RayStation 14.0.0.3338. Radiation Therapy Treatment Planning System.

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    UDI: 0735000201055620230630; GTIN: 07350002010556; Serial Numbers: 14.0.0.3338; Expiration Date: 2028-07-07.
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
RAYSEARCH LABORATORIES AB
Reason for Recall:
Potential for reported SSD to be too high.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
ongoing

Product Information

Full Description:

RayStation 14.0.0.3338. Radiation Therapy Treatment Planning System.

Product Codes/Lot Numbers:

UDI: 0735000201055620230630; GTIN: 07350002010556; Serial Numbers: 14.0.0.3338; Expiration Date: 2028-07-07.

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-1751-2024

Related Recalls

Potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for some ROI(s) which have no contours and either have a material override defined, or are of type Bolus, Fixation, or Support. In these cases, adding geometry to the ROI, subsequently modifying geometry, or removing the material override does not invalidate dose as intended.

Nov 28, 2025 Other Medical Devices View Details →

Potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for some ROI(s) which have no contours and either have a material override defined, or are of type Bolus, Fixation, or Support. In these cases, adding geometry to the ROI, subsequently modifying geometry, or removing the material override does not invalidate dose as intended.

Nov 28, 2025 Other Medical Devices View Details →

Potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for some ROI(s) which have no contours and either have a material override defined, or are of type Bolus, Fixation, or Support. In these cases, adding geometry to the ROI, subsequently modifying geometry, or removing the material override does not invalidate dose as intended.

Nov 28, 2025 Other Medical Devices View Details →