CRTD DTMB2QQ AMPLIA MRI QUAD OUS DF4, Model Number DTMB2QQ; Implantable Cardioverter Defibrillators
Class I - DangerousWhat Should You Do?
- Check if you have this product: GTIN 00643169720695, Lot Serial numbers: RPG619037S, RPG608482S, RPG611787S, RPG619305S, RPG609628S, RPG622247S, RPG619130S, RPG623603S, RPG610818S, RPG616992S, RPG619816S, RPG624535S, RPG614089S, RPG619704S, RPG619714S, RPG623002S, RPG624602S, RPG607260S, RPG607434S, RPG607977S, RPG608232S, RPG608831S, RPG609415S, RPG609836S, RPG609911S, RPG610009S, RPG610345S, RPG611061S, RPG611302S, RPG611328S, RPG611598S, RPG611600S, RPG611606S, RPG611666S, RPG611667S, RPG611668S, RPG612610S, RPG612884S, RPG612885S, RPG613206S, RPG613239S, RPG613578S, RPG613655S, RPG614353S, RPG614407S, RPG614439S, RPG614898S, RPG615028S, RPG617792S, RPG617839S, RPG618125S, RPG618648S, RPG618809S, RPG618811S, RPG619226S, RPG619228S, RPG619555S, RPG619776S, RPG619888S, RPG619890S, RPG619891S, RPG620540S, RPG620563S, RPG621028S, RPG621106S, RPG621190S, RPG621228S, RPG621272S, RPG621395S, RPG621680S, RPG621722S, RPG621724S, RPG621725S, RPG621730S, RPG621743S, RPG622038S, RPG622041S, RPG622085S, RPG622532S, RPG622533S, RPG622881S, RPG622905S, RPG622984S, RPG623287S, RPG623393S, RPG623681S, RPG623684S, RPG623693S, RPG624227S, RPG624640S, RPG624641S, RPG624699S, RPG624702S, RPG625103S, RPG625298S, RPG625299S, RPG625307S, RPG625310S, RPG625343S, RPG625704S, RPG625915S, RPG625919S, RPG626152S, RPG626157S, RPG626198S, RPG626210S, RPG626396S, RPG626399S, RPG626402S, RPG626560S, RPG626605S, RPG626606S, RPG618474S, RPG618477S, RPG618499S, RPG618602S, RPG618603S, RPG618611S, RPG618616S, RPG618617S, RPG618618S, RPG618619S, RPG618620S, RPG618621S, RPG618623S, RPG618624S, RPG618625S, RPG618626S, RPG618636S, RPG618637S, RPG618639S, RPG618640S, RPG618642S, RPG618643S, RPG618646S, RPG618647S, RPG620107S, RPG620115S, RPG620118S, RPG620120S, RPG620122S, RPG620160S, RPG620165S, RPG620166S, RPG620171S, RPG620177S, RPG620179S, RPG620185S, RPG620186S, RPG620189S, RPG620194S, RPG620200S, RPG620202S, RPG620213S, RPG620258S, RPG620270S, RPG620272S, RPG623171S, RPG623176S, RPG623180S, RPG623183S, RPG623184S, RPG623193S, RPG623197S, RPG623199S, RPG623200S, RPG623202S, RPG623204S, RPG623209S, RPG623213S, RPG623214S, RPG623215S, RPG623221S, RPG623228S, RPG623236S, RPG623247S, RPG623251S, RPG623253S, RPG625563S, RPG625608S, RPG625609S, RPG625616S, RPG625617S, RPG625621S, RPG625623S, RPG625624S, RPG625625S, RPG625630S, RPG625645S, RPG625650S, RPG625662S, RPG625673S, RPG625682S, RPG625703S, RPG625714S, RPG625715S, RPG625716S, RPG625736S, RPG625738S, RPG626407S, RPG626416S, RPG626417S, RPG626421S, RPG626423S, RPG626425S, RPG626426S, RPG626433S, RPG626435S, RPG626437S, RPG626441S, RPG626442S, RPG626443S, RPG626444S, RPG626450S, RPG626451S, RPG626453S, RPG626454S, RPG626468S, RPG626472S, RPG626475S, RPG626483S, RPG626487S, RPG626492S, RPG626502S, RPG626504S, RPG626505S, RPG626509S, RPG626516S, RPG626519S, RPG626521S, RPG626523S, RPG626525S, RPG626527S, RPG626528S, RPG626529S, RPG626531S, RPG626532S, RPG626535S, RPG626540S, RPG611844S, RPG611846S, RPG613809S, RPG616044S, RPG616046S, RPG616384S, RPG616764S, RPG623927S, RPG624385S, RPG608667S, RPG608668S, RPG608669S, RPG609041S, RPG609233S, RPG619334S, RPG619360S, RPG611019S, RPG611036S, RPG611894S, RPG614000S, RPG614199S, RPG614399S, RPG614560S, RPG614576S, RPG614934S, RPG615091S, RPG615105S, RPG616049S, RPG616438S, RPG617449S, RPG618128S, RPG618285S, RPG619117S, RPG619207S, RPG622790S, RPG622794S, RPG622800S, RPG622974S, RPG623018S, RPG623024S, RPG623772S, RPG623976S, RPG626039S, RPG626603S, RPG611022S, RPG611023S, RPG612375S, RPG612395S, RPG612888S, RPG612955S, RPG614106S, RPG614181S, RPG614184S, RPG614189S, RPG614191S, RPG614195S, RPG614202S, RPG614238S, RPG614241S, RPG614245S, RPG614401S, RPG614591S, RPG614593S, RPG614601S, RPG615087S, RPG615568S, RPG616022S, RPG616028S, RPG616037S, RPG616077S, RPG616321S, RPG616800S, RPG618065S, RPG618346S, RPG618790S, RPG619779S, RPG619785S, RPG619895S, RPG620126S, RPG620129S, RPG620130S, RPG622061S, RPG622796S, RPG622798S, RPG622801S, RPG622968S, RPG623112S, RPG623782S, RPG626081S, RPG626086S, RPG626091S, RPG613261S, RPG615706S, RPG616001S, RPG618928S, RPG622092S, RPG623088S, RPG623107S, RPG626006S, RPG615704S, RPG622267S, RPG626014S, RPG612037S, RPG612039S, RPG612042S, RPG612048S, RPG612061S, RPG612066S, RPG612116S, RPG612976S, RPG613251S, RPG615692S, RPG615707S, RPG615708S, RPG615709S, RPG615716S, RPG615738S, RPG616493S, RPG618849S, RPG618850S, RPG618897S, RPG618898S, RPG618905S, RPG618955S, RPG619701S, RPG620275S, RPG620282S, RPG620296S, RPG620327S, RPG620349S, RPG620383S, RPG620435S, RPG620466S, RPG620481S, RPG620492S, RPG620500S, RPG620636S, RPG621779S, RPG621802S, RPG621805S, RPG622149S, RPG622239S, RPG622995S, RPG623089S, RPG623093S, RPG623096S, RPG623104S, RPG623105S, RPG623296S, RPG623585S, RPG624013S, RPG624014S, RPG624025S, RPG624151S, RPG624162S, RPG612089S, RPG614070S, RPG616323S, RPG616445S, RPG616485S, RPG620380S, RPG620382S, RPG624021S, RPG611998S, RPG612100S, RPG613790S, RPG615765S, RPG616459S, RPG620350S, RPG621377S, RPG621681S, RPG621763S, RPG621882S, RPG624012S, RPG612421S, RPG613058S, RPG613263S, RPG614040S, RPG618972S, RPG620335S, RPG621786S, RPG623090S, RPG623584S, RPG612045S, RPG616042S, RPG616446S, RPG616449S, RPG617672S, RPG618963S, RPG620389S, RPG620497S, RPG622242S, RPG623085S, RPG624246S, RPG626011S, RPG618812S, RPG618907S, RPG618913S, RPG620215S, RPG620295S, RPG620359S, RPG620424S, RPG621773S, RPG621787S, RPG622258S, RPG622743S, RPG624282S, RPG611991S, RPG618892S, RPG620338S, RPG620361S, RPG621767S, RPG615711S, RPG616487S, RPG620443S, RPG622084S, RPG624167S, RPG613976S, RPG614013S, RPG615705S, RPG616452S, RPG616814S, RPG618855S, RPG618938S, RPG619696S, RPG615710S, RPG616817S, RPG618374S, RPG618791S, RPG618977S, RPG620346S, RPG620348S, RPG620351S, RPG621794S, RPG622246S, RPG623099S, RPG611974S, RPG618902S, RPG621800S, RPG621885S, RPG622240S, RPG622249S, RPG622263S, RPG619286S, RPG619585S, RPG620238S, RPG620247S, RPG620362S, RPG620376S, RPG620379S, RPG622742S, RPG622744S, RPG623472S, RPG623481S, RPG623557S, RPG624234S, RPG624251S, RPG618931S, RPG618943S, RPG618945S, RPG618947S, RPG623579S, RPG624249S, RPG616325S, RPG616349S, RPG616492S, RPG622082S, RPG623095S, RPG611995S, RPG618954S, RPG622253S, RPG624247S, RPG617674S, RPG617700S, RPG618941S, RPG619271S, RPG619281S, RPG619282S, RPG620375S, RPG622257S, RPG622260S, RPG622262S, RPG622270S, RPG622271S, RPG622513S, RPG622735S, RPG623297S, RPG623299S, RPG623456S, RPG623586S, RPG623720S, RPG623722S, RPG623723S, RPG624023S, RPG624028S, RPG624032S, RPG624225S, RPG619248S, RPG626280S, RPG607290S, RPG607449S, RPG607450S, RPG609351S, RPG609352S, RPG613090S, RPG613246S, RPG618649S, RPG619015S, RPG619344S, RPG620574S, RPG621332S, RPG624409S, RPG607198S, RPG615946S, RPG619245S, RPG619848S, RPG619920S, RPG620413S, RPG620797S, RPG622187S, RPG622330S, RPG607228S, RPG607233S, RPG607275S, RPG607346S, RPG607926S, RPG608143S, RPG608161S, RPG608957S, RPG609319S, RPG609509S, RPG609538S, RPG609614S, RPG611681S, RPG612356S, RPG613010S, RPG613253S, RPG613313S, RPG613921S, RPG614090S, RPG614329S, RPG614674S, RPG615144S, RPG616054S, RPG616636S, RPG616662S, RPG616706S, RPG617462S, RPG617609S, RPG618794S, RPG618795S, RPG619149S, RPG619709S, RPG619878S, RPG619924S, RPG620039S, RPG621928S, RPG622298S, RPG622322S, RPG622456S, RPG624404S, RPG625095S, RPG625098S, RPG625331S, RPG626331S, RPG626470S, RPG610095S, RPG614654S, RPG615936S, RPG617198S, RPG617633S, RPG618445S, RPG620816S, RPG622261S, RPG614544S, RPG615827S, RPG620109S, RPG621570S, RPG621972S, RPG622277S, RPG624205S, RPG626332S, RPG607345S, RPG607574S, RPG607628S, RPG607995S, RPG608135S, RPG608419S, RPG608624S, RPG608652S, RPG609367S, RPG609625S, RPG609990S, RPG610687S, RPG610695S, RPG610749S, RPG611366S, RPG611457S, RPG611544S, RPG612598S, RPG612777S, RPG612779S, RPG612984S, RPG613105S, RPG613357S, RPG613581S, RPG613703S, RPG613960S, RPG614007S, RPG614330S, RPG614540S, RPG614656S, RPG615032S, RPG615206S, RPG615875S, RPG615934S, RPG618415S, RPG619682S, RPG619705S, RPG619828S, RPG619876S, RPG619929S, RPG619965S, RPG620252S, RPG620777S, RPG620794S, RPG621391S, RPG621405S, RPG621887S, RPG622114S, RPG622140S, RPG623217S, RPG623384S, RPG623435S, RPG623710S, RPG624049S, RPG624258S, RPG624599S, RPG624934S, RPG625463S, RPG625737S, RPG625800S, RPG625809S, RPG625926S, RPG626247S, RPG626260S, RPG626292S, RPG626350S, RPG626352S, RPG626491S, RPG626614S, RPG607551S, RPG607196S, RPG608516S, RPG608956S, RPG609200S, RPG610250S, RPG611333S, RPG612578S, RPG614691S, RPG615139S, RPG615943S, RPG618264S, RPG619012S, RPG619208S, RPG619449S, RPG619608S, RPG619609S, RPG619792S, RPG622730S, RPG622829S, RPG622924S, RPG623106S, RPG623521S, RPG624882S, RPG625043S, RPG625053S, RPG625285S, RPG626132S, RPG621701S, RPG622336S, RPG622628S, RPG623078S, RPG623350S, RPG623512S, RPG623908S, RPG624635S, RPG624877S, RPG625025S, RPG625351S, RPG625585S, RPG625797S, RPG626209S, RPG626348S, RPG626356S, RPG626650S, RPG607626S, RPG607927S, RPG608136S, RPG608287S, RPG608779S, RPG609366S, RPG609945S, RPG610130S, RPG610256S, RPG610907S, RPG610912S, RPG611413S, RPG611985S, RPG612579S, RPG613600S, RPG613623S, RPG613789S, RPG613933S, RPG614318S, RPG614352S, RPG614403S, RPG615080S, RPG615376S, RPG615876S, RPG615938S, RPG616053S, RPG616056S, RPG616584S, RPG616853S, RPG617644S, RPG617720S, RPG618692S, RPG618781S, RPG619329S, RPG620251S, RPG620423S, RPG620462S, RPG621017S, RPG621616S, RPG621775S, RPG621913S, RPG622319S, RPG622323S, RPG622380S, RPG622830S, RPG622902S, RPG622966S, RPG623207S, RPG623312S, RPG623375S, RPG623394S, RPG623396S, RPG623398S, RPG623870S, RPG624260S, RPG624342S, RPG624548S, RPG625042S, RPG625498S, RPG625765S, RPG625783S, RPG625801S, RPG625917S, RPG626455S, RPG607274S, RPG626488S, RPG607255S, RPG610481S, RPG610906S, RPG610997S, RPG611191S, RPG612005S, RPG612022S, RPG616408S, RPG617180S, RPG617528S, RPG617833S, RPG618449S, RPG618450S, RPG618513S, RPG619405S, RPG620964S, RPG621659S, RPG621662S, RPG621832S, RPG621997S, RPG622286S, RPG626408S, RPG626661S, RPG612574S, RPG612809S, RPG616266S, RPG614290S, RPG621822S, RPG617620S, RPG614676S, RPG618784S, RPG622374S, RPG608203S, RPG612059S, RPG612155S, RPG612160S, RPG613221S, RPG613579S, RPG614381S, RPG614464S, RPG614637S, RPG620793S, RPG621702S, RPG622269S, RPG623565S, RPG623907S, RPG608237S, RPG608801S, RPG621611S, RPG617259S, RPG619840S, RPG620019S, RPG617253S, RPG613615S, RPG615994S, RPG620583S, RPG620603S, RPG620609S, RPG607675S, RPG607208S, RPG607510S, RPG607868S, RPG609682S, RPG610018S, RPG610377S, RPG610879S, RPG610926S, RPG611411S, RPG613213S, RPG613734S, RPG614258S, RPG614387S, RPG614431S, RPG614713S, RPG615383S, RPG615400S, RPG616361S, RPG616479S, RPG616789S, RPG616791S, RPG617621S, RPG618349S, RPG618638S, RPG618859S, RPG620567S, RPG620920S, RPG621396S, RPG621782S, RPG622053S, RPG624308S, RPG626143S, RPG626678S, RPG616700S, RPG616701S, RPG617499S, RPG617501S, RPG617502S, RPG607990S, RPG615603S, RPG616236S, RPG618482S, RPG619849S, RPG619862S, RPG625679S, RPG625775S, RPG616176S, RPG616180S, RPG616394S, RPG616395S, RPG617127S, RPG617134S, RPG617137S, RPG617439S, RPG617440S, RPG618048S, RPG618150S, RPG618368S, RPG619024S, RPG619339S, RPG619385S, RPG620075S, RPG620825S, RPG621335S, RPG621419S, RPG621486S, RPG621511S, RPG621548S, RPG621604S, RPG621964S, RPG622004S, RPG622529S, RPG622654S, RPG622655S, RPG622896S, RPG623307S, RPG623690S, RPG623692S, RPG623702S, RPG623711S, RPG623712S, RPG623875S, RPG623944S, RPG624009S, RPG624374S, RPG624551S, RPG624959S, RPG625185S, RPG626139S, RPG626141S, RPG626298S, RPG607288S, RPG609118S, RPG609342S, RPG610507S, RPG611632S, RPG611874S, RPG612647S, RPG612672S, RPG612951S, RPG615215S, RPG616825S, RPG617999S, RPG618508S, RPG619628S, RPG621101S, RPG621238S, RPG622777S, RPG623281S, RPG624262S, RPG624279S, RPG624431S, RPG624505S, RPG624751S, RPG624952S, RPG625598S, RPG625750S, RPG625909S, RPG626431S, RPG623622S, RPG607728S, RPG607230S, RPG607556S, RPG607637S, RPG607930S, RPG608902S, RPG609942S, RPG609960S, RPG611733S, RPG612808S, RPG613355S, RPG613646S, RPG613955S, RPG614256S, RPG614303S, RPG615967S, RPG616195S, RPG617787S, RPG619927S, RPG621606S, RPG624278S, RPG626136S, RPG616194S, RPG616470S, RPG619016S, RPG620568S, RPG620978S, RPG621579S, RPG622276S, RPG614378S, RPG615950S, RPG624958S, RPG620332S, RPG621057S, RPG622452S, RPG623724S, RPG624413S, RPG624547S, RPG626161S, RPG616769S, RPG620827S, RPG625657S, RPG616283S, RPG624209S, RPG611607S, RPG614001S, RPG615398S, RPG616430S, RPG616745S, RPG618538S, RPG618561S, RPG618578S, RPG621601S, RPG621664S, RPG625792S, RPG610704S, RPG615893S, RPG617974S, RPG618514S, RPG618786S, RPG620986S, RPG622396S, RPG622773S, RPG624504S, RPG626118S, RPG626162S, RPG614084S, RPG622441S, RPG615545S, RPG616614S, RPG616836S, RPG617309S, RPG618376S, RPG619781S, RPG620326S, RPG621750S, RPG621966S, RPG622971S, RPG623079S, RPG624271S, RPG625755S, RPG625834S, RPG626199S, RPG626354S, RPG626559S, RPG608714S, RPG609947S, RPG610877S, RPG610920S, RPG611364S, RPG613729S, RPG615025S, RPG616613S, RPG616744S, RPG619138S, RPG619171S, RPG619851S, RPG619937S, RPG619960S, RPG622329S, RPG623865S, RPG623934S, RPG625044S, RPG625543S, RPG625825S, RPG626123S, RPG626200S, RPG626554S, RPG615373S, RPG616182S, RPG616184S, RPG616187S, RPG616650S, RPG616651S, RPG617681S, RPG617682S, RPG620464S, RPG620465S, RPG621865S, RPG621871S, RPG622372S, RPG622373S, RPG622985S, RPG622986S, RPG614771S, RPG614849S, RPG615955S, RPG619020S, RPG619447S, RPG620637S, RPG621720S, RPG625367S, RPG607206S, RPG607635S, RPG608163S, RPG608207S, RPG608679S, RPG608684S, RPG608695S, RPG608727S, RPG608953S, RPG609293S, RPG609570S, RPG609843S, RPG609941S, RPG610153S, RPG610844S, RPG611130S, RPG611482S, RPG612176S, RPG613481S, RPG614640S, RPG615295S, RPG616126S, RPG620592S, RPG620762S, RPG621755S, RPG621784S, RPG621895S, RPG622470S, RPG622942S, RPG623985S, RPG625159S, RPG625788S, RPG626318S, RPG607213S, RPG622542S, RPG623234S, RPG623597S, RPG624083S, RPG625660S, RPG626203S, RPG626218S, RPG607568S, RPG609914S, RPG611175S, RPG611185S, RPG611458S, RPG615184S, RPG616484S, RPG617093S, RPG617553S, RPG622045S, RPG624402S, RPG625108S, RPG625545S, RPG610165S, RPG608206S, RPG608434S, RPG608697S, RPG608699S, RPG610502S, RPG610911S, RPG610913S, RPG610994S, RPG611159S, RPG613007S, RPG614383S, RPG615119S, RPG617638S, RPG618267S, RPG618793S, RPG618863S, RPG623940S, RPG625643S, RPG613364S, RPG614289S, RPG614613S, RPG614969S, RPG615605S, RPG615745S, RPG616397S, RPG616410S, RPG617135S, RPG617461S, RPG617527S, RPG618852S, RPG618853S, RPG618997S, RPG619038S, RPG619554S, RPG619603S, RPG619625S, RPG619708S, RPG620601S, RPG620778S, RPG620780S, RPG620959S, RPG621019S, RPG621404S, RPG621557S, RPG621635S, RPG622377S, RPG622455S, RPG622734S, RPG622874S, RPG623233S, RPG623490S, RPG623881S, RPG624891S, RPG625422S, RPG625479S, RPG625752S, RPG625786S, RPG607199S, RPG607223S, RPG607317S, RPG609059S, RPG609119S, RPG609380S, RPG609544S, RPG609810S, RPG610076S, RPG610141S, RPG611626S, RPG611658S, RPG614386S, RPG614525S, RPG615764S, RPG616585S, RPG616652S, RPG617098S, RPG617246S, RPG617267S, RPG618386S, RPG618580S, RPG618841S, RPG619006S, RPG619919S, RPG621632S, RPG621669S, RPG621709S, RPG621765S, RPG622110S, RPG622221S, RPG622359S, RPG622806S, RPG623220S, RPG623771S, RPG624004S, RPG624192S, RPG624268S, RPG624598S, RPG624949S, RPG625239S, RPG626404S, RPG626500S, RPG614354S, RPG616412S, RPG617189S, RPG617563S, RPG617893S, RPG618119S, RPG619031S, RPG619634S, RPG619844S, RPG620467S, RPG620975S, RPG623540S, RPG623991S, RPG625428S, RPG625816S, RPG626664S, RPG614280S, RPG622466S, RPG623861S, RPG625536S, RPG607247S, RPG620977S, RPG621718S, RPG626418S, RPG617316S, RPG618489S, RPG625100S, RPG607604S, RPG612047S, RPG622860S, RPG626252S, RPG607492S, RPG614080S, RPG614730S, RPG617103S, RPG620741S, RPG621876S, RPG621902S, RPG623888S, RPG625073S, RPG625839S, RPG625918S, RPG626641S, RPG618944S, RPG621572S, RPG622119S, RPG622446S, RPG623879S, RPG614997S, RPG615073S, RPG607555S, RPG608562S, RPG608563S, RPG611594S, RPG612597S, RPG612645S, RPG614679S, RPG615735S, RPG615921S, RPG616463S, RPG617707S, RPG617719S, RPG618843S, RPG619218S, RPG619534S, RPG619650S, RPG620204S, RPG620428S, RPG621209S, RPG621607S, RPG621666S, RPG621919S, RPG624434S, RPG626174S, RPG607265S, RPG617139S, RPG624357S, RPG618572S, RPG621408S, RPG622068S, RPG622293S, RPG622891S, RPG623068S, RPG624156S, RPG613812S, RPG615178S, RPG617643S, RPG617704S, RPG617737S, RPG617976S, RPG618067S, RPG618703S, RPG618798S, RPG618857S, RPG619654S, RPG619748S, RPG621059S, RPG621487S, RPG621527S, RPG622535S, RPG622784S, RPG622921S, RPG623428S, RPG623759S, RPG624006S, RPG624399S, RPG624661S, RPG624857S, RPG625550S, RPG625611S, RPG626374S, RPG621956S, RPG623478S, RPG624223S, RPG625811S, RPG626358S, RPG619566S, RPG621340S, RPG621536S, RPG621585S, RPG622304S, RPG623211S, RPG623936S, RPG624186S, RPG624805S, RPG625009S, RPG625646S, RPG625756S, RPG626480S, RPG626566S, RPG626633S, RPG611308S, RPG611934S, RPG613780S, RPG615988S, RPG616401S, RPG616448S, RPG617190S, RPG618754S, RPG619215S, RPG622301S, RPG622344S, RPG622355S, RPG623004S, RPG625653S, RPG608696S, RPG610429S, RPG615483S, RPG614319S, RPG618804S, RPG619150S, RPG619668S, RPG624615S, RPG616574S, RPG609495S, RPG610984S, RPG611944S, RPG614924S, RPG622477S, RPG622989S, RPG625326S, RPG626393S, RPG613086S, RPG622929S, RPG625699S, RPG618472S, RPG618473S, RPG619259S, RPG610019S, RPG612679S, RPG613631S, RPG614712S, RPG619707S, RPG623473S, RPG616499S, RPG617534S, RPG618761S, RPG619233S, RPG609721S, RPG609075S, RPG611027S, RPG611337S, RPG614252S, RPG614878S, RPG618134S, RPG619794S, RPG608118S, RPG610164S, RPG611369S, RPG612572S, RPG612630S, RPG618245S, RPG622485S, RPG615952S, RPG616886S, RPG617281S, RPG617293S, RPG617567S, RPG617598S, RPG617611S, RPG618133S, RPG618351S, RPG618584S, RPG618727S, RPG618813S, RPG619273S, RPG619784S, RPG619952S, RPG620024S, RPG620805S, RPG620822S, RPG620946S, RPG621005S, RPG621322S, RPG621450S, RPG621451S, RPG621466S, RPG621538S, RPG621540S, RPG621576S, RPG621580S, RPG621806S, RPG621810S, RPG621954S, RPG621970S, RPG622033S, RPG622063S, RPG622152S, RPG622212S, RPG622442S, RPG622717S, RPG622771S, RPG622785S, RPG622789S, RPG622791S, RPG624483S, RPG624981S, RPG625011S, RPG625176S, RPG625342S, RPG625444S, RPG625666S, RPG625810S, RPG625910S, RPG625938S, RPG626581S, RPG626666S, RPG608646S, RPG609704S, RPG609970S, RPG612756S, RPG613110S, RPG614630S, RPG624029S, RPG622987S, RPG622988S, RPG617751S, RPG618797S, RPG619702S, RPG620067S, RPG620624S, RPG622007S, RPG607945S, RPG610103S, RPG610968S, RPG611324S, RPG612837S, RPG613173S, RPG613343S, RPG613348S, RPG613788S, RPG617891S, RPG618541S, RPG620456S, RPG620457S, RPG622101S, RPG625685S, RPG611295S, RPG612486S, RPG615992S, RPG616931S, RPG618250S, RPG619943S, RPG620795S, RPG621400S, RPG621592S, RPG622900S, RPG623414S, RPG625145S, RPG625396S, RPG625458S, RPG626248S, RPG626347S, RPG626355S, RPG626366S, RPG607491S, RPG607595S, RPG608246S, RPG608691S, RPG610036S, RPG610482S, RPG610916S, RPG611165S, RPG611661S, RPG611877S, RPG612506S, RPG612952S, RPG613636S, RPG615126S, RPG615601S, RPG615935S, RPG616243S, RPG617157S, RPG617176S, RPG617204S, RPG617312S, RPG618509S, RPG618600S, RPG619165S, RPG620562S, RPG620968S, RPG621899S, RPG622126S, RPG624256S, RPG624736S, RPG625675S, RPG625677S, RPG626440S, RPG626464S, RPG615966S, RPG609339S, RPG611368S, RPG612484S, RPG613843S, RPG608070S, RPG608214S, RPG608221S, RPG610908S, RPG611487S, RPG611615S, RPG612649S, RPG612728S, RPG612931S, RPG613128S, RPG613816S, RPG614687S, RPG616653S, RPG616687S, RPG617177S, RPG617579S, RPG617850S, RPG617935S, RPG619027S, RPG619769S, RPG620222S, RPG620476S, RPG621529S, RPG621825S, RPG622060S, RPG622351S, RPG622354S, RPG625690S, RPG626128S, RPG626144S, RPG626346S, RPG612440S, RPG613793S, RPG614661S, RPG614698S, RPG614723S, RPG616511S, RPG616567S, RPG616830S, RPG617713S, RPG618019S, RPG618732S, RPG621583S, RPG622376S, RPG622379S, RPG622726S, RPG623493S, RPG623905S, RPG624514S, RPG625179S, RPG626556S, RPG626564S, RPG617315S, RPG619157S, RPG619676S, RPG607178S, RPG607764S, RPG607881S, RPG609717S, RPG610874S, RPG612405S, RPG614102S, RPG615742S, RPG618296S, RPG620604S, RPG622248S, RPG624649S, RPG625060S, RPG615009S, RPG610443S, RPG610875S, RPG612425S, RPG612508S, RPG612526S, RPG612532S, RPG612691S, RPG613042S, RPG613140S, RPG613141S, RPG613330S, RPG613471S, RPG613562S, RPG615127S, RPG615422S, RPG615425S, RPG615741S, RPG615890S, RPG616012S, RPG616193S, RPG616228S, RPG616293S, RPG616575S, RPG616717S, RPG617482S, RPG617623S, RPG618117S, RPG618507S, RPG618585S, RPG618782S, RPG619698S, RPG620042S, RPG620381S, RPG621376S, RPG621456S, RPG621637S, RPG621826S, RPG622111S, RPG622278S, RPG622281S, RPG623272S, RPG623516S, RPG623519S, RPG623652S, RPG623895S, RPG623984S, RPG624325S, RPG624401S, RPG625632S, RPG625892S, RPG626255S, RPG608480S, RPG615949S, RPG626652S, RPG607224S, RPG607245S, RPG607383S, RPG607786S, RPG609176S, RPG610692S, RPG611965S, RPG611975S, RPG612054S, RPG613252S, RPG614185S, RPG614428S, RPG614974S, RPG615163S, RPG615751S, RPG616237S, RPG616421S, RPG616440S, RPG617465S, RPG617758S, RPG618004S, RPG619010S, RPG611198S, RPG613093S, RPG618118S, RPG607599S, RPG607940S, RPG608754S, RPG608896S, RPG608961S, RPG613962S, RPG614376S, RPG609965S, RPG610517S, RPG611976S, RPG613340S, RPG617121S, RPG617123S, RPG620432S, RPG621553S, RPG623373S, RPG626215S, RPG614837S, RPG615426S, RPG617101S, RPG618428S, RPG620979S, RPG621689S, RPG622112S, RPG622737S, RPG624665S, RPG625855S, RPG621783S, RPG608951S, RPG613264S, RPG615794S, RPG620169S, RPG621225S, RPG623372S, RPG626386S, RPG620436S, RPG609496S, RPG610005S, RPG610996S, RPG611688S, RPG613970S, RPG614943S, RPG615728S, RPG616601S, RPG617690S, RPG619159S, RPG622026S, RPG622321S, RPG623466S, RPG623967S, RPG624711S, RPG624712S, RPG624862S, RPG624866S, RPG625096S, RPG625097S, RPG626205S, RPG626207S, RPG626226S, RPG626424S, RPG626550S, RPG609964S, RPG615382S, RPG610424S, RPG610494S, RPG616773S, RPG617601S, RPG621764S, RPG622324S, RPG623280S, RPG623524S, RPG626397S, RPG626538S, RPG613373S, RPG614940S, RPG616771S, RPG617329S, RPG611404S, RPG612618S, RPG613967S, RPG622486S, RPG623187S, RPG623864S, RPG610902S, RPG610922S, RPG612519S, RPG618377S, RPG619825S, RPG610275S, RPG617463S, RPG620236S, RPG610016S, RPG610073S, RPG611414S, RPG612555S, RPG612903S, RPG613362S, RPG613716S, RPG614096S, RPG614249S, RPG615822S, RPG615868S, RPG616390S, RPG617328S, RPG618122S, RPG619632S, RPG620062S, RPG623103S, RPG623922S, RPG624418S, RPG625147S, RPG625281S, RPG625627S, RPG626342S, RPG610986S, RPG613722S, RPG610787S, RPG611054S, RPG611055S, RPG611456S, RPG614710S, RPG615132S, RPG615225S, RPG616733S, RPG617136S, RPG617498S, RPG617524S, RPG618257S, RPG607720S, RPG609597S, RPG621020S, RPG608821S, RPG614312S, RPG615226S, RPG615392S, RPG618138S, RPG620525S, RPG621257S, RPG622568S, RPG622570S, RPG609081S, RPG609594S, RPG609920S, RPG610501S, RPG611503S, RPG621418S, RPG617597S, RPG608569S, RPG609321S, RPG613671S, RPG616888S, RPG617263S, RPG618391S, RPG618821S, RPG619590S, RPG619633S, RPG620598S, RPG621055S, RPG626311S, RPG610630S, RPG612862S, RPG615397S, RPG619881S, RPG619887S, RPG622219S, RPG609067S, RPG609069S, RPG624622S, RPG624869S, RPG607203S, RPG607438S, RPG607898S, RPG608156S, RPG608223S, RPG608234S, RPG608396S, RPG609088S, RPG609295S, RPG609876S, RPG610091S, RPG610426S, RPG610506S, RPG610838S, RPG611031S, RPG611317S, RPG611325S, RPG611444S, RPG611677S, RPG611987S, RPG612151S, RPG612400S, RPG612499S, RPG612853S, RPG613443S, RPG613583S, RPG615275S, RPG616168S, RPG616226S, RPG617200S, RPG617280S, RPG617768S, RPG618366S, RPG618389S, RPG618463S, RPG621063S, RPG621356S, RPG621586S, RPG621634S, RPG623341S, RPG624519S, RPG624638S, RPG625942S, RPG626486S, RPG626287S, RPG626465S, RPG608422S, RPG609068S, RPG609084S, RPG610021S, RPG610412S, RPG610433S, RPG610628S, RPG610974S, RPG610989S, RPG611630S, RPG611672S, RPG612812S, RPG612854S, RPG613202S, RPG613624S, RPG614672S, RPG616373S, RPG617307S, RPG619651S, RPG620426S, RPG620734S, RPG624941S, RPG625089S, RPG625634S, RPG625663S, RPG609851S, RPG614975S, RPG624515S, RPG625833S, RPG607732S, RPG609197S, RPG610735S, RPG611011S, RPG612576S, RPG613020S, RPG616762S, RPG621226S, RPG609360S, RPG609502S, RPG622553S, RPG608757S, RPG612786S, RPG619973S, RPG611825S, RPG611827S, RPG611829S, RPG611831S, RPG619431S, RPG619558S, RPG619560S, RPG619565S, RPG607480S, RPG607899S, RPG609072S, RPG610270S, RPG610829S, RPG612591S, RPG612599S, RPG613356S, RPG618748S, RPG619043S, RPG619321S, RPG620041S, RPG620643S, RPG621629S, RPG622160S, RPG623257S, RPG623987S, RPG611816S, RPG611958S, RPG613219S, RPG614382S, RPG614605S, RPG614980S, RPG618387S, RPG618388S, RPG619658S, RPG619703S, RPG620638S, RPG620857S, RPG621921S, RPG621948S, RPG622013S, RPG622227S, RPG622506S, RPG622609S, RPG622728S, RPG623238S, RPG623566S, RPG623913S, RPG623992S, RPG624609S, RPG624646S, RPG625721S, RPG625733S, RPG625831S, RPG626117S, RPG626447S, RPG608884S, RPG610921S, RPG610992S, RPG611021S, RPG611401S, RPG612401S, RPG613307S, RPG613586S, RPG613922S, RPG614755S, RPG614839S, RPG615942S, RPG616281S, RPG616356S, RPG617158S, RPG617472S, RPG618549S, RPG618719S, RPG619014S, RPG619691S, RPG619880S, RPG620910S, RPG621041S, RPG622150S, RPG622347S, RPG623427S, RPG623902S, RPG624366S, RPG624522S, RPG625224S, RPG625724S, RPG625734S, RPG625837S, RPG610279S, RPG611162S, RPG619673S, RPG623445S, RPG614642S, RPG614646S, RPG614650S, RPG614652S, RPG614901S, RPG614903S, RPG619039S, RPG613616S, RPG613617S, RPG613619S, RPG613620S, RPG616153S, RPG616154S, RPG617202S, RPG617991S, RPG617992S, RPG617995S, RPG623262S, RPG626572S, RPG610400S, RPG624838S, RPG608849S, RPG611795S, RPG612364S, RPG612500S, RPG614293S, RPG614477S, RPG615863S, RPG617412S, RPG619422S, RPG620569S, RPG621639S, RPG622445S, RPG623164S, RPG623288S, RPG625227S, RPG625546S, RPG626553S, RPG610985S, RPG613122S, RPG626188S, RPG611278S, RPG607204S, RPG607548S, RPG607549S, RPG607716S, RPG607729S, RPG607806S, RPG607907S, RPG608255S, RPG608405S, RPG608510S, RPG608748S, RPG610949S, RPG610958S, RPG612921S, RPG617438S, RPG620072S, RPG620354S, RPG622880S, RPG624392S, RPG624595S, RPG624704S, RPG618047S, RPG607193S, RPG607316S, RPG607487S, RPG607562S, RPG607606S, RPG607618S, RPG608362S, RPG608505S, RPG608916S, RPG609140S, RPG609285S, RPG609327S, RPG609548S, RPG609581S, RPG609852S, RPG610124S, RPG610219S, RPG610278S, RPG610305S, RPG610470S, RPG610471S, RPG610734S, RPG611315S, RPG611338S, RPG611502S, RPG611580S, RPG612110S, RPG612114S, RPG612468S, RPG612482S, RPG612529S, RPG612546S, RPG612596S, RPG612675S, RPG612684S, RPG613819S, RPG614187S, RPG614314S, RPG614465S, RPG615257S, RPG615288S, RPG615439S, RPG615951S, RPG616402S, RPG616569S, RPG616727S, RPG616839S, RPG616889S, RPG617297S, RPG617514S, RPG617639S, RPG618350S, RPG618384S, RPG618454S, RPG618660S, RPG618910S, RPG619962S, RPG619981S, RPG619982S, RPG620073S, RPG620156S, RPG620518S, RPG620960S, RPG621085S, RPG621289S, RPG621353S, RPG622081S, RPG622491S, RPG622534S, RPG622908S, RPG623526S, RPG624368S, RPG624378S, RPG624432S, RPG625523S, RPG625857S, RPG626007S, RPG626258S, RPG623465S, RPG625094S, RPG619048S, RPG625228S, RPG611671S, RPG614111S, RPG618563S, RPG618836S, RPG620421S, RPG620851S, RPG622592S, RPG625534S, RPG625922S, RPG616416S, RPG617562S, RPG617857S, RPG624707S, RPG614636S, RPG615989S, RPG616506S, RPG617294S, RPG617858S, RPG618361S, RPG618471S, RPG618551S, RPG619517S, RPG620033S, RPG620298S, RPG621568S, RPG621672S, RPG622083S, RPG622342S, RPG622405S, RPG623321S, RPG623461S, RPG623762S, RPG625300S, RPG625696S, RPG625743S, RPG625840S, RPG625904S, RPG626222S, RPG626239S, RPG626361S, RPG626398S, RPG612811S, RPG613937S, RPG614389S, RPG614972S, RPG615262S, RPG615311S, RPG615319S, RPG615385S, RPG615769S, RPG615980S, RPG616353S, RPG616507S, RPG616786S, RPG616866S, RPG617166S, RPG617542S, RPG617841S, RPG618691S, RPG619085S, RPG619142S, RPG619249S, RPG619413S, RPG619453S, RPG619522S, RPG619579S, RPG619730S, RPG619845S, RPG620046S, RPG620119S, RPG621462S, RPG621512S, RPG621900S, RPG621978S, RPG622969S, RPG622975S, RPG623218S, RPG623360S, RPG624149S, RPG624252S, RPG624424S, RPG624517S, RPG624591S, RPG625057S, RPG625827S, RPG626217S, RPG626264S, RPG626343S, RPG626524S, RPG612790S, RPG616015S, RPG618772S, RPG619605S, RPG625652S, RPG606230S, RPG608139S, RPG608642S, RPG608918S, RPG609015S, RPG609105S, RPG609486S, RPG611746S, RPG612424S, RPG612480S, RPG612483S, RPG612571S, RPG612794S, RPG612863S, RPG613006S, RPG613040S, RPG613154S, RPG613648S, RPG614668S, RPG615321S, RPG615722S, RPG616887S, RPG617314S, RPG617466S, RPG617506S, RPG617663S, RPG618747S, RPG618767S, RPG618815S, RPG619030S, RPG619592S, RPG619839S, RPG620111S, RPG621014S, RPG621711S, RPG622039S, RPG622349S, RPG622353S, RPG622712S, RPG623535S, RPG623739S, RPG623764S, RPG624637S, RPG624706S, RPG624858S, RPG624973S, RPG625015S, RPG625246S, RPG625513S, RPG625565S, RPG625782S, RPG625874S, RPG625898S, RPG626134S, RPG626284S, RPG626345S, RPG626405S, RPG626457S, RPG617164S, RPG612492S, RPG608144S, RPG608690S, RPG609083S, RPG609094S, RPG609925S, RPG609981S, RPG611172S, RPG611479S, RPG612124S, RPG612943S, RPG613314S, RPG613728S, RPG614695S, RPG616260S, RPG616289S, RPG616759S, RPG617884S, RPG618560S, RPG618807S, RPG619137S, RPG619739S, RPG619777S, RPG620409S, RPG620594S, RPG620761S, RPG620790S, RPG621008S, RPG621317S, RPG621319S, RPG621544S, RPG621626S, RPG621708S, RPG621751S, RPG621874S, RPG622020S, RPG622042S, RPG622043S, RPG622473S, RPG623003S, RPG623515S, RPG623625S, RPG623775S, RPG624051S, RPG624254S, RPG625070S, RPG625071S, RPG625126S, RPG625157S, RPG625293S, RPG625317S, RPG625456S, RPG625503S, RPG625802S, RPG626154S, RPG626274S, RPG626392S, RPG626459S, RPG626479S, RPG626571S, RPG608171S, RPG609928S, RPG610254S, RPG615878S, RPG621619S, RPG624579S, RPG624685S, RPG625700S, RPG626631S, RPG624000S, RPG625850S, RPG607586S, RPG608455S, RPG608746S, RPG608829S, RPG609858S, RPG610020S, RPG610144S, RPG611443S, RPG611860S, RPG613127S, RPG613156S, RPG613194S, RPG613968S, RPG614900S, RPG615869S, RPG616232S, RPG616571S, RPG617183S, RPG617503S, RPG620919S, RPG623117S, RPG624311S, RPG624940S, RPG625804S, RPG625899S, RPG607215S, RPG609398S, RPG611154S, RPG615120S, RPG615492S, RPG618301S, RPG620715S, RPG620751S, RPG621023S, RPG621628S, RPG622312S, RPG622325S, RPG623878S, RPG624667S, RPG625064S, RPG625480S, RPG625888S, RPG626202S, RPG626306S, RPG626557S, RPG626415S, RPG607202S, RPG607209S, RPG607640S, RPG607744S, RPG607816S, RPG607817S, RPG607830S, RPG608191S, RPG608364S, RPG608488S, RPG608644S, RPG608693S, RPG608725S, RPG608790S, RPG608968S, RPG608969S, RPG609021S, RPG609190S, RPG609320S, RPG609390S, RPG609608S, RPG609897S, RPG609898S, RPG609900S, RPG610022S, RPG610068S, RPG610114S, RPG610134S, RPG610182S, RPG610261S, RPG610263S, RPG610282S, RPG610298S, RPG610395S, RPG610436S, RPG610442S, RPG610448S, RPG610479S, RPG610489S, RPG610512S, RPG610529S, RPG610625S, RPG610722S, RPG610725S, RPG610872S, RPG610948S, RPG610965S, RPG610966S, RPG611006S, RPG611056S, RPG611106S, RPG611173S, RPG611241S, RPG611280S, RPG611463S, RPG611472S, RPG611513S, RPG611587S, RPG611622S, RPG611659S, RPG611696S, RPG611739S, RPG611917S, RPG611927S, RPG611957S, RPG611990S, RPG612444S, RPG612452S, RPG612518S, RPG612609S, RPG612631S, RPG612693S, RPG612807S, RPG612974S, RPG613076S, RPG613204S, RPG613316S, RPG613374S, RPG613408S, RPG613461S, RPG613476S, RPG613540S, RPG613612S, RPG613632S, RPG613706S, RPG613739S, RPG614207S, RPG614208S, RPG614405S, RPG621285S, RPG610873S, RPG612608S, RPG613618S, RPG614666S, RPG618014S, RPG613226S, RPG613660S, RPG613662S, RPG613665S, RPG613672S, RPG613674S, RPG613676S, RPG613680S, RPG613685S, RPG613688S, RPG613692S, RPG613693S, RPG618946S, RPG618948S, RPG618950S, RPG618951S, RPG618986S, RPG618988S, RPG618989S, RPG618990S, RPG618991S, RPG618992S, RPG624036S, RPG624037S, RPG624039S, RPG624040S, RPG624042S, RPG624046S, RPG624050S, RPG624055S, RPG624058S, RPG624059S, RPG624060S, RPG624061S, RPG624064S, RPG624066S, RPG624073S, RPG624076S, RPG624081S, RPG624084S, RPG624138S, RPG624142S, RPG607261S, RPG610325S, RPG610612S, RPG611310S, RPG611351S, RPG611410S, RPG611440S, RPG611494S, RPG611813S, RPG612430S, RPG612495S, RPG612729S, RPG612914S, RPG612934S, RPG613094S, RPG613108S, RPG614644S, RPG614745S, RPG614759S, RPG614760S, RPG615233S, RPG615791S, RPG615874S, RPG615981S, RPG616109S, RPG616411S, RPG616443S, RPG616591S, RPG616686S, RPG616704S, RPG616709S, RPG616838S, RPG617178S, RPG617188S, RPG617490S, RPG617712S, RPG618569S, RPG619026S, RPG619028S, RPG620012S, RPG620214S, RPG620253S, RPG620477S, RPG620485S, RPG620487S, RPG620781S, RPG620916S, RPG621004S, RPG621521S, RPG621831S, RPG621877S, RPG622123S, RPG622169S, RPG622327S, RPG622343S, RPG622469S, RPG622620S, RPG623123S, RPG623911S, RPG625295S, RPG625903S, RP
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
- Reason for Recall:
- There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
CRTD DTMB2QQ AMPLIA MRI QUAD OUS DF4, Model Number DTMB2QQ; Implantable Cardioverter Defibrillators
Product Codes/Lot Numbers:
GTIN 00643169720695, Lot Serial numbers: RPG619037S, RPG608482S, RPG611787S, RPG619305S, RPG609628S, RPG622247S, RPG619130S, RPG623603S, RPG610818S, RPG616992S, RPG619816S, RPG624535S, RPG614089S, RPG619704S, RPG619714S, RPG623002S, RPG624602S, RPG607260S, RPG607434S, RPG607977S, RPG608232S, RPG608831S, RPG609415S, RPG609836S, RPG609911S, RPG610009S, RPG610345S, RPG611061S, RPG611302S, RPG611328S, RPG611598S, RPG611600S, RPG611606S, RPG611666S, RPG611667S, RPG611668S, RPG612610S, RPG612884S, RPG612885S, RPG613206S, RPG613239S, RPG613578S, RPG613655S, RPG614353S, RPG614407S, RPG614439S, RPG614898S, RPG615028S, RPG617792S, RPG617839S, RPG618125S, RPG618648S, RPG618809S, RPG618811S, RPG619226S, RPG619228S, RPG619555S, RPG619776S, RPG619888S, RPG619890S, RPG619891S, RPG620540S, RPG620563S, RPG621028S, RPG621106S, RPG621190S, RPG621228S, RPG621272S, RPG621395S, RPG621680S, RPG621722S, RPG621724S, RPG621725S, RPG621730S, RPG621743S, RPG622038S, RPG622041S, RPG622085S, RPG622532S, RPG622533S, RPG622881S, RPG622905S, RPG622984S, RPG623287S, RPG623393S, RPG623681S, RPG623684S, RPG623693S, RPG624227S, RPG624640S, RPG624641S, RPG624699S, RPG624702S, RPG625103S, RPG625298S, RPG625299S, RPG625307S, RPG625310S, RPG625343S, RPG625704S, RPG625915S, RPG625919S, RPG626152S, RPG626157S, RPG626198S, RPG626210S, RPG626396S, RPG626399S, RPG626402S, RPG626560S, RPG626605S, RPG626606S, RPG618474S, RPG618477S, RPG618499S, RPG618602S, RPG618603S, RPG618611S, RPG618616S, RPG618617S, RPG618618S, RPG618619S, RPG618620S, RPG618621S, RPG618623S, RPG618624S, RPG618625S, RPG618626S, RPG618636S, RPG618637S, RPG618639S, RPG618640S, RPG618642S, RPG618643S, RPG618646S, RPG618647S, RPG620107S, RPG620115S, RPG620118S, RPG620120S, RPG620122S, RPG620160S, RPG620165S, RPG620166S, RPG620171S, RPG620177S, RPG620179S, RPG620185S, RPG620186S, RPG620189S, RPG620194S, RPG620200S, RPG620202S, RPG620213S, RPG620258S, RPG620270S, RPG620272S, RPG623171S, RPG623176S, RPG623180S, RPG623183S, RPG623184S, RPG623193S, RPG623197S, RPG623199S, RPG623200S, RPG623202S, RPG623204S, RPG623209S, RPG623213S, RPG623214S, RPG623215S, RPG623221S, RPG623228S, RPG623236S, RPG623247S, RPG623251S, RPG623253S, RPG625563S, RPG625608S, RPG625609S, RPG625616S, RPG625617S, RPG625621S, RPG625623S, RPG625624S, RPG625625S, RPG625630S, RPG625645S, RPG625650S, RPG625662S, RPG625673S, RPG625682S, RPG625703S, RPG625714S, RPG625715S, RPG625716S, RPG625736S, RPG625738S, RPG626407S, RPG626416S, RPG626417S, RPG626421S, RPG626423S, RPG626425S, RPG626426S, RPG626433S, RPG626435S, RPG626437S, RPG626441S, RPG626442S, RPG626443S, RPG626444S, RPG626450S, RPG626451S, RPG626453S, RPG626454S, RPG626468S, RPG626472S, RPG626475S, RPG626483S, RPG626487S, RPG626492S, RPG626502S, RPG626504S, RPG626505S, RPG626509S, RPG626516S, RPG626519S, RPG626521S, RPG626523S, RPG626525S, RPG626527S, RPG626528S, RPG626529S, RPG626531S, RPG626532S, RPG626535S, RPG626540S, RPG611844S, RPG611846S, RPG613809S, RPG616044S, RPG616046S, RPG616384S, RPG616764S, RPG623927S, RPG624385S, RPG608667S, RPG608668S, RPG608669S, RPG609041S, RPG609233S, RPG619334S, RPG619360S, RPG611019S, RPG611036S, RPG611894S, RPG614000S, RPG614199S, RPG614399S, RPG614560S, RPG614576S, RPG614934S, RPG615091S, RPG615105S, RPG616049S, RPG616438S, RPG617449S, RPG618128S, RPG618285S, RPG619117S, RPG619207S, RPG622790S, RPG622794S, RPG622800S, RPG622974S, RPG623018S, RPG623024S, RPG623772S, RPG623976S, RPG626039S, RPG626603S, RPG611022S, RPG611023S, RPG612375S, RPG612395S, RPG612888S, RPG612955S, RPG614106S, RPG614181S, RPG614184S, RPG614189S, RPG614191S, RPG614195S, RPG614202S, RPG614238S, RPG614241S, RPG614245S, RPG614401S, RPG614591S, RPG614593S, RPG614601S, RPG615087S, RPG615568S, RPG616022S, RPG616028S, RPG616037S, RPG616077S, RPG616321S, RPG616800S, RPG618065S, RPG618346S, RPG618790S, RPG619779S, RPG619785S, RPG619895S, RPG620126S, RPG620129S, RPG620130S, RPG622061S, RPG622796S, RPG622798S, RPG622801S, RPG622968S, RPG623112S, RPG623782S, RPG626081S, RPG626086S, RPG626091S, RPG613261S, RPG615706S, RPG616001S, RPG618928S, RPG622092S, RPG623088S, RPG623107S, RPG626006S, RPG615704S, RPG622267S, RPG626014S, RPG612037S, RPG612039S, RPG612042S, RPG612048S, RPG612061S, RPG612066S, RPG612116S, RPG612976S, RPG613251S, RPG615692S, RPG615707S, RPG615708S, RPG615709S, RPG615716S, RPG615738S, RPG616493S, RPG618849S, RPG618850S, RPG618897S, RPG618898S, RPG618905S, RPG618955S, RPG619701S, RPG620275S, RPG620282S, RPG620296S, RPG620327S, RPG620349S, RPG620383S, RPG620435S, RPG620466S, RPG620481S, RPG620492S, RPG620500S, RPG620636S, RPG621779S, RPG621802S, RPG621805S, RPG622149S, RPG622239S, RPG622995S, RPG623089S, RPG623093S, RPG623096S, RPG623104S, RPG623105S, RPG623296S, RPG623585S, RPG624013S, RPG624014S, RPG624025S, RPG624151S, RPG624162S, RPG612089S, RPG614070S, RPG616323S, RPG616445S, RPG616485S, RPG620380S, RPG620382S, RPG624021S, RPG611998S, RPG612100S, RPG613790S, RPG615765S, RPG616459S, RPG620350S, RPG621377S, RPG621681S, RPG621763S, RPG621882S, RPG624012S, RPG612421S, RPG613058S, RPG613263S, RPG614040S, RPG618972S, RPG620335S, RPG621786S, RPG623090S, RPG623584S, RPG612045S, RPG616042S, RPG616446S, RPG616449S, RPG617672S, RPG618963S, RPG620389S, RPG620497S, RPG622242S, RPG623085S, RPG624246S, RPG626011S, RPG618812S, RPG618907S, RPG618913S, RPG620215S, RPG620295S, RPG620359S, RPG620424S, RPG621773S, RPG621787S, RPG622258S, RPG622743S, RPG624282S, RPG611991S, RPG618892S, RPG620338S, RPG620361S, RPG621767S, RPG615711S, RPG616487S, RPG620443S, RPG622084S, RPG624167S, RPG613976S, RPG614013S, RPG615705S, RPG616452S, RPG616814S, RPG618855S, RPG618938S, RPG619696S, RPG615710S, RPG616817S, RPG618374S, RPG618791S, RPG618977S, RPG620346S, RPG620348S, RPG620351S, RPG621794S, RPG622246S, RPG623099S, RPG611974S, RPG618902S, RPG621800S, RPG621885S, RPG622240S, RPG622249S, RPG622263S, RPG619286S, RPG619585S, RPG620238S, RPG620247S, RPG620362S, RPG620376S, RPG620379S, RPG622742S, RPG622744S, RPG623472S, RPG623481S, RPG623557S, RPG624234S, RPG624251S, RPG618931S, RPG618943S, RPG618945S, RPG618947S, RPG623579S, RPG624249S, RPG616325S, RPG616349S, RPG616492S, RPG622082S, RPG623095S, RPG611995S, RPG618954S, RPG622253S, RPG624247S, RPG617674S, RPG617700S, RPG618941S, RPG619271S, RPG619281S, RPG619282S, RPG620375S, RPG622257S, RPG622260S, RPG622262S, RPG622270S, RPG622271S, RPG622513S, RPG622735S, RPG623297S, RPG623299S, RPG623456S, RPG623586S, RPG623720S, RPG623722S, RPG623723S, RPG624023S, RPG624028S, RPG624032S, RPG624225S, RPG619248S, RPG626280S, RPG607290S, RPG607449S, RPG607450S, RPG609351S, RPG609352S, RPG613090S, RPG613246S, RPG618649S, RPG619015S, RPG619344S, RPG620574S, RPG621332S, RPG624409S, RPG607198S, RPG615946S, RPG619245S, RPG619848S, RPG619920S, RPG620413S, RPG620797S, RPG622187S, RPG622330S, RPG607228S, RPG607233S, RPG607275S, RPG607346S, RPG607926S, RPG608143S, RPG608161S, RPG608957S, RPG609319S, RPG609509S, RPG609538S, RPG609614S, RPG611681S, RPG612356S, RPG613010S, RPG613253S, RPG613313S, RPG613921S, RPG614090S, RPG614329S, RPG614674S, RPG615144S, RPG616054S, RPG616636S, RPG616662S, RPG616706S, RPG617462S, RPG617609S, RPG618794S, RPG618795S, RPG619149S, RPG619709S, RPG619878S, RPG619924S, RPG620039S, RPG621928S, RPG622298S, RPG622322S, RPG622456S, RPG624404S, RPG625095S, RPG625098S, RPG625331S, RPG626331S, RPG626470S, RPG610095S, RPG614654S, RPG615936S, RPG617198S, RPG617633S, RPG618445S, RPG620816S, RPG622261S, RPG614544S, RPG615827S, RPG620109S, RPG621570S, RPG621972S, RPG622277S, RPG624205S, RPG626332S, RPG607345S, RPG607574S, RPG607628S, RPG607995S, RPG608135S, RPG608419S, RPG608624S, RPG608652S, RPG609367S, RPG609625S, RPG609990S, RPG610687S, RPG610695S, RPG610749S, RPG611366S, RPG611457S, RPG611544S, RPG612598S, RPG612777S, RPG612779S, RPG612984S, RPG613105S, RPG613357S, RPG613581S, RPG613703S, RPG613960S, RPG614007S, RPG614330S, RPG614540S, RPG614656S, RPG615032S, RPG615206S, RPG615875S, RPG615934S, RPG618415S, RPG619682S, RPG619705S, RPG619828S, RPG619876S, RPG619929S, RPG619965S, RPG620252S, RPG620777S, RPG620794S, RPG621391S, RPG621405S, RPG621887S, RPG622114S, RPG622140S, RPG623217S, RPG623384S, RPG623435S, RPG623710S, RPG624049S, RPG624258S, RPG624599S, RPG624934S, RPG625463S, RPG625737S, RPG625800S, RPG625809S, RPG625926S, RPG626247S, RPG626260S, RPG626292S, RPG626350S, RPG626352S, RPG626491S, RPG626614S, RPG607551S, RPG607196S, RPG608516S, RPG608956S, RPG609200S, RPG610250S, RPG611333S, RPG612578S, RPG614691S, RPG615139S, RPG615943S, RPG618264S, RPG619012S, RPG619208S, RPG619449S, RPG619608S, RPG619609S, RPG619792S, RPG622730S, RPG622829S, RPG622924S, RPG623106S, RPG623521S, RPG624882S, RPG625043S, RPG625053S, RPG625285S, RPG626132S, RPG621701S, RPG622336S, RPG622628S, RPG623078S, RPG623350S, RPG623512S, RPG623908S, RPG624635S, RPG624877S, RPG625025S, RPG625351S, RPG625585S, RPG625797S, RPG626209S, RPG626348S, RPG626356S, RPG626650S, RPG607626S, RPG607927S, RPG608136S, RPG608287S, RPG608779S, RPG609366S, RPG609945S, RPG610130S, RPG610256S, RPG610907S, RPG610912S, RPG611413S, RPG611985S, RPG612579S, RPG613600S, RPG613623S, RPG613789S, RPG613933S, RPG614318S, RPG614352S, RPG614403S, RPG615080S, RPG615376S, RPG615876S, RPG615938S, RPG616053S, RPG616056S, RPG616584S, RPG616853S, RPG617644S, RPG617720S, RPG618692S, RPG618781S, RPG619329S, RPG620251S, RPG620423S, RPG620462S, RPG621017S, RPG621616S, RPG621775S, RPG621913S, RPG622319S, RPG622323S, RPG622380S, RPG622830S, RPG622902S, RPG622966S, RPG623207S, RPG623312S, RPG623375S, RPG623394S, RPG623396S, RPG623398S, RPG623870S, RPG624260S, RPG624342S, RPG624548S, RPG625042S, RPG625498S, RPG625765S, RPG625783S, RPG625801S, RPG625917S, RPG626455S, RPG607274S, RPG626488S, RPG607255S, RPG610481S, RPG610906S, RPG610997S, RPG611191S, RPG612005S, RPG612022S, RPG616408S, RPG617180S, RPG617528S, RPG617833S, RPG618449S, RPG618450S, RPG618513S, RPG619405S, RPG620964S, RPG621659S, RPG621662S, RPG621832S, RPG621997S, RPG622286S, RPG626408S, RPG626661S, RPG612574S, RPG612809S, RPG616266S, RPG614290S, RPG621822S, RPG617620S, RPG614676S, RPG618784S, RPG622374S, RPG608203S, RPG612059S, RPG612155S, RPG612160S, RPG613221S, RPG613579S, RPG614381S, RPG614464S, RPG614637S, RPG620793S, RPG621702S, RPG622269S, RPG623565S, RPG623907S, RPG608237S, RPG608801S, RPG621611S, RPG617259S, RPG619840S, RPG620019S, RPG617253S, RPG613615S, RPG615994S, RPG620583S, RPG620603S, RPG620609S, RPG607675S, RPG607208S, RPG607510S, RPG607868S, RPG609682S, RPG610018S, RPG610377S, RPG610879S, RPG610926S, RPG611411S, RPG613213S, RPG613734S, RPG614258S, RPG614387S, RPG614431S, RPG614713S, RPG615383S, RPG615400S, RPG616361S, RPG616479S, RPG616789S, RPG616791S, RPG617621S, RPG618349S, RPG618638S, RPG618859S, RPG620567S, RPG620920S, RPG621396S, RPG621782S, RPG622053S, RPG624308S, RPG626143S, RPG626678S, RPG616700S, RPG616701S, RPG617499S, RPG617501S, RPG617502S, RPG607990S, RPG615603S, RPG616236S, RPG618482S, RPG619849S, RPG619862S, RPG625679S, RPG625775S, RPG616176S, RPG616180S, RPG616394S, RPG616395S, RPG617127S, RPG617134S, RPG617137S, RPG617439S, RPG617440S, RPG618048S, RPG618150S, RPG618368S, RPG619024S, RPG619339S, RPG619385S, RPG620075S, RPG620825S, RPG621335S, RPG621419S, RPG621486S, RPG621511S, RPG621548S, RPG621604S, RPG621964S, RPG622004S, RPG622529S, RPG622654S, RPG622655S, RPG622896S, RPG623307S, RPG623690S, RPG623692S, RPG623702S, RPG623711S, RPG623712S, RPG623875S, RPG623944S, RPG624009S, RPG624374S, RPG624551S, RPG624959S, RPG625185S, RPG626139S, RPG626141S, RPG626298S, RPG607288S, RPG609118S, RPG609342S, RPG610507S, RPG611632S, RPG611874S, RPG612647S, RPG612672S, RPG612951S, RPG615215S, RPG616825S, RPG617999S, RPG618508S, RPG619628S, RPG621101S, RPG621238S, RPG622777S, RPG623281S, RPG624262S, RPG624279S, RPG624431S, RPG624505S, RPG624751S, RPG624952S, RPG625598S, RPG625750S, RPG625909S, RPG626431S, RPG623622S, RPG607728S, RPG607230S, RPG607556S, RPG607637S, RPG607930S, RPG608902S, RPG609942S, RPG609960S, RPG611733S, RPG612808S, RPG613355S, RPG613646S, RPG613955S, RPG614256S, RPG614303S, RPG615967S, RPG616195S, RPG617787S, RPG619927S, RPG621606S, RPG624278S, RPG626136S, RPG616194S, RPG616470S, RPG619016S, RPG620568S, RPG620978S, RPG621579S, RPG622276S, RPG614378S, RPG615950S, RPG624958S, RPG620332S, RPG621057S, RPG622452S, RPG623724S, RPG624413S, RPG624547S, RPG626161S, RPG616769S, RPG620827S, RPG625657S, RPG616283S, RPG624209S, RPG611607S, RPG614001S, RPG615398S, RPG616430S, RPG616745S, RPG618538S, RPG618561S, RPG618578S, RPG621601S, RPG621664S, RPG625792S, RPG610704S, RPG615893S, RPG617974S, RPG618514S, RPG618786S, RPG620986S, RPG622396S, RPG622773S, RPG624504S, RPG626118S, RPG626162S, RPG614084S, RPG622441S, RPG615545S, RPG616614S, RPG616836S, RPG617309S, RPG618376S, RPG619781S, RPG620326S, RPG621750S, RPG621966S, RPG622971S, RPG623079S, RPG624271S, RPG625755S, RPG625834S, RPG626199S, RPG626354S, RPG626559S, RPG608714S, RPG609947S, RPG610877S, RPG610920S, RPG611364S, RPG613729S, RPG615025S, RPG616613S, RPG616744S, RPG619138S, RPG619171S, RPG619851S, RPG619937S, RPG619960S, RPG622329S, RPG623865S, RPG623934S, RPG625044S, RPG625543S, RPG625825S, RPG626123S, RPG626200S, RPG626554S, RPG615373S, RPG616182S, RPG616184S, RPG616187S, RPG616650S, RPG616651S, RPG617681S, RPG617682S, RPG620464S, RPG620465S, RPG621865S, RPG621871S, RPG622372S, RPG622373S, RPG622985S, RPG622986S, RPG614771S, RPG614849S, RPG615955S, RPG619020S, RPG619447S, RPG620637S, RPG621720S, RPG625367S, RPG607206S, RPG607635S, RPG608163S, RPG608207S, RPG608679S, RPG608684S, RPG608695S, RPG608727S, RPG608953S, RPG609293S, RPG609570S, RPG609843S, RPG609941S, RPG610153S, RPG610844S, RPG611130S, RPG611482S, RPG612176S, RPG613481S, RPG614640S, RPG615295S, RPG616126S, RPG620592S, RPG620762S, RPG621755S, RPG621784S, RPG621895S, RPG622470S, RPG622942S, RPG623985S, RPG625159S, RPG625788S, RPG626318S, RPG607213S, RPG622542S, RPG623234S, RPG623597S, RPG624083S, RPG625660S, RPG626203S, RPG626218S, RPG607568S, RPG609914S, RPG611175S, RPG611185S, RPG611458S, RPG615184S, RPG616484S, RPG617093S, RPG617553S, RPG622045S, RPG624402S, RPG625108S, RPG625545S, RPG610165S, RPG608206S, RPG608434S, RPG608697S, RPG608699S, RPG610502S, RPG610911S, RPG610913S, RPG610994S, RPG611159S, RPG613007S, RPG614383S, RPG615119S, RPG617638S, RPG618267S, RPG618793S, RPG618863S, RPG623940S, RPG625643S, RPG613364S, RPG614289S, RPG614613S, RPG614969S, RPG615605S, RPG615745S, RPG616397S, RPG616410S, RPG617135S, RPG617461S, RPG617527S, RPG618852S, RPG618853S, RPG618997S, RPG619038S, RPG619554S, RPG619603S, RPG619625S, RPG619708S, RPG620601S, RPG620778S, RPG620780S, RPG620959S, RPG621019S, RPG621404S, RPG621557S, RPG621635S, RPG622377S, RPG622455S, RPG622734S, RPG622874S, RPG623233S, RPG623490S, RPG623881S, RPG624891S, RPG625422S, RPG625479S, RPG625752S, RPG625786S, RPG607199S, RPG607223S, RPG607317S, RPG609059S, RPG609119S, RPG609380S, RPG609544S, RPG609810S, RPG610076S, RPG610141S, RPG611626S, RPG611658S, RPG614386S, RPG614525S, RPG615764S, RPG616585S, RPG616652S, RPG617098S, RPG617246S, RPG617267S, RPG618386S, RPG618580S, RPG618841S, RPG619006S, RPG619919S, RPG621632S, RPG621669S, RPG621709S, RPG621765S, RPG622110S, RPG622221S, RPG622359S, RPG622806S, RPG623220S, RPG623771S, RPG624004S, RPG624192S, RPG624268S, RPG624598S, RPG624949S, RPG625239S, RPG626404S, RPG626500S, RPG614354S, RPG616412S, RPG617189S, RPG617563S, RPG617893S, RPG618119S, RPG619031S, RPG619634S, RPG619844S, RPG620467S, RPG620975S, RPG623540S, RPG623991S, RPG625428S, RPG625816S, RPG626664S, RPG614280S, RPG622466S, RPG623861S, RPG625536S, RPG607247S, RPG620977S, RPG621718S, RPG626418S, RPG617316S, RPG618489S, RPG625100S, RPG607604S, RPG612047S, RPG622860S, RPG626252S, RPG607492S, RPG614080S, RPG614730S, RPG617103S, RPG620741S, RPG621876S, RPG621902S, RPG623888S, RPG625073S, RPG625839S, RPG625918S, RPG626641S, RPG618944S, RPG621572S, RPG622119S, RPG622446S, RPG623879S, RPG614997S, RPG615073S, RPG607555S, RPG608562S, RPG608563S, RPG611594S, RPG612597S, RPG612645S, RPG614679S, RPG615735S, RPG615921S, RPG616463S, RPG617707S, RPG617719S, RPG618843S, RPG619218S, RPG619534S, RPG619650S, RPG620204S, RPG620428S, RPG621209S, RPG621607S, RPG621666S, RPG621919S, RPG624434S, RPG626174S, RPG607265S, RPG617139S, RPG624357S, RPG618572S, RPG621408S, RPG622068S, RPG622293S, RPG622891S, RPG623068S, RPG624156S, RPG613812S, RPG615178S, RPG617643S, RPG617704S, RPG617737S, RPG617976S, RPG618067S, RPG618703S, RPG618798S, RPG618857S, RPG619654S, RPG619748S, RPG621059S, RPG621487S, RPG621527S, RPG622535S, RPG622784S, RPG622921S, RPG623428S, RPG623759S, RPG624006S, RPG624399S, RPG624661S, RPG624857S, RPG625550S, RPG625611S, RPG626374S, RPG621956S, RPG623478S, RPG624223S, RPG625811S, RPG626358S, RPG619566S, RPG621340S, RPG621536S, RPG621585S, RPG622304S, RPG623211S, RPG623936S, RPG624186S, RPG624805S, RPG625009S, RPG625646S, RPG625756S, RPG626480S, RPG626566S, RPG626633S, RPG611308S, RPG611934S, RPG613780S, RPG615988S, RPG616401S, RPG616448S, RPG617190S, RPG618754S, RPG619215S, RPG622301S, RPG622344S, RPG622355S, RPG623004S, RPG625653S, RPG608696S, RPG610429S, RPG615483S, RPG614319S, RPG618804S, RPG619150S, RPG619668S, RPG624615S, RPG616574S, RPG609495S, RPG610984S, RPG611944S, RPG614924S, RPG622477S, RPG622989S, RPG625326S, RPG626393S, RPG613086S, RPG622929S, RPG625699S, RPG618472S, RPG618473S, RPG619259S, RPG610019S, RPG612679S, RPG613631S, RPG614712S, RPG619707S, RPG623473S, RPG616499S, RPG617534S, RPG618761S, RPG619233S, RPG609721S, RPG609075S, RPG611027S, RPG611337S, RPG614252S, RPG614878S, RPG618134S, RPG619794S, RPG608118S, RPG610164S, RPG611369S, RPG612572S, RPG612630S, RPG618245S, RPG622485S, RPG615952S, RPG616886S, RPG617281S, RPG617293S, RPG617567S, RPG617598S, RPG617611S, RPG618133S, RPG618351S, RPG618584S, RPG618727S, RPG618813S, RPG619273S, RPG619784S, RPG619952S, RPG620024S, RPG620805S, RPG620822S, RPG620946S, RPG621005S, RPG621322S, RPG621450S, RPG621451S, RPG621466S, RPG621538S, RPG621540S, RPG621576S, RPG621580S, RPG621806S, RPG621810S, RPG621954S, RPG621970S, RPG622033S, RPG622063S, RPG622152S, RPG622212S, RPG622442S, RPG622717S, RPG622771S, RPG622785S, RPG622789S, RPG622791S, RPG624483S, RPG624981S, RPG625011S, RPG625176S, RPG625342S, RPG625444S, RPG625666S, RPG625810S, RPG625910S, RPG625938S, RPG626581S, RPG626666S, RPG608646S, RPG609704S, RPG609970S, RPG612756S, RPG613110S, RPG614630S, RPG624029S, RPG622987S, RPG622988S, RPG617751S, RPG618797S, RPG619702S, RPG620067S, RPG620624S, RPG622007S, RPG607945S, RPG610103S, RPG610968S, RPG611324S, RPG612837S, RPG613173S, RPG613343S, RPG613348S, RPG613788S, RPG617891S, RPG618541S, RPG620456S, RPG620457S, RPG622101S, RPG625685S, RPG611295S, RPG612486S, RPG615992S, RPG616931S, RPG618250S, RPG619943S, RPG620795S, RPG621400S, RPG621592S, RPG622900S, RPG623414S, RPG625145S, RPG625396S, RPG625458S, RPG626248S, RPG626347S, RPG626355S, RPG626366S, RPG607491S, RPG607595S, RPG608246S, RPG608691S, RPG610036S, RPG610482S, RPG610916S, RPG611165S, RPG611661S, RPG611877S, RPG612506S, RPG612952S, RPG613636S, RPG615126S, RPG615601S, RPG615935S, RPG616243S, RPG617157S, RPG617176S, RPG617204S, RPG617312S, RPG618509S, RPG618600S, RPG619165S, RPG620562S, RPG620968S, RPG621899S, RPG622126S, RPG624256S, RPG624736S, RPG625675S, RPG625677S, RPG626440S, RPG626464S, RPG615966S, RPG609339S, RPG611368S, RPG612484S, RPG613843S, RPG608070S, RPG608214S, RPG608221S, RPG610908S, RPG611487S, RPG611615S, RPG612649S, RPG612728S, RPG612931S, RPG613128S, RPG613816S, RPG614687S, RPG616653S, RPG616687S, RPG617177S, RPG617579S, RPG617850S, RPG617935S, RPG619027S, RPG619769S, RPG620222S, RPG620476S, RPG621529S, RPG621825S, RPG622060S, RPG622351S, RPG622354S, RPG625690S, RPG626128S, RPG626144S, RPG626346S, RPG612440S, RPG613793S, RPG614661S, RPG614698S, RPG614723S, RPG616511S, RPG616567S, RPG616830S, RPG617713S, RPG618019S, RPG618732S, RPG621583S, RPG622376S, RPG622379S, RPG622726S, RPG623493S, RPG623905S, RPG624514S, RPG625179S, RPG626556S, RPG626564S, RPG617315S, RPG619157S, RPG619676S, RPG607178S, RPG607764S, RPG607881S, RPG609717S, RPG610874S, RPG612405S, RPG614102S, RPG615742S, RPG618296S, RPG620604S, RPG622248S, RPG624649S, RPG625060S, RPG615009S, RPG610443S, RPG610875S, RPG612425S, RPG612508S, RPG612526S, RPG612532S, RPG612691S, RPG613042S, RPG613140S, RPG613141S, RPG613330S, RPG613471S, RPG613562S, RPG615127S, RPG615422S, RPG615425S, RPG615741S, RPG615890S, RPG616012S, RPG616193S, RPG616228S, RPG616293S, RPG616575S, RPG616717S, RPG617482S, RPG617623S, RPG618117S, RPG618507S, RPG618585S, RPG618782S, RPG619698S, RPG620042S, RPG620381S, RPG621376S, RPG621456S, RPG621637S, RPG621826S, RPG622111S, RPG622278S, RPG622281S, RPG623272S, RPG623516S, RPG623519S, RPG623652S, RPG623895S, RPG623984S, RPG624325S, RPG624401S, RPG625632S, RPG625892S, RPG626255S, RPG608480S, RPG615949S, RPG626652S, RPG607224S, RPG607245S, RPG607383S, RPG607786S, RPG609176S, RPG610692S, RPG611965S, RPG611975S, RPG612054S, RPG613252S, RPG614185S, RPG614428S, RPG614974S, RPG615163S, RPG615751S, RPG616237S, RPG616421S, RPG616440S, RPG617465S, RPG617758S, RPG618004S, RPG619010S, RPG611198S, RPG613093S, RPG618118S, RPG607599S, RPG607940S, RPG608754S, RPG608896S, RPG608961S, RPG613962S, RPG614376S, RPG609965S, RPG610517S, RPG611976S, RPG613340S, RPG617121S, RPG617123S, RPG620432S, RPG621553S, RPG623373S, RPG626215S, RPG614837S, RPG615426S, RPG617101S, RPG618428S, RPG620979S, RPG621689S, RPG622112S, RPG622737S, RPG624665S, RPG625855S, RPG621783S, RPG608951S, RPG613264S, RPG615794S, RPG620169S, RPG621225S, RPG623372S, RPG626386S, RPG620436S, RPG609496S, RPG610005S, RPG610996S, RPG611688S, RPG613970S, RPG614943S, RPG615728S, RPG616601S, RPG617690S, RPG619159S, RPG622026S, RPG622321S, RPG623466S, RPG623967S, RPG624711S, RPG624712S, RPG624862S, RPG624866S, RPG625096S, RPG625097S, RPG626205S, RPG626207S, RPG626226S, RPG626424S, RPG626550S, RPG609964S, RPG615382S, RPG610424S, RPG610494S, RPG616773S, RPG617601S, RPG621764S, RPG622324S, RPG623280S, RPG623524S, RPG626397S, RPG626538S, RPG613373S, RPG614940S, RPG616771S, RPG617329S, RPG611404S, RPG612618S, RPG613967S, RPG622486S, RPG623187S, RPG623864S, RPG610902S, RPG610922S, RPG612519S, RPG618377S, RPG619825S, RPG610275S, RPG617463S, RPG620236S, RPG610016S, RPG610073S, RPG611414S, RPG612555S, RPG612903S, RPG613362S, RPG613716S, RPG614096S, RPG614249S, RPG615822S, RPG615868S, RPG616390S, RPG617328S, RPG618122S, RPG619632S, RPG620062S, RPG623103S, RPG623922S, RPG624418S, RPG625147S, RPG625281S, RPG625627S, RPG626342S, RPG610986S, RPG613722S, RPG610787S, RPG611054S, RPG611055S, RPG611456S, RPG614710S, RPG615132S, RPG615225S, RPG616733S, RPG617136S, RPG617498S, RPG617524S, RPG618257S, RPG607720S, RPG609597S, RPG621020S, RPG608821S, RPG614312S, RPG615226S, RPG615392S, RPG618138S, RPG620525S, RPG621257S, RPG622568S, RPG622570S, RPG609081S, RPG609594S, RPG609920S, RPG610501S, RPG611503S, RPG621418S, RPG617597S, RPG608569S, RPG609321S, RPG613671S, RPG616888S, RPG617263S, RPG618391S, RPG618821S, RPG619590S, RPG619633S, RPG620598S, RPG621055S, RPG626311S, RPG610630S, RPG612862S, RPG615397S, RPG619881S, RPG619887S, RPG622219S, RPG609067S, RPG609069S, RPG624622S, RPG624869S, RPG607203S, RPG607438S, RPG607898S, RPG608156S, RPG608223S, RPG608234S, RPG608396S, RPG609088S, RPG609295S, RPG609876S, RPG610091S, RPG610426S, RPG610506S, RPG610838S, RPG611031S, RPG611317S, RPG611325S, RPG611444S, RPG611677S, RPG611987S, RPG612151S, RPG612400S, RPG612499S, RPG612853S, RPG613443S, RPG613583S, RPG615275S, RPG616168S, RPG616226S, RPG617200S, RPG617280S, RPG617768S, RPG618366S, RPG618389S, RPG618463S, RPG621063S, RPG621356S, RPG621586S, RPG621634S, RPG623341S, RPG624519S, RPG624638S, RPG625942S, RPG626486S, RPG626287S, RPG626465S, RPG608422S, RPG609068S, RPG609084S, RPG610021S, RPG610412S, RPG610433S, RPG610628S, RPG610974S, RPG610989S, RPG611630S, RPG611672S, RPG612812S, RPG612854S, RPG613202S, RPG613624S, RPG614672S, RPG616373S, RPG617307S, RPG619651S, RPG620426S, RPG620734S, RPG624941S, RPG625089S, RPG625634S, RPG625663S, RPG609851S, RPG614975S, RPG624515S, RPG625833S, RPG607732S, RPG609197S, RPG610735S, RPG611011S, RPG612576S, RPG613020S, RPG616762S, RPG621226S, RPG609360S, RPG609502S, RPG622553S, RPG608757S, RPG612786S, RPG619973S, RPG611825S, RPG611827S, RPG611829S, RPG611831S, RPG619431S, RPG619558S, RPG619560S, RPG619565S, RPG607480S, RPG607899S, RPG609072S, RPG610270S, RPG610829S, RPG612591S, RPG612599S, RPG613356S, RPG618748S, RPG619043S, RPG619321S, RPG620041S, RPG620643S, RPG621629S, RPG622160S, RPG623257S, RPG623987S, RPG611816S, RPG611958S, RPG613219S, RPG614382S, RPG614605S, RPG614980S, RPG618387S, RPG618388S, RPG619658S, RPG619703S, RPG620638S, RPG620857S, RPG621921S, RPG621948S, RPG622013S, RPG622227S, RPG622506S, RPG622609S, RPG622728S, RPG623238S, RPG623566S, RPG623913S, RPG623992S, RPG624609S, RPG624646S, RPG625721S, RPG625733S, RPG625831S, RPG626117S, RPG626447S, RPG608884S, RPG610921S, RPG610992S, RPG611021S, RPG611401S, RPG612401S, RPG613307S, RPG613586S, RPG613922S, RPG614755S, RPG614839S, RPG615942S, RPG616281S, RPG616356S, RPG617158S, RPG617472S, RPG618549S, RPG618719S, RPG619014S, RPG619691S, RPG619880S, RPG620910S, RPG621041S, RPG622150S, RPG622347S, RPG623427S, RPG623902S, RPG624366S, RPG624522S, RPG625224S, RPG625724S, RPG625734S, RPG625837S, RPG610279S, RPG611162S, RPG619673S, RPG623445S, RPG614642S, RPG614646S, RPG614650S, RPG614652S, RPG614901S, RPG614903S, RPG619039S, RPG613616S, RPG613617S, RPG613619S, RPG613620S, RPG616153S, RPG616154S, RPG617202S, RPG617991S, RPG617992S, RPG617995S, RPG623262S, RPG626572S, RPG610400S, RPG624838S, RPG608849S, RPG611795S, RPG612364S, RPG612500S, RPG614293S, RPG614477S, RPG615863S, RPG617412S, RPG619422S, RPG620569S, RPG621639S, RPG622445S, RPG623164S, RPG623288S, RPG625227S, RPG625546S, RPG626553S, RPG610985S, RPG613122S, RPG626188S, RPG611278S, RPG607204S, RPG607548S, RPG607549S, RPG607716S, RPG607729S, RPG607806S, RPG607907S, RPG608255S, RPG608405S, RPG608510S, RPG608748S, RPG610949S, RPG610958S, RPG612921S, RPG617438S, RPG620072S, RPG620354S, RPG622880S, RPG624392S, RPG624595S, RPG624704S, RPG618047S, RPG607193S, RPG607316S, RPG607487S, RPG607562S, RPG607606S, RPG607618S, RPG608362S, RPG608505S, RPG608916S, RPG609140S, RPG609285S, RPG609327S, RPG609548S, RPG609581S, RPG609852S, RPG610124S, RPG610219S, RPG610278S, RPG610305S, RPG610470S, RPG610471S, RPG610734S, RPG611315S, RPG611338S, RPG611502S, RPG611580S, RPG612110S, RPG612114S, RPG612468S, RPG612482S, RPG612529S, RPG612546S, RPG612596S, RPG612675S, RPG612684S, RPG613819S, RPG614187S, RPG614314S, RPG614465S, RPG615257S, RPG615288S, RPG615439S, RPG615951S, RPG616402S, RPG616569S, RPG616727S, RPG616839S, RPG616889S, RPG617297S, RPG617514S, RPG617639S, RPG618350S, RPG618384S, RPG618454S, RPG618660S, RPG618910S, RPG619962S, RPG619981S, RPG619982S, RPG620073S, RPG620156S, RPG620518S, RPG620960S, RPG621085S, RPG621289S, RPG621353S, RPG622081S, RPG622491S, RPG622534S, RPG622908S, RPG623526S, RPG624368S, RPG624378S, RPG624432S, RPG625523S, RPG625857S, RPG626007S, RPG626258S, RPG623465S, RPG625094S, RPG619048S, RPG625228S, RPG611671S, RPG614111S, RPG618563S, RPG618836S, RPG620421S, RPG620851S, RPG622592S, RPG625534S, RPG625922S, RPG616416S, RPG617562S, RPG617857S, RPG624707S, RPG614636S, RPG615989S, RPG616506S, RPG617294S, RPG617858S, RPG618361S, RPG618471S, RPG618551S, RPG619517S, RPG620033S, RPG620298S, RPG621568S, RPG621672S, RPG622083S, RPG622342S, RPG622405S, RPG623321S, RPG623461S, RPG623762S, RPG625300S, RPG625696S, RPG625743S, RPG625840S, RPG625904S, RPG626222S, RPG626239S, RPG626361S, RPG626398S, RPG612811S, RPG613937S, RPG614389S, RPG614972S, RPG615262S, RPG615311S, RPG615319S, RPG615385S, RPG615769S, RPG615980S, RPG616353S, RPG616507S, RPG616786S, RPG616866S, RPG617166S, RPG617542S, RPG617841S, RPG618691S, RPG619085S, RPG619142S, RPG619249S, RPG619413S, RPG619453S, RPG619522S, RPG619579S, RPG619730S, RPG619845S, RPG620046S, RPG620119S, RPG621462S, RPG621512S, RPG621900S, RPG621978S, RPG622969S, RPG622975S, RPG623218S, RPG623360S, RPG624149S, RPG624252S, RPG624424S, RPG624517S, RPG624591S, RPG625057S, RPG625827S, RPG626217S, RPG626264S, RPG626343S, RPG626524S, RPG612790S, RPG616015S, RPG618772S, RPG619605S, RPG625652S, RPG606230S, RPG608139S, RPG608642S, RPG608918S, RPG609015S, RPG609105S, RPG609486S, RPG611746S, RPG612424S, RPG612480S, RPG612483S, RPG612571S, RPG612794S, RPG612863S, RPG613006S, RPG613040S, RPG613154S, RPG613648S, RPG614668S, RPG615321S, RPG615722S, RPG616887S, RPG617314S, RPG617466S, RPG617506S, RPG617663S, RPG618747S, RPG618767S, RPG618815S, RPG619030S, RPG619592S, RPG619839S, RPG620111S, RPG621014S, RPG621711S, RPG622039S, RPG622349S, RPG622353S, RPG622712S, RPG623535S, RPG623739S, RPG623764S, RPG624637S, RPG624706S, RPG624858S, RPG624973S, RPG625015S, RPG625246S, RPG625513S, RPG625565S, RPG625782S, RPG625874S, RPG625898S, RPG626134S, RPG626284S, RPG626345S, RPG626405S, RPG626457S, RPG617164S, RPG612492S, RPG608144S, RPG608690S, RPG609083S, RPG609094S, RPG609925S, RPG609981S, RPG611172S, RPG611479S, RPG612124S, RPG612943S, RPG613314S, RPG613728S, RPG614695S, RPG616260S, RPG616289S, RPG616759S, RPG617884S, RPG618560S, RPG618807S, RPG619137S, RPG619739S, RPG619777S, RPG620409S, RPG620594S, RPG620761S, RPG620790S, RPG621008S, RPG621317S, RPG621319S, RPG621544S, RPG621626S, RPG621708S, RPG621751S, RPG621874S, RPG622020S, RPG622042S, RPG622043S, RPG622473S, RPG623003S, RPG623515S, RPG623625S, RPG623775S, RPG624051S, RPG624254S, RPG625070S, RPG625071S, RPG625126S, RPG625157S, RPG625293S, RPG625317S, RPG625456S, RPG625503S, RPG625802S, RPG626154S, RPG626274S, RPG626392S, RPG626459S, RPG626479S, RPG626571S, RPG608171S, RPG609928S, RPG610254S, RPG615878S, RPG621619S, RPG624579S, RPG624685S, RPG625700S, RPG626631S, RPG624000S, RPG625850S, RPG607586S, RPG608455S, RPG608746S, RPG608829S, RPG609858S, RPG610020S, RPG610144S, RPG611443S, RPG611860S, RPG613127S, RPG613156S, RPG613194S, RPG613968S, RPG614900S, RPG615869S, RPG616232S, RPG616571S, RPG617183S, RPG617503S, RPG620919S, RPG623117S, RPG624311S, RPG624940S, RPG625804S, RPG625899S, RPG607215S, RPG609398S, RPG611154S, RPG615120S, RPG615492S, RPG618301S, RPG620715S, RPG620751S, RPG621023S, RPG621628S, RPG622312S, RPG622325S, RPG623878S, RPG624667S, RPG625064S, RPG625480S, RPG625888S, RPG626202S, RPG626306S, RPG626557S, RPG626415S, RPG607202S, RPG607209S, RPG607640S, RPG607744S, RPG607816S, RPG607817S, RPG607830S, RPG608191S, RPG608364S, RPG608488S, RPG608644S, RPG608693S, RPG608725S, RPG608790S, RPG608968S, RPG608969S, RPG609021S, RPG609190S, RPG609320S, RPG609390S, RPG609608S, RPG609897S, RPG609898S, RPG609900S, RPG610022S, RPG610068S, RPG610114S, RPG610134S, RPG610182S, RPG610261S, RPG610263S, RPG610282S, RPG610298S, RPG610395S, RPG610436S, RPG610442S, RPG610448S, RPG610479S, RPG610489S, RPG610512S, RPG610529S, RPG610625S, RPG610722S, RPG610725S, RPG610872S, RPG610948S, RPG610965S, RPG610966S, RPG611006S, RPG611056S, RPG611106S, RPG611173S, RPG611241S, RPG611280S, RPG611463S, RPG611472S, RPG611513S, RPG611587S, RPG611622S, RPG611659S, RPG611696S, RPG611739S, RPG611917S, RPG611927S, RPG611957S, RPG611990S, RPG612444S, RPG612452S, RPG612518S, RPG612609S, RPG612631S, RPG612693S, RPG612807S, RPG612974S, RPG613076S, RPG613204S, RPG613316S, RPG613374S, RPG613408S, RPG613461S, RPG613476S, RPG613540S, RPG613612S, RPG613632S, RPG613706S, RPG613739S, RPG614207S, RPG614208S, RPG614405S, RPG621285S, RPG610873S, RPG612608S, RPG613618S, RPG614666S, RPG618014S, RPG613226S, RPG613660S, RPG613662S, RPG613665S, RPG613672S, RPG613674S, RPG613676S, RPG613680S, RPG613685S, RPG613688S, RPG613692S, RPG613693S, RPG618946S, RPG618948S, RPG618950S, RPG618951S, RPG618986S, RPG618988S, RPG618989S, RPG618990S, RPG618991S, RPG618992S, RPG624036S, RPG624037S, RPG624039S, RPG624040S, RPG624042S, RPG624046S, RPG624050S, RPG624055S, RPG624058S, RPG624059S, RPG624060S, RPG624061S, RPG624064S, RPG624066S, RPG624073S, RPG624076S, RPG624081S, RPG624084S, RPG624138S, RPG624142S, RPG607261S, RPG610325S, RPG610612S, RPG611310S, RPG611351S, RPG611410S, RPG611440S, RPG611494S, RPG611813S, RPG612430S, RPG612495S, RPG612729S, RPG612914S, RPG612934S, RPG613094S, RPG613108S, RPG614644S, RPG614745S, RPG614759S, RPG614760S, RPG615233S, RPG615791S, RPG615874S, RPG615981S, RPG616109S, RPG616411S, RPG616443S, RPG616591S, RPG616686S, RPG616704S, RPG616709S, RPG616838S, RPG617178S, RPG617188S, RPG617490S, RPG617712S, RPG618569S, RPG619026S, RPG619028S, RPG620012S, RPG620214S, RPG620253S, RPG620477S, RPG620485S, RPG620487S, RPG620781S, RPG620916S, RPG621004S, RPG621521S, RPG621831S, RPG621877S, RPG622123S, RPG622169S, RPG622327S, RPG622343S, RPG622469S, RPG622620S, RPG623123S, RPG623911S, RPG625295S, RPG625903S, RP
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1768-2023
Related Recalls
ICD-DR DDBB2D1 EVERA XT IS1/DF1 INTL, Model Number DDBB2D1; Implantable Cardioverter Defibrillators
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.
ICD-DR DDMC3D1 EVERA MRI S OUS/US DF1, Model Number DDMC3D1; Implantable Cardioverter Defibrillators
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.
ICD DDPA2D1G COBALT XT DR MRI DF1 GOLD, Model Number DDPA2D1G; Implantable Cardioverter Defibrillators
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.