BD Vacutainer Buffered Sodium Citrate: (9NC) Blood Collection Tubes, 4.5 mL, 13 x 75 mm, Mfd: Becton Dickinson and Company Product Usage: The BD Vacutainer Safety Coagulation tube is an evacuated blood collection Tube that provides means of collecting, transporting and processing blood in a closed tube. The buffered sodium citrate additive provides an anti-coagulated specimen that may be used for clinical laboratory coagulation assays and determinations.
Class I - DangerousWhat Should You Do?
- Check if you have this product: Catalog Number: 367716 - Lot numbers: 5251673 with expiry 3/31/2017; 6064542 with expiry 9/30/17; 6187520 with expiry 1/31/2018 and 6253669 with expiy 3/31/18 Catalog Number: 369714 - Lot numbers: 5272777 with expiry 4/30/17; 5336909 with expiry 6/30/17; 5357632 with expiry 7/31/17; 6040953 and 6064634 both with expiry 9/30/17; 6090572 with expiry 10/31/17; 6124932 with expiry 11/30/17; 6173983 with expiry 12/31/17; 6187600 with expiry 1/31/18 and 6216656 with expiry 2/28/18 Lot numbers 6064542, 6187520 and 6253669 - limited to distribution outside the U.S. market. UDI for lot 6253669 is DI (01)30382903677161, PI (17)180331(10)6253669(30)0100 and lot 6216656 is DI (01)30382903697145, PI (17)180228(10)6216656(30)0100; for all other lots UDI is not applicable based on manufacture date.
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Becton Dickinson & Company
- Reason for Recall:
- BD is initiating this product correction of multiple lots of BD Vacutainer Buffered Sodium Citrate: (9NC) Blood Collection Tube (13x75 mm x 4.5 mL). A limited number of tubes in these lots may exhibit stopper pullout from the tube as the user removes the needle from the stopper following specimen collection.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
BD Vacutainer Buffered Sodium Citrate: (9NC) Blood Collection Tubes, 4.5 mL, 13 x 75 mm, Mfd: Becton Dickinson and Company Product Usage: The BD Vacutainer Safety Coagulation tube is an evacuated blood collection Tube that provides means of collecting, transporting and processing blood in a closed tube. The buffered sodium citrate additive provides an anti-coagulated specimen that may be used for clinical laboratory coagulation assays and determinations.
Product Codes/Lot Numbers:
Catalog Number: 367716 - Lot numbers: 5251673 with expiry 3/31/2017; 6064542 with expiry 9/30/17; 6187520 with expiry 1/31/2018 and 6253669 with expiy 3/31/18 Catalog Number: 369714 - Lot numbers: 5272777 with expiry 4/30/17; 5336909 with expiry 6/30/17; 5357632 with expiry 7/31/17; 6040953 and 6064634 both with expiry 9/30/17; 6090572 with expiry 10/31/17; 6124932 with expiry 11/30/17; 6173983 with expiry 12/31/17; 6187600 with expiry 1/31/18 and 6216656 with expiry 2/28/18 Lot numbers 6064542, 6187520 and 6253669 - limited to distribution outside the U.S. market. UDI for lot 6253669 is DI (01)30382903677161, PI (17)180331(10)6253669(30)0100 and lot 6216656 is DI (01)30382903697145, PI (17)180228(10)6216656(30)0100; for all other lots UDI is not applicable based on manufacture date.
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1746-2017
Related Recalls
BD Luer Tip Caps; Catalog Number(s) or Model Number(s): 308341;
Becton Dickinson & Company
BD has confirmed through internal testing that certain device trays failed routine biocompatibility testing.
PosiFlush Prefilled Saline Syringe with General Pump Compatibility (10mL Syringe with 10mL Saline Fill), Catalog Number 306547
Becton Dickinson & Company
An incorrect stopper was used in nine lots of 10mL BD PosiFlush Normal Saline Flush Syringes with General Pump Compatibility. The use of BD PosiFlush Prefilled Flush Syringes with this stopper causes an increase in injection force, which has the potential to trigger alarms if used with a syringe pump.
Affected infusion sets labeled as free of DEHP actually do contain di(2-ethylhexyl) phthalate (DEHP).