InterStim(TM) System, Model Numbers: a) TH90G01 b) TH90GFA c) TH90G02 d) TH90G03 Product Usage: The Medtronic Model A510 Clinician application (app) is intended for use with the HH90 Handset and TM90 Communicator to program, adjust, and troubleshoot the Medtronic Models 3023 and 3058 InterStim neurostimulators for sacral neuromodulation therapy. The clinician uses the Clinician app to program settings for the patient. The A510 Clinician app, HH90 Handset, TM90 Communicator along with the A520 Patient app are only sold as a kit (TH90).
Class I - DangerousWhat Should You Do?
- Check if you have this product: Model Numbers/UDI: a) TH90G01/00763000058005 b) TH90GFA/00763000187231 c) TH90G02/00763000192259, 00763000192266, 00763000192273, 00763000192280, 00763000192297, 00763000192303, 00763000192310 d) TH90G03/00763000192310 ALL LOT/SERIAL NUMBERS
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Medtronic Neuromodulation
- Reason for Recall:
- There is a potential for an unexpected increase in stimulation during InterStim programming with the A10 Clinician Application (on Medtronic's smart programmer).
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
InterStim(TM) System, Model Numbers: a) TH90G01 b) TH90GFA c) TH90G02 d) TH90G03 Product Usage: The Medtronic Model A510 Clinician application (app) is intended for use with the HH90 Handset and TM90 Communicator to program, adjust, and troubleshoot the Medtronic Models 3023 and 3058 InterStim neurostimulators for sacral neuromodulation therapy. The clinician uses the Clinician app to program settings for the patient. The A510 Clinician app, HH90 Handset, TM90 Communicator along with the A520 Patient app are only sold as a kit (TH90).
Product Codes/Lot Numbers:
Model Numbers/UDI: a) TH90G01/00763000058005 b) TH90GFA/00763000187231 c) TH90G02/00763000192259, 00763000192266, 00763000192273, 00763000192280, 00763000192297, 00763000192303, 00763000192310 d) TH90G03/00763000192310 ALL LOT/SERIAL NUMBERS
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1774-2019
Related Recalls
Enhanced Verify Evaluation Handset (CFN HH90130FA)
Medtronic Neuromodulation
Evaluation handsets may not be able to communicate with the neurostimulator due to handsets having a faster processing speed than prior handsets. An inability to connect may cause the therapy to stop being delivered and a system error will be displayed. Patient may experience return of underlying disease symptoms (e.g. overactive bladder, urinary retention, or fecal incontinence) based on their response to stimulation during the evaluation period.
Stimulation RC Clinician Programmer Application, Model A71400, when used with the Inceptiv (Model 977119) and Intellis Pro (Model 977118) neurostimulators
Medtronic Neuromodulation
There is a software issue that can permanently disable communication with an implantable neurostimulator (INS) when a specific intraoperative programming sequence is used.
A820 myPTM Software Application associated with Medtronic SynchroMed Pump and Infusion System.
Medtronic Neuromodulation
Product complaints were received describing the A820 myPTM app taking longer than expected for patients to interact with their implantable pump.