Welch Allyn ProBP 2400 Digital Blood Pressure Device #2400, REF 901096, Rx ONLY, Made in China --- Distributed by Welch Allyn Inc., 4341 State Street Road, Skaneateles Falls, NY 13153 --- Microlife Corporation 9F, 431, RuiGuang Road, NeiHu, Taipei, 11492, Taiwan, R.O.C. --- Microlife AG Espenstrasse 139 9443 Widnau, Switzerland
Class I - DangerousWhat Should You Do?
- Check if you have this product: Lot Number 0715: Serial Number range (21)07150001 - (21)07150620; Lot Number 1215: Serial Number range (21)12150001 - (21)12150500; --- Note: (21) is NOT part of the serial number. It is the Global Unique Device Identification (GUDI) Application Identifier that precedes the serial number on any GUDI compliant label.
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Welch Allyn Inc
- Reason for Recall:
- A defective electrical component in the battery charging circuit in two lots of the ProBP 2400 may potentially cause an over-voltage battery charging condition when the device is connected to the external power supply/battery charger. This potential defect resides with the device, not the battery. The over-voltage condition can result in high NiMH battery temperatures that, in some cases, can reach levels sufficient to cause melting of the plastic battery door and other adjacent plastic and foam device components.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Welch Allyn ProBP 2400 Digital Blood Pressure Device #2400, REF 901096, Rx ONLY, Made in China --- Distributed by Welch Allyn Inc., 4341 State Street Road, Skaneateles Falls, NY 13153 --- Microlife Corporation 9F, 431, RuiGuang Road, NeiHu, Taipei, 11492, Taiwan, R.O.C. --- Microlife AG Espenstrasse 139 9443 Widnau, Switzerland
Product Codes/Lot Numbers:
Lot Number 0715: Serial Number range (21)07150001 - (21)07150620; Lot Number 1215: Serial Number range (21)12150001 - (21)12150500; --- Note: (21) is NOT part of the serial number. It is the Global Unique Device Identification (GUDI) Application Identifier that precedes the serial number on any GUDI compliant label.
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1777-2017
Related Recalls
If the device is exposed to fluid ingress and is used before the cleaning fluid has completely dried, there is a risk of the device overheating potentially causing a burn to the user or the ear canal of the patient
Potential Risk of fluid ingress to the power supplies, resulting in risk of an internal electrical short and possible electrical shock
Potential Risk of fluid ingress to the power supplies, resulting in risk of an internal electrical short and possible electrical shock