ICD-VR DVAB1D1 VISIA AF US IS1/DF1, Model Number DVAB1D1; Implantable Cardioverter Defibrillators
Class I - DangerousWhat Should You Do?
- Check if you have this product: GTIN 00643169720619, Lot Serial Numbers: BWN203363H, BWN203588H, BWN203428H, BWN203504H, BWN203530H, BWN203214H, BWN203258H, BWN203259H, BWN203304H, BWN203477H, BWN203665H, BWN203423H, BWN203403H, BWN203327H, BWN203518H, BWN203790H, BWN203174H, BWN203624H, BWN203750H, BWN203763H, BWN203786H, BWN203745H, BWN203526H, BWN203813H, BWN203168H, BWN203236H, BWN203374H, BWN203463H, BWN203640H, BWN203213H, BWN203431H, BWN203492H, BWN203585H, BWN203436H, BWN203524H, BWN203528H, BWN203809H, BWN203432H, BWN203231H, BWN203199H, BWN203163H, BWN203193H, BWN203306H, BWN203310H, BWN203352H, BWN203373H, BWN203433H, BWN203575H, BWN203611H, BWN203616H, BWN203689H, BWN203732H, BWN203739H, BWN203749H, BWN203796H, BWN203799H, BWN203803H, BWN203181H, BWN203642H, BWN203202H, BWN203758H, BWN203671H, BWN203769H, BWN203357H, BWN203248H, BWN203253H, BWN203448H, BWN203664H, BWN203733H, BWN203260H, BWN203347H, BWN203408H, BWN203422H, BWN203535H, BWN203582H, BWN203712H, BWN203770H, BWN203697H, BWN203205H, BWN203410H, BWN203590H, BWN203332H, BWN203381H, BWN203430H, BWN203511H, BWN203737H, BWN203355H, BWN203703H, BWN203625H, BWN203380H, BWN203498H, BWN203620H, BWN203622H, BWN203810H, BWN203681H, BWN203685H, BWN203434H, BWN203653H, BWN203564H, BWN203169H, BWN203188H, BWN203186H, BWN203318H, BWN203325H, BWN203544H, BWN203468H, BWN203727H, BWN203283H, BWN203693H, BWN203667H, BWN203180H, BWN203660H, BWN203584H, BWN203386H, BWN203806H, BWN203543H, BWN203417H, BWN203395H, BWN203549H, BWN203279H, BWN203326H, BWN203706H, BWN203726H, BWN203744H, BWN203583H, BWN203365H, BWN203556H, BWN203709H, BWN203730H, BWN203203H, BWN203443H, BWN203243H, BWN203331H, BWN203315H, BWN203419H, BWN203643H, BWN203760H, BWN203520H, BWN203607H, BWN203710H, BWN203540H, BWN203265H, BWN203293H, BWN203299H, BWN203517H, BWN203354H, BWN203371H, BWN203398H, BWN203273H, BWN203369H, BWN203450H, BWN203474H, BWN203694H, BWN203774H, BWN203288H, BWN203316H, BWN203487H, BWN203538H, BWN203553H, BWN203555H, BWN203618H, BWN203294H, BWN203690H, BWN203531H, BWN203159H, BWN203731H, BWN203787H, BWN203254H, BWN203480H, BWN203599H, BWN203157H, BWN203246H, BWN203581H, BWN203776H, BWN203211H, BWN203451H, BWN203789H, BWN203233H, BWN203356H, BWN203808H, BWN203212H, BWN203191H, BWN203447H, BWN203473H, BWN203413H, BWN203414H, BWN203415H, BWN203416H, BWN203486H, BWN203489H, BWN203516H, BWN203688H, BWN203815H, BWN203235H, BWN203648H, BWN203167H, BWN203702H, BWN203290H, BWN203296H, BWN203183H, BWN203781H, BWN203284H, BWN203635H, BWN203456H, BWN203472H, BWN203580H, BWN203691H, BWN203794H, BWN203382H, BWN203229H, BWN203247H, BWN203256H, BWN203264H, BWN203274H, BWN203277H, BWN203302H, BWN203532H, BWN202561H, BWN203228H, BWN203250H, BWN203268H, BWN203269H, BWN203270H, BWN203275H, BWN203292H, BWN203420H, BWN203440H, BWN203509H, BWN203548H, BWN203615H, BWN203656H, BWN203680H, BWN203716H, BWN203722H, BWN203771H, BWN203782H, BWN203404H, BWN203657H, BWN203788H, BWN203192H, BWN203397H, BWN203461H, BWN203513H, BWN203215H, BWN203578H, BWN203720H, BWN203747H, BWN203773H, BWN203683H, BWN203409H, BWN203597H, BWN203539H, BWN203351H, BWN203334H, BWN203639H, BWN203714H, BWN203728H, BWN203654H, BWN203695H, BWN203340H, BWN203343H, BWN203601H, BWN203336H, BWN203406H, BWN203178H, BWN203196H, BWN203696H, BWN203429H, BWN203378H, BWN203185H, BWN203562H, BWN203223H, BWN203641H, BWN203718H, BWN203319H, BWN203508H, BWN203603H, BWN203692H, BWN203777H, BWN203800H, BWN203400H, BWN203627H, BWN203208H, BWN203224H, BWN203179H, BWN203482H, BWN203418H, BWN203375H, BWN203658H, BWN203367H, BWN203701H, BWN203301H, BWN203445H, BWN203234H, BWN203478H, BWN203672H, BWN203673H, BWN203725H, BWN203757H, BWN203263H, BWN203500H, BWN203298H, BWN203467H, BWN203778H, BWN203546H, BWN203795H, BWN203682H, BWN203605H, BWN203242H, BWN203335H, BWN203170H, BWN203469H, BWN203804H, BWN203190H, BWN203278H, BWN203700H, BWN203743H, BWN203748H, BWN203762H, BWN203766H, BWN203322H, BWN203437H, BWN203438H, BWN203439H, BWN203204H, BWN203172H, BWN203182H, BWN203314H, BWN203324H, BWN203402H, BWN203529H, BWN203566H, BWN203592H, BWN203594H, BWN203595H, BWN203705H, BWN203734H, BWN203441H, BWN203444H, BWN203717H, BWN203194H, BWN203483H, BWN203527H, BWN203577H, BWN203662H, BWN203499H, BWN203217H, BWN203736H, BWN203545H, BWN203266H, BWN202560H, BWN203389H, BWN203589H, BWN203659H, BWN203670H, BWN203166H, BWN203591H, BWN203779H, BWN203465H, BWN203614H, BWN203661H, BWN203198H, BWN203554H, BWN203219H, BWN203617H, BWN203551H, BWN203312H, BWN203792H, BWN203515H, BWN203390H, BWN203189H, BWN203462H, BWN203241H, BWN203704H, BWN203328H, BWN203305H, BWN203323H, BWN203568H, BWN203557H, BWN203384H, BWN203164H, BWN203333H, BWN203519H, BWN203707H, BWN203244H, BWN203307H, BWN203547H, BWN203421H, BWN203377H, BWN203501H, BWN203227H, BWN203764H, BWN203721H, BWN203220H, BWN203724H, BWN203184H, BWN203177H, BWN203245H, BWN203262H, BWN203542H, BWN203613H, BWN203649H, BWN203699H, BWN203173H, BWN203491H, BWN203295H, BWN203756H, BWN203338H, BWN203514H, BWN203161H, BWN203396H, BWN203647H, BWN203506H, BWN203507H, BWN203353H, BWN203715H, BWN203249H, BWN203723H, BWN203485H, BWN203572H, BWN203646H, BWN203678H, BWN203496H, BWN203240H, BWN203560H
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
- Reason for Recall:
- There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
ICD-VR DVAB1D1 VISIA AF US IS1/DF1, Model Number DVAB1D1; Implantable Cardioverter Defibrillators
Product Codes/Lot Numbers:
GTIN 00643169720619, Lot Serial Numbers: BWN203363H, BWN203588H, BWN203428H, BWN203504H, BWN203530H, BWN203214H, BWN203258H, BWN203259H, BWN203304H, BWN203477H, BWN203665H, BWN203423H, BWN203403H, BWN203327H, BWN203518H, BWN203790H, BWN203174H, BWN203624H, BWN203750H, BWN203763H, BWN203786H, BWN203745H, BWN203526H, BWN203813H, BWN203168H, BWN203236H, BWN203374H, BWN203463H, BWN203640H, BWN203213H, BWN203431H, BWN203492H, BWN203585H, BWN203436H, BWN203524H, BWN203528H, BWN203809H, BWN203432H, BWN203231H, BWN203199H, BWN203163H, BWN203193H, BWN203306H, BWN203310H, BWN203352H, BWN203373H, BWN203433H, BWN203575H, BWN203611H, BWN203616H, BWN203689H, BWN203732H, BWN203739H, BWN203749H, BWN203796H, BWN203799H, BWN203803H, BWN203181H, BWN203642H, BWN203202H, BWN203758H, BWN203671H, BWN203769H, BWN203357H, BWN203248H, BWN203253H, BWN203448H, BWN203664H, BWN203733H, BWN203260H, BWN203347H, BWN203408H, BWN203422H, BWN203535H, BWN203582H, BWN203712H, BWN203770H, BWN203697H, BWN203205H, BWN203410H, BWN203590H, BWN203332H, BWN203381H, BWN203430H, BWN203511H, BWN203737H, BWN203355H, BWN203703H, BWN203625H, BWN203380H, BWN203498H, BWN203620H, BWN203622H, BWN203810H, BWN203681H, BWN203685H, BWN203434H, BWN203653H, BWN203564H, BWN203169H, BWN203188H, BWN203186H, BWN203318H, BWN203325H, BWN203544H, BWN203468H, BWN203727H, BWN203283H, BWN203693H, BWN203667H, BWN203180H, BWN203660H, BWN203584H, BWN203386H, BWN203806H, BWN203543H, BWN203417H, BWN203395H, BWN203549H, BWN203279H, BWN203326H, BWN203706H, BWN203726H, BWN203744H, BWN203583H, BWN203365H, BWN203556H, BWN203709H, BWN203730H, BWN203203H, BWN203443H, BWN203243H, BWN203331H, BWN203315H, BWN203419H, BWN203643H, BWN203760H, BWN203520H, BWN203607H, BWN203710H, BWN203540H, BWN203265H, BWN203293H, BWN203299H, BWN203517H, BWN203354H, BWN203371H, BWN203398H, BWN203273H, BWN203369H, BWN203450H, BWN203474H, BWN203694H, BWN203774H, BWN203288H, BWN203316H, BWN203487H, BWN203538H, BWN203553H, BWN203555H, BWN203618H, BWN203294H, BWN203690H, BWN203531H, BWN203159H, BWN203731H, BWN203787H, BWN203254H, BWN203480H, BWN203599H, BWN203157H, BWN203246H, BWN203581H, BWN203776H, BWN203211H, BWN203451H, BWN203789H, BWN203233H, BWN203356H, BWN203808H, BWN203212H, BWN203191H, BWN203447H, BWN203473H, BWN203413H, BWN203414H, BWN203415H, BWN203416H, BWN203486H, BWN203489H, BWN203516H, BWN203688H, BWN203815H, BWN203235H, BWN203648H, BWN203167H, BWN203702H, BWN203290H, BWN203296H, BWN203183H, BWN203781H, BWN203284H, BWN203635H, BWN203456H, BWN203472H, BWN203580H, BWN203691H, BWN203794H, BWN203382H, BWN203229H, BWN203247H, BWN203256H, BWN203264H, BWN203274H, BWN203277H, BWN203302H, BWN203532H, BWN202561H, BWN203228H, BWN203250H, BWN203268H, BWN203269H, BWN203270H, BWN203275H, BWN203292H, BWN203420H, BWN203440H, BWN203509H, BWN203548H, BWN203615H, BWN203656H, BWN203680H, BWN203716H, BWN203722H, BWN203771H, BWN203782H, BWN203404H, BWN203657H, BWN203788H, BWN203192H, BWN203397H, BWN203461H, BWN203513H, BWN203215H, BWN203578H, BWN203720H, BWN203747H, BWN203773H, BWN203683H, BWN203409H, BWN203597H, BWN203539H, BWN203351H, BWN203334H, BWN203639H, BWN203714H, BWN203728H, BWN203654H, BWN203695H, BWN203340H, BWN203343H, BWN203601H, BWN203336H, BWN203406H, BWN203178H, BWN203196H, BWN203696H, BWN203429H, BWN203378H, BWN203185H, BWN203562H, BWN203223H, BWN203641H, BWN203718H, BWN203319H, BWN203508H, BWN203603H, BWN203692H, BWN203777H, BWN203800H, BWN203400H, BWN203627H, BWN203208H, BWN203224H, BWN203179H, BWN203482H, BWN203418H, BWN203375H, BWN203658H, BWN203367H, BWN203701H, BWN203301H, BWN203445H, BWN203234H, BWN203478H, BWN203672H, BWN203673H, BWN203725H, BWN203757H, BWN203263H, BWN203500H, BWN203298H, BWN203467H, BWN203778H, BWN203546H, BWN203795H, BWN203682H, BWN203605H, BWN203242H, BWN203335H, BWN203170H, BWN203469H, BWN203804H, BWN203190H, BWN203278H, BWN203700H, BWN203743H, BWN203748H, BWN203762H, BWN203766H, BWN203322H, BWN203437H, BWN203438H, BWN203439H, BWN203204H, BWN203172H, BWN203182H, BWN203314H, BWN203324H, BWN203402H, BWN203529H, BWN203566H, BWN203592H, BWN203594H, BWN203595H, BWN203705H, BWN203734H, BWN203441H, BWN203444H, BWN203717H, BWN203194H, BWN203483H, BWN203527H, BWN203577H, BWN203662H, BWN203499H, BWN203217H, BWN203736H, BWN203545H, BWN203266H, BWN202560H, BWN203389H, BWN203589H, BWN203659H, BWN203670H, BWN203166H, BWN203591H, BWN203779H, BWN203465H, BWN203614H, BWN203661H, BWN203198H, BWN203554H, BWN203219H, BWN203617H, BWN203551H, BWN203312H, BWN203792H, BWN203515H, BWN203390H, BWN203189H, BWN203462H, BWN203241H, BWN203704H, BWN203328H, BWN203305H, BWN203323H, BWN203568H, BWN203557H, BWN203384H, BWN203164H, BWN203333H, BWN203519H, BWN203707H, BWN203244H, BWN203307H, BWN203547H, BWN203421H, BWN203377H, BWN203501H, BWN203227H, BWN203764H, BWN203721H, BWN203220H, BWN203724H, BWN203184H, BWN203177H, BWN203245H, BWN203262H, BWN203542H, BWN203613H, BWN203649H, BWN203699H, BWN203173H, BWN203491H, BWN203295H, BWN203756H, BWN203338H, BWN203514H, BWN203161H, BWN203396H, BWN203647H, BWN203506H, BWN203507H, BWN203353H, BWN203715H, BWN203249H, BWN203723H, BWN203485H, BWN203572H, BWN203646H, BWN203678H, BWN203496H, BWN203240H, BWN203560H
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1789-2023
Related Recalls
ICD-DR DDBB2D1 EVERA XT IS1/DF1 INTL, Model Number DDBB2D1; Implantable Cardioverter Defibrillators
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.
ICD-DR DDMC3D1 EVERA MRI S OUS/US DF1, Model Number DDMC3D1; Implantable Cardioverter Defibrillators
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.
ICD DDPA2D1G COBALT XT DR MRI DF1 GOLD, Model Number DDPA2D1G; Implantable Cardioverter Defibrillators
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.