RayStation 13.0.0.1547, 13.1.0.144 & 13.1.1.89. Radiation Therapy Treatment Planning System.
Class I - DangerousWhat Should You Do?
- Check if you have this product: UDI: 0735000201054920220616, 0735000201067920221007 and 0735000201073020230913; GTIN: 07350002010549, 07350002010679 and 07350002010730; Serial Numbers: 13.0.0.1547, 13.1.0.144 and 13.1.1.89; Expiration Date: 2028-10-05.
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- RAYSEARCH LABORATORIES AB
- Reason for Recall:
- Potential for reported SSD to be too high.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
RayStation 13.0.0.1547, 13.1.0.144 & 13.1.1.89. Radiation Therapy Treatment Planning System.
Product Codes/Lot Numbers:
UDI: 0735000201054920220616, 0735000201067920221007 and 0735000201073020230913; GTIN: 07350002010549, 07350002010679 and 07350002010730; Serial Numbers: 13.0.0.1547, 13.1.0.144 and 13.1.1.89; Expiration Date: 2028-10-05.
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1750-2024
Related Recalls
Brand Name: RayStation Product Name: RayStation/RayPlan Model/Catalog Number: 17.0.0, 17.0.1 Software Version: RayStation v2025, RayStation v2025 SP1 Product Description: Radiation Therapy Treatment Planning System Component: No
RAYSEARCH LABORATORIES AB
Potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for some ROI(s) which have no contours and either have a material override defined, or are of type Bolus, Fixation, or Support. In these cases, adding geometry to the ROI, subsequently modifying geometry, or removing the material override does not invalidate dose as intended.
Potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for some ROI(s) which have no contours and either have a material override defined, or are of type Bolus, Fixation, or Support. In these cases, adding geometry to the ROI, subsequently modifying geometry, or removing the material override does not invalidate dose as intended.
Brand Name: RayStation Product Name: RayStation/RayPlan Model/Catalog Number: 15.2.0, 15.1.3 Software Version: RayStation 2024A SP1, RayStation 2024A SP2 Product Description: Radiation Therapy Treatment Planning System Component: No
RAYSEARCH LABORATORIES AB
Potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for some ROI(s) which have no contours and either have a material override defined, or are of type Bolus, Fixation, or Support. In these cases, adding geometry to the ROI, subsequently modifying geometry, or removing the material override does not invalidate dose as intended.