BrightView XCT model number 882482 BrightView and BrightView X are gamma cameras designed for single or dual detector nuclear imaging accommodating a broad range of Emission Computed Tomography (ECT) studies. The cameras can be used to perform planar static, dynamic, gated, and total body studies, as well as circular-orbit and noncircular orbit SPECT studies, gated SPECT (circular and noncircular)studies, computer-programmed protocol strings, and reference scans (dual detectors only). BrightView XCT is a gamma camera for Single Photon Emission Computed Tomography (SPECT) and integrates with an attenuation device consisting of flat panel x-ray imaging components. BrightView XCT produces non-attenuation corrected SPECT images and attenuation corrected SPECT images with an x-ray transmission data that may also be used for scatter correction. The nuclear medicine images and BrightView XCT images may be registered and displayed in a fused format (overlaid in the same orientation) to provide anatomical localization of the nuclear medicine data. The BrightView XCT imaging system should only be used by trained healthcare professionals.
Class I - DangerousWhat Should You Do?
- Check if you have this product: 6000203, 6000396, 6000229, 6000195, 6000204, 6000402, 6000237, 6000093, 6000206, 6000022, 6000416, 6000330, 6000130, 6000339, 6000276, 6000086, 6000051, 6000351, 6000188, 6000065, 6000380, 6000013, 6000090, 6000370, 6000025, 6000236, 6000129, 6000121, 6000342, 6000367, 6000256, 6000101, 6000400, 6000371, 6000336, 6000168 , 6000418, 6000359, 6000421, 6000120, 6000299, 6000046, 6340010, 6000252, 6000428, 6000413, 6000230, 6000049, 6000231,, 6000151, 6000369, 6000244, 6000403, 6000405, 6000143, 6000110, 6000285, 6000312, 6000160, 6000344, 6000394, 6000283, 6000388, 6000146, 6000232, 6000335, 6000356, 6000140, 6000401, 6000408, 6340011, 6000348, 6000277, 6000245, 6000419, 6000177, 6000386, 6000390, 6000360, 6000114, 6000028, 6000042, 6000382, 6000275, 6000024, 6000002, 6000311, 6000213, 6000062, 6000363
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
β οΈ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Philips Medical Systems (Cleveland) Inc
- Reason for Recall:
- The brake hub was not engaging with the gear box shaft due to an error in the assembly of components within the brake hub mechanism after service, causing the detector to backdrive.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
BrightView XCT model number 882482 BrightView and BrightView X are gamma cameras designed for single or dual detector nuclear imaging accommodating a broad range of Emission Computed Tomography (ECT) studies. The cameras can be used to perform planar static, dynamic, gated, and total body studies, as well as circular-orbit and noncircular orbit SPECT studies, gated SPECT (circular and noncircular)studies, computer-programmed protocol strings, and reference scans (dual detectors only). BrightView XCT is a gamma camera for Single Photon Emission Computed Tomography (SPECT) and integrates with an attenuation device consisting of flat panel x-ray imaging components. BrightView XCT produces non-attenuation corrected SPECT images and attenuation corrected SPECT images with an x-ray transmission data that may also be used for scatter correction. The nuclear medicine images and BrightView XCT images may be registered and displayed in a fused format (overlaid in the same orientation) to provide anatomical localization of the nuclear medicine data. The BrightView XCT imaging system should only be used by trained healthcare professionals.
Product Codes/Lot Numbers:
6000203, 6000396, 6000229, 6000195, 6000204, 6000402, 6000237, 6000093, 6000206, 6000022, 6000416, 6000330, 6000130, 6000339, 6000276, 6000086, 6000051, 6000351, 6000188, 6000065, 6000380, 6000013, 6000090, 6000370, 6000025, 6000236, 6000129, 6000121, 6000342, 6000367, 6000256, 6000101, 6000400, 6000371, 6000336, 6000168 , 6000418, 6000359, 6000421, 6000120, 6000299, 6000046, 6340010, 6000252, 6000428, 6000413, 6000230, 6000049, 6000231,, 6000151, 6000369, 6000244, 6000403, 6000405, 6000143, 6000110, 6000285, 6000312, 6000160, 6000344, 6000394, 6000283, 6000388, 6000146, 6000232, 6000335, 6000356, 6000140, 6000401, 6000408, 6340011, 6000348, 6000277, 6000245, 6000419, 6000177, 6000386, 6000390, 6000360, 6000114, 6000028, 6000042, 6000382, 6000275, 6000024, 6000002, 6000311, 6000213, 6000062, 6000363
Distribution:
Nationwide distribution
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1751-2017
Related Recalls
Due to a software issue, there is a potential image error of the Region of Interest for expansion/contraction for HFP (Head First Prone), FFS (Feet First Supine) and FFP (Feet First Prone) orientations.
Pinnacle 3 with TumorLOC, (870258) Radiation Therapy Planning System Software
Philips Medical Systems (Cleveland)
Due to software issue, Radiation Therapy Planning system may provide incorrect dataset calculations when performing the "Stopping Power Ratio" (SPR) ,
Pinnacle3 Radiation Therapy Planning System, Model numbers 870231 and 870237.
Philips Medical Systems (Cleveland)
When computing a radiation dose in the system, the exported dose information is incorrect when there is more than one beam attached to the prescription and certain options are selected on the Edit Prescription screen.