ICD COBALT XT VR MRI DF4, Model Number DVPA2D4; Implantable Cardioverter Defibrillators
Class I - DangerousWhat Should You Do?
- Check if you have this product: GTIN 00763000178468. Lot Serial numbers: RSD605765S, RSD605767S, RSD607880S, RSD607881S, RSD607886S, RSD607890S, RSD603551S, RSD603552S, RSD603553S, RSD603554S, RSD603555S, RSD600536S, RSD601541S, RSD602283S, RSD602524S, RSD602542S, RSD602924S, RSD603303S, RSD603396S, RSD607136S, RSD608177S, RSD608547S, RSD606272S, RSD607000S, RSD607711S, RSD607712S, RSD602036S, RSD602645S, RSD602659S, RSD603096S, RSD603670S, RSD604815S, RSD608193S, RSD605918S, RSD605919S, RSD605920S, RSD605921S, RSD605922S, RSD605923S, RSD605924S, RSD605926S, RSD605927S, RSD605929S, RSD605930S, RSD605931S, RSD605932S, RSD605933S, RSD604238S, RSD604294S, RSD604312S, RSD604313S, RSD604314S, RSD604614S, RSD604616S, RSD604710S, RSD604721S, RSD604725S, RSD605925S, RSD607788S, RSD607789S, RSD607790S, RSD607791S, RSD607792S, RSD607793S, RSD607794S, RSD607795S, RSD607797S, RSD607800S, RSD607801S, RSD607802S, RSD609083S, RSD609084S, RSD609085S, RSD605068S, RSD608525S, RSD609065S, RSD609066S, RSD609067S, RSD609069S, RSD609071S, RSD609072S, RSD609073S, RSD609075S, RSD609079S, RSD609080S, RSD609081S, RSD609082S, RSD601603S, RSD601960S, RSD602778S, RSD603181S, RSD603321S, RSD603672S, RSD603879S, RSD606554S, RSD608293S, RSD600145S, RSD601926S, RSD602728S, RSD603152S, RSD604065S, RSD607875S, RSD603750S, RSD606677S, RSD600100S, RSD600200S, RSD601455S, RSD601551S, RSD601552S, RSD601750S, RSD602233S, RSD608150S, RSD609043S, RSD608404S, RSD608422S, RSD609843S, RSD609845S, RSD609846S, RSD603825S, RSD603826S, RSD609842S, RSD610592S, RSD610593S, RSD608739S, RSD608718S, RSD606638S, RSD606653S, RSD608253S, RSD601449S, RSD601450S, RSD603168S, RSD610535S, RSD600188S, RSD605289S, RSD605802S, RSD606161S, RSD608659S, RSD608720S, RSD603329S, RSD604030S, RSD605997S, RSD605998S, RSD608060S, RSD605348S, RSD605787S, RSD605790S, RSD606704S, RSD608634S, RSD608795S, RSD608855S, RSD609276S, RSD609277S, RSD609278S, RSD600209S, RSD600214S, RSD600366S, RSD600475S, RSD600484S, RSD600485S, RSD600499S, RSD600513S, RSD600525S, RSD600531S, RSD602048S, RSD602376S, RSD602378S, RSD602600S, RSD603398S, RSD604018S, RSD604559S, RSD605190S, RSD610520S, RSD610522S, RSD607657S, RSD600316S, RSD611020S, RSD603301S, RSD608946S, RSD610202S, RSD601558S, RSD601934S, RSD605328S, RSD605330S, RSD605333S, RSD605771S, RSD606050S, RSD606273S, RSD606413S, RSD606708S, RSD606948S, RSD606950S, RSD607114S, RSD607319S, RSD607703S, RSD608135S, RSD608453S, RSD608458S, RSD608560S, RSD608561S, RSD608563S, RSD609935S, RSD610123S, RSD600348S, RSD600349S, RSD600554S, RSD600609S, RSD602325S, RSD602582S, RSD602587S, RSD603101S, RSD603127S, RSD603297S, RSD603309S, RSD603363S, RSD603386S, RSD604519S, RSD605221S, RSD609936S, RSD605349S, RSD605799S, RSD600498S, RSD600501S, RSD600526S, RSD601840S, RSD608211S, RSD610933S, RSD605143S, RSD605747S, RSD605759S, RSD605762S, RSD605960S, RSD606078S, RSD606179S, RSD606836S, RSD606837S, RSD606838S, RSD607819S, RSD607823S, RSD608173S, RSD608239S, RSD608278S, RSD608555S, RSD608617S, RSD608618S, RSD608619S, RSD608622S, RSD608919S, RSD609186S, RSD610047S, RSD610081S, RSD610126S, RSD603694S, RSD603695S, RSD603696S, RSD603698S, RSD603706S, RSD603707S, RSD603840S, RSD603846S, RSD603884S, RSD604095S, RSD604096S, RSD604097S, RSD604220S, RSD604577S, RSD604767S, RSD605144S, RSD610136S, RSD610470S, RSD605466S, RSD605725S, RSD606088S, RSD606090S, RSD606732S, RSD606733S, RSD606734S, RSD603662S, RSD607692S, RSD608434S, RSD608630S, RSD608728S, RSD602666S, RSD603643S, RSD604822S, RSD607310S, RSD610154S, RSD602490S, RSD604734S, RSD604748S, RSD605724S, RSD607667S, RSD607668S, RSD607669S, RSD608120S, RSD608121S, RSD604717S, RSD607834S, RSD608947S, RSD610041S, RSD605304S, RSD602534S, RSD604804S, RSD605138S, RSD600615S, RSD600103S, RSD600106S, RSD600138S, RSD600150S, RSD602411S, RSD602402S, RSD602430S, RSD601435S, RSD602346S, RSD602407S, RSD602450S, RSD602684S, RSD603008S, RSD607163S, RSD610016S, RSD608939S, RSD604595S, RSD604597S, RSD604598S, RSD604601S, RSD604603S, RSD604605S, RSD604606S, RSD604607S, RSD604608S, RSD604610S, RSD604611S, RSD604612S, RSD608938S, RSD609017S, RSD609018S, RSD609019S, RSD609020S, RSD609021S, RSD609022S, RSD607651S, RSD607654S, RSD607658S, RSD607660S, RSD607662S, RSD600216S, RSD602212S, RSD602603S, RSD602741S, RSD602742S, RSD602743S, RSD602746S, RSD603040S, RSD603939S, RSD604478S, RSD610015S, RSD605331S, RSD605447S, RSD605464S, RSD606053S, RSD607493S, RSD608136S, RSD610193S, RSD605225S, RSD608809S, RSD608878S, RSD603094S, RSD606072S, RSD606735S, RSD608266S, RSD608267S, RSD608895S, RSD609991S, RSD604327S, RSD606969S, RSD607868S, RSD608624S, RSD606976S, RSD602299S, RSD603345S, RSD603611S, RSD605726S, RSD605727S, RSD606425S, RSD606531S, RSD606668S, RSD606669S, RSD607069S, RSD607070S, RSD607339S, RSD607341S, RSD607503S, RSD607596S, RSD607916S, RSD608341S, RSD608607S, RSD608608S, RSD608752S, RSD608753S, RSD609056S, RSD609057S, RSD610227S, RSD610233S, RSD610534S, RSD610766S, RSD610769S, RSD605353S, RSD605354S, RSD605896S, RSD605897S, RSD605913S, RSD605914S, RSD606280S, RSD606281S, RSD606754S, RSD607125S, RSD607582S, RSD608861S, RSD600193S, RSD600419S, RSD600441S, RSD600446S, RSD600579S, RSD601268S, RSD601392S, RSD601801S, RSD601842S, RSD601964S, RSD602058S, RSD602059S, RSD602204S, RSD602239S, RSD602384S, RSD602406S, RSD602526S, RSD602552S, RSD602780S, RSD602875S, RSD602877S, RSD603421S, RSD603956S, RSD603977S, RSD603979S, RSD604044S, RSD604049S, RSD604110S, RSD604530S, RSD604757S, RSD604806S, RSD604893S, RSD604961S, RSD605191S, RSD611147S, RSD611148S, RSD608157S, RSD603622S, RSD608807S, RSD608639S, RSD610025S, RSD610162S, RSD608872S, RSD602497S, RSD604475S, RSD604477S, RSD605139S, RSD601935S, RSD602389S, RSD605323S, RSD605803S, RSD600239S, RSD601555S, RSD602326S, RSD602328S, RSD602380S, RSD602393S, RSD602456S, RSD602469S, RSD602618S, RSD603120S, RSD605758S, RSD605761S, RSD608913S, RSD608914S, RSD608995S, RSD608999S, RSD609270S, RSD605142S, RSD608996S, RSD608997S, RSD608998S, RSD609000S, RSD601877S, RSD608645S, RSD610033S, RSD600319S, RSD601434S, RSD601878S, RSD602282S, RSD602731S, RSD608867S, RSD602485S, RSD602493S, RSD602630S, RSD602706S, RSD602773S, RSD602930S, RSD602931S, RSD603001S, RSD603004S, RSD603215S, RSD603248S, RSD603251S, RSD603366S, RSD603367S, RSD602350S, RSD602479S, RSD602705S, RSD608488S, RSD602500S, RSD607532S, RSD603731S, RSD603768S, RSD608096S, RSD608097S, RSD608098S, RSD608099S, RSD609811S, RSD609813S, RSD609814S, RSD609810S, RSD609812S, RSD609815S, RSD605355S, RSD605356S, RSD606265S, RSD606266S, RSD606696S, RSD606729S, RSD606939S, RSD607107S, RSD607230S, RSD607231S, RSD607544S, RSD608141S, RSD608197S, RSD608430S, RSD608769S, RSD609914S, RSD610118S, RSD610293S, RSD600223S, RSD600231S, RSD600233S, RSD600250S, RSD600266S, RSD600270S, RSD600326S, RSD600330S, RSD600364S, RSD600422S, RSD600565S, RSD601792S, RSD601936S, RSD601937S, RSD602145S, RSD602166S, RSD602213S, RSD602297S, RSD602354S, RSD602686S, RSD602900S, RSD602993S, RSD603021S, RSD603858S, RSD603866S, RSD603872S, RSD603933S, RSD604008S, RSD604013S, RSD604208S, RSD604214S, RSD604759S, RSD604760S, RSD605192S, RSD607305S, RSD600386S, RSD600607S, RSD602675S, RSD602992S, RSD603898S, RSD605915S, RSD605935S, RSD606256S, RSD606258S, RSD601882S, RSD601884S, RSD601885S, RSD601886S, RSD601888S, RSD604502S, RSD604503S, RSD605157S, RSD605158S, RSD605934S, RSD609016S, RSD605368S, RSD605369S, RSD606782S, RSD606803S, RSD606840S, RSD606841S, RSD607292S, RSD607891S, RSD608080S, RSD608081S, RSD608082S, RSD608219S, RSD608337S, RSD608570S, RSD608783S, RSD608924S, RSD609007S, RSD610166S, RSD610591S, RSD605365S, RSD605902S, RSD606974S, RSD607833S, RSD608176S, RSD608754S, RSD602015S, RSD608831S, RSD607380S, RSD606233S, RSD606247S, RSD609704S, RSD609705S, RSD602458S, RSD602509S, RSD602713S, RSD604029S, RSD604537S, RSD607708S, RSD601793S, RSD602521S, RSD603070S, RSD603735S, RSD608586S, RSD602540S, RSD605774S, RSD606194S, RSD606270S, RSD606701S, RSD606703S, RSD606997S, RSD607110S, RSD607487S, RSD607554S, RSD607556S, RSD607557S, RSD607694S, RSD608203S, RSD608559S, RSD608730S, RSD608773S, RSD609925S, RSD609930S, RSD610097S, RSD610098S, RSD610181S, RSD610300S, RSD600259S, RSD602585S, RSD602606S, RSD602620S, RSD603610S, RSD603646S, RSD610475S, RSD608880S, RSD609052S, RSD600401S, RSD602545S, RSD603326S, RSD603652S, RSD601410S, RSD601412S, RSD609103S, RSD609135S, RSD609138S, RSD609145S, RSD608129S, RSD608647S, RSD600199S, RSD601363S, RSD601372S, RSD601904S, RSD603745S, RSD605322S, RSD606414S, RSD606415S, RSD606688S, RSD606736S, RSD607173S, RSD608190S, RSD608268S, RSD608602S, RSD600311S, RSD600563S, RSD602336S, RSD602397S, RSD602621S, RSD602772S, RSD603074S, RSD603975S, RSD605218S, RSD610568S, RSD610569S, RSD600222S, RSD600581S, RSD602120S, RSD602870S, RSD603207S, RSD607237S, RSD602661S, RSD604820S, RSD609209S, RSD609210S, RSD609212S, RSD603016S, RSD603250S, RSD607321S, RSD608921S, RSD611122S, RSD608859S, RSD610763S, RSD600370S, RSD608772S, RSD607061S, RSD607223S, RSD607311S, RSD608155S, RSD608243S, RSD608262S, RSD602539S, RSD603741S, RSD603751S, RSD603844S, RSD608128S, RSD610119S, RSD600489S, RSD610515S, RSD607829S, RSD600371S, RSD600460S, RSD601354S, RSD601828S, RSD602462S, RSD602682S, RSD603196S, RSD604021S, RSD605989S, RSD605990S, RSD609635S, RSD609636S, RSD608204S, RSD603876S, RSD605351S, RSD606900S, RSD608759S, RSD610223S, RSD600210S, RSD601389S, RSD603005S, RSD603039S, RSD604534S, RSD604780S, RSD606238S, RSD606489S, RSD606490S, RSD606657S, RSD606983S, RSD607459S, RSD607460S, RSD607811S, RSD607812S, RSD607813S, RSD608362S, RSD608609S, RSD608610S, RSD608611S, RSD608613S, RSD608614S, RSD608750S, RSD608751S, RSD609046S, RSD609047S, RSD610130S, RSD610131S, RSD610134S, RSD604121S, RSD604122S, RSD604123S, RSD604124S, RSD604126S, RSD604128S, RSD604130S, RSD604131S, RSD604132S, RSD604133S, RSD604134S, RSD604135S, RSD604138S, RSD604139S, RSD604141S, RSD604142S, RSD604147S, RSD604682S, RSD604684S, RSD604690S, RSD604691S, RSD604692S, RSD604693S, RSD604694S, RSD604695S, RSD604697S, RSD604699S, RSD604700S, RSD604702S, RSD604703S, RSD604704S, RSD604705S, RSD604707S, RSD604873S, RSD604875S, RSD604876S, RSD604877S, RSD608612S, RSD608615S, RSD610082S, RSD610129S, RSD610132S, RSD610133S, RSD606294S, RSD606297S, RSD606427S, RSD606973S, RSD607131S, RSD607222S, RSD607225S, RSD607570S, RSD607571S, RSD608183S, RSD600185S, RSD600215S, RSD600217S, RSD600261S, RSD600299S, RSD600332S, RSD600333S, RSD600338S, RSD600357S, RSD600360S, RSD600383S, RSD600414S, RSD600416S, RSD600434S, RSD600436S, RSD600451S, RSD600552S, RSD600559S, RSD600597S, RSD600606S, RSD601366S, RSD601440S, RSD601559S, RSD601596S, RSD601819S, RSD601859S, RSD601871S, RSD601924S, RSD601925S, RSD602480S, RSD602679S, RSD602918S, RSD603033S, RSD603064S, RSD603153S, RSD603255S, RSD603547S, RSD603868S, RSD604533S, RSD609191S, RSD609194S, RSD609197S, RSD609201S, RSD609192S, RSD609195S, RSD609196S, RSD609198S, RSD609199S, RSD605826S, RSD605827S, RSD606225S, RSD606437S, RSD606438S, RSD606690S, RSD606901S, RSD607027S, RSD607251S, RSD607258S, RSD607346S, RSD607362S, RSD607364S, RSD607756S, RSD608134S, RSD608188S, RSD608189S, RSD608625S, RSD608626S, RSD608688S, RSD608786S, RSD608832S, RSD600532S, RSD601814S, RSD601830S, RSD601833S, RSD601836S, RSD601906S, RSD602123S, RSD602134S, RSD602361S, RSD602451S, RSD602678S, RSD602880S, RSD603230S, RSD603233S, RSD603235S, RSD603236S, RSD603277S, RSD603278S, RSD603299S, RSD603385S, RSD603518S, RSD603587S, RSD603717S, RSD604514S, RSD605069S, RSD605231S, RSD601391S, RSD602236S, RSD604771S, RSD609968S, RSD601569S, RSD606747S, RSD603034S, RSD605292S, RSD605392S, RSD605816S, RSD606063S, RSD606146S, RSD606201S, RSD606441S, RSD606484S, RSD606485S, RSD606743S, RSD606818S, RSD607017S, RSD607091S, RSD607094S, RSD607594S, RSD607602S, RSD608793S, RSD600124S, RSD600184S, RSD600189S, RSD600191S, RSD600194S, RSD600452S, RSD601332S, RSD601334S, RSD601365S, RSD602760S, RSD602761S, RSD603035S, RSD603044S, RSD603404S, RSD603405S, RSD603407S, RSD603914S, RSD603915S, RSD604892S, RSD605891S, RSD607159S, RSD610125S, RSD608071S, RSD608072S, RSD608073S, RSD608074S, RSD608075S, RSD607160S, RSD608516S, RSD608545S, RSD609014S, RSD609015S, RSD610128S, RSD610231S, RSD610854S, RSD600262S, RSD600505S, RSD601983S, RSD602719S, RSD603043S, RSD603346S, RSD603690S, RSD603954S, RSD603955S, RSD610153S, RSD606475S, RSD609149S, RSD601368S, RSD602917S, RSD601853S, RSD609053S, RSD610443S, RSD610474S, RSD610476S, RSD610477S, RSD610952S, RSD610949S, RSD605950S, RSD605281S, RSD606816S, RSD603129S, RSD601844S, RSD607702S, RSD600608S, RSD601426S, RSD602063S, RSD602096S, RSD602588S, RSD602629S, RSD603012S, RSD603037S, RSD603331S, RSD603548S, RSD604517S, RSD603392S, RSD605302S, RSD605905S, RSD606040S, RSD606296S, RSD606914S, RSD607129S, RSD607130S, RSD607158S, RSD607312S, RSD607830S, RSD600324S, RSD600398S, RSD601820S, RSD601841S, RSD602251S, RSD602475S, RSD602755S, RSD602800S, RSD603154S, RSD603182S, RSD603887S, RSD604047S, RSD604741S, RSD605313S, RSD605314S, RSD605315S, RSD605316S, RSD605317S, RSD605885S, RSD605886S, RSD605887S, RSD605888S, RSD605889S, RSD603828S, RSD603842S, RSD603889S, RSD603894S, RSD603895S, RSD603896S, RSD604258S, RSD604259S, RSD604300S, RSD604301S, RSD604303S, RSD604334S, RSD604509S, RSD604510S, RSD604511S, RSD604512S, RSD604513S, RSD604569S, RSD604584S, RSD604585S, RSD604586S, RSD606491S, RSD606493S, RSD606507S, RSD606508S, RSD605909S, RSD602486S, RSD603086S, RSD601568S, RSD605312S, RSD606822S, RSD607055S, RSD607135S, RSD607352S, RSD608182S, RSD608643S, RSD608760S, RSD610032S, RSD610194S, RSD600329S, RSD600379S, RSD602288S, RSD602418S, RSD602494S, RSD602564S, RSD602566S, RSD602613S, RSD602866S, RSD602980S, RSD603253S, RSD603867S, RSD603869S, RSD603870S, RSD604026S, RSD604532S, RSD604772S, RSD604520S, RSD606674S, RSD605340S, RSD606048S, RSD606190S, RSD606191S, RSD606220S, RSD606697S, RSD606938S, RSD607105S, RSD607314S, RSD607317S, RSD607542S, RSD607547S, RSD604575S, RSD604819S, RSD604859S, RSD609281S, RSD609282S, RSD609284S, RSD609279S, RSD609280S, RSD609283S, RSD600218S, RSD600246S, RSD600315S, RSD600365S, RSD602691S, RSD602711S, RSD603191S, RSD603633S, RSD603684S, RSD608144S, RSD606678S, RSD603088S, RSD608427S, RSD608724S, RSD608725S, RSD602011S, RSD602013S, RSD602135S, RSD602136S, RSD602137S, RSD605417S, RSD608564S, RSD609223S, RSD610094S, RSD610452S, RSD609894S, RSD606726S, RSD607156S, RSD607681S, RSD608231S, RSD610273S, RSD600447S, RSD600604S, RSD600605S, RSD601378S, RSD601379S, RSD601380S, RSD602258S, RSD602460S, RSD602730S, RSD610157S, RSD605382S, RSD606057S, RSD606738S, RSD606758S, RSD606994S, RSD607111S, RSD607263S, RSD608275S, RSD601405S, RSD602285S, RSD603418S, RSD603679S, RSD606177S, RSD606615S, RSD610172S, RSD601852S, RSD602427S, RSD603186S, RSD603382S, RSD603637S, RSD603926S, RSD606418S, RSD606448S, RSD606477S, RSD606641S, RSD610524S, RSD608584S, RSD608963S, RSD610190S, RSD600284S, RSD600472S, RSD600582S, RSD600596S, RSD603065S, RSD603066S, RSD603730S, RSD610564S, RSD605805S, RSD602031S, RSD603383S, RSD603674S, RSD605115S, RSD606783S, RSD606784S, RSD606785S, RSD606786S, RSD606790S, RSD606793S, RSD606797S, RSD606798S, RSD606801S, RSD606805S, RSD606806S, RSD606807S, RSD606808S, RSD606833S, RSD606834S, RSD606843S, RSD608748S, RSD608749S, RSD608784S, RSD608798S, RSD609224S, RSD606788S, RSD606835S, RSD608714S, RSD609225S, RSD610870S, RSD610871S, RSD606021S, RSD606022S, RSD606023S, RSD606024S, RSD606026S, RSD606028S, RSD606032S, RSD606033S, RSD606035S, RSD609207S, RSD606479S, RSD608234S, RSD608779S, RSD608808S, RSD603080S, RSD606207S, RSD606766S, RSD608159S, RSD601800S, RSD602491S, RSD603172S, RSD603184S, RSD605815S, RSD608695S, RSD604560S, RSD604884S, RSD606084S, RSD606761S, RSD607599S, RSD605291S, RSD605809S, RSD606163S, RSD607770S, RSD608979S, RSD600267S, RSD601456S, RSD601457S, RSD602067S, RSD607761S, RSD610935S, RSD611144S, RSD607810S, RSD607837S, RSD607838S, RSD607840S, RSD608257S, RSD610055S, RSD610057S, RSD608869S, RSD601415S, RSD601421S, RSD601550S, RSD602287S, RSD603614S, RSD604019S, RSD606221S, RSD607723S, RSD608685S, RSD606913S, RSD607458S, RSD607572S, RSD600394S, RSD600448S, RSD601923S, RSD604761S, RSD606086S, RSD606626S, RSD609042S, RSD603656S, RSD602391S, RSD605811S, RSD605825S, RSD608192S, RSD608702S, RSD608788S, RSD605819S, RSD606202S, RSD607092S, RSD608576S, RSD601897S, RSD606410S, RSD607300S, RSD608269S, RSD610214S, RSD610243S, RSD601916S, RSD610551S, RSD605391S, RSD606442S, RSD606705S, RSD607766S, RSD607769S, RSD603389S, RSD604508S, RSD604959S, RSD605109S, RSD605189S, RSD600511S, RSD602665S, RSD610346S, RSD608172S, RSD607533S, RSD601542S, RSD602499S, RSD602570S, RSD602608S, RSD603015S, RSD603208S, RSD604230S, RSD604239S, RSD604861S, RSD604862S, RSD604864S, RSD606894S, RSD607995S, RSD603762S, RSD610961S, RSD610962S, RSD610964S, RSD603330S, RSD604090S, RSD610790S, RSD600478S, RSD600481S, RSD602699S, RSD605252S, RSD605743S, RSD605876S, RSD605877S, RSD606158S, RSD606290S, RSD608652S, RSD608936S, RSD600555S, RSD602319S, RSD602913S, RSD602984S, RSD603173S, RSD603736S, RSD604060S, RSD604062S, RSD604544S, RSD604856S, RSD611046S, RSD607161S, RSD607581S, RSD600507S, RSD603274S, RSD610211S, RSD602248S, RSD604046S, RSD605949S, RSD607782S, RSD610770S, RSD600387S, RSD600483S, RSD602447S, RSD602517S, RSD603399S, RSD607360S, RSD608207S, RSD608233S, RSD608600S, RSD608692S, RSD609901S, RSD610159S, RSD607133S, RSD601575S, RSD601586S, RSD602271S, RSD602925S, RSD603773S, RSD602215S, RSD603589S, RSD610628S, RSD603743S, RSD604196S, RSD603263S, RSD606424S, RSD606896S, RSD607809S, RSD602206S, RSD602217S, RSD602694S, RSD602937S, RSD603861S, RSD608948S, RSD604740S, RSD606143S, RSD606226S, RSD607592S, RSD602432S, RSD608866S, RSD610222S, RSD601413S, RSD601873S, RSD602092S, RSD602932S, RSD603549S, RSD603964S, RSD610853S, RSD606222S, RSD606634S, RSD603974S, RSD607099S, RSD610782S, RSD605244S, RSD605344S, RSD605352S, RSD605367S, RSD605390S, RSD605476S, RSD605754S, RSD605800S, RSD606097S, RSD606156S, RSD606160S, RSD606200S, RSD606443S, RSD606482S, RSD606640S, RSD606642S, RSD606762S, RSD607327S, RSD607330S, RSD607497S, RSD607498S, RSD607583S, RSD607595S, RSD607771S, RSD607779S, RSD608163S, RSD608305S, RSD608494S, RSD608499S, RSD608721S, RSD608746S, RSD608747S, RSD610528S, RSD610786S, RSD600092S, RSD600256S, RSD600275S, RSD600345S, RSD600482S, RSD600553S, RSD600585S, RSD602108S, RSD602320S, RSD602518S, RSD602595S, RSD602642S, RSD602717S, RSD602720S, RSD602737S, RSD603031S, RSD603324S, RSD603361S, RSD603377S, RSD603419S, RSD603638S, RSD603647S, RSD603710S, RSD603969S, RSD604528S, RSD604558S, RSD604578S, RSD604891S, RSD605114S, RSD602421S, RSD605385S, RSD606694S, RSD607296S, RSD607566S, RSD608640S, RSD610027S, RSD610497S, RSD610498S, RSD602457S, RSD602747S, RSD607524S, RSD606142S, RSD604943S, RSD601839S, RSD603234S, RSD610106S, RSD606778S, RSD607098S, RSD607283S, RSD607326S, RSD607584S, RSD607764S, RSD608214S, RSD608803S, RSD608974S, RSD608975S, RSD608976S, RSD610350S, RSD605780S, RSD600305S, RSD601958S, RSD602465S, RSD602886S, RSD602919S, RSD602942S, RSD603201S, RSD605016S, RSD600257S, RSD607104S, RSD607106S, RSD602461S, RSD602714S, RSD605741S, RSD605742S, RSD605744S, RSD607935S, RSD607936S, RSD607937S, RSD607939S, RSD604223S, RSD604224S, RSD604226S, RSD604869S, RSD604871S, RSD604872S, RSD600449S, RSD600454S, RSD600584S, RSD601447S, RSD601452S, RSD601556S, RSD601560S, RSD602084S, RSD602345S, RSD603661S, RSD610029S, RSD605939S, RSD607284S, RSD606199S, RSD606772S, RSD607097S, RSD607100S, RSD607598S, RSD607601S, RSD607762S, RSD607772S, RSD607773S, RSD608654S, RSD608658S, RSD608717S, RSD608719S, RSD608927S, RSD608928S, RSD608656S, RSD610526S, RSD610639S, RSD610747S, RSD608687S, RSD609895S, RSD608493S, RSD600437S, RSD600456S, RSD603219S, RSD603562S, RSD606157S, RSD606763S, RSD607279S, RSD607328S, RSD608079S, RSD608313S, RSD608585S, RSD608828S, RSD610114S, RSD610221S, RSD600307S, RSD603256S, RSD603601S, RSD604043S, RSD604587S, RSD611062S, RSD605775S, RSD606449S, RSD600263S, RSD601386S, RSD602303S, RSD602314S, RSD602626S, RSD602704S, RSD603855S, RSD604465S, RSD605159S, RSD603820S, RSD604556S, RSD604857S, RSD606504S, RSD607568S, RSD607715S, RSD608621S, RSD600292S, RSD601594S, RSD601893S, RSD602175S, RSD603414S, RSD603099S, RSD603292S, RSD603294S, RSD603355S, RSD605778S, RSD606094S, RSD606951S, RSD606999S, RSD607116S, RSD603880S, RSD603881S, RSD603936S, RSD608237S, RSD608681S, RSD607745S, RSD605361S, RSD605770S, RSD606274S, RSD602245S, RSD606683S, RSD609973S, RSD606295S, RSD606910S, RSD610539S, RSD606076S, RSD603246S, RSD608738S, RSD603957S, RSD604751S, RSD604753S, RSD607749S, RSD608704S, RSD609939S, RSD610103S, RSD610324S, RSD603026S, RSD603724S, RSD604811S, RSD604942S, RSD609213S, RSD609214S, RSD610137S, RSD607113S, RSD604033S, RSD605729S, RSD600286S, RSD600291S, RSD600298S, RSD600301S, RSD600393S, RSD600535S, RSD600598S, RSD602203S, RSD602434S, RSD602472S, RSD602610S, RSD602611S, RSD602700S, RSD602735S, RSD602756S, RSD602921S, RSD602933S, RSD603725S, RSD603848S, RSD603849S, RSD604068S, RSD604471S, RSD604522S, RSD604570S, RSD604889S, RSD606065S, RSD606219S, RSD606474S, RSD606942S, RSD607371S, RSD607685S, RSD609996S, RSD608577S, RSD610096S, RSD602176S, RSD602198S, RSD602658S, RSD603131S, RSD603149S, RSD603150S, RSD603612S, RSD604504S, RSD605387S, RSD606469S, RSD606728S, RSD608147S, RSD609227S, RSD602513S, RSD603176S, RSD604749S, RSD605951S, RSD606450S, RSD604874S, RSD606411S, RSD606451S, RSD606452S, RSD603774S, RSD603778S, RSD605419S, RSD605420S, RSD605422S, RSD605424S, RSD605425S, RSD605426S, RSD605441S, RSD605444S, RSD606245S, RSD606248S, RSD606791S, RSD607274S, RSD607291S, RSD607320S, RSD607519S, RSD608076S, RSD608077S, RSD608078S, RSD608595S, RSD608833S, RSD608954S, RSD604205S, RSD605141S, RSD608937S, RSD610224S, RSD610438S, RSD605377S, RSD605378S, RSD605903S, RSD606647S, RSD606842S, RSD608761S, RSD608822S, RSD608987S, RSD609883S, RSD610212S, RSD600213S, RSD602710S, RSD602908S, RSD603577S, RSD610638S, RSD606172S, RSD608215S, RSD608513S, RSD603617S, RSD609095S, RSD609096S, RSD609098S, RSD609099S, RSD609097S, RSD605409S, RSD606282S, RSD606613S, RSD606825S, RSD606922S, RSD606923S, RSD607123S, RSD607124S, RSD607127S, RSD607128S, RSD608100S, RSD608594S, RSD608881S, RSD600566S, RSD600572S, RSD602093S, RSD602097S, RSD602121S, RSD602197S, RSD602395S, RSD602614S, RSD602874S, RSD602889S, RSD603147S, RSD603381S, RSD603729S, RSD603968S, RSD604535S, RSD604776S, RSD605013S, RSD605102S, RSD605136S, RSD605195S, RSD603349S, RSD605363S, RSD605907S, RSD606920S, RSD608185S, RSD608868S, RSD610019S, RSD610576S, RSD601557S, RSD602723S, RSD602724S, RSD602725S, RSD603058S, RSD603334S, RSD603348S, RSD603949S, RSD604215S, RSD604536S, RSD604752S, RSD604789S, RSD610048S, RSD610836S, RSD610840S, RSD600377S, RSD601367S, RSD610548S, RSD603911S, RSD606926S, RSD607139S, RSD607140S, RSD607659S, RSD607784S, RSD607785S, RSD602797S, RSD607787S, RSD609023S, RSD609024S, RSD609025S, RSD609026S, RSD605910S, RSD605911S, RSD605912S, RSD602309S, RSD602975S, RSD602981S, RSD602982S, RSD602110S, RSD605880S, RSD606361S, RSD607901S, RSD607908S, RSD609236S, RSD609237S, RSD609238S, RSD605360S, RSD606755S, RSD606756S, RSD607120S, RSD602722S, RSD603701S, RSD600391S, RSD602511S, RSD608340S, RSD609721S, RSD604518S, RSD605036S, RSD606083S, RSD606259S, RSD606467S, RSD606614S, RSD607153S, RSD607670S, RSD610271S, RSD602182S, RSD602801S, RSD602804S, RSD602869S, RSD602907S, RSD603000S, RSD603145S, RSD603356S, RSD603641S, RSD603704S, RSD603728S, RSD603865S, RSD604066S, RSD604111S, RSD604113S, RSD604118S, RSD604885S, RSD608087S, RSD608646S, RSD608734S, RSD600198S, RSD601909S, RSD602481S, RSD610360S, RSD602423S, RSD607597S, RSD600558S, RSD602085S, RSD603970S, RSD606205S, RSD606481S, RSD608649S, RSD608844S, RSD608967S, RSD602453S, RSD603057S, RSD603103S, RSD603124S, RSD603283S, RSD604063S, RSD604064S, RSD606283S, RSD606918S, RSD610850S, RSD600203S, RSD600399S, RSD602094S, RSD603746S, RSD604779S, RSD605196S, RSD605394S, RSD606372S, RSD606749S, RSD607288S, RSD607525S, RSD608222S, RSD608589S, RSD608983S, RSD609882S, RSD610236S, RSD610453S, RSD608110S, RSD608112S, RSD608113S, RSD608114S, RSD608115S, RSD606183S, RSD606197S, RSD606496S, RSD606637S, RSD606902S, RSD607090S, RSD607591S, RSD607757S, RSD608834S, RSD610554S, RSD601987S, RSD602138S, RSD602195S, RSD602317S, RSD602318S, RSD602424S, RSD602425S, RSD602428S, RSD603030S, RSD603703S, RSD603719S, RSD603834S, RSD603857S, RSD604886S, RSD605219S, RSD610920S, RSD606223S, RSD607019S, RSD607359S, RSD608187S, RSD608274S, RSD609978S, RSD609829S, RSD609830S, RSD609831S, RSD604218S, RSD604225S, RSD606059S, RSD606173S, RSD606278S, RSD606727S, RSD606937S, RSD607299S, RSD607454S, RSD607736S, RSD608637S, RSD608929S, RSD608971S, RSD610018S, RSD610284S, RSD610295S, RSD610574S, RSD610666S, RSD600545S, RSD602781S, RSD603051S, RSD604525S, RSD604883S, RSD610665S, RSD610875S, RSD611026S, RSD606611S, RSD608737S, RSD608758S, RSD608863S, RSD600235S, RSD602355S, RSD605132S, RSD607064S, RSD607065S, RSD607066S, RSD607298S, RSD607395S, RSD607567S, RSD608578S, RSD608641S, RSD610026S, RSD610127S, RSD610307S, RSD610309S, RSD610509S, RSD610502S, RSD610513S, RSD610310S, RSD605783S, RSD606137S, RSD606725S, RSD607535S, RSD607678S, RSD608762S, RSD608764S, RSD610454S, RSD610455S, RSD601377S, RSD602281S, RSD602388S, RSD602410S, RSD603546S, RSD603875S, RSD604515S, RSD605179S, RSD606506S, RSD608549S, RSD608700S, RSD610461S, RSD610562S, RSD610884S, RSD607087S, RSD607754S, RSD607313S, RSD608744S, RSD608706S, RSD607345S, RSD607579S, RSD608597S, RSD601390S, RSD603962S, RSD606494S, RSD610939S, RSD608858S, RSD610523S, RSD610938S, RSD604775S, RSD607141S, RSD606151S, RSD610545S, RSD606056S, RSD606367S, RSD606741S, RSD600529S, RSD601931S, RSD602106S, RSD602905S, RSD603097S, RSD603619S, RSD603734S, RSD607289S, RSD605018S, RSD602733S, RSD602785S, RSD602149S, RSD604058S, RSD610043S, RSD610042S, RSD606144S, RSD606416S, RSD606689S, RSD608530S, RSD608596S, RSD600232S, RSD600344S, RSD601425S, RSD602533S, RSD603420S, RSD604016S, RSD605320S, RSD605375S, RSD605781S, RSD606263S, RSD600430S, RSD600539S, RSD601359S, RSD602091S, RSD602095S, RSD602692S, RSD604954S, RSD603100S, RSD603873S, RSD606810S, RSD606811S, RSD609100S, RSD609101S, RSD605480S, RSD606060S, RSD607539S, RSD600590S, RSD601863S, RSD601865S, RSD601891S, RSD602433S, RSD602512S, RSD602594S, RSD602752S, RSD602782S, RSD602788S, RSD603567S, RSD603618S, RSD603625S, RSD603629S, RSD603758S, RSD604860S, RSD605045S, RSD607108S, RSD607109S, RSD608771S, RSD610449S, RSD600586S, RSD600588S, RSD600589S, RSD602604S, RSD602687S, RSD602890S, RSD603045S, RSD603046S, RSD603265S, RSD603336S, RSD603342S, RSD603749S, RSD604202S, RSD611039S, RSD611056S, RSD604524S, RSD608528S, RSD603041S, RSD605199S, RSD606277S, RSD606630S, RSD607522S, RSD608224S, RSD606230S, RSD606433S, RSD606434S, RSD606681S, RSD607469S, RSD608132S, RSD608605S, RSD609694S, RSD609950S, RSD610092S, RSD610178S, RSD610232S, RSD602209S, RSD602634S, RSD603020S, RSD605358S, RSD608591S, RSD604802S, RSD606210S, RSD608092S, RSD610447S, RSD610876S, RSD600201S, RSD600435S, RSD601914S, RSD601985S, RSD603391S, RSD606176S, RSD607752S, RSD604927S, RSD605301S, RSD605410S, RSD606612S, RSD606967S, RSD607054S, RSD607309S, RSD608179S, RSD608272S, RSD603171S, RSD603360S, RSD603929S, RSD603948S, RSD604901S, RSD606011S, RSD605894S, RSD608873S, RSD610068S, RSD600224S, RSD602344S, RSD603068S, RSD605916S, RSD607747S, RSD607748S, RSD602220S, RSD601361S, RSD602471S, RSD606073S, RSD609956S, RSD606954S, RSD608935S, RSD603702S, RSD604554S, RSD604747S, RSD608705S, RSD601373S, RSD605167S, RSD606285S, RSD610610S, RSD610612S, RSD605942S, RSD605943S, RSD605944S, RSD605945S, RSD605946S, RSD605947S, RSD607623S, RSD607624S, RSD608167S, RSD608168S, RSD608169S, RSD608171S, RSD609045S, RSD610071S, RSD610072S, RSD610608S, RSD610613S, RSD605463S, RSD605794S, RSD606145S, RSD606181S, RSD607144S, RSD607342S, RSD607727S, RSD607734S, RSD608871S, RSD610030S, RSD610421S, RSD602504S, RSD603747S, RSD603748S, RSD604807S, RSD605177S, RSD607726S, RSD610422S, RSD610611S, RSD608223S, RSD600570S, RSD603335S, RSD610626S, RSD606930S, RSD606932S, RSD606933S, RSD606984S, RSD607729S, RSD607732S, RSD607820S, RSD607821S, RSD607822S, RSD606141S, RSD606262S, RSD606422S, RSD606423S, RSD606470S, RSD606676S, RSD607302S, RSD608552S, RSD608553S, RSD610087S, RSD610089S, RSD600443S, RSD601880S, RSD602146S, RSD602196S, RSD602262S, RSD602396S, RSD602972S, RSD602986S, RSD603077S, RSD603083S, RSD603658S, RSD603668S, RSD603671S, RSD603732S, RSD603937S, RSD604571S, RSD604956S, RSD605183S, RSD605234S, RSD610088S, RSD610090S, RSD610889S, RSD601336S, RSD605786S, RSD606070S, RSD606152S, RSD606631S, RSD606692S, RSD606693S, RSD606730S, RSD607067S, RSD607068S, RSD607145S, RSD607538S, RSD608505S, RSD608642S, RSD601591S, RSD602592S, RSD602624S, RSD602662S, RSD603055S, RSD603305S, RSD603306S, RSD603576S, RSD610872S, RSD611063S, RSD610623S, RSD607126S, RSD601564S, RSD601850S, RSD602617S, RSD605393S, RSD606648S, RSD603190S, RSD608918S, RSD604770S, RSD609055S, RSD602576S, RSD602579S, RSD603091S, RSD603093S, RSD604549S, RSD604552S, RSD605818S, RSD606731S, RSD606991S, RSD607562S, RSD610095S, RSD602051S, RSD602414S, RSD602416S, RSD602712S, RSD603287S, RSD603384S, RSD603395S, RSD603415S, RSD603913S, RSD605206S, RSD611149S, RSD607318S, RSD602404S, RSD601988S, RSD603054S, RSD606895S, RSD610363S, RSD610368S, RSD610386S, RSD610532S, RSD600346S, RSD602242S, RSD602941S, RSD603388S, RSD610530S, RSD610533S, RSD608648S, RSD600600S, RSD601576S, RSD602589S, RSD602935S, RSD607998S, RSD607999S, RSD608000S, RSD608001S, RSD608002S, RSD608003S, RSD608004S, RSD609233S, RSD609234S, RSD609235S, RSD602605S, RSD602669S, RSD602673S, RSD602676S, RSD602685S, RSD603678S, RSD603680S, RSD603683S, RSD603686S, RSD603687S, RSD603688S, RSD603689S, RSD603697S, RSD603699S, RSD603700S, RSD604227S, RSD604228S, RSD604235S, RSD604241S, RSD604244S, RSD604247S, RSD604251S, RSD609228S, RSD609229S, RSD609230S, RSD609231S, RSD609232S, RSD605764S, RSD605766S, RSD606629S, RSD607115S, RSD607264S, RSD607265S, RSD607268S, RSD607569S, RSD602256S, RSD602744S, RSD603535S, RSD607142S, RSD601956S, RSD602701S, RSD603243S, RSD603279S, RSD603522S, RSD604014S, RSD608804S, RSD606150S, RSD606180S, RSD604038S, RSD607350S, RSD607351S, RSD608904S, RSD608956S, RSD608957S, RSD602012S, RSD602358S, RSD602359S, RSD604297S, RSD604299S, RSD604719S, RSD604723S, RSD604726S, RSD605121S, RSD607349S, RSD608408S, RSD608581S, RSD608903S, RSD606644S, RSD606718S, RSD604059S, RSD600098S, RSD606706S, RSD602266S, RSD606287S, RSD605296S, RSD606169S, RSD606739S, RSD606799S, RSD607629S, RSD608796S, RSD610200S, RSD610343S, RSD600592S, RSD601335S, RSD602377S, RSD602381S, RSD602702S, RSD603558S, RSD603920S, RSD604958S, RSD610932S, RSD605793S, RSD605804S, RSD606149S, RSD606649S, RSD606804S, RSD606921S, RSD607286S, RSD608210S, RSD608333S, RSD608506S, RSD608636S, RSD608805S, RSD608949S, RSD600367S, RSD601563S, RSD602101S, RSD602394S, RSD603006S, RSD603315S, RSD603599S, RSD611019S, RSD610201S, RSD606074S, RSD608127S, RSD607778S, RSD610527S, RSD608777S, RSD608813S, RSD609089S, RSD607563S, RSD610863S, RSD608806S, RSD610014S, RSD606812S, RSD610519S, RSD607608S, RSD608277S, RSD608534S, RSD600450S, RSD601548S, RSD603756S, RSD605984S, RSD607621S, RSD607731S, RSD602583S, RSD611010S, RSD608933S, RSD608934S, RSD610174S, RSD610175S, RSD610315S, RSD608091S, RSD602155S, RSD603156S, RSD608265S, RSD610912S, RSD609027S, RSD607357S, RSD602904S, RSD604055S, RSD607393S, RSD610397S, RSD605188S, RSD608830S, RSD606665S, RSD607152S, RSD608689S, RSD608988S, RSD602887S, RSD606989S, RSD600186S, RSD600279S, RSD600280S, RSD601382S, RSD601384S, RSD601385S, RSD601387S, RSD601401S, RSD601566S, RSD601892S, RSD601921S, RSD601967S, RSD601968S, RSD601975S, RSD602227S, RSD602569S, RSD602695S, RSD603166S, RSD603249S, RSD603976S, RSD605892S, RSD605904S, RSD606286S, RSD606824S, RSD606831S, RSD606908S, RSD607118S, RSD607119S, RSD607574S, RSD607578S, RSD608153S, RSD608154S, RSD608156S, RSD608164S, RSD608165S, RSD608256S, RSD608258S, RSD608259S, RSD608527S, RSD608592S, RSD610031S, RSD610051S, RSD610052S, RSD610423S, RSD610424S, RSD610425S, RSD610062S, RSD610070S, RSD610802S, RSD610830S, RSD611017S, RSD609692S, RSD609693S, RSD600322S, RSD600350S, RSD600439S, RSD602315S, RSD603934S, RSD610913S, RSD606486S, RSD608247S, RSD608657S, RSD602052S, RSD603049S, RSD603623S, RSD603624S, RSD603841S, RSD605300S, RSD607575S, RSD607841S, RSD607921S, RSD608151S, RSD608546S, RSD610138S, RSD610140S, RSD603270S, RSD601847S, RSD602593S, RSD603347S, RSD603752S, RSD605740S, RSD605948S, RSD603897S, RSD604307S, RSD605288S, RSD605294S, RSD605295S, RSD605350S, RSD605380S, RSD605475S, RSD606148S, RSD606203S, RSD606980S, RSD607076S, RSD607275S, RSD608218S, RSD608635S, RSD608740S, RSD608945S, RS
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
- Reason for Recall:
- There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
ICD COBALT XT VR MRI DF4, Model Number DVPA2D4; Implantable Cardioverter Defibrillators
Product Codes/Lot Numbers:
GTIN 00763000178468. Lot Serial numbers: RSD605765S, RSD605767S, RSD607880S, RSD607881S, RSD607886S, RSD607890S, RSD603551S, RSD603552S, RSD603553S, RSD603554S, RSD603555S, RSD600536S, RSD601541S, RSD602283S, RSD602524S, RSD602542S, RSD602924S, RSD603303S, RSD603396S, RSD607136S, RSD608177S, RSD608547S, RSD606272S, RSD607000S, RSD607711S, RSD607712S, RSD602036S, RSD602645S, RSD602659S, RSD603096S, RSD603670S, RSD604815S, RSD608193S, RSD605918S, RSD605919S, RSD605920S, RSD605921S, RSD605922S, RSD605923S, RSD605924S, RSD605926S, RSD605927S, RSD605929S, RSD605930S, RSD605931S, RSD605932S, RSD605933S, RSD604238S, RSD604294S, RSD604312S, RSD604313S, RSD604314S, RSD604614S, RSD604616S, RSD604710S, RSD604721S, RSD604725S, RSD605925S, RSD607788S, RSD607789S, RSD607790S, RSD607791S, RSD607792S, RSD607793S, RSD607794S, RSD607795S, RSD607797S, RSD607800S, RSD607801S, RSD607802S, RSD609083S, RSD609084S, RSD609085S, RSD605068S, RSD608525S, RSD609065S, RSD609066S, RSD609067S, RSD609069S, RSD609071S, RSD609072S, RSD609073S, RSD609075S, RSD609079S, RSD609080S, RSD609081S, RSD609082S, RSD601603S, RSD601960S, RSD602778S, RSD603181S, RSD603321S, RSD603672S, RSD603879S, RSD606554S, RSD608293S, RSD600145S, RSD601926S, RSD602728S, RSD603152S, RSD604065S, RSD607875S, RSD603750S, RSD606677S, RSD600100S, RSD600200S, RSD601455S, RSD601551S, RSD601552S, RSD601750S, RSD602233S, RSD608150S, RSD609043S, RSD608404S, RSD608422S, RSD609843S, RSD609845S, RSD609846S, RSD603825S, RSD603826S, RSD609842S, RSD610592S, RSD610593S, RSD608739S, RSD608718S, RSD606638S, RSD606653S, RSD608253S, RSD601449S, RSD601450S, RSD603168S, RSD610535S, RSD600188S, RSD605289S, RSD605802S, RSD606161S, RSD608659S, RSD608720S, RSD603329S, RSD604030S, RSD605997S, RSD605998S, RSD608060S, RSD605348S, RSD605787S, RSD605790S, RSD606704S, RSD608634S, RSD608795S, RSD608855S, RSD609276S, RSD609277S, RSD609278S, RSD600209S, RSD600214S, RSD600366S, RSD600475S, RSD600484S, RSD600485S, RSD600499S, RSD600513S, RSD600525S, RSD600531S, RSD602048S, RSD602376S, RSD602378S, RSD602600S, RSD603398S, RSD604018S, RSD604559S, RSD605190S, RSD610520S, RSD610522S, RSD607657S, RSD600316S, RSD611020S, RSD603301S, RSD608946S, RSD610202S, RSD601558S, RSD601934S, RSD605328S, RSD605330S, RSD605333S, RSD605771S, RSD606050S, RSD606273S, RSD606413S, RSD606708S, RSD606948S, RSD606950S, RSD607114S, RSD607319S, RSD607703S, RSD608135S, RSD608453S, RSD608458S, RSD608560S, RSD608561S, RSD608563S, RSD609935S, RSD610123S, RSD600348S, RSD600349S, RSD600554S, RSD600609S, RSD602325S, RSD602582S, RSD602587S, RSD603101S, RSD603127S, RSD603297S, RSD603309S, RSD603363S, RSD603386S, RSD604519S, RSD605221S, RSD609936S, RSD605349S, RSD605799S, RSD600498S, RSD600501S, RSD600526S, RSD601840S, RSD608211S, RSD610933S, RSD605143S, RSD605747S, RSD605759S, RSD605762S, RSD605960S, RSD606078S, RSD606179S, RSD606836S, RSD606837S, RSD606838S, RSD607819S, RSD607823S, RSD608173S, RSD608239S, RSD608278S, RSD608555S, RSD608617S, RSD608618S, RSD608619S, RSD608622S, RSD608919S, RSD609186S, RSD610047S, RSD610081S, RSD610126S, RSD603694S, RSD603695S, RSD603696S, RSD603698S, RSD603706S, RSD603707S, RSD603840S, RSD603846S, RSD603884S, RSD604095S, RSD604096S, RSD604097S, RSD604220S, RSD604577S, RSD604767S, RSD605144S, RSD610136S, RSD610470S, RSD605466S, RSD605725S, RSD606088S, RSD606090S, RSD606732S, RSD606733S, RSD606734S, RSD603662S, RSD607692S, RSD608434S, RSD608630S, RSD608728S, RSD602666S, RSD603643S, RSD604822S, RSD607310S, RSD610154S, RSD602490S, RSD604734S, RSD604748S, RSD605724S, RSD607667S, RSD607668S, RSD607669S, RSD608120S, RSD608121S, RSD604717S, RSD607834S, RSD608947S, RSD610041S, RSD605304S, RSD602534S, RSD604804S, RSD605138S, RSD600615S, RSD600103S, RSD600106S, RSD600138S, RSD600150S, RSD602411S, RSD602402S, RSD602430S, RSD601435S, RSD602346S, RSD602407S, RSD602450S, RSD602684S, RSD603008S, RSD607163S, RSD610016S, RSD608939S, RSD604595S, RSD604597S, RSD604598S, RSD604601S, RSD604603S, RSD604605S, RSD604606S, RSD604607S, RSD604608S, RSD604610S, RSD604611S, RSD604612S, RSD608938S, RSD609017S, RSD609018S, RSD609019S, RSD609020S, RSD609021S, RSD609022S, RSD607651S, RSD607654S, RSD607658S, RSD607660S, RSD607662S, RSD600216S, RSD602212S, RSD602603S, RSD602741S, RSD602742S, RSD602743S, RSD602746S, RSD603040S, RSD603939S, RSD604478S, RSD610015S, RSD605331S, RSD605447S, RSD605464S, RSD606053S, RSD607493S, RSD608136S, RSD610193S, RSD605225S, RSD608809S, RSD608878S, RSD603094S, RSD606072S, RSD606735S, RSD608266S, RSD608267S, RSD608895S, RSD609991S, RSD604327S, RSD606969S, RSD607868S, RSD608624S, RSD606976S, RSD602299S, RSD603345S, RSD603611S, RSD605726S, RSD605727S, RSD606425S, RSD606531S, RSD606668S, RSD606669S, RSD607069S, RSD607070S, RSD607339S, RSD607341S, RSD607503S, RSD607596S, RSD607916S, RSD608341S, RSD608607S, RSD608608S, RSD608752S, RSD608753S, RSD609056S, RSD609057S, RSD610227S, RSD610233S, RSD610534S, RSD610766S, RSD610769S, RSD605353S, RSD605354S, RSD605896S, RSD605897S, RSD605913S, RSD605914S, RSD606280S, RSD606281S, RSD606754S, RSD607125S, RSD607582S, RSD608861S, RSD600193S, RSD600419S, RSD600441S, RSD600446S, RSD600579S, RSD601268S, RSD601392S, RSD601801S, RSD601842S, RSD601964S, RSD602058S, RSD602059S, RSD602204S, RSD602239S, RSD602384S, RSD602406S, RSD602526S, RSD602552S, RSD602780S, RSD602875S, RSD602877S, RSD603421S, RSD603956S, RSD603977S, RSD603979S, RSD604044S, RSD604049S, RSD604110S, RSD604530S, RSD604757S, RSD604806S, RSD604893S, RSD604961S, RSD605191S, RSD611147S, RSD611148S, RSD608157S, RSD603622S, RSD608807S, RSD608639S, RSD610025S, RSD610162S, RSD608872S, RSD602497S, RSD604475S, RSD604477S, RSD605139S, RSD601935S, RSD602389S, RSD605323S, RSD605803S, RSD600239S, RSD601555S, RSD602326S, RSD602328S, RSD602380S, RSD602393S, RSD602456S, RSD602469S, RSD602618S, RSD603120S, RSD605758S, RSD605761S, RSD608913S, RSD608914S, RSD608995S, RSD608999S, RSD609270S, RSD605142S, RSD608996S, RSD608997S, RSD608998S, RSD609000S, RSD601877S, RSD608645S, RSD610033S, RSD600319S, RSD601434S, RSD601878S, RSD602282S, RSD602731S, RSD608867S, RSD602485S, RSD602493S, RSD602630S, RSD602706S, RSD602773S, RSD602930S, RSD602931S, RSD603001S, RSD603004S, RSD603215S, RSD603248S, RSD603251S, RSD603366S, RSD603367S, RSD602350S, RSD602479S, RSD602705S, RSD608488S, RSD602500S, RSD607532S, RSD603731S, RSD603768S, RSD608096S, RSD608097S, RSD608098S, RSD608099S, RSD609811S, RSD609813S, RSD609814S, RSD609810S, RSD609812S, RSD609815S, RSD605355S, RSD605356S, RSD606265S, RSD606266S, RSD606696S, RSD606729S, RSD606939S, RSD607107S, RSD607230S, RSD607231S, RSD607544S, RSD608141S, RSD608197S, RSD608430S, RSD608769S, RSD609914S, RSD610118S, RSD610293S, RSD600223S, RSD600231S, RSD600233S, RSD600250S, RSD600266S, RSD600270S, RSD600326S, RSD600330S, RSD600364S, RSD600422S, RSD600565S, RSD601792S, RSD601936S, RSD601937S, RSD602145S, RSD602166S, RSD602213S, RSD602297S, RSD602354S, RSD602686S, RSD602900S, RSD602993S, RSD603021S, RSD603858S, RSD603866S, RSD603872S, RSD603933S, RSD604008S, RSD604013S, RSD604208S, RSD604214S, RSD604759S, RSD604760S, RSD605192S, RSD607305S, RSD600386S, RSD600607S, RSD602675S, RSD602992S, RSD603898S, RSD605915S, RSD605935S, RSD606256S, RSD606258S, RSD601882S, RSD601884S, RSD601885S, RSD601886S, RSD601888S, RSD604502S, RSD604503S, RSD605157S, RSD605158S, RSD605934S, RSD609016S, RSD605368S, RSD605369S, RSD606782S, RSD606803S, RSD606840S, RSD606841S, RSD607292S, RSD607891S, RSD608080S, RSD608081S, RSD608082S, RSD608219S, RSD608337S, RSD608570S, RSD608783S, RSD608924S, RSD609007S, RSD610166S, RSD610591S, RSD605365S, RSD605902S, RSD606974S, RSD607833S, RSD608176S, RSD608754S, RSD602015S, RSD608831S, RSD607380S, RSD606233S, RSD606247S, RSD609704S, RSD609705S, RSD602458S, RSD602509S, RSD602713S, RSD604029S, RSD604537S, RSD607708S, RSD601793S, RSD602521S, RSD603070S, RSD603735S, RSD608586S, RSD602540S, RSD605774S, RSD606194S, RSD606270S, RSD606701S, RSD606703S, RSD606997S, RSD607110S, RSD607487S, RSD607554S, RSD607556S, RSD607557S, RSD607694S, RSD608203S, RSD608559S, RSD608730S, RSD608773S, RSD609925S, RSD609930S, RSD610097S, RSD610098S, RSD610181S, RSD610300S, RSD600259S, RSD602585S, RSD602606S, RSD602620S, RSD603610S, RSD603646S, RSD610475S, RSD608880S, RSD609052S, RSD600401S, RSD602545S, RSD603326S, RSD603652S, RSD601410S, RSD601412S, RSD609103S, RSD609135S, RSD609138S, RSD609145S, RSD608129S, RSD608647S, RSD600199S, RSD601363S, RSD601372S, RSD601904S, RSD603745S, RSD605322S, RSD606414S, RSD606415S, RSD606688S, RSD606736S, RSD607173S, RSD608190S, RSD608268S, RSD608602S, RSD600311S, RSD600563S, RSD602336S, RSD602397S, RSD602621S, RSD602772S, RSD603074S, RSD603975S, RSD605218S, RSD610568S, RSD610569S, RSD600222S, RSD600581S, RSD602120S, RSD602870S, RSD603207S, RSD607237S, RSD602661S, RSD604820S, RSD609209S, RSD609210S, RSD609212S, RSD603016S, RSD603250S, RSD607321S, RSD608921S, RSD611122S, RSD608859S, RSD610763S, RSD600370S, RSD608772S, RSD607061S, RSD607223S, RSD607311S, RSD608155S, RSD608243S, RSD608262S, RSD602539S, RSD603741S, RSD603751S, RSD603844S, RSD608128S, RSD610119S, RSD600489S, RSD610515S, RSD607829S, RSD600371S, RSD600460S, RSD601354S, RSD601828S, RSD602462S, RSD602682S, RSD603196S, RSD604021S, RSD605989S, RSD605990S, RSD609635S, RSD609636S, RSD608204S, RSD603876S, RSD605351S, RSD606900S, RSD608759S, RSD610223S, RSD600210S, RSD601389S, RSD603005S, RSD603039S, RSD604534S, RSD604780S, RSD606238S, RSD606489S, RSD606490S, RSD606657S, RSD606983S, RSD607459S, RSD607460S, RSD607811S, RSD607812S, RSD607813S, RSD608362S, RSD608609S, RSD608610S, RSD608611S, RSD608613S, RSD608614S, RSD608750S, RSD608751S, RSD609046S, RSD609047S, RSD610130S, RSD610131S, RSD610134S, RSD604121S, RSD604122S, RSD604123S, RSD604124S, RSD604126S, RSD604128S, RSD604130S, RSD604131S, RSD604132S, RSD604133S, RSD604134S, RSD604135S, RSD604138S, RSD604139S, RSD604141S, RSD604142S, RSD604147S, RSD604682S, RSD604684S, RSD604690S, RSD604691S, RSD604692S, RSD604693S, RSD604694S, RSD604695S, RSD604697S, RSD604699S, RSD604700S, RSD604702S, RSD604703S, RSD604704S, RSD604705S, RSD604707S, RSD604873S, RSD604875S, RSD604876S, RSD604877S, RSD608612S, RSD608615S, RSD610082S, RSD610129S, RSD610132S, RSD610133S, RSD606294S, RSD606297S, RSD606427S, RSD606973S, RSD607131S, RSD607222S, RSD607225S, RSD607570S, RSD607571S, RSD608183S, RSD600185S, RSD600215S, RSD600217S, RSD600261S, RSD600299S, RSD600332S, RSD600333S, RSD600338S, RSD600357S, RSD600360S, RSD600383S, RSD600414S, RSD600416S, RSD600434S, RSD600436S, RSD600451S, RSD600552S, RSD600559S, RSD600597S, RSD600606S, RSD601366S, RSD601440S, RSD601559S, RSD601596S, RSD601819S, RSD601859S, RSD601871S, RSD601924S, RSD601925S, RSD602480S, RSD602679S, RSD602918S, RSD603033S, RSD603064S, RSD603153S, RSD603255S, RSD603547S, RSD603868S, RSD604533S, RSD609191S, RSD609194S, RSD609197S, RSD609201S, RSD609192S, RSD609195S, RSD609196S, RSD609198S, RSD609199S, RSD605826S, RSD605827S, RSD606225S, RSD606437S, RSD606438S, RSD606690S, RSD606901S, RSD607027S, RSD607251S, RSD607258S, RSD607346S, RSD607362S, RSD607364S, RSD607756S, RSD608134S, RSD608188S, RSD608189S, RSD608625S, RSD608626S, RSD608688S, RSD608786S, RSD608832S, RSD600532S, RSD601814S, RSD601830S, RSD601833S, RSD601836S, RSD601906S, RSD602123S, RSD602134S, RSD602361S, RSD602451S, RSD602678S, RSD602880S, RSD603230S, RSD603233S, RSD603235S, RSD603236S, RSD603277S, RSD603278S, RSD603299S, RSD603385S, RSD603518S, RSD603587S, RSD603717S, RSD604514S, RSD605069S, RSD605231S, RSD601391S, RSD602236S, RSD604771S, RSD609968S, RSD601569S, RSD606747S, RSD603034S, RSD605292S, RSD605392S, RSD605816S, RSD606063S, RSD606146S, RSD606201S, RSD606441S, RSD606484S, RSD606485S, RSD606743S, RSD606818S, RSD607017S, RSD607091S, RSD607094S, RSD607594S, RSD607602S, RSD608793S, RSD600124S, RSD600184S, RSD600189S, RSD600191S, RSD600194S, RSD600452S, RSD601332S, RSD601334S, RSD601365S, RSD602760S, RSD602761S, RSD603035S, RSD603044S, RSD603404S, RSD603405S, RSD603407S, RSD603914S, RSD603915S, RSD604892S, RSD605891S, RSD607159S, RSD610125S, RSD608071S, RSD608072S, RSD608073S, RSD608074S, RSD608075S, RSD607160S, RSD608516S, RSD608545S, RSD609014S, RSD609015S, RSD610128S, RSD610231S, RSD610854S, RSD600262S, RSD600505S, RSD601983S, RSD602719S, RSD603043S, RSD603346S, RSD603690S, RSD603954S, RSD603955S, RSD610153S, RSD606475S, RSD609149S, RSD601368S, RSD602917S, RSD601853S, RSD609053S, RSD610443S, RSD610474S, RSD610476S, RSD610477S, RSD610952S, RSD610949S, RSD605950S, RSD605281S, RSD606816S, RSD603129S, RSD601844S, RSD607702S, RSD600608S, RSD601426S, RSD602063S, RSD602096S, RSD602588S, RSD602629S, RSD603012S, RSD603037S, RSD603331S, RSD603548S, RSD604517S, RSD603392S, RSD605302S, RSD605905S, RSD606040S, RSD606296S, RSD606914S, RSD607129S, RSD607130S, RSD607158S, RSD607312S, RSD607830S, RSD600324S, RSD600398S, RSD601820S, RSD601841S, RSD602251S, RSD602475S, RSD602755S, RSD602800S, RSD603154S, RSD603182S, RSD603887S, RSD604047S, RSD604741S, RSD605313S, RSD605314S, RSD605315S, RSD605316S, RSD605317S, RSD605885S, RSD605886S, RSD605887S, RSD605888S, RSD605889S, RSD603828S, RSD603842S, RSD603889S, RSD603894S, RSD603895S, RSD603896S, RSD604258S, RSD604259S, RSD604300S, RSD604301S, RSD604303S, RSD604334S, RSD604509S, RSD604510S, RSD604511S, RSD604512S, RSD604513S, RSD604569S, RSD604584S, RSD604585S, RSD604586S, RSD606491S, RSD606493S, RSD606507S, RSD606508S, RSD605909S, RSD602486S, RSD603086S, RSD601568S, RSD605312S, RSD606822S, RSD607055S, RSD607135S, RSD607352S, RSD608182S, RSD608643S, RSD608760S, RSD610032S, RSD610194S, RSD600329S, RSD600379S, RSD602288S, RSD602418S, RSD602494S, RSD602564S, RSD602566S, RSD602613S, RSD602866S, RSD602980S, RSD603253S, RSD603867S, RSD603869S, RSD603870S, RSD604026S, RSD604532S, RSD604772S, RSD604520S, RSD606674S, RSD605340S, RSD606048S, RSD606190S, RSD606191S, RSD606220S, RSD606697S, RSD606938S, RSD607105S, RSD607314S, RSD607317S, RSD607542S, RSD607547S, RSD604575S, RSD604819S, RSD604859S, RSD609281S, RSD609282S, RSD609284S, RSD609279S, RSD609280S, RSD609283S, RSD600218S, RSD600246S, RSD600315S, RSD600365S, RSD602691S, RSD602711S, RSD603191S, RSD603633S, RSD603684S, RSD608144S, RSD606678S, RSD603088S, RSD608427S, RSD608724S, RSD608725S, RSD602011S, RSD602013S, RSD602135S, RSD602136S, RSD602137S, RSD605417S, RSD608564S, RSD609223S, RSD610094S, RSD610452S, RSD609894S, RSD606726S, RSD607156S, RSD607681S, RSD608231S, RSD610273S, RSD600447S, RSD600604S, RSD600605S, RSD601378S, RSD601379S, RSD601380S, RSD602258S, RSD602460S, RSD602730S, RSD610157S, RSD605382S, RSD606057S, RSD606738S, RSD606758S, RSD606994S, RSD607111S, RSD607263S, RSD608275S, RSD601405S, RSD602285S, RSD603418S, RSD603679S, RSD606177S, RSD606615S, RSD610172S, RSD601852S, RSD602427S, RSD603186S, RSD603382S, RSD603637S, RSD603926S, RSD606418S, RSD606448S, RSD606477S, RSD606641S, RSD610524S, RSD608584S, RSD608963S, RSD610190S, RSD600284S, RSD600472S, RSD600582S, RSD600596S, RSD603065S, RSD603066S, RSD603730S, RSD610564S, RSD605805S, RSD602031S, RSD603383S, RSD603674S, RSD605115S, RSD606783S, RSD606784S, RSD606785S, RSD606786S, RSD606790S, RSD606793S, RSD606797S, RSD606798S, RSD606801S, RSD606805S, RSD606806S, RSD606807S, RSD606808S, RSD606833S, RSD606834S, RSD606843S, RSD608748S, RSD608749S, RSD608784S, RSD608798S, RSD609224S, RSD606788S, RSD606835S, RSD608714S, RSD609225S, RSD610870S, RSD610871S, RSD606021S, RSD606022S, RSD606023S, RSD606024S, RSD606026S, RSD606028S, RSD606032S, RSD606033S, RSD606035S, RSD609207S, RSD606479S, RSD608234S, RSD608779S, RSD608808S, RSD603080S, RSD606207S, RSD606766S, RSD608159S, RSD601800S, RSD602491S, RSD603172S, RSD603184S, RSD605815S, RSD608695S, RSD604560S, RSD604884S, RSD606084S, RSD606761S, RSD607599S, RSD605291S, RSD605809S, RSD606163S, RSD607770S, RSD608979S, RSD600267S, RSD601456S, RSD601457S, RSD602067S, RSD607761S, RSD610935S, RSD611144S, RSD607810S, RSD607837S, RSD607838S, RSD607840S, RSD608257S, RSD610055S, RSD610057S, RSD608869S, RSD601415S, RSD601421S, RSD601550S, RSD602287S, RSD603614S, RSD604019S, RSD606221S, RSD607723S, RSD608685S, RSD606913S, RSD607458S, RSD607572S, RSD600394S, RSD600448S, RSD601923S, RSD604761S, RSD606086S, RSD606626S, RSD609042S, RSD603656S, RSD602391S, RSD605811S, RSD605825S, RSD608192S, RSD608702S, RSD608788S, RSD605819S, RSD606202S, RSD607092S, RSD608576S, RSD601897S, RSD606410S, RSD607300S, RSD608269S, RSD610214S, RSD610243S, RSD601916S, RSD610551S, RSD605391S, RSD606442S, RSD606705S, RSD607766S, RSD607769S, RSD603389S, RSD604508S, RSD604959S, RSD605109S, RSD605189S, RSD600511S, RSD602665S, RSD610346S, RSD608172S, RSD607533S, RSD601542S, RSD602499S, RSD602570S, RSD602608S, RSD603015S, RSD603208S, RSD604230S, RSD604239S, RSD604861S, RSD604862S, RSD604864S, RSD606894S, RSD607995S, RSD603762S, RSD610961S, RSD610962S, RSD610964S, RSD603330S, RSD604090S, RSD610790S, RSD600478S, RSD600481S, RSD602699S, RSD605252S, RSD605743S, RSD605876S, RSD605877S, RSD606158S, RSD606290S, RSD608652S, RSD608936S, RSD600555S, RSD602319S, RSD602913S, RSD602984S, RSD603173S, RSD603736S, RSD604060S, RSD604062S, RSD604544S, RSD604856S, RSD611046S, RSD607161S, RSD607581S, RSD600507S, RSD603274S, RSD610211S, RSD602248S, RSD604046S, RSD605949S, RSD607782S, RSD610770S, RSD600387S, RSD600483S, RSD602447S, RSD602517S, RSD603399S, RSD607360S, RSD608207S, RSD608233S, RSD608600S, RSD608692S, RSD609901S, RSD610159S, RSD607133S, RSD601575S, RSD601586S, RSD602271S, RSD602925S, RSD603773S, RSD602215S, RSD603589S, RSD610628S, RSD603743S, RSD604196S, RSD603263S, RSD606424S, RSD606896S, RSD607809S, RSD602206S, RSD602217S, RSD602694S, RSD602937S, RSD603861S, RSD608948S, RSD604740S, RSD606143S, RSD606226S, RSD607592S, RSD602432S, RSD608866S, RSD610222S, RSD601413S, RSD601873S, RSD602092S, RSD602932S, RSD603549S, RSD603964S, RSD610853S, RSD606222S, RSD606634S, RSD603974S, RSD607099S, RSD610782S, RSD605244S, RSD605344S, RSD605352S, RSD605367S, RSD605390S, RSD605476S, RSD605754S, RSD605800S, RSD606097S, RSD606156S, RSD606160S, RSD606200S, RSD606443S, RSD606482S, RSD606640S, RSD606642S, RSD606762S, RSD607327S, RSD607330S, RSD607497S, RSD607498S, RSD607583S, RSD607595S, RSD607771S, RSD607779S, RSD608163S, RSD608305S, RSD608494S, RSD608499S, RSD608721S, RSD608746S, RSD608747S, RSD610528S, RSD610786S, RSD600092S, RSD600256S, RSD600275S, RSD600345S, RSD600482S, RSD600553S, RSD600585S, RSD602108S, RSD602320S, RSD602518S, RSD602595S, RSD602642S, RSD602717S, RSD602720S, RSD602737S, RSD603031S, RSD603324S, RSD603361S, RSD603377S, RSD603419S, RSD603638S, RSD603647S, RSD603710S, RSD603969S, RSD604528S, RSD604558S, RSD604578S, RSD604891S, RSD605114S, RSD602421S, RSD605385S, RSD606694S, RSD607296S, RSD607566S, RSD608640S, RSD610027S, RSD610497S, RSD610498S, RSD602457S, RSD602747S, RSD607524S, RSD606142S, RSD604943S, RSD601839S, RSD603234S, RSD610106S, RSD606778S, RSD607098S, RSD607283S, RSD607326S, RSD607584S, RSD607764S, RSD608214S, RSD608803S, RSD608974S, RSD608975S, RSD608976S, RSD610350S, RSD605780S, RSD600305S, RSD601958S, RSD602465S, RSD602886S, RSD602919S, RSD602942S, RSD603201S, RSD605016S, RSD600257S, RSD607104S, RSD607106S, RSD602461S, RSD602714S, RSD605741S, RSD605742S, RSD605744S, RSD607935S, RSD607936S, RSD607937S, RSD607939S, RSD604223S, RSD604224S, RSD604226S, RSD604869S, RSD604871S, RSD604872S, RSD600449S, RSD600454S, RSD600584S, RSD601447S, RSD601452S, RSD601556S, RSD601560S, RSD602084S, RSD602345S, RSD603661S, RSD610029S, RSD605939S, RSD607284S, RSD606199S, RSD606772S, RSD607097S, RSD607100S, RSD607598S, RSD607601S, RSD607762S, RSD607772S, RSD607773S, RSD608654S, RSD608658S, RSD608717S, RSD608719S, RSD608927S, RSD608928S, RSD608656S, RSD610526S, RSD610639S, RSD610747S, RSD608687S, RSD609895S, RSD608493S, RSD600437S, RSD600456S, RSD603219S, RSD603562S, RSD606157S, RSD606763S, RSD607279S, RSD607328S, RSD608079S, RSD608313S, RSD608585S, RSD608828S, RSD610114S, RSD610221S, RSD600307S, RSD603256S, RSD603601S, RSD604043S, RSD604587S, RSD611062S, RSD605775S, RSD606449S, RSD600263S, RSD601386S, RSD602303S, RSD602314S, RSD602626S, RSD602704S, RSD603855S, RSD604465S, RSD605159S, RSD603820S, RSD604556S, RSD604857S, RSD606504S, RSD607568S, RSD607715S, RSD608621S, RSD600292S, RSD601594S, RSD601893S, RSD602175S, RSD603414S, RSD603099S, RSD603292S, RSD603294S, RSD603355S, RSD605778S, RSD606094S, RSD606951S, RSD606999S, RSD607116S, RSD603880S, RSD603881S, RSD603936S, RSD608237S, RSD608681S, RSD607745S, RSD605361S, RSD605770S, RSD606274S, RSD602245S, RSD606683S, RSD609973S, RSD606295S, RSD606910S, RSD610539S, RSD606076S, RSD603246S, RSD608738S, RSD603957S, RSD604751S, RSD604753S, RSD607749S, RSD608704S, RSD609939S, RSD610103S, RSD610324S, RSD603026S, RSD603724S, RSD604811S, RSD604942S, RSD609213S, RSD609214S, RSD610137S, RSD607113S, RSD604033S, RSD605729S, RSD600286S, RSD600291S, RSD600298S, RSD600301S, RSD600393S, RSD600535S, RSD600598S, RSD602203S, RSD602434S, RSD602472S, RSD602610S, RSD602611S, RSD602700S, RSD602735S, RSD602756S, RSD602921S, RSD602933S, RSD603725S, RSD603848S, RSD603849S, RSD604068S, RSD604471S, RSD604522S, RSD604570S, RSD604889S, RSD606065S, RSD606219S, RSD606474S, RSD606942S, RSD607371S, RSD607685S, RSD609996S, RSD608577S, RSD610096S, RSD602176S, RSD602198S, RSD602658S, RSD603131S, RSD603149S, RSD603150S, RSD603612S, RSD604504S, RSD605387S, RSD606469S, RSD606728S, RSD608147S, RSD609227S, RSD602513S, RSD603176S, RSD604749S, RSD605951S, RSD606450S, RSD604874S, RSD606411S, RSD606451S, RSD606452S, RSD603774S, RSD603778S, RSD605419S, RSD605420S, RSD605422S, RSD605424S, RSD605425S, RSD605426S, RSD605441S, RSD605444S, RSD606245S, RSD606248S, RSD606791S, RSD607274S, RSD607291S, RSD607320S, RSD607519S, RSD608076S, RSD608077S, RSD608078S, RSD608595S, RSD608833S, RSD608954S, RSD604205S, RSD605141S, RSD608937S, RSD610224S, RSD610438S, RSD605377S, RSD605378S, RSD605903S, RSD606647S, RSD606842S, RSD608761S, RSD608822S, RSD608987S, RSD609883S, RSD610212S, RSD600213S, RSD602710S, RSD602908S, RSD603577S, RSD610638S, RSD606172S, RSD608215S, RSD608513S, RSD603617S, RSD609095S, RSD609096S, RSD609098S, RSD609099S, RSD609097S, RSD605409S, RSD606282S, RSD606613S, RSD606825S, RSD606922S, RSD606923S, RSD607123S, RSD607124S, RSD607127S, RSD607128S, RSD608100S, RSD608594S, RSD608881S, RSD600566S, RSD600572S, RSD602093S, RSD602097S, RSD602121S, RSD602197S, RSD602395S, RSD602614S, RSD602874S, RSD602889S, RSD603147S, RSD603381S, RSD603729S, RSD603968S, RSD604535S, RSD604776S, RSD605013S, RSD605102S, RSD605136S, RSD605195S, RSD603349S, RSD605363S, RSD605907S, RSD606920S, RSD608185S, RSD608868S, RSD610019S, RSD610576S, RSD601557S, RSD602723S, RSD602724S, RSD602725S, RSD603058S, RSD603334S, RSD603348S, RSD603949S, RSD604215S, RSD604536S, RSD604752S, RSD604789S, RSD610048S, RSD610836S, RSD610840S, RSD600377S, RSD601367S, RSD610548S, RSD603911S, RSD606926S, RSD607139S, RSD607140S, RSD607659S, RSD607784S, RSD607785S, RSD602797S, RSD607787S, RSD609023S, RSD609024S, RSD609025S, RSD609026S, RSD605910S, RSD605911S, RSD605912S, RSD602309S, RSD602975S, RSD602981S, RSD602982S, RSD602110S, RSD605880S, RSD606361S, RSD607901S, RSD607908S, RSD609236S, RSD609237S, RSD609238S, RSD605360S, RSD606755S, RSD606756S, RSD607120S, RSD602722S, RSD603701S, RSD600391S, RSD602511S, RSD608340S, RSD609721S, RSD604518S, RSD605036S, RSD606083S, RSD606259S, RSD606467S, RSD606614S, RSD607153S, RSD607670S, RSD610271S, RSD602182S, RSD602801S, RSD602804S, RSD602869S, RSD602907S, RSD603000S, RSD603145S, RSD603356S, RSD603641S, RSD603704S, RSD603728S, RSD603865S, RSD604066S, RSD604111S, RSD604113S, RSD604118S, RSD604885S, RSD608087S, RSD608646S, RSD608734S, RSD600198S, RSD601909S, RSD602481S, RSD610360S, RSD602423S, RSD607597S, RSD600558S, RSD602085S, RSD603970S, RSD606205S, RSD606481S, RSD608649S, RSD608844S, RSD608967S, RSD602453S, RSD603057S, RSD603103S, RSD603124S, RSD603283S, RSD604063S, RSD604064S, RSD606283S, RSD606918S, RSD610850S, RSD600203S, RSD600399S, RSD602094S, RSD603746S, RSD604779S, RSD605196S, RSD605394S, RSD606372S, RSD606749S, RSD607288S, RSD607525S, RSD608222S, RSD608589S, RSD608983S, RSD609882S, RSD610236S, RSD610453S, RSD608110S, RSD608112S, RSD608113S, RSD608114S, RSD608115S, RSD606183S, RSD606197S, RSD606496S, RSD606637S, RSD606902S, RSD607090S, RSD607591S, RSD607757S, RSD608834S, RSD610554S, RSD601987S, RSD602138S, RSD602195S, RSD602317S, RSD602318S, RSD602424S, RSD602425S, RSD602428S, RSD603030S, RSD603703S, RSD603719S, RSD603834S, RSD603857S, RSD604886S, RSD605219S, RSD610920S, RSD606223S, RSD607019S, RSD607359S, RSD608187S, RSD608274S, RSD609978S, RSD609829S, RSD609830S, RSD609831S, RSD604218S, RSD604225S, RSD606059S, RSD606173S, RSD606278S, RSD606727S, RSD606937S, RSD607299S, RSD607454S, RSD607736S, RSD608637S, RSD608929S, RSD608971S, RSD610018S, RSD610284S, RSD610295S, RSD610574S, RSD610666S, RSD600545S, RSD602781S, RSD603051S, RSD604525S, RSD604883S, RSD610665S, RSD610875S, RSD611026S, RSD606611S, RSD608737S, RSD608758S, RSD608863S, RSD600235S, RSD602355S, RSD605132S, RSD607064S, RSD607065S, RSD607066S, RSD607298S, RSD607395S, RSD607567S, RSD608578S, RSD608641S, RSD610026S, RSD610127S, RSD610307S, RSD610309S, RSD610509S, RSD610502S, RSD610513S, RSD610310S, RSD605783S, RSD606137S, RSD606725S, RSD607535S, RSD607678S, RSD608762S, RSD608764S, RSD610454S, RSD610455S, RSD601377S, RSD602281S, RSD602388S, RSD602410S, RSD603546S, RSD603875S, RSD604515S, RSD605179S, RSD606506S, RSD608549S, RSD608700S, RSD610461S, RSD610562S, RSD610884S, RSD607087S, RSD607754S, RSD607313S, RSD608744S, RSD608706S, RSD607345S, RSD607579S, RSD608597S, RSD601390S, RSD603962S, RSD606494S, RSD610939S, RSD608858S, RSD610523S, RSD610938S, RSD604775S, RSD607141S, RSD606151S, RSD610545S, RSD606056S, RSD606367S, RSD606741S, RSD600529S, RSD601931S, RSD602106S, RSD602905S, RSD603097S, RSD603619S, RSD603734S, RSD607289S, RSD605018S, RSD602733S, RSD602785S, RSD602149S, RSD604058S, RSD610043S, RSD610042S, RSD606144S, RSD606416S, RSD606689S, RSD608530S, RSD608596S, RSD600232S, RSD600344S, RSD601425S, RSD602533S, RSD603420S, RSD604016S, RSD605320S, RSD605375S, RSD605781S, RSD606263S, RSD600430S, RSD600539S, RSD601359S, RSD602091S, RSD602095S, RSD602692S, RSD604954S, RSD603100S, RSD603873S, RSD606810S, RSD606811S, RSD609100S, RSD609101S, RSD605480S, RSD606060S, RSD607539S, RSD600590S, RSD601863S, RSD601865S, RSD601891S, RSD602433S, RSD602512S, RSD602594S, RSD602752S, RSD602782S, RSD602788S, RSD603567S, RSD603618S, RSD603625S, RSD603629S, RSD603758S, RSD604860S, RSD605045S, RSD607108S, RSD607109S, RSD608771S, RSD610449S, RSD600586S, RSD600588S, RSD600589S, RSD602604S, RSD602687S, RSD602890S, RSD603045S, RSD603046S, RSD603265S, RSD603336S, RSD603342S, RSD603749S, RSD604202S, RSD611039S, RSD611056S, RSD604524S, RSD608528S, RSD603041S, RSD605199S, RSD606277S, RSD606630S, RSD607522S, RSD608224S, RSD606230S, RSD606433S, RSD606434S, RSD606681S, RSD607469S, RSD608132S, RSD608605S, RSD609694S, RSD609950S, RSD610092S, RSD610178S, RSD610232S, RSD602209S, RSD602634S, RSD603020S, RSD605358S, RSD608591S, RSD604802S, RSD606210S, RSD608092S, RSD610447S, RSD610876S, RSD600201S, RSD600435S, RSD601914S, RSD601985S, RSD603391S, RSD606176S, RSD607752S, RSD604927S, RSD605301S, RSD605410S, RSD606612S, RSD606967S, RSD607054S, RSD607309S, RSD608179S, RSD608272S, RSD603171S, RSD603360S, RSD603929S, RSD603948S, RSD604901S, RSD606011S, RSD605894S, RSD608873S, RSD610068S, RSD600224S, RSD602344S, RSD603068S, RSD605916S, RSD607747S, RSD607748S, RSD602220S, RSD601361S, RSD602471S, RSD606073S, RSD609956S, RSD606954S, RSD608935S, RSD603702S, RSD604554S, RSD604747S, RSD608705S, RSD601373S, RSD605167S, RSD606285S, RSD610610S, RSD610612S, RSD605942S, RSD605943S, RSD605944S, RSD605945S, RSD605946S, RSD605947S, RSD607623S, RSD607624S, RSD608167S, RSD608168S, RSD608169S, RSD608171S, RSD609045S, RSD610071S, RSD610072S, RSD610608S, RSD610613S, RSD605463S, RSD605794S, RSD606145S, RSD606181S, RSD607144S, RSD607342S, RSD607727S, RSD607734S, RSD608871S, RSD610030S, RSD610421S, RSD602504S, RSD603747S, RSD603748S, RSD604807S, RSD605177S, RSD607726S, RSD610422S, RSD610611S, RSD608223S, RSD600570S, RSD603335S, RSD610626S, RSD606930S, RSD606932S, RSD606933S, RSD606984S, RSD607729S, RSD607732S, RSD607820S, RSD607821S, RSD607822S, RSD606141S, RSD606262S, RSD606422S, RSD606423S, RSD606470S, RSD606676S, RSD607302S, RSD608552S, RSD608553S, RSD610087S, RSD610089S, RSD600443S, RSD601880S, RSD602146S, RSD602196S, RSD602262S, RSD602396S, RSD602972S, RSD602986S, RSD603077S, RSD603083S, RSD603658S, RSD603668S, RSD603671S, RSD603732S, RSD603937S, RSD604571S, RSD604956S, RSD605183S, RSD605234S, RSD610088S, RSD610090S, RSD610889S, RSD601336S, RSD605786S, RSD606070S, RSD606152S, RSD606631S, RSD606692S, RSD606693S, RSD606730S, RSD607067S, RSD607068S, RSD607145S, RSD607538S, RSD608505S, RSD608642S, RSD601591S, RSD602592S, RSD602624S, RSD602662S, RSD603055S, RSD603305S, RSD603306S, RSD603576S, RSD610872S, RSD611063S, RSD610623S, RSD607126S, RSD601564S, RSD601850S, RSD602617S, RSD605393S, RSD606648S, RSD603190S, RSD608918S, RSD604770S, RSD609055S, RSD602576S, RSD602579S, RSD603091S, RSD603093S, RSD604549S, RSD604552S, RSD605818S, RSD606731S, RSD606991S, RSD607562S, RSD610095S, RSD602051S, RSD602414S, RSD602416S, RSD602712S, RSD603287S, RSD603384S, RSD603395S, RSD603415S, RSD603913S, RSD605206S, RSD611149S, RSD607318S, RSD602404S, RSD601988S, RSD603054S, RSD606895S, RSD610363S, RSD610368S, RSD610386S, RSD610532S, RSD600346S, RSD602242S, RSD602941S, RSD603388S, RSD610530S, RSD610533S, RSD608648S, RSD600600S, RSD601576S, RSD602589S, RSD602935S, RSD607998S, RSD607999S, RSD608000S, RSD608001S, RSD608002S, RSD608003S, RSD608004S, RSD609233S, RSD609234S, RSD609235S, RSD602605S, RSD602669S, RSD602673S, RSD602676S, RSD602685S, RSD603678S, RSD603680S, RSD603683S, RSD603686S, RSD603687S, RSD603688S, RSD603689S, RSD603697S, RSD603699S, RSD603700S, RSD604227S, RSD604228S, RSD604235S, RSD604241S, RSD604244S, RSD604247S, RSD604251S, RSD609228S, RSD609229S, RSD609230S, RSD609231S, RSD609232S, RSD605764S, RSD605766S, RSD606629S, RSD607115S, RSD607264S, RSD607265S, RSD607268S, RSD607569S, RSD602256S, RSD602744S, RSD603535S, RSD607142S, RSD601956S, RSD602701S, RSD603243S, RSD603279S, RSD603522S, RSD604014S, RSD608804S, RSD606150S, RSD606180S, RSD604038S, RSD607350S, RSD607351S, RSD608904S, RSD608956S, RSD608957S, RSD602012S, RSD602358S, RSD602359S, RSD604297S, RSD604299S, RSD604719S, RSD604723S, RSD604726S, RSD605121S, RSD607349S, RSD608408S, RSD608581S, RSD608903S, RSD606644S, RSD606718S, RSD604059S, RSD600098S, RSD606706S, RSD602266S, RSD606287S, RSD605296S, RSD606169S, RSD606739S, RSD606799S, RSD607629S, RSD608796S, RSD610200S, RSD610343S, RSD600592S, RSD601335S, RSD602377S, RSD602381S, RSD602702S, RSD603558S, RSD603920S, RSD604958S, RSD610932S, RSD605793S, RSD605804S, RSD606149S, RSD606649S, RSD606804S, RSD606921S, RSD607286S, RSD608210S, RSD608333S, RSD608506S, RSD608636S, RSD608805S, RSD608949S, RSD600367S, RSD601563S, RSD602101S, RSD602394S, RSD603006S, RSD603315S, RSD603599S, RSD611019S, RSD610201S, RSD606074S, RSD608127S, RSD607778S, RSD610527S, RSD608777S, RSD608813S, RSD609089S, RSD607563S, RSD610863S, RSD608806S, RSD610014S, RSD606812S, RSD610519S, RSD607608S, RSD608277S, RSD608534S, RSD600450S, RSD601548S, RSD603756S, RSD605984S, RSD607621S, RSD607731S, RSD602583S, RSD611010S, RSD608933S, RSD608934S, RSD610174S, RSD610175S, RSD610315S, RSD608091S, RSD602155S, RSD603156S, RSD608265S, RSD610912S, RSD609027S, RSD607357S, RSD602904S, RSD604055S, RSD607393S, RSD610397S, RSD605188S, RSD608830S, RSD606665S, RSD607152S, RSD608689S, RSD608988S, RSD602887S, RSD606989S, RSD600186S, RSD600279S, RSD600280S, RSD601382S, RSD601384S, RSD601385S, RSD601387S, RSD601401S, RSD601566S, RSD601892S, RSD601921S, RSD601967S, RSD601968S, RSD601975S, RSD602227S, RSD602569S, RSD602695S, RSD603166S, RSD603249S, RSD603976S, RSD605892S, RSD605904S, RSD606286S, RSD606824S, RSD606831S, RSD606908S, RSD607118S, RSD607119S, RSD607574S, RSD607578S, RSD608153S, RSD608154S, RSD608156S, RSD608164S, RSD608165S, RSD608256S, RSD608258S, RSD608259S, RSD608527S, RSD608592S, RSD610031S, RSD610051S, RSD610052S, RSD610423S, RSD610424S, RSD610425S, RSD610062S, RSD610070S, RSD610802S, RSD610830S, RSD611017S, RSD609692S, RSD609693S, RSD600322S, RSD600350S, RSD600439S, RSD602315S, RSD603934S, RSD610913S, RSD606486S, RSD608247S, RSD608657S, RSD602052S, RSD603049S, RSD603623S, RSD603624S, RSD603841S, RSD605300S, RSD607575S, RSD607841S, RSD607921S, RSD608151S, RSD608546S, RSD610138S, RSD610140S, RSD603270S, RSD601847S, RSD602593S, RSD603347S, RSD603752S, RSD605740S, RSD605948S, RSD603897S, RSD604307S, RSD605288S, RSD605294S, RSD605295S, RSD605350S, RSD605380S, RSD605475S, RSD606148S, RSD606203S, RSD606980S, RSD607076S, RSD607275S, RSD608218S, RSD608635S, RSD608740S, RSD608945S, RS
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1819-2023
Related Recalls
ICD-DR DDBB2D1 EVERA XT IS1/DF1 INTL, Model Number DDBB2D1; Implantable Cardioverter Defibrillators
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.
ICD-DR DDMC3D1 EVERA MRI S OUS/US DF1, Model Number DDMC3D1; Implantable Cardioverter Defibrillators
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.
ICD DDPA2D1G COBALT XT DR MRI DF1 GOLD, Model Number DDPA2D1G; Implantable Cardioverter Defibrillators
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.