600267S, RSF600843S, RSF601555S, RSF601574S, RSF602637S, RSF602700S, RSF602711S, RSF603054S, RSF603201S, RSF603369S, RSF603789S, RSF603918S, RSF603932S, RSF603955S, RSF603991S, RSF604098S, RSF604188S, RSF604766S, RSF600666S, RSF600949S, RSF602672S, RSF602988S, RSF603864S, RSF602619S, RSF607247S, RSF601449S, RSF603223S, RSF602423S, RSF606943S, RSF600395S, RSF600427S, RSF600504S, RSF601231S, RSF601278S, RSF601542S, RSF602802S, RSF603945S, RSF600309S, RSF600314S, RSF600560S, RSF600747S, RSF602327S, RSF602329S, RSF602411S, RSF605810S, RSF605849S, RSF605851S, RSF605852S, RSF605853S, RSF606197S, RSF606198S, RSF606200S, RSF606201S, RSF607098S, RSF607099S, RSF607101S, RSF607114S, RSF607116S, RSF607951S, RSF607952S, RSF601109S, RSF602897S, RSF603060S, RSF603062S, RSF603063S, RSF603065S, RSF603066S, RSF603069S, RSF603070S, RSF603071S, RSF603107S, RSF603639S, RSF603676S, RSF603801S, RSF603803S, RSF603807S, RSF603808S, RSF603809S, RSF603811S, RSF603812S, RSF605366S, RSF607097S, RSF607392S, RSF607393S, RSF607953S, RSF607954S, RSF607955S, RSF607956S, RSF607957S, RSF607958S, RSF607959S, RSF607960S, RSF607961S, RSF607962S, RSF607963S, RSF607964S, RSF605650S, RSF600398S, RSF600399S, RSF600401S, RSF600407S, RSF600743S, RSF600836S, RSF600838S, RSF600839S, RSF601261S, RSF601306S, RSF601314S, RSF601532S, RSF603106S, RSF603800S, RSF605678S, RSF607074S, RSF602425S, RSF602741S, RSF605956S, RSF603576S, RSF600511S, RSF600512S, RSF602372S, RSF602751S, RSF602752S, RSF602808S, RSF602932S, RSF602969S, RSF602974S, RSF603132S, RSF603362S, RSF603538S, RSF603697S, RSF603876S, RSF604012S, RSF604020S, RSF604173S, RSF604205S, RSF604344S, RSF604481S, RSF604482S, RSF604485S, RSF604500S, RSF602512S, RSF601491S, RSF606519S, RSF607271S, RSF602427S, RSF602859S, RSF602958S, RSF603675S, RSF600274S, RSF602553S, RSF607212S, RSF603430S, RSF603632S, RSF603653S, RSF608036S, RSF600142S, RSF600670S, RSF603625S, RSF601240S, RSF600532S, RSF600739S, RSF600741S, RSF604032S, RSF600967S, RSF605007S, RSF605008S, RSF606286S, RSF607865S, RSF608022S, RSF600209S, RSF600346S, RSF601075S, RSF601096S, RSF602605S, RSF602628S, RSF602780S, RSF603266S, RSF603522S, RSF603529S, RSF603491S, RSF607387S, RSF600668S, RSF602575S, RSF603512S, RSF607942S, RSF607943S, RSF607944S, RSF607945S, RSF607946S, RSF605684S, RSF605950S, RSF605962S, RSF606173S, RSF606228S, RSF606229S, RSF606230S, RSF606264S, RSF606265S, RSF606503S, RSF606504S, RSF606521S, RSF606963S, RSF607077S, RSF607266S, RSF607283S, RSF607377S, RSF600466S, RSF600605S, RSF600812S, RSF600951S, RSF601232S, RSF601235S, RSF601238S, RSF601242S, RSF601246S, RSF601262S, RSF601559S, RSF602462S, RSF602522S, RSF602524S, RSF602526S, RSF602527S, RSF602745S, RSF602748S, RSF602749S, RSF602809S, RSF603349S, RSF603494S, RSF603619S, RSF603738S, RSF603879S, RSF603935S, RSF603938S, RSF603971S, RSF603976S, RSF604055S, RSF604153S, RSF604170S, RSF604181S, RSF604255S, RSF604440S, RSF604488S, RSF604496S, RSF604668S, RSF604669S, RSF604670S, RSF604723S, RSF604821S, RSF604826S, RSF604904S, RSF605020S, RSF605976S, RSF600413S, RSF602638S, RSF602720S, RSF602841S, RSF602869S, RSF603429S, RSF603445S, RSF603940S, RSF600558S, RSF600927S, RSF601191S, RSF601213S, RSF602873S, RSF600431S, RSF600432S, RSF601120S, RSF602844S, RSF603226S, RSF604822S, RSF600598S, RSF607735S, RSF601333S, RSF602868S, RSF600374S, RSF601241S, RSF603784S, RSF600319S, RSF602511S, RSF600377S, RSF600926S, RSF601343S, RSF602862S, RSF604216S, RSF605676S, RSF606882S, RSF607027S, RSF607772S, RSF607780S, RSF607791S, RSF607927S, RSF608001S, RSF600555S, RSF600640S, RSF600658S, RSF600685S, RSF600686S, RSF600795S, RSF601228S, RSF601264S, RSF603099S, RSF603352S, RSF603397S, RSF603424S, RSF603541S, RSF603623S, RSF603708S, RSF603856S, RSF604176S, RSF604231S, RSF604310S, RSF604952S, RSF604990S, RSF605035S, RSF607771S, RSF600393S, RSF601014S, RSF603943S, RSF601162S, RSF602364S, RSF602928S, RSF600699S, RSF601070S, RSF601139S, RSF601154S, RSF601402S, RSF602784S, RSF604245S, RSF604258S, RSF604052S, RSF605870S, RSF601448S, RSF602609S, RSF604200S, RSF601272S, RSF602961S, RSF604661S, RSF605660S, RSF605710S, RSF605925S, RSF605926S, RSF606169S, RSF600392S, RSF600424S, RSF600521S, RSF600619S, RSF600626S, RSF600637S, RSF600638S, RSF600667S, RSF600684S, RSF600773S, RSF600989S, RSF601253S, RSF601386S, RSF601558S, RSF602518S, RSF602573S, RSF602613S, RSF602733S, RSF602806S, RSF603050S, RSF603384S, RSF603407S, RSF603461S, RSF603463S, RSF603567S, RSF603637S, RSF603924S, RSF604165S, RSF604207S, RSF604285S, RSF604323S, RSF604758S, RSF601406S, RSF602716S, RSF603263S, RSF601281S, RSF602696S, RSF603869S, RSF605015S, RSF605681S, RSF605758S, RSF606175S, RSF600438S, RSF600635S, RSF600902S, RSF600993S, RSF601036S, RSF601230S, RSF601403S, RSF601429S, RSF603488S, RSF604316S, RSF604327S, RSF605046S, RSF605528S, RSF600663S, RSF600665S, RSF601087S, RSF600179S, RSF600198S, RSF600243S, RSF600332S, RSF600341S, RSF600813S, RSF600994S, RSF601039S, RSF601322S, RSF602468S, RSF602785S, RSF603179S, RSF605338S, RSF605649S, RSF605920S, RSF606053S, RSF606317S, RSF606482S, RSF607034S, RSF607038S, RSF607040S, RSF607041S, RSF607776S, RSF607778S, RSF607800S, RSF600340S, RSF600525S, RSF600925S, RSF601445S, RSF601553S, RSF602585S, RSF602612S, RSF602670S, RSF603374S, RSF603377S, RSF603389S, RSF603612S, RSF603615S, RSF603626S, RSF603719S, RSF603774S, RSF603903S, RSF604053S, RSF604238S, RSF604947S, RSF604951S, RSF604978S, RSF604999S, RSF607765S, RSF608049S, RSF608050S, RSF600354S, RSF600355S, RSF600919S, RSF601063S, RSF601177S, RSF601374S, RSF602833S, RSF603692S, RSF603972S, RSF604224S, RSF604225S, RSF604898S, RSF601286S, RSF600653S, RSF601027S, RSF602354S, RSF603383S, RSF603388S, RSF603448S, RSF603525S, RSF604787S, RSF605034S, RSF605970S, RSF606059S, RSF600295S, RSF600368S, RSF603951S, RSF604229S, RSF603867S, RSF600594S, RSF600791S, RSF601263S, RSF602692S, RSF604635S, RSF607065S, RSF606226S, RSF606270S, RSF607236S, RSF607334S, RSF607372S, RSF604250S, RSF604278S, RSF604279S, RSF607022S, RSF600161S, RSF600956S, RSF601234S, RSF605641S, RSF606181S, RSF607348S, RSF607350S, RSF601376S, RSF603469S, RSF604027S, RSF602439S, RSF605639S, RSF601104S, RSF602721S, RSF606987S, RSF600498S, RSF600770S, RSF607898S, RSF603423S, RSF603479S, RSF603548S, RSF605947S, RSF600754S, RSF600809S, RSF600942S, RSF601133S, RSF601200S, RSF603043S, RSF603849S, RSF604460S, RSF601320S, RSF603148S, RSF603218S, RSF603690S, RSF604455S, RSF602938S, RSF607880S, RSF600936S, RSF601136S, RSF601194S, RSF602181S, RSF602852S, RSF603192S, RSF603363S, RSF603591S, RSF603705S, RSF606009S, RSF606010S, RSF606019S, RSF607316S, RSF603556S, RSF605087S, RSF605088S, RSF605089S, RSF605340S, RSF605526S, RSF607317S, RSF607320S, RSF608046S, RSF600568S, RSF603090S, RSF604230S, RSF605617S, RSF600217S, RSF601082S, RSF601083S, RSF601143S, RSF602320S, RSF603193S, RSF603599S, RSF600535S, RSF603120S, RSF601000S, RSF601584S, RSF607972S, RSF607973S, RSF607976S, RSF607977S, RSF607978S, RSF607970S, RSF607971S, RSF607974S, RSF607975S, RSF607979S, RSF605655S, RSF605878S, RSF606234S, RSF606489S, RSF606491S, RSF607195S, RSF607196S, RSF607197S, RSF607198S, RSF607199S, RSF607200S, RSF607275S, RSF600147S, RSF600148S, RSF600190S, RSF600342S, RSF600347S, RSF600351S, RSF600365S, RSF600517S, RSF600590S, RSF600682S, RSF600683S, RSF600846S, RSF600860S, RSF600984S, RSF600992S, RSF601029S, RSF601166S, RSF601188S, RSF601295S, RSF601303S, RSF601310S, RSF602085S, RSF602159S, RSF602164S, RSF602179S, RSF602767S, RSF602815S, RSF602857S, RSF602870S, RSF602882S, RSF603116S, RSF603187S, RSF603663S, RSF603835S, RSF603873S, RSF604261S, RSF604262S, RSF604400S, RSF604837S, RSF605012S, RSF605262S, RSF605898S, RSF605899S, RSF605900S, RSF605901S, RSF606209S, RSF606210S, RSF606322S, RSF600311S, RSF600344S, RSF600559S, RSF600588S, RSF600600S, RSF600695S, RSF600702S, RSF600719S, RSF600800S, RSF600819S, RSF600831S, RSF600972S, RSF601182S, RSF601259S, RSF601396S, RSF601401S, RSF601489S, RSF602339S, RSF602593S, RSF602621S, RSF602627S, RSF602645S, RSF602647S, RSF602773S, RSF602774S, RSF602776S, RSF602940S, RSF603359S, RSF603375S, RSF603408S, RSF603506S, RSF603561S, RSF603745S, RSF603777S, RSF603859S, RSF603878S, RSF603905S, RSF603919S, RSF604167S, RSF604177S, RSF604208S, RSF604233S, RSF604426S, RSF604450S, RSF604477S, RSF604657S, RSF604796S, RSF605014S, RSF605038S, RSF605048S, RSF605507S, RSF605533S, RSF605534S, RSF605535S, RSF605562S, RSF605564S, RSF605518S, RSF600516S, RSF603536S, RSF604325S, RSF606968S, RSF607886S, RSF600227S, RSF600272S, RSF600457S, RSF600837S, RSF600939S, RSF601071S, RSF601112S, RSF603606S, RSF607336S, RSF600434S, RSF601390S, RSF604269S, RSF600513S, RSF601332S, RSF603173S, RSF600536S, RSF600829S, RSF601138S, RSF602403S, RSF603129S, RSF605693S, RSF605921S, RSF605922S, RSF606208S, RSF606315S, RSF607364S, RSF603681S, RSF604469S, RSF604689S, RSF604792S, RSF605027S, RSF605215S, RSF605508S, RSF605566S, RSF602923S, RSF605683S, RSF602519S, RSF603728S, RSF604265S, RSF602537S, RSF600259S, RSF600369S, RSF603779S, RSF605961S, RSF605992S, RSF600164S, RSF606237S, RSF602975S, RSF603825S, RSF603828S, RSF604138S, RSF602695S, RSF600248S, RSF600766S, RSF600172S, RSF600173S, RSF601312S, RSF603552S, RSF607717S, RSF600182S, RSF604198S, RSF603573S, RSF603644S, RSF604024S, RSF604118S, RSF605002S, RSF600562S, RSF600816S, RSF605859S, RSF606269S, RSF606331S, RSF606516S, RSF600163S, RSF600888S, RSF602430S, RSF602830S, RSF603126S, RSF603515S, RSF603547S, RSF603558S, RSF603577S, RSF603766S, RSF603881S, RSF603887S, RSF603954S, RSF604685S, RSF604696S, RSF604763S, RSF605010S, RSF605671S, RSF607045S, RSF607279S, RSF607895S, RSF602163S, RSF602793S, RSF603820S, RSF604252S, RSF604281S, RSF608055S, RSF601308S, RSF604249S, RSF600439S, RSF600714S, RSF600715S, RSF600734S, RSF600778S, RSF600856S, RSF601342S, RSF602722S, RSF603023S, RSF603581S, RSF603871S, RSF604096S, RSF604162S, RSF603588S, RSF604839S, RSF604998S, RSF607939S, RSF603364S, RSF603365S, RSF603974S, RSF604847S, RSF604901S, RSF607201S, RSF607851S, RSF608040S, RSF600885S, RSF600907S, RSF601576S, RSF602730S, RSF603234S, RSF603555S, RSF604185S, RSF604967S, RSF605239S, RSF603628S, RSF603629S, RSF600425S, RSF605647S, RSF605648S, RSF605737S, RSF605756S, RSF605913S, RSF605914S, RSF605917S, RSF606054S, RSF606055S, RSF606058S, RSF606178S, RSF606179S, RSF606204S, RSF606205S, RSF606247S, RSF606248S, RSF606249S, RSF606250S, RSF606319S, RSF606481S, RSF606485S, RSF606591S, RSF606878S, RSF606879S, RSF606883S, RSF607026S, RSF607748S, RSF607749S, RSF607752S, RSF607754S, RSF607755S, RSF607756S, RSF607760S, RSF607932S, RSF607933S, RSF608047S, RSF600330S, RSF600363S, RSF600423S, RSF600461S, RSF600491S, RSF600520S, RSF600522S, RSF600726S, RSF600727S, RSF600730S, RSF600746S, RSF600768S, RSF600798S, RSF600857S, RSF601221S, RSF601284S, RSF602513S, RSF602538S, RSF602542S, RSF603031S, RSF603144S, RSF603351S, RSF603353S, RSF603381S, RSF603495S, RSF603571S, RSF603608S, RSF603616S, RSF603967S, RSF604069S, RSF604166S, RSF604247S, RSF604318S, RSF604443S, RSF604475S, RSF604646S, RSF604672S, RSF604690S, RSF604790S, RSF604960S, RSF604963S, RSF605000S, RSF605192S, RSF605193S, RSF605219S, RSF605487S, RSF605490S, RSF605504S, RSF605506S, RSF605538S, RSF605542S, RSF605551S, RSF605553S, RSF605554S, RSF608000S, RSF607638S, RSF600300S, RSF600929S, RSF603427S, RSF603460S, RSF607893S, RSF606212S, RSF607042S, RSF607043S, RSF607297S, RSF607394S, RSF607703S, RSF600197S, RSF600912S, RSF600945S, RSF602643S, RSF602677S, RSF602851S, RSF602856S, RSF603235S, RSF603391S, RSF603560S, RSF604008S, RSF604104S, RSF604212S, RSF606185S, RSF606186S, RSF606243S, RSF606524S, RSF607202S, RSF600629S, RSF600646S, RSF601095S, RSF601111S, RSF601454S, RSF602890S, RSF603185S, RSF603583S, RSF603992S, RSF605227S, RSF605503S, RSF607728S, RSF600184S, RSF600735S, RSF601124S, RSF607937S, RSF607938S, RSF607936S, RSF605960S, RSF606253S, RSF606512S, RSF606522S, RSF607078S, RSF607228S, RSF607344S, RSF600194S, RSF600871S, RSF601458S, RSF601536S, RSF602440S, RSF602688S, RSF602824S, RSF602864S, RSF603145S, RSF603434S, RSF603907S, RSF604490S, RSF604667S, RSF600609S, RSF600614S, RSF603481S, RSF600744S, RSF604257S, RSF606254S, RSF606280S, RSF606281S, RSF606958S, RSF607230S, RSF607391S, RSF601005S, RSF602636S, RSF602686S, RSF602964S, RSF602966S, RSF602971S, RSF602978S, RSF603882S, RSF603884S, RSF603894S, RSF604199S, RSF604744S, RSF607871S, RSF601134S, RSF601090S, RSF603952S, RSF605229S, RSF603704S, RSF607874S, RSF608011S, RSF605589S, RSF605672S, RSF605673S, RSF605871S, RSF605872S, RSF606213S, RSF606215S, RSF606246S, RSF606495S, RSF606944S, RSF607060S, RSF607363S, RSF600921S, RSF601016S, RSF601051S, RSF601123S, RSF602554S, RSF602582S, RSF602614S, RSF602639S, RSF603119S, RSF603453S, RSF603487S, RSF603715S, RSF603862S, RSF604011S, RSF604253S, RSF604259S, RSF604299S, RSF604419S, RSF604833S, RSF605198S, RSF605407S, RSF605532S, RSF607061S, RSF605682S, RSF600265S, RSF600405S, RSF602556S, RSF602576S, RSF602811S, RSF603380S, RSF603393S, RSF603419S, RSF603421S, RSF603508S, RSF603621S, RSF603739S, RSF603877S, RSF603906S, RSF604276S, RSF604414S, RSF604693S, RSF605039S, RSF605235S, RSF605493S, RSF600143S, RSF600644S, RSF601220S, RSF601565S, RSF600202S, RSF600381S, RSF600480S, RSF607249S, RSF607379S, RSF607894S, RSF608090S, RSF603127S, RSF604678S, RSF603636S, RSF603997S, RSF605942S, RSF600264S, RSF601335S, RSF603442S, RSF603843S, RSF603617S, RSF604682S, RSF602900S, RSF604864S, RSF606483S, RSF607079S, RSF607751S, RSF600678S, RSF604452S, RSF606954S, RSF602566S, RSF606274S, RSF600624S, RSF602625S, RSF600280S, RSF600913S, RSF601099S, RSF602682S, RSF603688S, RSF602371S, RSF602800S, RSF603385S, RSF603387S, RSF603590S, RSF603679S, RSF603860S, RSF604828S, RSF604972S, RSF605001S, RSF605026S, RSF605218S, RSF605220S, RSF606170S, RSF601435S, RSF607881S, RSF600859S, RSF606222S, RSF606956S, RSF600595S, RSF602724S, RSF602397S, RSF605675S, RSF603141S, RSF606492S, RSF604189S, RSF604881S, RSF605197S, RSF607303S, RSF600288S, RSF600764S, RSF601037S, RSF603707S, RSF603725S, RSF605919S, RSF607788S, RSF607789S, RSF607790S, RSF600601S, RSF600604S, RSF600606S, RSF600851S, RSF601216S, RSF604144S, RSF604145S, RSF605009S, RSF605234S, RSF605409S, RSF605552S, RSF605986S, RSF604458S, RSF604068S, RSF604076S, RSF604077S, RSF604078S, RSF605106S, RSF605107S, RSF600799S, RSF601381S, RSF602567S, RSF603039S, RSF603040S, RSF603390S, RSF603394S, RSF603431S, RSF603610S, RSF603872S, RSF603874S, RSF604126S, RSF604495S, RSF604778S, RSF604965S, RSF605216S, RSF605486S, RSF605753S, RSF605854S, RSF608026S, RSF608035S, RSF608056S, RSF605709S, RSF605963S, RSF606171S, RSF606174S, RSF606176S, RSF606255S, RSF607075S, RSF600361S, RSF600496S, RSF600950S, RSF603401S, RSF603446S, RSF603753S, RSF604097S, RSF604161S, RSF604286S, RSF604836S, RSF608092S, RSF605949S, RSF605954S, RSF606182S, RSF606183S, RSF606242S, RSF608002S, RSF600660S, RSF601243S, RSF601252S, RSF601452S, RSF601562S, RSF601583S, RSF602187S, RSF602399S, RSF602828S, RSF602832S, RSF602840S, RSF602879S, RSF605969S, RSF600738S, RSF608012S, RSF607245S, RSF607246S, RSF607021S, RSF607918S, RSF600487S, RSF606520S, RSF600448S, RSF605703S, RSF602659S, RSF600464S, RSF602525S, RSF602747S, RSF602803S, RSF600497S, RSF600796S, RSF601294S, RSF602457S, RSF602697S, RSF603594S, RSF604137S, RSF605051S, RSF601315S, RSF601316S, RSF601341S, RSF603029S, RSF603056S, RSF604707S, RSF602352S, RSF602853S, RSF603791S, RSF600997S, RSF600175S, RSF600178S, RSF600200S, RSF600388S, RSF601157S, RSF601159S, RSF601160S, RSF601161S, RSF601476S, RSF602756S, RSF603414S, RSF603706S, RSF604184S, RSF601441S, RSF601457S, RSF601363S, RSF601364S, RSF601368S, RSF600218S, RSF601126S, RSF603998S, RSF604445S, RSF607396S, RSF602875S, RSF605952S, RSF600308S, RSF600495S, RSF600553S, RSF600578S, RSF600579S, RSF600881S, RSF600916S, RSF600995S, RSF601219S, RSF601273S, RSF602594S, RSF602760S, RSF603896S, RSF604275S, RSF604454S, RSF601351S, RSF601367S, RSF600187S, RSF605147S, RSF600904S, RSF604768S, RSF605500S, RSF607413S, RSF603956S, RSF607090S, RSF607221S, RSF607407S, RSF607899S, RSF605545S, RSF608005S, RSF605642S, RSF606172S, RSF600642S, RSF601399S, RSF603044S, RSF603474S, RSF603578S, RSF605055S, RSF606511S, RSF601257S, RSF602791S, RSF605971S, RSF607070S, RSF607285S, RSF607358S, RSF600485S, RSF600784S, RSF601020S, RSF601163S, RSF601361S, RSF601460S, RSF601472S, RSF602269S, RSF602801S, RSF602929S, RSF603635S, RSF603953S, RSF607354S, RSF604641S, RSF605951S, RSF601418S, RSF603462S, RSF601195S, RSF604256S, RSF602521S, RSF600479S, RSF600551S, RSF601180S, RSF601199S, RSF601590S, RSF602189S, RSF602198S, RSF602199S, RSF603118S, RSF603554S, RSF603582S, RSF603671S, RSF604072S, RSF604123S, RSF604242S, RSF604014S, RSF603579S, RSF600893S, RSF603117S, RSF604066S, RSF600239S, RSF600541S, RSF603010S, RSF603095S, RSF603110S, RSF604322S, RSF604438S, RSF606332S, RSF600486S, RSF600519S, RSF601015S, RSF602959S, RSF603087S, RSF603959S, RSF605694S, RSF606238S, RSF601072S, RSF603370S, RSF605484S, RSF605238S, RSF603140S, RSF601197S, RSF601383S, RSF603094S, RSF603231S, RSF603233S, RSF603498S, RSF603667S, RSF607410S, RSF607411S, RSF602982S, RSF603250S, RSF603251S, RSF605686S, RSF605989S, RSF607866S, RSF600249S, RSF600285S, RSF601032S, RSF601034S, RSF601040S, RSF601041S, RSF601042S, RSF601046S, RSF601317S, RSF601412S, RSF602406S, RSF602633S, RSF602673S, RSF602675S, RSF602754S, RSF602814S, RSF602871S, RSF603121S, RSF603123S, RSF603124S, RSF603125S, RSF603128S, RSF603168S, RSF603220S, RSF603721S, RSF603795S, RSF603840S, RSF604459S, RSF600168S, RSF601385S, RSF600266S, RSF600718S, RSF600882S, RSF600803S, RSF600245S, RSF600326S, RSF600416S, RSF600584S, RSF600634S, RSF600978S, RSF601233S, RSF602487S, RSF602545S, RSF602768S, RSF603028S, RSF603531S, RSF600268S, RSF602821S, RSF604017S, RSF602726S, RSF602805S, RSF603673S, RSF603698S, RSF605272S, RSF605627S, RSF605653S, RSF606341S, RSF600316S, RSF600378S, RSF600415S, RSF600447S, RSF600556S, RSF600580S, RSF600582S, RSF600750S, RSF601304S, RSF601440S, RSF602766S, RSF602798S, RSF602907S, RSF603485S, RSF603524S, RSF603674S, RSF603768S, RSF604421S, RSF604764S, RSF605043S, RSF605491S, RSF600366S, RSF601023S, RSF601271S, RSF601371S, RSF602942S, RSF604132S, RSF604183S, RSF607290S, RSF601237S, RSF601348S, RSF603350S, RSF603780S, RSF605369S, RSF605413S, RSF602772S, RSF600772S, RSF602520S, RSF602640S, RSF602867S, RSF602919S, RSF603597S, RSF603888S, RSF603904S, RSF603989S, RSF603483S, RSF600533S, RSF604709S, RSF604713S, RSF604714S, RSF607746S, RSF600465S, RSF606890S, RSF606995S, RSF607203S, RSF600186S, RSF600199S, RSF600201S, RSF600230S, RSF600287S, RSF600293S, RSF600468S, RSF600469S, RSF601156S, RSF601578S, RSF602986S, RSF603021S, RSF603361S, RSF605217S, RSF600231S, RSF600286S, RSF600294S, RSF600317S, RSF600470S, RSF600810S, RSF600981S, RSF601004S, RSF601302S, RSF601485S, RSF602325S, RSF603459S, RSF606257S, RSF607054S, RSF600449S, RSF601248S, RSF602536S, RSF602539S, RSF602563S, RSF602642S, RSF602655S, RSF603420S, RSF603477S, RSF603503S, RSF603624S, RSF604241S, RSF604428S, RSF604706S, RSF604774S, RSF604775S, RSF604843S, RSF607210S, RSF600923S, RSF600930S, RSF601114S, RSF601115S, RSF601276S, RSF602365S, RSF602825S, RSF603410S, RSF603422S, RSF603786S, RSF603787S, RSF603790S, RSF603796S, RSF604650S, RSF604835S, RSF605029S, RSF605053S, RSF605226S, RSF607272S, RSF605883S, RSF600278S, RSF600301S, RSF600349S, RSF600362S, RSF600625S, RSF600792S, RSF600834S, RSF601010S, RSF601044S, RSF601130S, RSF601378S, RSF601455S, RSF601466S, RSF602161S, RSF602581S, RSF602583S, RSF602762S, RSF602878S, RSF603355S, RSF603836S, RSF604270S, RSF604695S, RSF605569S, RSF608085S, RSF605575S, RSF605880S, RSF606184S, RSF606926S, RSF600196S, RSF600277S, RSF601300S, RSF602548S, RSF602679S, RSF602731S, RSF602787S, RSF602866S, RSF603194S, RSF603486S, RSF603642S, RSF603797S, RSF603822S, RSF603845S, RSF603899S, RSF604074S, RSF604127S, RSF604141S, RSF604501S, RSF604769S, RSF605195S, RSF603085S, RSF603086S, RSF605033S, RSF600305S, RSF600542S, RSF600583S, RSF600586S, RSF600620S, RSF600690S, RSF601026S, RSF601218S, RSF601225S, RSF600509S, RSF600547S, RSF600565S, RSF600758S, RSF600760S, RSF600765S, RSF600776S, RSF600877S, RSF600879S, RSF600918S, RSF605646S, RSF605892S, RSF605906S, RSF605907S, RSF606190S, RSF606219S, RSF606240S, RSF606539S, RSF606988S, RSF607310S, RSF607339S, RSF607340S, RSF607355S, RSF607408S, RSF607409S, RSF608008S, RSF600236S, RSF600983S, RSF601293S, RSF601463S, RSF602624S, RSF602649S, RSF602650S, RSF602656S, RSF602687S, RSF602850S, RSF602972S, RSF602973S, RSF603025S, RSF603245S, RSF603247S, RSF603666S, RSF603730S, RSF604021S, RSF604474S, RSF604493S, RSF604644S, RSF604773S, RSF604941S, RSF604974S, RSF605222S, RSF605439S, RSF605441S, RSF605494S, RSF605498S, RSF605557S, RSF604626S, RSF604715S, RSF604716S, RSF605095S, RSF605099S, RSF605105S, RSF605108S, RSF605109S, RSF605110S, RSF605111S, RSF607115S, RSF600987S, RSF601043S, RSF601045S, RSF601047S, RSF601050S, RSF601052S, RSF601054S, RSF601055S, RSF601056S, RSF601057S, RSF601059S, RSF601066S, RSF601067S, RSF601068S, RSF601069S, RSF601074S, RSF602941S, RSF602953S, RSF602954S, RSF602957S, RSF602960S, RSF602962S, RSF604001S, RSF604004S, RSF604007S, RSF606321S, RSF608096S, RSF600574S, RSF600700S, RSF603722S, RSF606064S, RSF606065S, RSF603936S, RSF603975S, RSF605953S, RSF605957S, RSF606999S, RSF600693S, RSF600823S, RSF601389S, RSF601552S, RSF602488S, RSF602788S, RSF600226S, RSF601018S, RSF601110S, RSF601270S, RSF606517S, RSF600244S, RSF604067S, RSF607288S, RSF607346S, RSF601210S, RSF601292S, RSF601478S, RSF602704S, RSF603830S, RSF604226S, RSF605058S, RSF603038S, RSF606490S, RSF607047S, RSF607223S, RSF607269S, RSF607295S, RSF607381S, RSF603798S, RSF602740S, RSF603433S, RSF603609S, RSF604171S, RSF603059S, RSF604057S, RSF605668S, RSF607693S, RSF601236S, RSF601465S, RSF601223S, RSF603834S, RSF605983S, RSF608038S, RSF603944S, RSF604203S, RSF600654S, RSF601268S, RSF604680S, RSF600691S, RSF603514S, RSF605887S, RSF606241S, RSF606991S, RSF607241S, RSF607273S, RSF607305S, RSF607398S, RSF607999S, RSF600952S, RSF600953S, RSF602729S, RSF602742S, RSF603133S, RSF603395S, RSF605236S, RSF605438S, RSF608126S, RSF605704S, RSF605965S, RSF606232S, RSF606537S, RSF602503S, RSF602735S, RSF604094S, RSF604288S, RSF604964S, RSF602265S, RSF600167S, RSF600174S, RSF600242S, RSF600292S, RSF600587S, RSF600592S, RSF600613S, RSF603207S, RSF603478S, RSF603564S, RSF602955S, RSF606923S, RSF602569S, RSF603930S, RSF604009S, RSF604061S, RSF606008S, RSF606480S, RSF606881S, RSF607036S, RSF600529S, RSF600618S, RSF602769S, RSF603358S, RSF603922S, RSF604131S, RSF604449S, RSF604651S, RSF604652S, RSF605541S, RSF605689S, RSF600575S, RSF603254S, RSF603241S, RSF603670S, RSF604085S, RSF604752S, RSF600499S, RSF601084S, RSF601325S, RSF604155S, RSF604274S, RSF603096S, RSF603101S, RSF603366S, RSF603928S, RSF604673S, RSF604679S, RSF600379S, RSF606268S, RSF600181S, RSF603530S, RSF602492S, RSF602661S, RSF603669S, RSF600193S, RSF602846S, RSF605346S, RSF601329S, RSF602550S, RSF603640S, RSF606180S, RSF606206S, RSF606880S, RSF607031S, RSF607250S, RSF600526S, RSF600673S, RSF600970S, RSF602568S, RSF602606S, RSF602622S, RSF603584S, RSF603770S, RSF603900S, RSF603979S, RSF604046S, RSF604676S, RSF604686S, RSF604741S, RSF605023S, RSF607929S, RSF604113S, RSF603819S, RSF601481S, RSF602765S, RSF605665S, RSF605228S, RSF600372S, RSF602167S, RSF602718S, RSF602737S, RSF602823S, RSF606966S, RSF601362S, RSF602517S, RSF604026S, RSF601298S, RSF606887S, RSF600752S, RSF604149S, RSF600647S, RSF605094S, RSF603451S, RSF603714S, RSF603438S, RSF603948S, RSF605044S, RSF605045S, RSF604063S, RSF603993S, RSF605667S, RSF605905S, RSF606187S, RSF606220S, RSF606258S, RSF606279S, RSF606326S, RSF606515S, RSF606989S, RSF607209S, RSF607248S, RSF607265S, RSF607293S, RSF607347S, RSF607369S, RSF607415S, RSF607416S, RSF608007S, RSF601201S, RSF601203S, RSF601204S, RSF601319S, RSF602377S, RSF602415S, RSF602467S, RSF602681S, RSF602736S, RSF602829S, RSF603001S, RSF603020S, RSF603115S, RSF603224S, RSF603502S, RSF603694S, RSF603857S, RSF603988S, RSF604075S, RSF604239S, RSF604384S, RSF604442S, RSF604944S, RSF605124S, RSF605194S, RSF605221S, RSF605237S, RSF605258S, RSF605546S, RSF605558S, RSF607353S, RSF600335S, RSF600336S, RSF600339S, RSF601340S, RSF602712S, RSF600304S, RSF600306S, RSF600307S, RSF600284S, RSF601006S, RSF603509S, RSF603732S, RSF606541S, RSF606538S, RSF606244S, RSF601297S, RSF608075S, RSF608111S, RSF600412S, RSF600418S, RSF600460S, RSF600462S, RSF601183S, RSF601347S, RSF601380S, RSF601470S, RSF601556S, RSF602669S, RSF602705S, RSF602796S, RSF605858S, RSF608025S, RSF608027S, RSF604982S, RSF601344S, RSF601451S, RSF605657S, RSF605868S, RSF605981S, RSF607341S, RSF600960S, RSF604193S, RSF604330S, RSF605004S, RSF606553S, RSF604140S, RSF602558S, RSF602561S, RSF600367S, RSF602424S, RSF605685S, RSF600188S, RSF600189S, RSF600276S, RSF600291S, RSF600482S, RSF600506S, RSF600514S, RSF600763S, RSF600944S, RSF601024S, RSF601146S, RSF601431S, RSF602410S, RSF602549S, RSF602693S, RSF602758S, RSF603137S, RSF603138S, RSF603143S, RSF603466S, RSF604264S, RS
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1821-2023
Related Recalls
ICD-DR DDBB2D1 EVERA XT IS1/DF1 INTL, Model Number DDBB2D1; Implantable Cardioverter Defibrillators
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.
ICD-DR DDMC3D1 EVERA MRI S OUS/US DF1, Model Number DDMC3D1; Implantable Cardioverter Defibrillators
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.
ICD DDPA2D1G COBALT XT DR MRI DF1 GOLD, Model Number DDPA2D1G; Implantable Cardioverter Defibrillators
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.