Atec Insignia Anterior Cervical Plate System, REF 136-0248 Insignia, ACP, 2-Level, 48 mm, Rx Only, Non-Sterile, UDI: (01)00190376268521

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    8609319R and 8663404R
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Alphatec Spine, Inc.
Reason for Recall:
There is a potential for the screw blocking mechanism on the anterior cervical plate system to disassociate intraoperatively/postoperatively or unlock postoperatively.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

Atec Insignia Anterior Cervical Plate System, REF 136-0248 Insignia, ACP, 2-Level, 48 mm, Rx Only, Non-Sterile, UDI: (01)00190376268521

Product Codes/Lot Numbers:

8609319R and 8663404R

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-1814-2021

Related Recalls

Calibrate CCX Interbody System for spinal fusion procedures: Implants: (1) Calibrate CCX-O, 7 x 10 x 25mm 5 degrees, Part #325-07102505, nonsterile; (2) Calibrate CCX-O, 7 x 10 x 25mm 10 degrees, Part #325-07102510, nonsterile; (3) Calibrate CCX-O, 7 x 10 x 25mm 15 degrees, Part #325-07102515, nonsterile; (4) Calibrate CCX-O, 7 X 10 X 30mm 5 degrees, Part #325-07103005, nonsterile; (5) Calibrate CCX-O, 7 x 10 x 30mm 10 degrees, Part #325-07103010, nonsterile; (6) Calibrate CCX-O, 7 x 10 x 30mm 15 degrees, Part #325-07103015, nonsterile; (7) Calibrate CCX-O, 7 x 10 x 35mm 10 degrees, Part #325-07103510, nonsterile; (8) Calibrate CCX-O, 7 x 10 x 35mm 15 degrees, Part #325-07103515, nonsterile; (9) Calibrate CCX-O, 9 x 10 x 25mm 5 degrees, Part #325-09102505, nonsterile; (10) Calibrate CCX-O, 9 x 10 x 25mm 10 degrees, Part #325-09102510, nonsterile; (11) Calibrate CCX-O, 9 x 10 x 30mm 5 degrees, Part #325-09103005, nonsterile; (12) Calibrate CCX-O, 9 x 10 x 30mm 10 degrees, Part #325-09103010, nonsterile; (13) Calibrate CCX-O, 11 x 10 x 30mm 10 degrees, Part #325-11103010, nonsterile; (14) Calibrate CCX-S, 7 x 10 x 22mm 5 degrees, Part #326-07102205, nonsterile; (15) Calibrate CCX-S, 7 x 10 x 22mm 10 degrees, Part #326-07102210, nonsterile; (16) Calibrate CCX-S, 7 x 10 x 22mm 15 degrees, Part #326-07102215, nonsterile; (17) Calibrate CCX-S, 7 x 10 x 26mm 5 degrees, Part #326-07102605, nonsterile; (18) Calibrate CCX-S, 7 x 10 x 26mm 10 degrees, Part #326-07102610, nonsterile; (19) Calibrate CCX-S, 7 x 10 x 26mm 15 degrees, Part #326-07102615, nonsterile; (20) Calibrate CCX-S, 7 x 10 x 30mm 10 degrees, Part #326-07103010, nonsterile; (21) Calibrate CCX-S, 9 x 10 x 22mm 5 degrees, Part #326-09102205, nonsterile; (22) Calibrate CCX-S, 9 x 10 x 22mm 10 degrees, Part #326-09102210, nonsterile; (23) Calibrate CCX-S, 9 x 10 x 26mm 5 degrees, Part #326-09102605, nonsterile; (24) Calibrate CCX-S, 9 x 10 x 26mm 10 degrees, Part #326-09102610, nonsterile; and (25) Calibrate CCX-S, 11 x 10 x 26mm 10 degrees, Part #326-11102610, nonsterile. Convenience kits containing affected implants: (1) Calibrate CCX Oblique Implants, REF CLCCXOIMPA, containing 24 implants/kit, nonsterile; and (2) Calibrate CCX Straight Implants, REF CLCCXSIMPA, containing 36 implants/kit, nonsterile.

Alphatec Spine

Class I - Dangerous

Complaints have been received regarding post-operative implant collapse.

Dec 15, 2024 Implants & Prosthetics Nationwide View Details →

Due to reports of intraoperative graft bolt implantation breakages.

Jan 26, 2024 Other Medical Devices Nationwide View Details →

Awl instrument adjustable drill button assembly assembled in the incorrect orientation, preventing locking feature from engaging with shaft at the desired set point; instruments lack ability to control awl depth by means of adjustable shaft advancement stop, which may lead to over insertion, dural tear, vascular/neurologic injury, adjacent tissue damage, increased operative time, revision surgery.