Viking M, L, and XL Mobile Lifts. Non-AC-powered patient lift.
Class I - DangerousWhat Should You Do?
- Check if you have this product: Viking¿ M Lift Model 2040035 (S/N 9 200 000 - 9 201 689) Model 2040015 (S/N 7 500 401 - 7 568 899) Model 2040005 (S/N 7 100 101 - 7 200 200) Viking¿ L Lift Model 2040004 (S/N 7 200 201 - 7 300 300) Viking¿ XL/Viking¿ 300 Lift Model 2040003 (S/N 800 001 - 804 999)
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Hill-Rom, Inc.
- Reason for Recall:
- Complaints (including one reported death) allegedly of the lift arm drifting down suddenly. If the lift arm assembly is manually lifted, the actuator can become damaged and get stuck in the highest position. If a patient is lifted into the sling while the actuator is stuck, there is a potential risk of a free fall of the patient, resulting in minor or potentially catastrophic injuries.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Viking M, L, and XL Mobile Lifts. Non-AC-powered patient lift.
Product Codes/Lot Numbers:
Viking¿ M Lift Model 2040035 (S/N 9 200 000 - 9 201 689) Model 2040015 (S/N 7 500 401 - 7 568 899) Model 2040005 (S/N 7 100 101 - 7 200 200) Viking¿ L Lift Model 2040004 (S/N 7 200 201 - 7 300 300) Viking¿ XL/Viking¿ 300 Lift Model 2040003 (S/N 800 001 - 804 999)
Distribution:
Nationwide distribution
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1818-2015
Related Recalls
Traverse rail carriage delivered with non-conforming screws are too short to ensure appropriate installation. There is potential that the screws are unable to secure the traverse rail appropriately and could loosen over time causing the rail system to fall, potentially causing injury to caregivers and/or patients
Missing bushing on Liko M220 and M230 causing wear and potential for patient falling
Missing bushing on Liko M220 and M230 causing wear and potential for patient falling