HEMO-10 Hemospray Endoscopic Hemostat
Class I - DangerousWhat Should You Do?
- Check if you have this product: All Lot numbers manufactured 01/16/2017 - 01/15/2020 of both the following: HEMO-7 Order Number G56572 HEMO-10 Order Number G21049
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Wilson-Cook Medical Inc.
- Reason for Recall:
- Complaints that the handle and/or activation knob on Hemostat devices have cracked or broken when the device is activated. In some cases, this has caused the carbon dioxide cartridge to exit the handle. Most of these complaints reported that the carbon dioxide cartridge exited the handle with minimal force. However, some complaints reported that the carbon dioxide cartridge exited the handle with the potential to project towards a user or a patient.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
HEMO-10 Hemospray Endoscopic Hemostat
Product Codes/Lot Numbers:
All Lot numbers manufactured 01/16/2017 - 01/15/2020 of both the following: HEMO-7 Order Number G56572 HEMO-10 Order Number G21049
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1746-2020
Related Recalls
Cook Medical HEMO-7-EU Hemospray Endoscopic Hemostat, REF G24663, 2.8 cm, sterile.
Wilson-Cook Medical
Specific lots of product have been manufactured with nonconforming raw materials resulting in the red activation knob cracking or breaking at the activation knob internal threading while the device is activated.
Specific lots of product have been manufactured with nonconforming raw materials.
QUANTUM TTC BILIARY BALLOON DILATOR, REF QBD-4X3, Order Number G22654; used to dilate strictures of the biliary tree
Wilson-Cook Medical
Nonconforming devices while manufactured correctly, do not comply to their design requirements. Potential harms associated with balloon detachment are endoscopic retrieval of an impacted or non-impacted object, bleeding caused by the retrieval device, or the incident can occur without injury and the object is left to pass naturally.