Apo B Reagent, REF: OSR6143
Class I - DangerousWhat Should You Do?
- Check if you have this product: Lot # 2631/UDI: 15099590010409
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Beckman Coulter Ireland, Inc.
- Reason for Recall:
- The product Value Assignment Sheet (VAS) is used in calibration of the device. An erroneous value was included in the VAS which has the potential to generate an incorrect APO B patient result which could exceed the total allowable error for the APO B assay (11.6%). The error results in results with a positive bias up to 42%. However, based on the mitigating factors the risk to health is considered to be negligible or limited.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
Apo B Reagent, REF: OSR6143
Product Codes/Lot Numbers:
Lot # 2631/UDI: 15099590010409
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1815-2025