QIAcube Connect MDx, Model No. 9003070

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    UDI-DI (GTIN): 04053228039129 Serial Numbers (U.S.) 30000 30009 30027 30039 30049 30061 30073 30002 30010 30028 30041 30050 30062 30075 30003 30132 30031 30042 30051 30063 30078 30004 30013 30032 30043 30053 30064 30080 30005 30016 30034 30044 30057 30065 30082 30006 30019 30035 30045 30058 30067 30089 30007 30021 30033 30046 30059 30069 30090 30008 30023 30037 30047 30060 30071 30093 30094 30095 30097 30102 30104 30109 30112 30114 30115 30117 30118 30119 30121 30124 30127 30128 30129 30130 30131 30133 30134 30136 30138 30139 30140 30141 30143 30145 30106 30025 30077 30054 30055 30066 30076
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Qiagen Sciences LLC
Reason for Recall:
During the "Load tip racks and enzymes" step of the run set-up, the info screen indicates "minimal volume to be loaded", but the indicated volume is instead the exact volume that should be loaded. Misunderstanding the guidance may lead to over diluted, under diluted, or improperly lysed sample, which in turn could lead to false negative or false positive results.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
ongoing

Product Information

Full Description:

QIAcube Connect MDx, Model No. 9003070

Product Codes/Lot Numbers:

UDI-DI (GTIN): 04053228039129 Serial Numbers (U.S.) 30000 30009 30027 30039 30049 30061 30073 30002 30010 30028 30041 30050 30062 30075 30003 30132 30031 30042 30051 30063 30078 30004 30013 30032 30043 30053 30064 30080 30005 30016 30034 30044 30057 30065 30082 30006 30019 30035 30045 30058 30067 30089 30007 30021 30033 30046 30059 30069 30090 30008 30023 30037 30047 30060 30071 30093 30094 30095 30097 30102 30104 30109 30112 30114 30115 30117 30118 30119 30121 30124 30127 30128 30129 30130 30131 30133 30134 30136 30138 30139 30140 30141 30143 30145 30106 30025 30077 30054 30055 30066 30076

Distribution:

Nationwide distribution

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-1746-2022

Related Recalls

Heater shaker module does not perform heating if the temperature is set below 40ΒΏC for static incubation steps when running IVD protocol QIAampΒΏ DSP DNA Mini Kit (cat. no. 61304), Protocol: Isolation of genomic DNA from Gram-positive bacteria (Script filename: Kt_name_Bacteria (Gram+) or rY'east_Enzymatic Lysis_ V2) which could potentially cause delayed or erroneous results depending on the downstream assay.

Jun 3, 2024 Diagnostic Equipment Nationwide View Details β†’