PREVI Isola Insrument and PREVI Isola Printer spares PSUs (power supply units), P1028888-001 PSU (used in Zebra LP2824 Plus printers) PSUs with the following date code; 0639XX through 1253XX (YYWWXX, where YY=Year, and WW=week).

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Instrument Serial Number AS180-00028 1035196 AS180-00067 1007580 AS180-00084 2055833 AS180-00105 2025461 AS180-00138 2025461 AS180-00179 1039938 AS180-00217 1033454 AS180-00248 1029262 AS180-00278 1031886 AS180-00300 1033400 AS180-00306 1034034 AS180-00326 1034190 AS180-00336 1041241 AS180-00339 2054793 AS180-00357 2055775 AS180-00381 1073849 AS180-00385 1054234 AS180-00398 1052053 AS180-00402 2055620 AS180-00405 1063739 AS180-00418 1042188 AS180-00420 1040343 AS180-00428 1030890 AS180-00453 1031472 AS180-00468 1043640 AS180-00476 1029272 AS180-00478 1053432 AS180-00479 1011691 AS180-00480 5001554 AS180-00481 1039588 AS180-00482 1034363 AS180-00483 1029990 AS180-00484 1049298 AS180-00485 AU01 AS180-00486 1047965 AS180-00487 1074074 AS180-00488 1000283 AS180-00489 1035594 AS180-00490 1047965 AS180-00491 1035549 AS180-00492 1070878 AS180-00493 1090075 AS180-00494 1039290 AS180-00495 1029148 AS180-00500 1007884 AS180-00509 1065153
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
bioMerieux, Inc.
Reason for Recall:
The recall has been initiated due to potential for overheating or a fire hazard associated with the Zebra LP2824 Plus printer PSU (power supply unit) located within the PREVI Isola instrument.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

PREVI Isola Insrument and PREVI Isola Printer spares PSUs (power supply units), P1028888-001 PSU (used in Zebra LP2824 Plus printers) PSUs with the following date code; 0639XX through 1253XX (YYWWXX, where YY=Year, and WW=week).

Product Codes/Lot Numbers:

Instrument Serial Number AS180-00028 1035196 AS180-00067 1007580 AS180-00084 2055833 AS180-00105 2025461 AS180-00138 2025461 AS180-00179 1039938 AS180-00217 1033454 AS180-00248 1029262 AS180-00278 1031886 AS180-00300 1033400 AS180-00306 1034034 AS180-00326 1034190 AS180-00336 1041241 AS180-00339 2054793 AS180-00357 2055775 AS180-00381 1073849 AS180-00385 1054234 AS180-00398 1052053 AS180-00402 2055620 AS180-00405 1063739 AS180-00418 1042188 AS180-00420 1040343 AS180-00428 1030890 AS180-00453 1031472 AS180-00468 1043640 AS180-00476 1029272 AS180-00478 1053432 AS180-00479 1011691 AS180-00480 5001554 AS180-00481 1039588 AS180-00482 1034363 AS180-00483 1029990 AS180-00484 1049298 AS180-00485 AU01 AS180-00486 1047965 AS180-00487 1074074 AS180-00488 1000283 AS180-00489 1035594 AS180-00490 1047965 AS180-00491 1035549 AS180-00492 1070878 AS180-00493 1090075 AS180-00494 1039290 AS180-00495 1029148 AS180-00500 1007884 AS180-00509 1065153

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-1793-2020

Related Recalls

VITEK 2 Gram-negative Susceptibility Cards Containing o Polymyxin B (pb02n) o Ceftazidime/avibactam (cza02n) o Meropenem/Vaborbactam (mev01n) o Imipenem (ipm05n) o Imipenem/Relebactam (ipr01n); CARD NAME/ PRODUCT REFERENCE: AST-N399 423701 AST-N402 423644 AST-N403 423645 AST-N404 423664 AST-N405 423864 AST-N406 423869 AST-N407 423870 AST-N409 423925 AST-N412 423936 AST-N413 423928 AST-N414 423933 AST-N415 423934 AST-N416 423935 AST-N417 423880 AST-N419 423948 AST-N421 424055 AST-N428 424320 AST-N429 424321 AST-N430 424322 AST-N433 424389 AST-N439 424501 AST-N440 424502 AST-N443 424541 AST-N444 424587 AST-N467 424857 AST-N476 424934 AST-N478 425052 AST-N480 425084 AST-N481 425085 AST-N485 425182 AST-N496 425265 AST-N800 423310 AST-N802 423706 AST-N804 424634 AST-N808 424711 AST-N810 424712 AST-N811 424713 AST-N813 424722 AST-N815 425054 AST-XN09 423425 AST-XN15 423829 AST-XN17 423673 AST-XN18 423874 AST-XN20 423947 AST-XN21 424197 AST-XN22 424199 AST-XN23 424198 AST-XN24 424351 AST-XN26 424500 AST-XN28 424586 AST-XN29 424604 AST-XN30 424639 AST-XN31 424640 AST-XN32 424678 AST-XN33 424723 AST-XN34 424752 AST-XN35 424810 AST-XN36 424860 AST-XN37 424888 AST-XN38 425074 AST-XN39 425086 AST-XN40 425095 AST-XN41 425204 AST-XN42 425309 ***Update 4/2/26 -- Additional product references not manufactured at time of initial notification. These products will not be distributed within the United States.*** AST-N484 Ref. 425181 AST-N486 Ref. 425183 AST-N491 Ref. 425206 AST-N494 Ref. 425224 AST-N485 Ref. 425182 AST-XN41 Ref. 425204 AST-XN42 Ref. 425309 AST-N504 Ref. 425491

Biomerieux

Class I - Dangerous

Potential risk of Quality Control failures and/or false resistant antibiotic results when testing isolates of Enterobacterales/Enterobacteriaceae species and/or Pseudomonas aeruginosa with the identified antibiotic formulations. Issue may lead to QC failure or false resistant results.

Jan 5, 2026 Infusion Pumps Nationwide View Details →

Potential for false resistant results for Colistin (cs02n) with multi-drug-resistant Pseudomonas aeruginosa and Acinetobacter baumannii complex used in specific cards

Jun 26, 2025 Diagnostic Equipment View Details →

Ceftriaxone (cro02n) concentration errors associated with specific recently manufactured test kits which would have the potential for for false susceptible results. The risk is present for isolates that have a result of MIC=0.5, 1, 2.

Mar 14, 2024 Diagnostic Equipment View Details →