Ingenia Ambition X - Model Number: 781356, Diagnostic MR system. ***Updated November 2023*** Added Model Number 782109 - Ingenia Ambition X

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    REF 781356/UDI-DI 00884838090040 Serial Numbers 47001 47002 47003 47009 47000 47012 47017 47013 47047 47040 47052 47015 47049 47056 47023 47019 47030 47053 47014 47035 47034 (removed from scope as of November 2023) 47020 47048 47080 47032 47057 47041 47072 47075 47091 47219 47184 47058 47258 47215 47197 47004 47026 47006 47007 47018 47027 47011 47037 47039 47029 47038 47043 47022 47044 47070 47062 47071 47054 ***Updated November 2023*** REF 781356/UDI-DI 00884838090040 Serial Numbers 47016 47045 47059 47067 47065 47082 47089 47033 47244 47085 47025 47060 47024 47063 47061 47051 47096 47097 47028 47094 47225 47064 47248 47074 47083 47187 47098 47174 47068 47190 47195 47227 47194 47245 47175 47221 47226 47088 47220 47270 47050 47084 47099 47093 47203 47275 47260 47264 47252 REF 782109/UDI-DI 00884838098350 Serial Numbers 47500, 47519, 47511
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Philips North America Llc
Reason for Recall:
The MR magnet is equipped with 4 lifting blocks, which are positioned on top of the magnet. These blocks are used to allow the magnets to be moved in or out by a lifting crane, e.g. during decommissioning or moving activities. A mechanical issue was detected on these hoisting interface blocks on top of the magnet.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
ongoing

Product Information

Full Description:

Ingenia Ambition X - Model Number: 781356, Diagnostic MR system. ***Updated November 2023*** Added Model Number 782109 - Ingenia Ambition X

Product Codes/Lot Numbers:

REF 781356/UDI-DI 00884838090040 Serial Numbers 47001 47002 47003 47009 47000 47012 47017 47013 47047 47040 47052 47015 47049 47056 47023 47019 47030 47053 47014 47035 47034 (removed from scope as of November 2023) 47020 47048 47080 47032 47057 47041 47072 47075 47091 47219 47184 47058 47258 47215 47197 47004 47026 47006 47007 47018 47027 47011 47037 47039 47029 47038 47043 47022 47044 47070 47062 47071 47054 ***Updated November 2023*** REF 781356/UDI-DI 00884838090040 Serial Numbers 47016 47045 47059 47067 47065 47082 47089 47033 47244 47085 47025 47060 47024 47063 47061 47051 47096 47097 47028 47094 47225 47064 47248 47074 47083 47187 47098 47174 47068 47190 47195 47227 47194 47245 47175 47221 47226 47088 47220 47270 47050 47084 47099 47093 47203 47275 47260 47264 47252 REF 782109/UDI-DI 00884838098350 Serial Numbers 47500, 47519, 47511

Distribution:

Distributed in: FL, GA, MI, MO, NH, NJ, NY, PA, SC, TX, VT

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-1795-2021