Temperature Sensing 100% Silicone Foley Catheter, Part Number 102201101663MD - Product Usage: is a 3 lumen silicone catheter with a sensor to measure temperature while draining the urinary bladder.
Class I - DangerousWhat Should You Do?
- Check if you have this product: Lot Numbers: V18007422 V18010291 V18023106 V18023107 V18023108 V18023109 V18025115 V18025117 V18025118 V18025119 V18026073 V18037032 V18026072 V18035228 V18035229 V18036131 V18037031 V18037033 V18037938 V19001678 V19001679 V19002478 V19002479 V19002480 V19013333 V19025876 V19026564 V19027234 V19027235 V19016726
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Degania Medical Devices Pvt. Ltd.
- Reason for Recall:
- Inadequate incoming inspection practices were utilized on thermistor sensors used in the production of the specified lots of temperature sensing catheters.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Temperature Sensing 100% Silicone Foley Catheter, Part Number 102201101663MD - Product Usage: is a 3 lumen silicone catheter with a sensor to measure temperature while draining the urinary bladder.
Product Codes/Lot Numbers:
Lot Numbers: V18007422 V18010291 V18023106 V18023107 V18023108 V18023109 V18025115 V18025117 V18025118 V18025119 V18026073 V18037032 V18026072 V18035228 V18035229 V18036131 V18037031 V18037033 V18037938 V19001678 V19001679 V19002478 V19002479 V19002480 V19013333 V19025876 V19026564 V19027234 V19027235 V19016726
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1783-2020
Related Recalls
Temperature Sensing 100% Silicone Foley Catheter, Part Number 102201100880MD - Product Usage: is a 3 lumen silicone catheter with a sensor to measure temperature while draining the urinary bladder.
Degania Medical Devices Pvt.
Inadequate incoming inspection practices were utilized on thermistor sensors used in the production of the specified lots of temperature sensing catheters.
Temperature Sensing 100% Silicone Foley Catheter, Part Number 102201101063MD - Product Usage: is a 3 lumen silicone catheter with a sensor to measure temperature while draining the urinary bladder.
Degania Medical Devices Pvt.
Inadequate incoming inspection practices were utilized on thermistor sensors used in the production of the specified lots of temperature sensing catheters.
Temperature Sensing 100% Silicone Foley Catheter, Part Number 102201101080MD - Product Usage: is a 3 lumen silicone catheter with a sensor to measure temperature while draining the urinary bladder.
Degania Medical Devices Pvt.
Inadequate incoming inspection practices were utilized on thermistor sensors used in the production of the specified lots of temperature sensing catheters.