Autosoft 90, model numbers 1000282, 1002817, 1002818, 1002819, 1002820, 1002821, 1002822, 1002823, and 1002824. Infusion sets used with Tandem Insulin Subcutaneous Infusion Pump.
Class I - DangerousWhat Should You Do?
- Check if you have this product: UDI-DI: 05705244018129. Model 1000282, lot numbers: 5381154 and 5381155; Model 1002817, lot numbers: 5372743, 5381082, 5381087, 5381083, 5381088, 5381091, 5381159, 5381084, 5381089, 5381160, 5381092, 5381094, 5381156, 5381157, 5381090, 5381158, 5372745, 5381085, 5381093, 5381086, 5381101,5372744, 5381095, 5381096, 5381097, 5381099, 5381098, 5381100, 5381162, 5381107, 5381103, 5381102, 5381104, 5381108, 5381164, 5381105, 5381109, 5381161, 5381163, 5381166, 5381106, 5381165, 5381169, 5381168, 5381172, 5381167, 5381170, 5381171, 5381173, 5381178, 5381175, 5381179, 5381180, 5381176, 5381174, 5381177, 5381181, 5381182, 5381183, 5381184, 5381186, 5381189, 5381197, 5381190, 5381192, and 5381194; Model 1002818, lot numbers: 5381110, 5381111, 5381112, 5381113, and 5381200; Model 1002819, lot numbers: 5372758, 5381114, 5372759, 5381115, 5372756, 5381117, 5381118, 5381121, 5381116, 5381120, 5381122, 5381205, 5381208, 5381202, 5381119, 5381204, 5381203, 5381207, 5381206, and 5381209. Model 1002820, lot numbers: 5372762, 5381123, 5381124, 5383835, 5381125, and 5381215; Model 1002821, lot numbers: 5381126, 5381127, 5381128, 5381129, 5381130, 5381133, 5381134, 5381132, 5381131, 5381218, 5381223, 5381219, 5381221, 5381220, 5381222, and 5381224; Model 1002822, lot numbers: 5381135, 5381136, and 5381228. Model 1002823, lot numbers: 5381139, 5381137, 5381138, 5381140, 5381141, 5381142, 5381144, 5381145, 5381146, 5381147, 5381143, 5381148, 5381149, 5381230, 5381229, 5381232, 5381231, 5381233, 5381234, 5381235, and 5381238; and Model, 1002824, lot numbers: 5376469 and 5381243.
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Unomedical A/S
- Reason for Recall:
- The IFU incorrectly states the insulin fill value of two sizes of cannula which can result in an initial insulin infusion to be fewer units than intended.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
Autosoft 90, model numbers 1000282, 1002817, 1002818, 1002819, 1002820, 1002821, 1002822, 1002823, and 1002824. Infusion sets used with Tandem Insulin Subcutaneous Infusion Pump.
Product Codes/Lot Numbers:
UDI-DI: 05705244018129. Model 1000282, lot numbers: 5381154 and 5381155; Model 1002817, lot numbers: 5372743, 5381082, 5381087, 5381083, 5381088, 5381091, 5381159, 5381084, 5381089, 5381160, 5381092, 5381094, 5381156, 5381157, 5381090, 5381158, 5372745, 5381085, 5381093, 5381086, 5381101,5372744, 5381095, 5381096, 5381097, 5381099, 5381098, 5381100, 5381162, 5381107, 5381103, 5381102, 5381104, 5381108, 5381164, 5381105, 5381109, 5381161, 5381163, 5381166, 5381106, 5381165, 5381169, 5381168, 5381172, 5381167, 5381170, 5381171, 5381173, 5381178, 5381175, 5381179, 5381180, 5381176, 5381174, 5381177, 5381181, 5381182, 5381183, 5381184, 5381186, 5381189, 5381197, 5381190, 5381192, and 5381194; Model 1002818, lot numbers: 5381110, 5381111, 5381112, 5381113, and 5381200; Model 1002819, lot numbers: 5372758, 5381114, 5372759, 5381115, 5372756, 5381117, 5381118, 5381121, 5381116, 5381120, 5381122, 5381205, 5381208, 5381202, 5381119, 5381204, 5381203, 5381207, 5381206, and 5381209. Model 1002820, lot numbers: 5372762, 5381123, 5381124, 5383835, 5381125, and 5381215; Model 1002821, lot numbers: 5381126, 5381127, 5381128, 5381129, 5381130, 5381133, 5381134, 5381132, 5381131, 5381218, 5381223, 5381219, 5381221, 5381220, 5381222, and 5381224; Model 1002822, lot numbers: 5381135, 5381136, and 5381228. Model 1002823, lot numbers: 5381139, 5381137, 5381138, 5381140, 5381141, 5381142, 5381144, 5381145, 5381146, 5381147, 5381143, 5381148, 5381149, 5381230, 5381229, 5381232, 5381231, 5381233, 5381234, 5381235, and 5381238; and Model, 1002824, lot numbers: 5376469 and 5381243.
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1810-2022
Related Recalls
VariSoft Infusion Set, Single Use.
Unomedical A/S
Specific VariSoftTM Infusion Sets were packaged using incorrect secondary packaging (boxes) branded "Trusteel Infusion Set".
Specific lots of Neria Soft Infusion Sets, which have been shipped into the US Market, have been shipped with the incorrect Instruction for Use (IFU) within the final pack. The IFU included is the IFU registered for the Rest of World (ROW) countries. The difference between the US IFU and the ROW IFU is the addition of an intended use section in the ROW IFU.
VariSoft infusion set, Model Number 1002828
Unomedical A/S
The Varisoft infusion set is used in conjunction with tandem insulin pumps. The defect reported was damage to the connector piece during manufacturing. This damage results in the connector becoming detached more easily from the insulin set than expected, requiring less force to disconnect than intended. Disconnection, when not detected, for example during sleep, leads to missed basal dosing. which in turn can result in elevated blood glucose and ketone level (nocturnal hyperglycemia). Diabetic ketoacidosis.hyperglycemia). Diabetic ketoacidosis.