CRTD DTMA1Q1 CLARIA MRI QUAD US DF1, Model Number DTMA1Q1; Implantable Cardioverter Defibrillators
Class I - DangerousWhat Should You Do?
- Check if you have this product: GTIN 00643169837669, Lot Serial numbers: RPS205974H, RPS208665H, RPS206712H, RPS207744H, RPS205038H, RPS205791H, RPS206225H, RPS208216H, RPS206084H, RPS203804H, RPS208198H, RPS208245H, RPS205937H, RPS205565H, RPS206114H, RPS206723H, RPS206730H, RPS206733H, RPS206734H, RPS206754H, RPS207303H, RPS207318H, RPS207338H, RPS207399H, RPS207703H, RPS207711H, RPS207722H, RPS208049H, RPS208505H, RPS208646H, RPS208679H, RPS205303H, RPS205918H, RPS203315H, RPS204385H, RPS205650H, RPS202760H, RPS203013H, RPS203034H, RPS203256H, RPS203610H, RPS205372H, RPS206383H, RPS206771H, RPS207603H, RPS207933H, RPS208110H, RPS208492H, RPS204009H, RPS202981H, RPS207563H, RPS208546H, RPS202245H, RPS203960H, RPS204342H, RPS204357H, RPS205958H, RPS207822H, RPS207840H, RPS207922H, RPS207923H, RPS207984H, RPS208255H, RPS208529H, RPS208575H, RPS208590H, RPS208615H, RPS203563H, RPS206128H, RPS206413H, RPS206762H, RPS207904H, RPS208608H, RPS202171H, RPS202841H, RPS202860H, RPS203565H, RPS204198H, RPS202792H, RPS202793H, RPS206515H, RPS202224H, RPS203192H, RPS203545H, RPS203910H, RPS204035H, RPS204166H, RPS204197H, RPS204394H, RPS204492H, RPS205560H, RPS206777H, RPS207146H, RPS207520H, RPS208166H, RPS204464H, RPS204776H, RPS207117H, RPS204003H, RPS205323H, RPS206063H, RPS208464H, RPS206252H, RPS202211H, RPS202602H, RPS202612H, RPS205006H, RPS205638H, RPS205639H, RPS205710H, RPS205998H, RPS207105H, RPS207368H, RPS207409H, RPS202030H, RPS202067H, RPS202507H, RPS206211H, RPS206605H, RPS207137H, RPS208591H, RPS202053H, RPS202511H, RPS202116H, RPS202210H, RPS202273H, RPS202274H, RPS202276H, RPS202307H, RPS202349H, RPS202427H, RPS202447H, RPS202498H, RPS202527H, RPS202601H, RPS202603H, RPS202605H, RPS203195H, RPS203197H, RPS203431H, RPS203432H, RPS203433H, RPS203435H, RPS203436H, RPS203438H, RPS203439H, RPS203440H, RPS203441H, RPS203442H, RPS203444H, RPS204202H, RPS204209H, RPS204210H, RPS204212H, RPS204215H, RPS204217H, RPS204223H, RPS204926H, RPS204978H, RPS204980H, RPS204983H, RPS204986H, RPS204987H, RPS204988H, RPS204992H, RPS204997H, RPS205262H, RPS205271H, RPS205276H, RPS205277H, RPS205665H, RPS205714H, RPS205717H, RPS205719H, RPS205720H, RPS205737H, RPS205739H, RPS205741H, RPS205742H, RPS205744H, RPS205750H, RPS205751H, RPS205752H, RPS205754H, RPS205755H, RPS205756H, RPS205757H, RPS205758H, RPS205759H, RPS206175H, RPS206179H, RPS206235H, RPS206367H, RPS206371H, RPS206374H, RPS206378H, RPS206379H, RPS207186H, RPS207187H, RPS207247H, RPS207248H, RPS207252H, RPS207253H, RPS207254H, RPS207255H, RPS207256H, RPS207258H, RPS207260H, RPS207264H, RPS207298H, RPS207393H, RPS207402H, RPS207574H, RPS207774H, RPS207994H, RPS208133H, RPS208237H, RPS208240H, RPS208242H, RPS208327H, RPS208457H, RPS207507H, RPS205911H, RPS205204H, RPS206761H, RPS207437H, RPS207712H, RPS204627H, RPS205509H, RPS205082H, RPS202759H, RPS203251H, RPS202130H, RPS202167H, RPS203220H, RPS204656H, RPS205410H, RPS205487H, RPS207695H, RPS207867H, RPS204362H, RPS208517H, RPS203913H, RPS205806H, RPS205807H, RPS205808H, RPS202277H, RPS204053H, RPS205975H, RPS206636H, RPS208079H, RPS208119H, RPS208349H, RPS204604H, RPS205654H, RPS205765H, RPS202490H, RPS203352H, RPS203816H, RPS206903H, RPS202237H, RPS202704H, RPS203037H, RPS203532H, RPS203707H, RPS203763H, RPS204225H, RPS205169H, RPS205364H, RPS205648H, RPS205731H, RPS205789H, RPS206089H, RPS206450H, RPS206537H, RPS207592H, RPS207719H, RPS208334H, RPS202531H, RPS203458H, RPS204162H, RPS205383H, RPS205492H, RPS208444H, RPS207494H, RPS202057H, RPS202202H, RPS202567H, RPS202849H, RPS202906H, RPS202916H, RPS203068H, RPS203332H, RPS203660H, RPS203726H, RPS203796H, RPS204204H, RPS204613H, RPS205849H, RPS205850H, RPS205851H, RPS205852H, RPS206153H, RPS206452H, RPS206455H, RPS206456H, RPS206459H, RPS206963H, RPS208177H, RPS203421H, RPS203524H, RPS203615H, RPS203770H, RPS204642H, RPS205009H, RPS205058H, RPS205134H, RPS205583H, RPS205990H, RPS206633H, RPS206756H, RPS206947H, RPS207473H, RPS207927H, RPS204248H, RPS203955H, RPS206451H, RPS207611H, RPS207945H, RPS202288H, RPS202290H, RPS203479H, RPS204022H, RPS204172H, RPS204173H, RPS205800H, RPS205801H, RPS205802H, RPS205803H, RPS205805H, RPS206527H, RPS206531H, RPS206540H, RPS208266H, RPS202776H, RPS203086H, RPS202768H, RPS203253H, RPS205214H, RPS207088H, RPS207485H, RPS207487H, RPS207488H, RPS207885H, RPS208313H, RPS202656H, RPS205497H, RPS207691H, RPS202052H, RPS202249H, RPS202955H, RPS203776H, RPS203928H, RPS205585H, RPS207002H, RPS207099H, RPS207354H, RPS207447H, RPS207689H, RPS202453H, RPS202547H, RPS203069H, RPS203115H, RPS203163H, RPS203246H, RPS203407H, RPS203971H, RPS204134H, RPS204135H, RPS204139H, RPS204140H, RPS205733H, RPS205951H, RPS206134H, RPS206165H, RPS206213H, RPS206303H, RPS207173H, RPS207184H, RPS207241H, RPS207243H, RPS207249H, RPS203602H, RPS203678H, RPS204908H, RPS206177H, RPS206267H, RPS206555H, RPS204760H, RPS205535H, RPS202140H, RPS203633H, RPS203635H, RPS204218H, RPS204219H, RPS204220H, RPS205880H, RPS205883H, RPS205884H, RPS206407H, RPS202761H, RPS203704H, RPS205380H, RPS202881H, RPS202738H, RPS202799H, RPS203812H, RPS204143H, RPS206221H, RPS206611H, RPS203092H, RPS203824H, RPS207474H, RPS207599H, RPS202781H, RPS204000H, RPS205399H, RPS205642H, RPS206625H, RPS205858H, RPS207910H, RPS202021H, RPS202074H, RPS202573H, RPS202574H, RPS207292H, RPS208244H, RPS208254H, RPS207297H, RPS202655H, RPS203077H, RPS203655H, RPS207436H, RPS205848H, RPS204773H, RPS205886H, RPS207053H, RPS207979H, RPS202035H, RPS202296H, RPS203753H, RPS204353H, RPS205506H, RPS205816H, RPS206044H, RPS206083H, RPS206648H, RPS206770H, RPS207492H, RPS207543H, RPS207914H, RPS208306H, RPS208167H, RPS208419H, RPS208658H, RPS205390H, RPS207954H, RPS208554H, RPS208555H, RPS202718H, RPS202910H, RPS204089H, RPS207768H, RPS207849H, RPS207837H, RPS208118H, RPS208413H, RPS205895H, RPS202976H, RPS203159H, RPS202200H, RPS203162H, RPS203295H, RPS205887H, RPS208633H, RPS203215H, RPS205831H, RPS206886H, RPS207554H, RPS202160H, RPS203327H, RPS207796H, RPS208406H, RPS206749H, RPS202228H, RPS202231H, RPS202874H, RPS203456H, RPS204874H, RPS205547H, RPS206690H, RPS204943H, RPS202400H, RPS203158H, RPS203178H, RPS203934H, RPS204147H, RPS204278H, RPS205362H, RPS205365H, RPS205429H, RPS205953H, RPS206106H, RPS206138H, RPS207815H, RPS207919H, RPS207987H, RPS208362H, RPS208576H, RPS202741H, RPS203951H, RPS206069H, RPS206348H, RPS204270H, RPS205421H, RPS206117H, RPS202314H, RPS202425H, RPS202582H, RPS202945H, RPS203330H, RPS203451H, RPS203566H, RPS203567H, RPS203612H, RPS203818H, RPS203952H, RPS204050H, RPS204071H, RPS204150H, RPS204348H, RPS204443H, RPS204954H, RPS205125H, RPS205163H, RPS205343H, RPS205375H, RPS205814H, RPS207931H, RPS207344H, RPS207397H, RPS208487H, RPS203072H, RPS203381H, RPS206250H, RPS206268H, RPS207607H, RPS207706H, RPS207860H, RPS204922H, RPS202020H, RPS202304H, RPS202982H, RPS203231H, RPS205552H, RPS207790H, RPS203265H, RPS205968H, RPS206533H, RPS206776H, RPS206826H, RPS202083H, RPS203533H, RPS204247H, RPS204860H, RPS206325H, RPS206621H, RPS207119H, RPS207633H, RPS203321H, RPS208161H, RPS206938H, RPS204403H, RPS207683H, RPS203168H, RPS205415H, RPS208169H, RPS205890H, RPS203239H, RPS203775H, RPS206631H, RPS202564H, RPS205300H, RPS205074H, RPS207635H, RPS206802H, RPS202055H, RPS202372H, RPS202636H, RPS202665H, RPS203134H, RPS203255H, RPS203339H, RPS203392H, RPS203741H, RPS203805H, RPS203838H, RPS203865H, RPS203889H, RPS203982H, RPS204015H, RPS204097H, RPS204231H, RPS204241H, RPS204296H, RPS204355H, RPS204441H, RPS204598H, RPS204759H, RPS204778H, RPS204955H, RPS205010H, RPS205079H, RPS205114H, RPS205215H, RPS205240H, RPS205242H, RPS205248H, RPS205289H, RPS205304H, RPS205345H, RPS205431H, RPS205511H, RPS205543H, RPS205569H, RPS205661H, RPS205992H, RPS206052H, RPS206062H, RPS206066H, RPS206192H, RPS206220H, RPS206253H, RPS206259H, RPS206292H, RPS206340H, RPS206423H, RPS206658H, RPS207170H, RPS207981H, RPS208012H, RPS208071H, RPS208138H, RPS208308H, RPS208343H, RPS208348H, RPS202127H, RPS203424H, RPS204387H, RPS206030H, RPS206543H, RPS206549H, RPS208171H, RPS208172H, RPS205977H, RPS206538H, RPS208047H, RPS206126H, RPS208669H, RPS202978H, RPS203083H, RPS203572H, RPS203871H, RPS204148H, RPS208491H, RPS204768H, RPS202460H, RPS202461H, RPS202463H, RPS203291H, RPS203597H, RPS204070H, RPS205211H, RPS205522H, RPS205545H, RPS205904H, RPS205905H, RPS206096H, RPS206301H, RPS206323H, RPS206590H, RPS206591H, RPS207069H, RPS207070H, RPS207143H, RPS207639H, RPS208001H, RPS208352H, RPS208510H, RPS208644H, RPS203028H, RPS202870H, RPS203325H, RPS204607H, RPS205130H, RPS205132H, RPS208057H, RPS202639H, RPS202643H, RPS202644H, RPS205020H, RPS205021H, RPS205187H, RPS205503H, RPS205558H, RPS205778H, RPS205780H, RPS205840H, RPS205853H, RPS205885H, RPS205967H, RPS208423H, RPS208424H, RPS203413H, RPS205862H, RPS202552H, RPS203284H, RPS203521H, RPS203830H, RPS203918H, RPS204008H, RPS204199H, RPS204203H, RPS206783H, RPS206892H, RPS208135H, RPS205908H, RPS207594H, RPS207824H, RPS207917H, RPS202977H, RPS206042H, RPS206620H, RPS206713H, RPS207673H, RPS207808H, RPS208563H, RPS205949H, RPS206794H, RPS207929H, RPS204229H, RPS204323H, RPS204298H, RPS205601H, RPS206005H, RPS206263H, RPS207068H, RPS202282H, RPS203110H, RPS204525H, RPS204119H, RPS204377H, RPS206855H, RPS207018H, RPS207197H, RPS203483H, RPS203484H, RPS203488H, RPS204176H, RPS204177H, RPS204178H, RPS204180H, RPS204181H, RPS204185H, RPS204189H, RPS204190H, RPS204192H, RPS207234H, RPS207238H, RPS207240H, RPS207242H, RPS207244H, RPS207245H, RPS207246H, RPS208314H, RPS208325H, RPS208329H, RPS208332H, RPS208333H, RPS203408H, RPS205446H, RPS207943H, RPS207051H, RPS207148H, RPS207993H, RPS202739H, RPS204118H, RPS205461H, RPS203944H, RPS205419H, RPS206635H, RPS206696H, RPS207893H, RPS208483H, RPS203758H, RPS206314H, RPS203643H, RPS203702H, RPS203903H, RPS203906H, RPS206772H, RPS207160H, RPS207273H, RPS208140H, RPS203249H, RPS203803H, RPS204787H, RPS205209H, RPS205357H, RPS206742H, RPS207832H, RPS208291H, RPS202722H, RPS204059H, RPS204615H, RPS205084H, RPS205095H, RPS207663H, RPS202753H, RPS202291H, RPS202714H, RPS202965H, RPS203216H, RPS203631H, RPS203987H, RPS206671H, RPS207377H, RPS207780H, RPS208123H, RPS205643H, RPS205331H, RPS207930H, RPS206587H, RPS208098H, RPS203081H, RPS207396H, RPS208541H, RPS206384H, RPS207530H, RPS204266H, RPS208605H, RPS203738H, RPS206240H, RPS206738H, RPS207519H, RPS207967H, RPS208249H, RPS208570H, RPS202444H, RPS202451H, RPS203460H, RPS205468H, RPS205603H, RPS205609H, RPS205909H, RPS206091H, RPS206890H, RPS206977H, RPS207387H, RPS208479H, RPS202450H, RPS208353H, RPS202113H, RPS204124H, RPS204343H, RPS204471H, RPS204795H, RPS205086H, RPS205162H, RPS205312H, RPS206228H, RPS206271H, RPS206289H, RPS206355H, RPS207277H, RPS207561H, RPS207616H, RPS208372H, RPS202470H, RPS202471H, RPS202551H, RPS202571H, RPS202590H, RPS202901H, RPS202969H, RPS203225H, RPS203226H, RPS203227H, RPS203230H, RPS203491H, RPS204507H, RPS204508H, RPS204970H, RPS204972H, RPS204976H, RPS205452H, RPS206015H, RPS206187H, RPS206725H, RPS203207H, RPS207517H, RPS207874H, RPS202682H, RPS202791H, RPS203425H, RPS206135H, RPS206657H, RPS207153H, RPS202125H, RPS202143H, RPS202213H, RPS203367H, RPS204418H, RPS205329H, RPS206285H, RPS207717H, RPS208579H, RPS202039H, RPS202128H, RPS203015H, RPS203709H, RPS203762H, RPS204222H, RPS204456H, RPS204779H, RPS204919H, RPS205319H, RPS205657H, RPS206858H, RPS206860H, RPS206861H, RPS207156H, RPS207438H, RPS207848H, RPS208403H, RPS208429H, RPS208514H, RPS208564H, RPS202157H, RPS202672H, RPS202775H, RPS203766H, RPS204280H, RPS205988H, RPS202136H, RPS202921H, RPS204700H, RPS205302H, RPS205559H, RPS207023H, RPS207086H, RPS207315H, RPS207707H, RPS207942H, RPS208022H, RPS208303H, RPS208475H, RPS202893H, RPS203943H, RPS205243H, RPS206049H, RPS203050H, RPS203101H, RPS208433H, RPS203085H, RPS203476H, RPS204920H, RPS204360H, RPS204489H, RPS204852H, RPS205614H, RPS202503H, RPS202762H, RPS203266H, RPS203492H, RPS205068H, RPS205580H, RPS205763H, RPS206295H, RPS207531H, RPS207541H, RPS207547H, RPS208176H, RPS208648H, RPS202668H, RPS203126H, RPS205210H, RPS208527H, RPS203359H, RPS203592H, RPS205863H, RPS206829H, RPS208489H, RPS204685H, RPS208331H, RPS203212H, RPS203931H, RPS204132H, RPS204380H, RPS204968H, RPS205678H, RPS207016H, RPS207596H, RPS205387H, RPS202223H, RPS206521H, RPS206779H, RPS204427H, RPS207880H, RPS208498H, RPS208567H, RPS202608H, RPS207613H, RPS208573H, RPS203935H, RPS202708H, RPS202727H, RPS203725H, RPS203734H, RPS204620H, RPS205512H, RPS206750H, RPS207782H, RPS208037H, RPS208446H, RPS203765H, RPS202510H, RPS203472H, RPS203473H, RPS205024H, RPS205025H, RPS206509H, RPS207414H, RPS204014H, RPS204111H, RPS206077H, RPS206670H, RPS206746H, RPS208295H, RPS202036H, RPS204224H, RPS206503H, RPS206506H, RPS207010H, RPS202755H, RPS204659H, RPS204780H, RPS205376H, RPS206449H, RPS206388H, RPS205655H, RPS202491H, RPS205600H, RPS206801H, RPS202794H, RPS202795H, RPS204286H, RPS204288H, RPS205088H, RPS206290H, RPS206517H, RPS206548H, RPS207045H, RPS208127H, RPS208128H, RPS208129H, RPS208130H, RPS208131H, RPS206087H, RPS202736H, RPS202737H, RPS202929H, RPS205237H, RPS205311H, RPS207968H, RPS205430H, RPS206200H, RPS206249H, RPS206333H, RPS206905H, RPS208039H, RPS208345H, RPS206276H, RPS206159H, RPS208032H, RPS203520H, RPS206744H, RPS202426H, RPS203410H, RPS203099H, RPS203169H, RPS203579H, RPS204450H, RPS205013H, RPS205288H, RPS206674H, RPS204052H, RPS204692H, RPS204940H, RPS205128H, RPS205178H, RPS206627H, RPS206640H, RPS208683H, RPS205051H, RPS205052H, RPS205607H, RPS207271H, RPS207843H, RPS207939H, RPS207953H, RPS208163H, RPS208584H, RPS205190H, RPS206215H, RPS206258H, RPS202772H, RPS202038H, RPS203623H, RPS205936H, RPS202079H, RPS202305H, RPS202654H, RPS202947H, RPS204010H, RPS204042H, RPS204145H, RPS204509H, RPS205228H, RPS205623H, RPS206072H, RPS206823H, RPS207109H, RPS207132H, RPS207525H, RPS207556H, RPS207620H, RPS207801H, RPS207803H, RPS207816H, RPS207817H, RPS208309H, RPS208360H, RPS208674H, RPS202675H, RPS202676H, RPS202680H, RPS204395H, RPS205282H, RPS205704H, RPS205705H, RPS205706H, RPS206370H, RPS206401H, RPS206402H, RPS207422H, RPS207425H, RPS207428H, RPS208006H, RPS208007H, RPS208640H, RPS208666H, RPS203426H, RPS203854H, RPS205534H, RPS206602H, RPS207677H, RPS207940H, RPS205233H, RPS208180H, RPS202315H, RPS206202H, RPS206307H, RPS206442H, RPS207938H, RPS208445H, RPS204428H, RPS208145H, RPS202871H, RPS203090H, RPS203114H, RPS205195H, RPS205316H, RPS206630H, RPS207161H, RPS202823H, RPS203537H, RPS204339H, RPS204537H, RPS206028H, RPS206830H, RPS208168H, RPS206971H, RPS203593H, RPS204754H, RPS204756H, RPS206288H, RPS207415H, RPS207421H, RPS202584H, RPS203586H, RPS203716H, RPS203789H, RPS203862H, RPS204879H, RPS205401H, RPS205698H, RPS205901H, RPS206644H, RPS206989H, RPS207022H, RPS207037H, RPS207073H, RPS207392H, RPS207443H, RPS207946H, RPS207978H, RPS208264H, RPS208271H, RPS208379H, RPS208437H, RPS208454H, RPS208482H, RPS208571H, RPS208614H, RPS206842H, RPS205318H, RPS205246H, RPS208422H, RPS202717H, RPS203147H, RPS203326H, RPS203703H, RPS203794H, RPS204397H, RPS206392H, RPS206535H, RPS207168H, RPS208187H, RPS204501H, RPS207061H, RPS207956H, RPS206603H, RPS208225H, RPS207532H, RPS203173H, RPS205245H, RPS205630H, RPS202032H, RPS205310H, RPS203933H, RPS204102H, RPS207969H, RPS204314H, RPS205416H, RPS205417H, RPS206243H, RPS207573H, RPS207999H, RPS202740H, RPS203228H, RPS204239H, RPS203283H, RPS204099H, RPS208054H, RPS208272H, RPS206216H, RPS204049H, RPS204528H, RPS206995H, RPS202179H, RPS202180H, RPS202409H, RPS203464H, RPS203841H, RPS205499H, RPS207046H, RPS205530H, RPS206614H, RPS203722H, RPS204766H, RPS202050H, RPS203744H, RPS203893H, RPS208574H, RPS204963H, RPS208301H, RPS202502H, RPS205458H, RPS205696H, RPS208322H, RPS203388H, RPS203713H, RPS202804H, RPS203554H, RPS205103H, RPS206394H, RPS206850H, RPS207915H, RPS208551H, RPS203904H, RPS205599H, RPS206330H, RPS206344H, RPS207806H, RPS205934H, RPS206716H, RPS206984H, RPS203175H, RPS206694H, RPS205221H, RPS207107H, RPS202525H, RPS203780H, RPS204493H, RPS208503H, RPS202247H, RPS202252H, RPS202327H, RPS202328H, RPS202355H, RPS203670H, RPS203671H, RPS203672H, RPS203679H, RPS203680H, RPS205701H, RPS205827H, RPS206726H, RPS207621H, RPS207789H, RPS205632H, RPS208185H, RPS208195H, RPS205615H, RPS208566H, RPS203535H, RPS203695H, RPS204789H, RPS204928H, RPS204921H, RPS202886H, RPS204396H, RPS206484H, RPS207720H, RPS207772H, RPS203309H, RPS206558H, RPS206758H, RPS208484H, RPS203613H, RPS202926H, RPS205984H, RPS208072H, RPS202756H, RPS208380H, RPS202812H, RPS203038H, RPS203261H, RPS203400H, RPS203676H, RPS203774H, RPS203778H, RPS204320H, RPS204444H, RPS204475H, RPS204476H, RPS204614H, RPS205145H, RPS205542H, RPS205611H, RPS207295H, RPS202558H, RPS206104H, RPS202092H, RPS202480H, RPS202963H, RPS206360H, RPS207017H, RPS203600H, RPS203965H, RPS206539H, RPS202782H, RPS202994H, RPS203132H, RPS203312H, RPS203614H, RPS204480H, RPS204936H, RPS205367H, RPS205388H, RPS205666H, RPS205669H, RPS205824H, RPS206198H, RPS206282H, RPS208061H, RPS208065H, RPS208153H, RPS203214H, RPS206422H, RPS206993H, RPS206113H, RPS202279H, RPS202399H, RPS202948H, RPS202967H, RPS203276H, RPS203591H, RPS203876H, RPS204912H, RPS205664H, RPS206110H, RPS206556H, RPS207391H, RPS206151H, RPS206182H, RPS205156H, RPS205875H, RPS206056H, RPS206542H, RPS206999H, RPS202185H, RPS202805H, RPS204131H, RPS204159H, RPS204160H, RPS202037H, RPS202056H, RPS202744H, RPS202826H, RPS202850H, RPS203414H, RPS203831H, RPS204606H, RPS205007H, RPS205427H, RPS205529H, RPS206024H, RPS206294H, RPS206683H, RPS207024H, RPS207510H, RPS207684H, RPS207734H, RPS207866H, RPS208623H, RPS206093H, RPS204934H, RPS205001H, RPS208548H, RPS202879H, RPS203937H, RPS204084H, RPS204616H, RPS205541H, RPS205916H, RPS205919H, RPS206057H, RPS206119H, RPS206120H, RPS207419H, RPS207615H, RPS207649H, RPS207672H, RPS208179H, RPS208511H, RPS208585H, RPS208580H, RPS208530H, RPS203868H, RPS203962H, RPS206642H, RPS205216H, RPS205727H, RPS205728H, RPS208616H, RPS203317H, RPS203364H, RPS203462H, RPS203651H, RPS207853H, RPS206659H, RPS205265H, RPS202743H, RPS204313H, RPS205989H, RPS207612H, RPS202044H, RPS202275H, RPS204404H, RPS205804H, RPS206766H, RPS208217H, RPS203048H, RPS203071H, RPS205332H, RPS207049H, RPS202807H, RPS203144H, RPS206103H, RPS206411H, RPS206541H, RPS208440H, RPS202746H, RPS206791H, RPS202187H, RPS202586H, RPS203154H, RPS205605H, RPS205893H, RPS206208H, RPS208204H, RPS208369H, RPS208549H, RPS207724H, RPS208366H, RPS202206H, RPS202209H, RPS203490H, RPS203898H, RPS203900H, RPS204490H, RPS204621H, RPS205546H, RPS205982H, RPS206306H, RPS207067H, RPS207638H, RPS208060H, RPS208207H, RPS203035H, RPS203834H, RPS203920H, RPS203950H, RPS205131H, RPS205227H, RPS205241H, RPS206245H, RPS206581H, RPS206819H, RPS207585H, RPS208090H, RPS208535H, RPS202196H, RPS208534H, RPS202219H, RPS202221H, RPS203791H, RPS204106H, RPS205396H, RPS208089H, RPS208147H, RPS206161H, RPS206606H, RPS206773H, RPS205226H, RPS206461H, RPS205402H, RPS202439H, RPS204929H, RPS205356H, RPS206900H, RPS207042H, RPS207370H, RPS207496H, RPS205142H, RPS206142H, RPS206191H, RPS206650H, RPS206810H, RPS207040H, RPS207321H, RPS207448H, RPS207508H, RPS208055H, RPS208456H, RPS204379H, RPS204381H, RPS204887H, RPS204979H, RPS205167H, RPS205266H, RPS205550H, RPS206287H, RPS208222H, RPS203293H, RPS203297H, RPS203849H, RPS204785H, RPS205915H, RPS206951H, RPS202078H, RPS203049H, RPS205556H, RPS206272H, RPS207162H, RPS207588H, RPS207348H, RPS208004H, RPS203371H, RPS203947H, RPS204371H, RPS205292H, RPS204383H, RPS204486H, RPS204781H, RPS208084H, RPS206397H, RPS207538H, RPS203329H, RPS207237H, RPS202777H, RPS203917H, RPS204251H, RPS204310H, RPS206080H, RPS206329H, RPS207899H, RPS204158H, RPS204228H, RPS202353H, RPS202434H, RPS202537H, RPS203041H, RPS203254H, RPS203344H, RPS203466H, RPS204051H, RPS204179H, RPS204281H, RPS204376H, RPS205151H, RPS205152H, RPS205634H, RPS205649H, RPS208542H, RPS205513H, RPS205869H, RPS208286H, RPS208550H, RPS203798H, RPS204317H, RPS202018H, RPS202788H, RPS203030H, RPS204274H, RPS205610H, RPS205933H, RPS205941H, RPS206214H, RPS206751H, RPS207052H, RPS207618H, RPS203938H, RPS206641H, RPS202734H, RPS202787H, RPS203948H, RPS207286H, RPS207516H, RPS207674H, RPS206933H, RPS203199H, RPS203727H, RPS207514H, RPS207572H, RPS208149H, RPS208261H, RPS208029H, RPS202233H, RPS203190H, RPS203570H, RPS203945H, RPS206932H, RPS208238H, RPS202607H, RPS203011H, RPS206502H, RPS208275H, RPS205507H, RPS205268H, RPS205400H, RPS208027H, RPS203252H, RPS204494H, RPS207728H, RPS202142H, RPS202469H, RPS202767H, RPS202907H, RPS203782H, RPS203964H, RPS203970H, RPS204123H, RPS204411H, RPS204474H, RPS204774H, RPS205408H, RPS205515H, RPS205970H, RPS206217H, RPS206261H, RPS206638H, RPS206765H, RPS207123H, RPS207996H, RPS208221H, RPS208603H, RPS208622H, RPS203700H, RPS205548H, RPS207961H, RPS208670H, RPS202151H, RPS203333H, RPS204275H, RPS204364H, RPS205251H, RPS206239H, RPS206885H, RPS207129H, RPS207511H, RPS202022H, RPS202133H, RPS202514H, RPS203172H, RPS203302H, RPS203319H, RPS203363H, RPS203689H, RPS204006H, RPS204182H, RPS204205H, RPS205244H, RPS205443H, RPS205924H, RPS205985H, RPS206327H, RPS206386H, RPS206534H, RPS206573H, RPS206676H, RPS206677H, RPS207071H, RPS207180H, RPS207626H, RPS207769H, RPS208036H, RPS208354H, RPS208367H, RPS208659H, RPS202664H, RPS202667H, RPS202671H, RPS205825H, RPS205826H, RPS206065H, RPS206078H, RPS206085H, RPS205275H, RPS202725H, RPS202545H, RPS202803H, RPS202891H, RPS203006H, RPS203268H, RPS203596H, RPS203747H, RPS204849H, RPS204850H, RPS205525H, RPS206001H, RPS207112H, RPS207963H, RPS207976H, RPS208008H, RPS208595H, RPS202885H, RPS205306H, RPS202268H, RPS202455H, RPS202475H, RPS202822H, RPS203118H, RPS203180H, RPS203354H, RPS203412H, RPS203526H, RPS203735H, RPS203878H, RPS203989H, RPS204116H, RPS204365H, RPS205234H, RPS205334H, RPS205366H, RPS205566H, RPS206876H, RPS207912H, RPS208397H, RPS208587H, RPS208685H, RPS202483H, RPS202524H, RPS203620H, RPS206205H, RPS208588H, RPS208636H, RPS202321H, RPS203082H, RPS203285H, RPS203547H, RPS204430H, RPS204432H, RPS204433H, RPS204436H, RPS204437H, RPS205712H, RPS206888H, RPS207287H, RPS207288H, RPS208115H, RPS208215H, RPS202800H, RPS202813H, RPS203232H, RPS203234H, RPS203866H, RPS205479H, RPS205551H, RPS206764H, RPS207169H, RPS207565H, RPS202964H, RPS203919H, RPS204183H, RPS207512H, RPS208011H, RPS202651H, RPS202660H, RPS202677H, RPS202688H, RPS202693H, RPS202697H, RPS202702H, RPS203493H, RPS203500H, RPS203749H, RPS205023H, RPS205029H, RPS205033H, RPS205035H, RPS205041H, RPS205042H, RPS205046H, RPS205048H, RPS205083H, RPS205538H, RPS205540H, RPS205596H, RPS205734H, RPS206265H, RPS206520H, RPS206687H, RPS206832H, RPS206834H, RPS206835H, RPS206907H, RPS207697H, RPS207758H, RPS207784H, RPS207807H, RPS207811H, RPS208000H, RPS208028H, RPS208033H, RPS208058H, RPS208175H, RPS208228H, RPS208297H, RPS208504H, RPS208516H, RPS208553H, RPS208594H, RPS208629H, RPS206686H, RPS208407H, RPS203953H, RPS204108H, RPS204424H, RPS206133H, RPS206815H, RPS207050H, RPS208137H, RPS208197H, RPS206334H, RPS206508H, RPS207122H, RPS205092H, RPS205099H, RPS205776H, RPS205872H, RPS207590H, RPS208209H, RPS208435H, RPS208520H, RPS208641H, RPS202071H, RPS205500H, RPS208448H, RPS208617H, RPS203057H, RPS203370H, RPS204797H, RPS206092H, RPS202176H, RPS202183H, RPS203489H, RPS203752H, RPS204488H, RPS205855H, RPS207515H, RPS208476H, RPS207829H, RPS208386H, RPS205347H, RPS207063H, RPS207864H, RPS208418H, RPS203399H, RPS203690H, RPS205670H, RPS205677H, RPS205680H, RPS205681H, RPS205964H, RPS205996H, RPS207185H, RPS207192H, RPS207206H, RPS207221H, RPS207228H, RPS207263H, RPS207269H, RPS202204H, RPS202306H, RPS202504H, RPS202528H, RPS202991H, RPS202993H, RPS203555H, RPS203748H, RPS203768H, RPS203777H, RPS203867H, RPS203875H, RPS204193H, RPS204214H, RPS204232H, RPS204332H, RPS204523H, RPS205220H, RPS205314H, RPS205378H, RPS205554H, RPS206011H, RPS206105H, RPS206703H, RPS207671H, RPS202516H, RPS202798H, RPS203288H, RPS203558H, RPS204506H, RPS205393H, RPS203024H, RPS202229H, RPS202433H, RPS202719H, RPS203017H, RPS203632H, RPS203983H, RPS203986H, RPS204155H, RPS204472H, RPS205105H, RPS205384H, RPS205589H, RPS205594H, RPS205628H, RPS205788H, RPS205820H, RPS207075H, RPS207301H, RPS207305H, RPS207307H, RPS207308H, RPS207309H, RPS207310H, RPS207312H, RPS207313H, RPS207453H, RPS208375H, RPS208664H, RPS202729H, RPS202962H, RPS203308H, RPS204590H, RPS204686H, RPS205363H, RPS205403H, RPS205944H, RPS206882H, RPS205361H, RPS208500H, RPS206376H, RPS206814H, RPS207116H, RPS208311H, RPS202182H, RPS204002H, RPS207163H, RPS206166H, RPS203076H, RPS203366H, RPS204309H, RPS206785H, RPS208357H, RPS204761H, RPS202526H, RPS203561H, RPS203578H, RPS206684H, RPS207011H, RPS208692H, RPS203136H, RPS204242H, RPS204435H, RPS206025H, RPS206453H, RPS207920H, RPS207133H, RPS208160H, RPS203003H, RPS203193H, RPS203792H, RPS205377H, RPS205494H, RPS205865H, RPS205948H, RPS206393H, RPS208470H, RPS202272H, RPS202588H, RPS202589H, RPS202591H, RPS202593H, RPS202594H, RPS202598H, RPS206183H, RPS206184H, RPS206186H, RPS206189H, RPS206194H, RPS207081H, RPS203377H, RPS204312H, RPS206132H, RPS206522H, RPS207430H, RPS208545H, RPS205168H, RPS208410H, RPS203260H, RPS203622H, RPS206736H, RPS207748H, RPS207884H, RPS207900H, RPS203881H, RPS204354H, RPS205412H, RPS205795H, RPS205747H, RPS202519H, RPS202121H, RPS202212H, RPS203104H, RPS204431H, RPS205428H, RPS205586H, RPS208474H, RPS203959H, RPS205864H, RPS205868H, RPS206445H, RPS207771H, RPS203684H, RPS204600H, RPS205424H, RPS206809H, RPS207730H, RPS208165H, RPS203117H, RPS204188H, RPS207195H, RPS208477H, RPS206922H, RPS202173H, RPS202227H, RPS203301H, RPS206036H, RPS206201H, RPS207007H, RPS207868H, RPS204868H, RPS208611H, RPS203840H, RPS204449H, RPS206021H, RPS208025H, RPS202859H, RPS203949H, RPS204120H, RPS204792H, RPS204914H, RPS202226H, RPS203724H, RPS204916H, RPS205008H, RPS206632H, RPS202784H, RPS205561H, RPS206784H, RPS202356H, RPS203886H, RPS206045H, RPS207077H, RPS207167H, RPS208631H, RPS206465H, RPS202033H, RPS203204H, RPS204532H, RPS204533H, RPS205002H, RPS205679H, RPS206054H, RPS206139H, RPS206300H, RPS207524H, RPS207571H, RPS207587H, RPS208653H, RPS203974H, RPS205625H, RPS206359H, RPS208070H, RPS204244H, RPS205892H, RPS207032H, RPS202424H, RPS206058H, RPS208359H, RPS203235H, RPS204261H, RPS205107H, RPS205118H, RPS205121H, RPS206067H, RPS206513H, RPS206645H, RPS208247H, RPS208625H, RPS203306H, RPS205987H, RPS206896H, RPS205379H, RPS207727H, RPS203858H, RPS204893H, RPS204895H, RPS204896H, RPS204897H, RPS204898H, RPS202029H, RPS203127H, RPS204311H, RPS205676H, RPS206398H, RPS206871H, RPS204044H, RPS205106H, RPS206098H, RPS208263H, RPS203026H, RPS203539H, RPS203540H, RPS205624H, RPS205963H, RPS206479H, RPS206488H, RPS206489H, RPS206649H, RPS206805H, RPS207000H, RPS207013H, RPS207403H, RPS207404H, RPS207405H, RPS207450H, RPS207451H, RPS208139H, RPS208250H, RPS204783H, RPS207036H, RPS203639H, RPS203654H, RPS203659H, RPS203663H, RPS203664H, RPS204909H, RPS205108H, RPS205267H, RPS206237H, RPS206298H, RPS207823H, RPS204142H, RPS202181H, RPS203280H, RPS203383H, RPS203511H, RPS202170H, RPS203837H, RPS206124H, RPS202508H, RPS205435H, RPS205966H, RPS206345H, RPS206463H, RPS207048H, RPS208465H, RPS203237H, RPS204457H, RPS205261H, RPS207079H, RPS206748H, RPS204121H, RPS202484H, RPS202647H, RPS202648H, RPS203346H, RPS203360H, RPS203496H, RPS203905H, RPS204324H, RPS204487H, RPS204966H, RPS205188H, RPS205206H, RPS206122H, RPS206904H, RPS208582H, RPS208672H, RPS202780H, RPS205019H, RPS202165H, RPS203386H, RPS203740H, RPS203969H, RPS203993H, RPS205441H, RPS205504H, RPS206583H, RPS206672H, RPS207140H, RPS207575H, RPS207725H, RPS203455H, RPS202626H, RPS202629H, RPS202641H, RPS204039H, RPS204283H, RPS207879H, RPS208660H, RPS202126H, RPS202905H, RPS202935H, RPS203274H, RPS203417H, RPS203872H, RPS204888H, RPS205295H, RPS205471H, RPS207952H, RPS203923H, RPS208040H, RPS208432H, RPS208533H, RPS204368H, RPS204871H, RPS205557H, RPS206446H, RPS207700H, RPS207701H, RPS207831H, RPS207928H, RPS206404H, RPS206406H, RPS202026H, RPS204268H, RPS205039H, RPS205902H, RPS208256H, RPS206375H, RPS202998H, RPS203429H, RPS206209H, RPS206238H, RPS206565H, RPS207540H, RPS202482H, RPS206840H, RPS203335H, RPS204930H, RPS204993H, RPS206880H, RPS202779H, RPS203176H, RPS204717H, RPS207465H, RPS207798H, RPS203043H, RPS203382H, RPS203723H, RPS205252H, RPS205285H, RPS206144H, RPS207072H, RPS207773H, RPS207779H, RPS208525H, RPS202721H, RPS202858H, RPS206516H, RPS203516H, RPS202487H, RPS205812H, RPS207102H, RPS206342H, RPS206884H, RPS207693H, RPS207903H, RPS202215H, RPS202316H, RPS202411H, RPS203454H, RPS203814H, RPS203988H, RPS204090H, RPS204330H, RPS204352H, RPS204416H, RPS204446H, RPS204512H, RPS205562H, RPS205945H, RPS206150H, RPS206244H, RPS206366H, RPS206466H, RPS206468H, RPS206469H, RPS206470H, RPS206474H, RPS206475H, RPS206481H, RPS206483H, RPS206490H, RPS206493H, RPS206494H, RPS206495H, RPS206498H, RPS206507H, RPS206510H, RPS206609H, RPS206782H, RPS206806H, RPS206865H, RPS206893H, RPS207183H, RPS207198H, RPS207200H, RPS207207H, RPS207208H, RPS207209H, RPS207214H, RPS207215H, RPS207218H, RPS207220H, RPS207225H, RPS207227H, RPS207229H, RPS207233H, RPS207745H, RPS207791H, RPS207909H, RPS208144H, RPS208296H, RPS208374H, RPS208390H, RPS208656H, RPS208684H, RPS203107H, RPS205136H, RPS205914H, RPS206149H, RPS207738H, RPS207844H, RPS208157H, RPS206678H, RPS206747H, RPS202445H, RPS203806H, RPS204520H, RPS205457H, RPS205913H, RPS206990H, RPS202925H, RPS202988H, RPS203167H, RPS204295H, RPS205947H, RPS206095H, RPS206131H, RPS206437H, RPS206440H, RPS207972H, RPS208461H, RPS208647H, RPS202825H, RPS203423H, RPS203585H, RPS203627H, RPS203839H, RPS203863H, RPS204016H, RPS204163H, RPS205358H, RPS206305H, RPS206976H, RPS207155H, RPS207549H, RPS207916H, RPS208050H, RPS208356H, RPS203337H, RPS203499H, RPS204027H, RPS207058H, RPS202902H, RPS203857H, RPS203860H, RPS206262H, RPS207359H, RPS207375H, RPS207385H, RPS208405H, RPS208507H, RPS202501H, RPS202368H, RPS202115H, RPS203394H, RPS203981H, RPS204337H, RPS206115H, RPS208205H, RPS208515H, RPS202754H, RPS202808H, RPS203331H, RPS204304H, RPS203852H, RPS205518H, RPS202205H, RPS202505H, RPS203843H, RPS204087H, RPS207362H, RPS207845H, RPS202432H, RPS208427H, RPS205219H, RPS206899H, RPS207523H, RPS207882H, RPS208621H, RPS205498H, RPS202100H, RPS202325H, RPS202600H, RPS202830H, RPS203336H, RPS203686H, RPS203737H, RPS203963H, RPS204017H, RPS204020H, RPS204292H, RPS204361H, RPS204398H, RPS204409H, RPS204453H, RPS204539H, RPS205253H, RPS205374H, RPS205575H, RPS205622H, RPS205693H, RPS205980H, RPS206040H, RPS206060H, RPS206108H, RPS206188H, RPS206197H, RPS206310H, RPS206339H, RPS206350H, RPS206551H, RPS206559H, RPS206566H, RPS206568H, RPS206575H, RPS206579H, RPS206584H, RPS206592H, RPS206700H, RPS206718H, RPS206798H, RPS206813H, RPS206847H, RPS206862H, RPS207216H, RPS207278H, RPS207337H, RPS207380H, RPS207390H, RPS207457H, RPS207458H, RPS207460H, RPS207461H, RPS207466H, RPS207470H, RPS207475H, RPS207477H, RPS207478H, RPS207871H, RPS208068H, RPS208073H, RPS208074H, RPS208075H, RPS208078H, RPS208080H, RPS208082H, RPS208083H, RPS208116H, RPS208467H, RPS208652H, RPS208654H, RPS205481H, RPS203155H, RPS204114H, RPS207039H, RPS207124H, RPS208035H, RPS208233H, RPS206576H, RPS203019H, RPS207505H, RPS202216H, RPS202898H, RPS202972H, RPS203390H, RPS203573H, RPS203696H, RPS203823H, RPS204246H, RPS204322H, RPS204335H, RPS204513H, RPS204542H, RPS204952H, RPS205217H, RPS205502H, RPS205619H, RPS205856H, RPS206034H, RPS206293H, RPS206433H, RPS206600H, RPS206833H, RPS207110H, RPS207560H, RPS207609H, RPS207856H, RPS208552H, RPS208645H, RPS204258H, RPS204935H, RPS206318H, RPS208526H, RPS203122H, RPS203467H, RPS203846H, RPS204540H, RPS204592H, RPS205822H, RPS206496H, RPS206499H, RPS207433H, RPS208392H, RPS202845H, RPS204406H, RPS205203H, RPS202566H, RPS203527H, RPS203577H, RPS205591H, RPS207004H, RPS208086H, RPS202705H, RPS205587H, RPS207604H, RPS202401H, RPS202896H, RPS202918H, RPS202986H, RPS203139H, RPS204091H, RPS204772H, RPS206482H, RPS206504H, RPS208103H, RPS207805H, RPS208384H, RPS203398H, RPS203584H, RPS205291H, RPS203146H, RPS204001H, RPS207506H, RPS203888H, RPS206662H, RPS203322H, RPS203697H, RPS205290H, RPS206680H, RPS203730H, RPS203877H, RPS203188H, RPS205531H, RPS204944H, RPS205149H, RPS205185H, RPS205590H, RPS207656H, RPS207870H, RPS208024H, RPS203217H, RPS204058H, RPS205054H, RPS205993H, RPS207662H, RPS202663H, RPS204985H, RPS207108H, RPS204853H, RPS207165H, RPS206741H, RPS207907H, RPS208267H, RPS204765H, RPS206357H, RPS207855H, RPS203379H, RPS206026H, RPS207842H, RPS208613H, RPS206881H, RPS204041H, RPS207983H, RPS202309H, RPS202765H, RPS202979H, RPS202995H, RPS203047H, RPS203345H, RPS203385H, RPS203767H, RPS205308H, RPS206948H, RP
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
- Reason for Recall:
- There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
CRTD DTMA1Q1 CLARIA MRI QUAD US DF1, Model Number DTMA1Q1; Implantable Cardioverter Defibrillators
Product Codes/Lot Numbers:
GTIN 00643169837669, Lot Serial numbers: RPS205974H, RPS208665H, RPS206712H, RPS207744H, RPS205038H, RPS205791H, RPS206225H, RPS208216H, RPS206084H, RPS203804H, RPS208198H, RPS208245H, RPS205937H, RPS205565H, RPS206114H, RPS206723H, RPS206730H, RPS206733H, RPS206734H, RPS206754H, RPS207303H, RPS207318H, RPS207338H, RPS207399H, RPS207703H, RPS207711H, RPS207722H, RPS208049H, RPS208505H, RPS208646H, RPS208679H, RPS205303H, RPS205918H, RPS203315H, RPS204385H, RPS205650H, RPS202760H, RPS203013H, RPS203034H, RPS203256H, RPS203610H, RPS205372H, RPS206383H, RPS206771H, RPS207603H, RPS207933H, RPS208110H, RPS208492H, RPS204009H, RPS202981H, RPS207563H, RPS208546H, RPS202245H, RPS203960H, RPS204342H, RPS204357H, RPS205958H, RPS207822H, RPS207840H, RPS207922H, RPS207923H, RPS207984H, RPS208255H, RPS208529H, RPS208575H, RPS208590H, RPS208615H, RPS203563H, RPS206128H, RPS206413H, RPS206762H, RPS207904H, RPS208608H, RPS202171H, RPS202841H, RPS202860H, RPS203565H, RPS204198H, RPS202792H, RPS202793H, RPS206515H, RPS202224H, RPS203192H, RPS203545H, RPS203910H, RPS204035H, RPS204166H, RPS204197H, RPS204394H, RPS204492H, RPS205560H, RPS206777H, RPS207146H, RPS207520H, RPS208166H, RPS204464H, RPS204776H, RPS207117H, RPS204003H, RPS205323H, RPS206063H, RPS208464H, RPS206252H, RPS202211H, RPS202602H, RPS202612H, RPS205006H, RPS205638H, RPS205639H, RPS205710H, RPS205998H, RPS207105H, RPS207368H, RPS207409H, RPS202030H, RPS202067H, RPS202507H, RPS206211H, RPS206605H, RPS207137H, RPS208591H, RPS202053H, RPS202511H, RPS202116H, RPS202210H, RPS202273H, RPS202274H, RPS202276H, RPS202307H, RPS202349H, RPS202427H, RPS202447H, RPS202498H, RPS202527H, RPS202601H, RPS202603H, RPS202605H, RPS203195H, RPS203197H, RPS203431H, RPS203432H, RPS203433H, RPS203435H, RPS203436H, RPS203438H, RPS203439H, RPS203440H, RPS203441H, RPS203442H, RPS203444H, RPS204202H, RPS204209H, RPS204210H, RPS204212H, RPS204215H, RPS204217H, RPS204223H, RPS204926H, RPS204978H, RPS204980H, RPS204983H, RPS204986H, RPS204987H, RPS204988H, RPS204992H, RPS204997H, RPS205262H, RPS205271H, RPS205276H, RPS205277H, RPS205665H, RPS205714H, RPS205717H, RPS205719H, RPS205720H, RPS205737H, RPS205739H, RPS205741H, RPS205742H, RPS205744H, RPS205750H, RPS205751H, RPS205752H, RPS205754H, RPS205755H, RPS205756H, RPS205757H, RPS205758H, RPS205759H, RPS206175H, RPS206179H, RPS206235H, RPS206367H, RPS206371H, RPS206374H, RPS206378H, RPS206379H, RPS207186H, RPS207187H, RPS207247H, RPS207248H, RPS207252H, RPS207253H, RPS207254H, RPS207255H, RPS207256H, RPS207258H, RPS207260H, RPS207264H, RPS207298H, RPS207393H, RPS207402H, RPS207574H, RPS207774H, RPS207994H, RPS208133H, RPS208237H, RPS208240H, RPS208242H, RPS208327H, RPS208457H, RPS207507H, RPS205911H, RPS205204H, RPS206761H, RPS207437H, RPS207712H, RPS204627H, RPS205509H, RPS205082H, RPS202759H, RPS203251H, RPS202130H, RPS202167H, RPS203220H, RPS204656H, RPS205410H, RPS205487H, RPS207695H, RPS207867H, RPS204362H, RPS208517H, RPS203913H, RPS205806H, RPS205807H, RPS205808H, RPS202277H, RPS204053H, RPS205975H, RPS206636H, RPS208079H, RPS208119H, RPS208349H, RPS204604H, RPS205654H, RPS205765H, RPS202490H, RPS203352H, RPS203816H, RPS206903H, RPS202237H, RPS202704H, RPS203037H, RPS203532H, RPS203707H, RPS203763H, RPS204225H, RPS205169H, RPS205364H, RPS205648H, RPS205731H, RPS205789H, RPS206089H, RPS206450H, RPS206537H, RPS207592H, RPS207719H, RPS208334H, RPS202531H, RPS203458H, RPS204162H, RPS205383H, RPS205492H, RPS208444H, RPS207494H, RPS202057H, RPS202202H, RPS202567H, RPS202849H, RPS202906H, RPS202916H, RPS203068H, RPS203332H, RPS203660H, RPS203726H, RPS203796H, RPS204204H, RPS204613H, RPS205849H, RPS205850H, RPS205851H, RPS205852H, RPS206153H, RPS206452H, RPS206455H, RPS206456H, RPS206459H, RPS206963H, RPS208177H, RPS203421H, RPS203524H, RPS203615H, RPS203770H, RPS204642H, RPS205009H, RPS205058H, RPS205134H, RPS205583H, RPS205990H, RPS206633H, RPS206756H, RPS206947H, RPS207473H, RPS207927H, RPS204248H, RPS203955H, RPS206451H, RPS207611H, RPS207945H, RPS202288H, RPS202290H, RPS203479H, RPS204022H, RPS204172H, RPS204173H, RPS205800H, RPS205801H, RPS205802H, RPS205803H, RPS205805H, RPS206527H, RPS206531H, RPS206540H, RPS208266H, RPS202776H, RPS203086H, RPS202768H, RPS203253H, RPS205214H, RPS207088H, RPS207485H, RPS207487H, RPS207488H, RPS207885H, RPS208313H, RPS202656H, RPS205497H, RPS207691H, RPS202052H, RPS202249H, RPS202955H, RPS203776H, RPS203928H, RPS205585H, RPS207002H, RPS207099H, RPS207354H, RPS207447H, RPS207689H, RPS202453H, RPS202547H, RPS203069H, RPS203115H, RPS203163H, RPS203246H, RPS203407H, RPS203971H, RPS204134H, RPS204135H, RPS204139H, RPS204140H, RPS205733H, RPS205951H, RPS206134H, RPS206165H, RPS206213H, RPS206303H, RPS207173H, RPS207184H, RPS207241H, RPS207243H, RPS207249H, RPS203602H, RPS203678H, RPS204908H, RPS206177H, RPS206267H, RPS206555H, RPS204760H, RPS205535H, RPS202140H, RPS203633H, RPS203635H, RPS204218H, RPS204219H, RPS204220H, RPS205880H, RPS205883H, RPS205884H, RPS206407H, RPS202761H, RPS203704H, RPS205380H, RPS202881H, RPS202738H, RPS202799H, RPS203812H, RPS204143H, RPS206221H, RPS206611H, RPS203092H, RPS203824H, RPS207474H, RPS207599H, RPS202781H, RPS204000H, RPS205399H, RPS205642H, RPS206625H, RPS205858H, RPS207910H, RPS202021H, RPS202074H, RPS202573H, RPS202574H, RPS207292H, RPS208244H, RPS208254H, RPS207297H, RPS202655H, RPS203077H, RPS203655H, RPS207436H, RPS205848H, RPS204773H, RPS205886H, RPS207053H, RPS207979H, RPS202035H, RPS202296H, RPS203753H, RPS204353H, RPS205506H, RPS205816H, RPS206044H, RPS206083H, RPS206648H, RPS206770H, RPS207492H, RPS207543H, RPS207914H, RPS208306H, RPS208167H, RPS208419H, RPS208658H, RPS205390H, RPS207954H, RPS208554H, RPS208555H, RPS202718H, RPS202910H, RPS204089H, RPS207768H, RPS207849H, RPS207837H, RPS208118H, RPS208413H, RPS205895H, RPS202976H, RPS203159H, RPS202200H, RPS203162H, RPS203295H, RPS205887H, RPS208633H, RPS203215H, RPS205831H, RPS206886H, RPS207554H, RPS202160H, RPS203327H, RPS207796H, RPS208406H, RPS206749H, RPS202228H, RPS202231H, RPS202874H, RPS203456H, RPS204874H, RPS205547H, RPS206690H, RPS204943H, RPS202400H, RPS203158H, RPS203178H, RPS203934H, RPS204147H, RPS204278H, RPS205362H, RPS205365H, RPS205429H, RPS205953H, RPS206106H, RPS206138H, RPS207815H, RPS207919H, RPS207987H, RPS208362H, RPS208576H, RPS202741H, RPS203951H, RPS206069H, RPS206348H, RPS204270H, RPS205421H, RPS206117H, RPS202314H, RPS202425H, RPS202582H, RPS202945H, RPS203330H, RPS203451H, RPS203566H, RPS203567H, RPS203612H, RPS203818H, RPS203952H, RPS204050H, RPS204071H, RPS204150H, RPS204348H, RPS204443H, RPS204954H, RPS205125H, RPS205163H, RPS205343H, RPS205375H, RPS205814H, RPS207931H, RPS207344H, RPS207397H, RPS208487H, RPS203072H, RPS203381H, RPS206250H, RPS206268H, RPS207607H, RPS207706H, RPS207860H, RPS204922H, RPS202020H, RPS202304H, RPS202982H, RPS203231H, RPS205552H, RPS207790H, RPS203265H, RPS205968H, RPS206533H, RPS206776H, RPS206826H, RPS202083H, RPS203533H, RPS204247H, RPS204860H, RPS206325H, RPS206621H, RPS207119H, RPS207633H, RPS203321H, RPS208161H, RPS206938H, RPS204403H, RPS207683H, RPS203168H, RPS205415H, RPS208169H, RPS205890H, RPS203239H, RPS203775H, RPS206631H, RPS202564H, RPS205300H, RPS205074H, RPS207635H, RPS206802H, RPS202055H, RPS202372H, RPS202636H, RPS202665H, RPS203134H, RPS203255H, RPS203339H, RPS203392H, RPS203741H, RPS203805H, RPS203838H, RPS203865H, RPS203889H, RPS203982H, RPS204015H, RPS204097H, RPS204231H, RPS204241H, RPS204296H, RPS204355H, RPS204441H, RPS204598H, RPS204759H, RPS204778H, RPS204955H, RPS205010H, RPS205079H, RPS205114H, RPS205215H, RPS205240H, RPS205242H, RPS205248H, RPS205289H, RPS205304H, RPS205345H, RPS205431H, RPS205511H, RPS205543H, RPS205569H, RPS205661H, RPS205992H, RPS206052H, RPS206062H, RPS206066H, RPS206192H, RPS206220H, RPS206253H, RPS206259H, RPS206292H, RPS206340H, RPS206423H, RPS206658H, RPS207170H, RPS207981H, RPS208012H, RPS208071H, RPS208138H, RPS208308H, RPS208343H, RPS208348H, RPS202127H, RPS203424H, RPS204387H, RPS206030H, RPS206543H, RPS206549H, RPS208171H, RPS208172H, RPS205977H, RPS206538H, RPS208047H, RPS206126H, RPS208669H, RPS202978H, RPS203083H, RPS203572H, RPS203871H, RPS204148H, RPS208491H, RPS204768H, RPS202460H, RPS202461H, RPS202463H, RPS203291H, RPS203597H, RPS204070H, RPS205211H, RPS205522H, RPS205545H, RPS205904H, RPS205905H, RPS206096H, RPS206301H, RPS206323H, RPS206590H, RPS206591H, RPS207069H, RPS207070H, RPS207143H, RPS207639H, RPS208001H, RPS208352H, RPS208510H, RPS208644H, RPS203028H, RPS202870H, RPS203325H, RPS204607H, RPS205130H, RPS205132H, RPS208057H, RPS202639H, RPS202643H, RPS202644H, RPS205020H, RPS205021H, RPS205187H, RPS205503H, RPS205558H, RPS205778H, RPS205780H, RPS205840H, RPS205853H, RPS205885H, RPS205967H, RPS208423H, RPS208424H, RPS203413H, RPS205862H, RPS202552H, RPS203284H, RPS203521H, RPS203830H, RPS203918H, RPS204008H, RPS204199H, RPS204203H, RPS206783H, RPS206892H, RPS208135H, RPS205908H, RPS207594H, RPS207824H, RPS207917H, RPS202977H, RPS206042H, RPS206620H, RPS206713H, RPS207673H, RPS207808H, RPS208563H, RPS205949H, RPS206794H, RPS207929H, RPS204229H, RPS204323H, RPS204298H, RPS205601H, RPS206005H, RPS206263H, RPS207068H, RPS202282H, RPS203110H, RPS204525H, RPS204119H, RPS204377H, RPS206855H, RPS207018H, RPS207197H, RPS203483H, RPS203484H, RPS203488H, RPS204176H, RPS204177H, RPS204178H, RPS204180H, RPS204181H, RPS204185H, RPS204189H, RPS204190H, RPS204192H, RPS207234H, RPS207238H, RPS207240H, RPS207242H, RPS207244H, RPS207245H, RPS207246H, RPS208314H, RPS208325H, RPS208329H, RPS208332H, RPS208333H, RPS203408H, RPS205446H, RPS207943H, RPS207051H, RPS207148H, RPS207993H, RPS202739H, RPS204118H, RPS205461H, RPS203944H, RPS205419H, RPS206635H, RPS206696H, RPS207893H, RPS208483H, RPS203758H, RPS206314H, RPS203643H, RPS203702H, RPS203903H, RPS203906H, RPS206772H, RPS207160H, RPS207273H, RPS208140H, RPS203249H, RPS203803H, RPS204787H, RPS205209H, RPS205357H, RPS206742H, RPS207832H, RPS208291H, RPS202722H, RPS204059H, RPS204615H, RPS205084H, RPS205095H, RPS207663H, RPS202753H, RPS202291H, RPS202714H, RPS202965H, RPS203216H, RPS203631H, RPS203987H, RPS206671H, RPS207377H, RPS207780H, RPS208123H, RPS205643H, RPS205331H, RPS207930H, RPS206587H, RPS208098H, RPS203081H, RPS207396H, RPS208541H, RPS206384H, RPS207530H, RPS204266H, RPS208605H, RPS203738H, RPS206240H, RPS206738H, RPS207519H, RPS207967H, RPS208249H, RPS208570H, RPS202444H, RPS202451H, RPS203460H, RPS205468H, RPS205603H, RPS205609H, RPS205909H, RPS206091H, RPS206890H, RPS206977H, RPS207387H, RPS208479H, RPS202450H, RPS208353H, RPS202113H, RPS204124H, RPS204343H, RPS204471H, RPS204795H, RPS205086H, RPS205162H, RPS205312H, RPS206228H, RPS206271H, RPS206289H, RPS206355H, RPS207277H, RPS207561H, RPS207616H, RPS208372H, RPS202470H, RPS202471H, RPS202551H, RPS202571H, RPS202590H, RPS202901H, RPS202969H, RPS203225H, RPS203226H, RPS203227H, RPS203230H, RPS203491H, RPS204507H, RPS204508H, RPS204970H, RPS204972H, RPS204976H, RPS205452H, RPS206015H, RPS206187H, RPS206725H, RPS203207H, RPS207517H, RPS207874H, RPS202682H, RPS202791H, RPS203425H, RPS206135H, RPS206657H, RPS207153H, RPS202125H, RPS202143H, RPS202213H, RPS203367H, RPS204418H, RPS205329H, RPS206285H, RPS207717H, RPS208579H, RPS202039H, RPS202128H, RPS203015H, RPS203709H, RPS203762H, RPS204222H, RPS204456H, RPS204779H, RPS204919H, RPS205319H, RPS205657H, RPS206858H, RPS206860H, RPS206861H, RPS207156H, RPS207438H, RPS207848H, RPS208403H, RPS208429H, RPS208514H, RPS208564H, RPS202157H, RPS202672H, RPS202775H, RPS203766H, RPS204280H, RPS205988H, RPS202136H, RPS202921H, RPS204700H, RPS205302H, RPS205559H, RPS207023H, RPS207086H, RPS207315H, RPS207707H, RPS207942H, RPS208022H, RPS208303H, RPS208475H, RPS202893H, RPS203943H, RPS205243H, RPS206049H, RPS203050H, RPS203101H, RPS208433H, RPS203085H, RPS203476H, RPS204920H, RPS204360H, RPS204489H, RPS204852H, RPS205614H, RPS202503H, RPS202762H, RPS203266H, RPS203492H, RPS205068H, RPS205580H, RPS205763H, RPS206295H, RPS207531H, RPS207541H, RPS207547H, RPS208176H, RPS208648H, RPS202668H, RPS203126H, RPS205210H, RPS208527H, RPS203359H, RPS203592H, RPS205863H, RPS206829H, RPS208489H, RPS204685H, RPS208331H, RPS203212H, RPS203931H, RPS204132H, RPS204380H, RPS204968H, RPS205678H, RPS207016H, RPS207596H, RPS205387H, RPS202223H, RPS206521H, RPS206779H, RPS204427H, RPS207880H, RPS208498H, RPS208567H, RPS202608H, RPS207613H, RPS208573H, RPS203935H, RPS202708H, RPS202727H, RPS203725H, RPS203734H, RPS204620H, RPS205512H, RPS206750H, RPS207782H, RPS208037H, RPS208446H, RPS203765H, RPS202510H, RPS203472H, RPS203473H, RPS205024H, RPS205025H, RPS206509H, RPS207414H, RPS204014H, RPS204111H, RPS206077H, RPS206670H, RPS206746H, RPS208295H, RPS202036H, RPS204224H, RPS206503H, RPS206506H, RPS207010H, RPS202755H, RPS204659H, RPS204780H, RPS205376H, RPS206449H, RPS206388H, RPS205655H, RPS202491H, RPS205600H, RPS206801H, RPS202794H, RPS202795H, RPS204286H, RPS204288H, RPS205088H, RPS206290H, RPS206517H, RPS206548H, RPS207045H, RPS208127H, RPS208128H, RPS208129H, RPS208130H, RPS208131H, RPS206087H, RPS202736H, RPS202737H, RPS202929H, RPS205237H, RPS205311H, RPS207968H, RPS205430H, RPS206200H, RPS206249H, RPS206333H, RPS206905H, RPS208039H, RPS208345H, RPS206276H, RPS206159H, RPS208032H, RPS203520H, RPS206744H, RPS202426H, RPS203410H, RPS203099H, RPS203169H, RPS203579H, RPS204450H, RPS205013H, RPS205288H, RPS206674H, RPS204052H, RPS204692H, RPS204940H, RPS205128H, RPS205178H, RPS206627H, RPS206640H, RPS208683H, RPS205051H, RPS205052H, RPS205607H, RPS207271H, RPS207843H, RPS207939H, RPS207953H, RPS208163H, RPS208584H, RPS205190H, RPS206215H, RPS206258H, RPS202772H, RPS202038H, RPS203623H, RPS205936H, RPS202079H, RPS202305H, RPS202654H, RPS202947H, RPS204010H, RPS204042H, RPS204145H, RPS204509H, RPS205228H, RPS205623H, RPS206072H, RPS206823H, RPS207109H, RPS207132H, RPS207525H, RPS207556H, RPS207620H, RPS207801H, RPS207803H, RPS207816H, RPS207817H, RPS208309H, RPS208360H, RPS208674H, RPS202675H, RPS202676H, RPS202680H, RPS204395H, RPS205282H, RPS205704H, RPS205705H, RPS205706H, RPS206370H, RPS206401H, RPS206402H, RPS207422H, RPS207425H, RPS207428H, RPS208006H, RPS208007H, RPS208640H, RPS208666H, RPS203426H, RPS203854H, RPS205534H, RPS206602H, RPS207677H, RPS207940H, RPS205233H, RPS208180H, RPS202315H, RPS206202H, RPS206307H, RPS206442H, RPS207938H, RPS208445H, RPS204428H, RPS208145H, RPS202871H, RPS203090H, RPS203114H, RPS205195H, RPS205316H, RPS206630H, RPS207161H, RPS202823H, RPS203537H, RPS204339H, RPS204537H, RPS206028H, RPS206830H, RPS208168H, RPS206971H, RPS203593H, RPS204754H, RPS204756H, RPS206288H, RPS207415H, RPS207421H, RPS202584H, RPS203586H, RPS203716H, RPS203789H, RPS203862H, RPS204879H, RPS205401H, RPS205698H, RPS205901H, RPS206644H, RPS206989H, RPS207022H, RPS207037H, RPS207073H, RPS207392H, RPS207443H, RPS207946H, RPS207978H, RPS208264H, RPS208271H, RPS208379H, RPS208437H, RPS208454H, RPS208482H, RPS208571H, RPS208614H, RPS206842H, RPS205318H, RPS205246H, RPS208422H, RPS202717H, RPS203147H, RPS203326H, RPS203703H, RPS203794H, RPS204397H, RPS206392H, RPS206535H, RPS207168H, RPS208187H, RPS204501H, RPS207061H, RPS207956H, RPS206603H, RPS208225H, RPS207532H, RPS203173H, RPS205245H, RPS205630H, RPS202032H, RPS205310H, RPS203933H, RPS204102H, RPS207969H, RPS204314H, RPS205416H, RPS205417H, RPS206243H, RPS207573H, RPS207999H, RPS202740H, RPS203228H, RPS204239H, RPS203283H, RPS204099H, RPS208054H, RPS208272H, RPS206216H, RPS204049H, RPS204528H, RPS206995H, RPS202179H, RPS202180H, RPS202409H, RPS203464H, RPS203841H, RPS205499H, RPS207046H, RPS205530H, RPS206614H, RPS203722H, RPS204766H, RPS202050H, RPS203744H, RPS203893H, RPS208574H, RPS204963H, RPS208301H, RPS202502H, RPS205458H, RPS205696H, RPS208322H, RPS203388H, RPS203713H, RPS202804H, RPS203554H, RPS205103H, RPS206394H, RPS206850H, RPS207915H, RPS208551H, RPS203904H, RPS205599H, RPS206330H, RPS206344H, RPS207806H, RPS205934H, RPS206716H, RPS206984H, RPS203175H, RPS206694H, RPS205221H, RPS207107H, RPS202525H, RPS203780H, RPS204493H, RPS208503H, RPS202247H, RPS202252H, RPS202327H, RPS202328H, RPS202355H, RPS203670H, RPS203671H, RPS203672H, RPS203679H, RPS203680H, RPS205701H, RPS205827H, RPS206726H, RPS207621H, RPS207789H, RPS205632H, RPS208185H, RPS208195H, RPS205615H, RPS208566H, RPS203535H, RPS203695H, RPS204789H, RPS204928H, RPS204921H, RPS202886H, RPS204396H, RPS206484H, RPS207720H, RPS207772H, RPS203309H, RPS206558H, RPS206758H, RPS208484H, RPS203613H, RPS202926H, RPS205984H, RPS208072H, RPS202756H, RPS208380H, RPS202812H, RPS203038H, RPS203261H, RPS203400H, RPS203676H, RPS203774H, RPS203778H, RPS204320H, RPS204444H, RPS204475H, RPS204476H, RPS204614H, RPS205145H, RPS205542H, RPS205611H, RPS207295H, RPS202558H, RPS206104H, RPS202092H, RPS202480H, RPS202963H, RPS206360H, RPS207017H, RPS203600H, RPS203965H, RPS206539H, RPS202782H, RPS202994H, RPS203132H, RPS203312H, RPS203614H, RPS204480H, RPS204936H, RPS205367H, RPS205388H, RPS205666H, RPS205669H, RPS205824H, RPS206198H, RPS206282H, RPS208061H, RPS208065H, RPS208153H, RPS203214H, RPS206422H, RPS206993H, RPS206113H, RPS202279H, RPS202399H, RPS202948H, RPS202967H, RPS203276H, RPS203591H, RPS203876H, RPS204912H, RPS205664H, RPS206110H, RPS206556H, RPS207391H, RPS206151H, RPS206182H, RPS205156H, RPS205875H, RPS206056H, RPS206542H, RPS206999H, RPS202185H, RPS202805H, RPS204131H, RPS204159H, RPS204160H, RPS202037H, RPS202056H, RPS202744H, RPS202826H, RPS202850H, RPS203414H, RPS203831H, RPS204606H, RPS205007H, RPS205427H, RPS205529H, RPS206024H, RPS206294H, RPS206683H, RPS207024H, RPS207510H, RPS207684H, RPS207734H, RPS207866H, RPS208623H, RPS206093H, RPS204934H, RPS205001H, RPS208548H, RPS202879H, RPS203937H, RPS204084H, RPS204616H, RPS205541H, RPS205916H, RPS205919H, RPS206057H, RPS206119H, RPS206120H, RPS207419H, RPS207615H, RPS207649H, RPS207672H, RPS208179H, RPS208511H, RPS208585H, RPS208580H, RPS208530H, RPS203868H, RPS203962H, RPS206642H, RPS205216H, RPS205727H, RPS205728H, RPS208616H, RPS203317H, RPS203364H, RPS203462H, RPS203651H, RPS207853H, RPS206659H, RPS205265H, RPS202743H, RPS204313H, RPS205989H, RPS207612H, RPS202044H, RPS202275H, RPS204404H, RPS205804H, RPS206766H, RPS208217H, RPS203048H, RPS203071H, RPS205332H, RPS207049H, RPS202807H, RPS203144H, RPS206103H, RPS206411H, RPS206541H, RPS208440H, RPS202746H, RPS206791H, RPS202187H, RPS202586H, RPS203154H, RPS205605H, RPS205893H, RPS206208H, RPS208204H, RPS208369H, RPS208549H, RPS207724H, RPS208366H, RPS202206H, RPS202209H, RPS203490H, RPS203898H, RPS203900H, RPS204490H, RPS204621H, RPS205546H, RPS205982H, RPS206306H, RPS207067H, RPS207638H, RPS208060H, RPS208207H, RPS203035H, RPS203834H, RPS203920H, RPS203950H, RPS205131H, RPS205227H, RPS205241H, RPS206245H, RPS206581H, RPS206819H, RPS207585H, RPS208090H, RPS208535H, RPS202196H, RPS208534H, RPS202219H, RPS202221H, RPS203791H, RPS204106H, RPS205396H, RPS208089H, RPS208147H, RPS206161H, RPS206606H, RPS206773H, RPS205226H, RPS206461H, RPS205402H, RPS202439H, RPS204929H, RPS205356H, RPS206900H, RPS207042H, RPS207370H, RPS207496H, RPS205142H, RPS206142H, RPS206191H, RPS206650H, RPS206810H, RPS207040H, RPS207321H, RPS207448H, RPS207508H, RPS208055H, RPS208456H, RPS204379H, RPS204381H, RPS204887H, RPS204979H, RPS205167H, RPS205266H, RPS205550H, RPS206287H, RPS208222H, RPS203293H, RPS203297H, RPS203849H, RPS204785H, RPS205915H, RPS206951H, RPS202078H, RPS203049H, RPS205556H, RPS206272H, RPS207162H, RPS207588H, RPS207348H, RPS208004H, RPS203371H, RPS203947H, RPS204371H, RPS205292H, RPS204383H, RPS204486H, RPS204781H, RPS208084H, RPS206397H, RPS207538H, RPS203329H, RPS207237H, RPS202777H, RPS203917H, RPS204251H, RPS204310H, RPS206080H, RPS206329H, RPS207899H, RPS204158H, RPS204228H, RPS202353H, RPS202434H, RPS202537H, RPS203041H, RPS203254H, RPS203344H, RPS203466H, RPS204051H, RPS204179H, RPS204281H, RPS204376H, RPS205151H, RPS205152H, RPS205634H, RPS205649H, RPS208542H, RPS205513H, RPS205869H, RPS208286H, RPS208550H, RPS203798H, RPS204317H, RPS202018H, RPS202788H, RPS203030H, RPS204274H, RPS205610H, RPS205933H, RPS205941H, RPS206214H, RPS206751H, RPS207052H, RPS207618H, RPS203938H, RPS206641H, RPS202734H, RPS202787H, RPS203948H, RPS207286H, RPS207516H, RPS207674H, RPS206933H, RPS203199H, RPS203727H, RPS207514H, RPS207572H, RPS208149H, RPS208261H, RPS208029H, RPS202233H, RPS203190H, RPS203570H, RPS203945H, RPS206932H, RPS208238H, RPS202607H, RPS203011H, RPS206502H, RPS208275H, RPS205507H, RPS205268H, RPS205400H, RPS208027H, RPS203252H, RPS204494H, RPS207728H, RPS202142H, RPS202469H, RPS202767H, RPS202907H, RPS203782H, RPS203964H, RPS203970H, RPS204123H, RPS204411H, RPS204474H, RPS204774H, RPS205408H, RPS205515H, RPS205970H, RPS206217H, RPS206261H, RPS206638H, RPS206765H, RPS207123H, RPS207996H, RPS208221H, RPS208603H, RPS208622H, RPS203700H, RPS205548H, RPS207961H, RPS208670H, RPS202151H, RPS203333H, RPS204275H, RPS204364H, RPS205251H, RPS206239H, RPS206885H, RPS207129H, RPS207511H, RPS202022H, RPS202133H, RPS202514H, RPS203172H, RPS203302H, RPS203319H, RPS203363H, RPS203689H, RPS204006H, RPS204182H, RPS204205H, RPS205244H, RPS205443H, RPS205924H, RPS205985H, RPS206327H, RPS206386H, RPS206534H, RPS206573H, RPS206676H, RPS206677H, RPS207071H, RPS207180H, RPS207626H, RPS207769H, RPS208036H, RPS208354H, RPS208367H, RPS208659H, RPS202664H, RPS202667H, RPS202671H, RPS205825H, RPS205826H, RPS206065H, RPS206078H, RPS206085H, RPS205275H, RPS202725H, RPS202545H, RPS202803H, RPS202891H, RPS203006H, RPS203268H, RPS203596H, RPS203747H, RPS204849H, RPS204850H, RPS205525H, RPS206001H, RPS207112H, RPS207963H, RPS207976H, RPS208008H, RPS208595H, RPS202885H, RPS205306H, RPS202268H, RPS202455H, RPS202475H, RPS202822H, RPS203118H, RPS203180H, RPS203354H, RPS203412H, RPS203526H, RPS203735H, RPS203878H, RPS203989H, RPS204116H, RPS204365H, RPS205234H, RPS205334H, RPS205366H, RPS205566H, RPS206876H, RPS207912H, RPS208397H, RPS208587H, RPS208685H, RPS202483H, RPS202524H, RPS203620H, RPS206205H, RPS208588H, RPS208636H, RPS202321H, RPS203082H, RPS203285H, RPS203547H, RPS204430H, RPS204432H, RPS204433H, RPS204436H, RPS204437H, RPS205712H, RPS206888H, RPS207287H, RPS207288H, RPS208115H, RPS208215H, RPS202800H, RPS202813H, RPS203232H, RPS203234H, RPS203866H, RPS205479H, RPS205551H, RPS206764H, RPS207169H, RPS207565H, RPS202964H, RPS203919H, RPS204183H, RPS207512H, RPS208011H, RPS202651H, RPS202660H, RPS202677H, RPS202688H, RPS202693H, RPS202697H, RPS202702H, RPS203493H, RPS203500H, RPS203749H, RPS205023H, RPS205029H, RPS205033H, RPS205035H, RPS205041H, RPS205042H, RPS205046H, RPS205048H, RPS205083H, RPS205538H, RPS205540H, RPS205596H, RPS205734H, RPS206265H, RPS206520H, RPS206687H, RPS206832H, RPS206834H, RPS206835H, RPS206907H, RPS207697H, RPS207758H, RPS207784H, RPS207807H, RPS207811H, RPS208000H, RPS208028H, RPS208033H, RPS208058H, RPS208175H, RPS208228H, RPS208297H, RPS208504H, RPS208516H, RPS208553H, RPS208594H, RPS208629H, RPS206686H, RPS208407H, RPS203953H, RPS204108H, RPS204424H, RPS206133H, RPS206815H, RPS207050H, RPS208137H, RPS208197H, RPS206334H, RPS206508H, RPS207122H, RPS205092H, RPS205099H, RPS205776H, RPS205872H, RPS207590H, RPS208209H, RPS208435H, RPS208520H, RPS208641H, RPS202071H, RPS205500H, RPS208448H, RPS208617H, RPS203057H, RPS203370H, RPS204797H, RPS206092H, RPS202176H, RPS202183H, RPS203489H, RPS203752H, RPS204488H, RPS205855H, RPS207515H, RPS208476H, RPS207829H, RPS208386H, RPS205347H, RPS207063H, RPS207864H, RPS208418H, RPS203399H, RPS203690H, RPS205670H, RPS205677H, RPS205680H, RPS205681H, RPS205964H, RPS205996H, RPS207185H, RPS207192H, RPS207206H, RPS207221H, RPS207228H, RPS207263H, RPS207269H, RPS202204H, RPS202306H, RPS202504H, RPS202528H, RPS202991H, RPS202993H, RPS203555H, RPS203748H, RPS203768H, RPS203777H, RPS203867H, RPS203875H, RPS204193H, RPS204214H, RPS204232H, RPS204332H, RPS204523H, RPS205220H, RPS205314H, RPS205378H, RPS205554H, RPS206011H, RPS206105H, RPS206703H, RPS207671H, RPS202516H, RPS202798H, RPS203288H, RPS203558H, RPS204506H, RPS205393H, RPS203024H, RPS202229H, RPS202433H, RPS202719H, RPS203017H, RPS203632H, RPS203983H, RPS203986H, RPS204155H, RPS204472H, RPS205105H, RPS205384H, RPS205589H, RPS205594H, RPS205628H, RPS205788H, RPS205820H, RPS207075H, RPS207301H, RPS207305H, RPS207307H, RPS207308H, RPS207309H, RPS207310H, RPS207312H, RPS207313H, RPS207453H, RPS208375H, RPS208664H, RPS202729H, RPS202962H, RPS203308H, RPS204590H, RPS204686H, RPS205363H, RPS205403H, RPS205944H, RPS206882H, RPS205361H, RPS208500H, RPS206376H, RPS206814H, RPS207116H, RPS208311H, RPS202182H, RPS204002H, RPS207163H, RPS206166H, RPS203076H, RPS203366H, RPS204309H, RPS206785H, RPS208357H, RPS204761H, RPS202526H, RPS203561H, RPS203578H, RPS206684H, RPS207011H, RPS208692H, RPS203136H, RPS204242H, RPS204435H, RPS206025H, RPS206453H, RPS207920H, RPS207133H, RPS208160H, RPS203003H, RPS203193H, RPS203792H, RPS205377H, RPS205494H, RPS205865H, RPS205948H, RPS206393H, RPS208470H, RPS202272H, RPS202588H, RPS202589H, RPS202591H, RPS202593H, RPS202594H, RPS202598H, RPS206183H, RPS206184H, RPS206186H, RPS206189H, RPS206194H, RPS207081H, RPS203377H, RPS204312H, RPS206132H, RPS206522H, RPS207430H, RPS208545H, RPS205168H, RPS208410H, RPS203260H, RPS203622H, RPS206736H, RPS207748H, RPS207884H, RPS207900H, RPS203881H, RPS204354H, RPS205412H, RPS205795H, RPS205747H, RPS202519H, RPS202121H, RPS202212H, RPS203104H, RPS204431H, RPS205428H, RPS205586H, RPS208474H, RPS203959H, RPS205864H, RPS205868H, RPS206445H, RPS207771H, RPS203684H, RPS204600H, RPS205424H, RPS206809H, RPS207730H, RPS208165H, RPS203117H, RPS204188H, RPS207195H, RPS208477H, RPS206922H, RPS202173H, RPS202227H, RPS203301H, RPS206036H, RPS206201H, RPS207007H, RPS207868H, RPS204868H, RPS208611H, RPS203840H, RPS204449H, RPS206021H, RPS208025H, RPS202859H, RPS203949H, RPS204120H, RPS204792H, RPS204914H, RPS202226H, RPS203724H, RPS204916H, RPS205008H, RPS206632H, RPS202784H, RPS205561H, RPS206784H, RPS202356H, RPS203886H, RPS206045H, RPS207077H, RPS207167H, RPS208631H, RPS206465H, RPS202033H, RPS203204H, RPS204532H, RPS204533H, RPS205002H, RPS205679H, RPS206054H, RPS206139H, RPS206300H, RPS207524H, RPS207571H, RPS207587H, RPS208653H, RPS203974H, RPS205625H, RPS206359H, RPS208070H, RPS204244H, RPS205892H, RPS207032H, RPS202424H, RPS206058H, RPS208359H, RPS203235H, RPS204261H, RPS205107H, RPS205118H, RPS205121H, RPS206067H, RPS206513H, RPS206645H, RPS208247H, RPS208625H, RPS203306H, RPS205987H, RPS206896H, RPS205379H, RPS207727H, RPS203858H, RPS204893H, RPS204895H, RPS204896H, RPS204897H, RPS204898H, RPS202029H, RPS203127H, RPS204311H, RPS205676H, RPS206398H, RPS206871H, RPS204044H, RPS205106H, RPS206098H, RPS208263H, RPS203026H, RPS203539H, RPS203540H, RPS205624H, RPS205963H, RPS206479H, RPS206488H, RPS206489H, RPS206649H, RPS206805H, RPS207000H, RPS207013H, RPS207403H, RPS207404H, RPS207405H, RPS207450H, RPS207451H, RPS208139H, RPS208250H, RPS204783H, RPS207036H, RPS203639H, RPS203654H, RPS203659H, RPS203663H, RPS203664H, RPS204909H, RPS205108H, RPS205267H, RPS206237H, RPS206298H, RPS207823H, RPS204142H, RPS202181H, RPS203280H, RPS203383H, RPS203511H, RPS202170H, RPS203837H, RPS206124H, RPS202508H, RPS205435H, RPS205966H, RPS206345H, RPS206463H, RPS207048H, RPS208465H, RPS203237H, RPS204457H, RPS205261H, RPS207079H, RPS206748H, RPS204121H, RPS202484H, RPS202647H, RPS202648H, RPS203346H, RPS203360H, RPS203496H, RPS203905H, RPS204324H, RPS204487H, RPS204966H, RPS205188H, RPS205206H, RPS206122H, RPS206904H, RPS208582H, RPS208672H, RPS202780H, RPS205019H, RPS202165H, RPS203386H, RPS203740H, RPS203969H, RPS203993H, RPS205441H, RPS205504H, RPS206583H, RPS206672H, RPS207140H, RPS207575H, RPS207725H, RPS203455H, RPS202626H, RPS202629H, RPS202641H, RPS204039H, RPS204283H, RPS207879H, RPS208660H, RPS202126H, RPS202905H, RPS202935H, RPS203274H, RPS203417H, RPS203872H, RPS204888H, RPS205295H, RPS205471H, RPS207952H, RPS203923H, RPS208040H, RPS208432H, RPS208533H, RPS204368H, RPS204871H, RPS205557H, RPS206446H, RPS207700H, RPS207701H, RPS207831H, RPS207928H, RPS206404H, RPS206406H, RPS202026H, RPS204268H, RPS205039H, RPS205902H, RPS208256H, RPS206375H, RPS202998H, RPS203429H, RPS206209H, RPS206238H, RPS206565H, RPS207540H, RPS202482H, RPS206840H, RPS203335H, RPS204930H, RPS204993H, RPS206880H, RPS202779H, RPS203176H, RPS204717H, RPS207465H, RPS207798H, RPS203043H, RPS203382H, RPS203723H, RPS205252H, RPS205285H, RPS206144H, RPS207072H, RPS207773H, RPS207779H, RPS208525H, RPS202721H, RPS202858H, RPS206516H, RPS203516H, RPS202487H, RPS205812H, RPS207102H, RPS206342H, RPS206884H, RPS207693H, RPS207903H, RPS202215H, RPS202316H, RPS202411H, RPS203454H, RPS203814H, RPS203988H, RPS204090H, RPS204330H, RPS204352H, RPS204416H, RPS204446H, RPS204512H, RPS205562H, RPS205945H, RPS206150H, RPS206244H, RPS206366H, RPS206466H, RPS206468H, RPS206469H, RPS206470H, RPS206474H, RPS206475H, RPS206481H, RPS206483H, RPS206490H, RPS206493H, RPS206494H, RPS206495H, RPS206498H, RPS206507H, RPS206510H, RPS206609H, RPS206782H, RPS206806H, RPS206865H, RPS206893H, RPS207183H, RPS207198H, RPS207200H, RPS207207H, RPS207208H, RPS207209H, RPS207214H, RPS207215H, RPS207218H, RPS207220H, RPS207225H, RPS207227H, RPS207229H, RPS207233H, RPS207745H, RPS207791H, RPS207909H, RPS208144H, RPS208296H, RPS208374H, RPS208390H, RPS208656H, RPS208684H, RPS203107H, RPS205136H, RPS205914H, RPS206149H, RPS207738H, RPS207844H, RPS208157H, RPS206678H, RPS206747H, RPS202445H, RPS203806H, RPS204520H, RPS205457H, RPS205913H, RPS206990H, RPS202925H, RPS202988H, RPS203167H, RPS204295H, RPS205947H, RPS206095H, RPS206131H, RPS206437H, RPS206440H, RPS207972H, RPS208461H, RPS208647H, RPS202825H, RPS203423H, RPS203585H, RPS203627H, RPS203839H, RPS203863H, RPS204016H, RPS204163H, RPS205358H, RPS206305H, RPS206976H, RPS207155H, RPS207549H, RPS207916H, RPS208050H, RPS208356H, RPS203337H, RPS203499H, RPS204027H, RPS207058H, RPS202902H, RPS203857H, RPS203860H, RPS206262H, RPS207359H, RPS207375H, RPS207385H, RPS208405H, RPS208507H, RPS202501H, RPS202368H, RPS202115H, RPS203394H, RPS203981H, RPS204337H, RPS206115H, RPS208205H, RPS208515H, RPS202754H, RPS202808H, RPS203331H, RPS204304H, RPS203852H, RPS205518H, RPS202205H, RPS202505H, RPS203843H, RPS204087H, RPS207362H, RPS207845H, RPS202432H, RPS208427H, RPS205219H, RPS206899H, RPS207523H, RPS207882H, RPS208621H, RPS205498H, RPS202100H, RPS202325H, RPS202600H, RPS202830H, RPS203336H, RPS203686H, RPS203737H, RPS203963H, RPS204017H, RPS204020H, RPS204292H, RPS204361H, RPS204398H, RPS204409H, RPS204453H, RPS204539H, RPS205253H, RPS205374H, RPS205575H, RPS205622H, RPS205693H, RPS205980H, RPS206040H, RPS206060H, RPS206108H, RPS206188H, RPS206197H, RPS206310H, RPS206339H, RPS206350H, RPS206551H, RPS206559H, RPS206566H, RPS206568H, RPS206575H, RPS206579H, RPS206584H, RPS206592H, RPS206700H, RPS206718H, RPS206798H, RPS206813H, RPS206847H, RPS206862H, RPS207216H, RPS207278H, RPS207337H, RPS207380H, RPS207390H, RPS207457H, RPS207458H, RPS207460H, RPS207461H, RPS207466H, RPS207470H, RPS207475H, RPS207477H, RPS207478H, RPS207871H, RPS208068H, RPS208073H, RPS208074H, RPS208075H, RPS208078H, RPS208080H, RPS208082H, RPS208083H, RPS208116H, RPS208467H, RPS208652H, RPS208654H, RPS205481H, RPS203155H, RPS204114H, RPS207039H, RPS207124H, RPS208035H, RPS208233H, RPS206576H, RPS203019H, RPS207505H, RPS202216H, RPS202898H, RPS202972H, RPS203390H, RPS203573H, RPS203696H, RPS203823H, RPS204246H, RPS204322H, RPS204335H, RPS204513H, RPS204542H, RPS204952H, RPS205217H, RPS205502H, RPS205619H, RPS205856H, RPS206034H, RPS206293H, RPS206433H, RPS206600H, RPS206833H, RPS207110H, RPS207560H, RPS207609H, RPS207856H, RPS208552H, RPS208645H, RPS204258H, RPS204935H, RPS206318H, RPS208526H, RPS203122H, RPS203467H, RPS203846H, RPS204540H, RPS204592H, RPS205822H, RPS206496H, RPS206499H, RPS207433H, RPS208392H, RPS202845H, RPS204406H, RPS205203H, RPS202566H, RPS203527H, RPS203577H, RPS205591H, RPS207004H, RPS208086H, RPS202705H, RPS205587H, RPS207604H, RPS202401H, RPS202896H, RPS202918H, RPS202986H, RPS203139H, RPS204091H, RPS204772H, RPS206482H, RPS206504H, RPS208103H, RPS207805H, RPS208384H, RPS203398H, RPS203584H, RPS205291H, RPS203146H, RPS204001H, RPS207506H, RPS203888H, RPS206662H, RPS203322H, RPS203697H, RPS205290H, RPS206680H, RPS203730H, RPS203877H, RPS203188H, RPS205531H, RPS204944H, RPS205149H, RPS205185H, RPS205590H, RPS207656H, RPS207870H, RPS208024H, RPS203217H, RPS204058H, RPS205054H, RPS205993H, RPS207662H, RPS202663H, RPS204985H, RPS207108H, RPS204853H, RPS207165H, RPS206741H, RPS207907H, RPS208267H, RPS204765H, RPS206357H, RPS207855H, RPS203379H, RPS206026H, RPS207842H, RPS208613H, RPS206881H, RPS204041H, RPS207983H, RPS202309H, RPS202765H, RPS202979H, RPS202995H, RPS203047H, RPS203345H, RPS203385H, RPS203767H, RPS205308H, RPS206948H, RP
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1755-2023
Related Recalls
ICD-DR DDBB2D1 EVERA XT IS1/DF1 INTL, Model Number DDBB2D1; Implantable Cardioverter Defibrillators
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.
ICD-DR DDMC3D1 EVERA MRI S OUS/US DF1, Model Number DDMC3D1; Implantable Cardioverter Defibrillators
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.
ICD DDPA2D1G COBALT XT DR MRI DF1 GOLD, Model Number DDPA2D1G; Implantable Cardioverter Defibrillators
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.