Alinity m Resp-4-Plex AMP Kit, List Number 09N79-096, a multiplex real-time reverse transcription (RT) polymerase chain reaction (PCR) test intended for the simultaneous qualitative detection and differentiation of RNA from SARS-CoV-2, influenza A virus (flu A), influenza B virus (flu B), and/or Respiratory Syncytial Virus (RSV) in anterior nasal or nasopharyngeal swab specimens collected by a healthcare provider (HCP), or in anterior nasal swab specimens that are self-collected at a healthcare location, from individuals suspected by their HCP of respiratory viral infection consistent with COVID-19.
Class I - DangerousWhat Should You Do?
- Check if you have this product: UDI-DI: 00884929049390; Lot Number: 409384
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Abbott Molecular, Inc.
- Reason for Recall:
- There were reports of an increase in reactive negative controls and false positive results.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
Alinity m Resp-4-Plex AMP Kit, List Number 09N79-096, a multiplex real-time reverse transcription (RT) polymerase chain reaction (PCR) test intended for the simultaneous qualitative detection and differentiation of RNA from SARS-CoV-2, influenza A virus (flu A), influenza B virus (flu B), and/or Respiratory Syncytial Virus (RSV) in anterior nasal or nasopharyngeal swab specimens collected by a healthcare provider (HCP), or in anterior nasal swab specimens that are self-collected at a healthcare location, from individuals suspected by their HCP of respiratory viral infection consistent with COVID-19.
Product Codes/Lot Numbers:
UDI-DI: 00884929049390; Lot Number: 409384
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1749-2025
Related Recalls
There were reports of an increase in reactive negative controls and false positive results.
The design of the Alinity m System Solutions drawer enclosure may allow liquid (liquid waste or system solutions) originating in the System Solutions drawer to flow beyond the footprint of the instrument and into the walking-path of the user.
Abbott Alinity m HPV AMP Kit, used with the Alinity m System, product codes: a) REF 09N15-095; b) REF 09N15-090; c) REF 09N15-091
Abbott Molecular
Abbott has identified an increase of incidences regarding Error Code (EC) 9198 (Positive control is non-reactive) while using the Alinity m HR HPV AMP Kit and Alinity m STI AMP Kit. Certain invalidated positive assay controls can be traced to iron leaching into the Alinity m Lysis Solution from the lysis transfer pump in the Alinity m System.