AquaC UNO H [120V, 60Hz, 1000VA, US-Version - a reverse osmosis unit intended for use with hemodialysis systems. Fresenius Part Number: 24-10VA-1 Vivonic Part Number: G02000970-US

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    UDI: (01)4251285200234(11)210628(21)1CHS5379 (01)4251285200234(11)160927(21)6CHS1270 (01)4251285200234(11)161109(21)6CHS1394 (01)4251285200234(11)161109(21)6CHS1435 (01)4251285200234(11)170131(21)7CHS1569 (01)4251285200234(11)170201(21)7CHS1583 (01)4251285200234(11)170213(21)7CHS1619 (01)4251285200234(11)170314(21)7CHS1641 (01)4251285200234(11)170519(21)7CHS1790 (01)4251285200234(11)170807(21)7CHS1953 (01)4251285200234(11)180326(21)8CHS2197 (01)4251285200234(11)181010(21)8CHS2608 (01)4251285200234(11)181102(21)8CHS2633 (01)4251285200234(11)181107(21)8CHS2644 Serial Numbers: 1CHS5379 6CHS1192 6CHS1194 6CHS1195 6CHS1196 6CHS1197 6CHS1198 6CHS1199 6CHS1200 6CHS1201 6CHS1233 6CHS1270 6CHS1394 6CHS1435 7CHS1569 7CHS1583 7CHS1619 7CHS1641 7CHS1790 7CHS1953 8CHS2197 8CHS2608 8CHS2633 8CHS2644
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Fresenius Medical Care Holdings, Inc.
Reason for Recall:
Internal ethemet cables are missing one or both ferrite cores therefore the electromagnetic compatibility of the device is not ensured
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
ongoing

Product Information

Full Description:

AquaC UNO H [120V, 60Hz, 1000VA, US-Version - a reverse osmosis unit intended for use with hemodialysis systems. Fresenius Part Number: 24-10VA-1 Vivonic Part Number: G02000970-US

Product Codes/Lot Numbers:

UDI: (01)4251285200234(11)210628(21)1CHS5379 (01)4251285200234(11)160927(21)6CHS1270 (01)4251285200234(11)161109(21)6CHS1394 (01)4251285200234(11)161109(21)6CHS1435 (01)4251285200234(11)170131(21)7CHS1569 (01)4251285200234(11)170201(21)7CHS1583 (01)4251285200234(11)170213(21)7CHS1619 (01)4251285200234(11)170314(21)7CHS1641 (01)4251285200234(11)170519(21)7CHS1790 (01)4251285200234(11)170807(21)7CHS1953 (01)4251285200234(11)180326(21)8CHS2197 (01)4251285200234(11)181010(21)8CHS2608 (01)4251285200234(11)181102(21)8CHS2633 (01)4251285200234(11)181107(21)8CHS2644 Serial Numbers: 1CHS5379 6CHS1192 6CHS1194 6CHS1195 6CHS1196 6CHS1197 6CHS1198 6CHS1199 6CHS1200 6CHS1201 6CHS1233 6CHS1270 6CHS1394 6CHS1435 7CHS1569 7CHS1583 7CHS1619 7CHS1641 7CHS1790 7CHS1953 8CHS2197 8CHS2608 8CHS2633 8CHS2644

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-1769-2022