Connecting Tube - Female to Male Luer Lock, HPCT8.8-100-M-FM Product Usage: The connecting tube is intended to be a conduit between a fluid reservoir and a vascular catheter in the patient. The connecting tube may be used during angiographic procedures to connect radiographic contrast power injector to a vascular catheter, arterial pressure monitoring, and fluid infusion.
Class I - DangerousWhat Should You Do?
- Check if you have this product: Lots NS7835000, NS7835001, NS7846433, NS7853877, NS7853880, NS7866660, NS7875342, NS7898129, NS7898130, NS7898131, NS7928009, NS7928014, NS7940349
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Cook Inc.
- Reason for Recall:
- Cap-adapter assembly of the device may potentially not be adequately tightened, possibly leading to fluid leakage from or air aspiration into the connecting tube.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Connecting Tube - Female to Male Luer Lock, HPCT8.8-100-M-FM Product Usage: The connecting tube is intended to be a conduit between a fluid reservoir and a vascular catheter in the patient. The connecting tube may be used during angiographic procedures to connect radiographic contrast power injector to a vascular catheter, arterial pressure monitoring, and fluid infusion.
Product Codes/Lot Numbers:
Lots NS7835000, NS7835001, NS7846433, NS7853877, NS7853880, NS7866660, NS7875342, NS7898129, NS7898130, NS7898131, NS7928009, NS7928014, NS7940349
Distribution:
Distributed in: AL, CA, GA, NE, NV, NY, TX
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1755-2018
Related Recalls
Specific lots may contain black debris on the exterior of the inner stylet and/or on the interior of the outer cannula. Local inflammatory reactions may occur if an affected product is used.
Flexor Check-Flo Introducer Ansel Modification mislabeled. Products packaged with a label stating 6FR may contain a 7FR device and vice versa