CRT-D DTBA2QQ VIVA QUAD XT IS4/DF4 INTL, Model Number DTBA2QQ; Implantable Cardioverter Defibrillators
Class I - DangerousWhat Should You Do?
- Check if you have this product: GTIN 00643169720169, Lot Serial Numbers: BLG624311S, BLG624755S, BLG625177S, BLG625181S, BLG625182S, BLG625188S, BLG625193S, BLG625324S, BLG625325S, BLG625434S, BLG625596S, BLG625598S, BLG625606S, BLG625608S, BLG625636S, BLG625647S, BLG625702S, BLG625740S, BLG625741S, BLG625745S, BLG625962S, BLG626105S, BLG626141S, BLG624936S, BLG624987S, BLG624995S, BLG625164S, BLG625167S, BLG625168S, BLG625242S, BLG625247S, BLG625565S, BLG625566S, BLG625568S, BLG625573S, BLG625616S, BLG625619S, BLG625620S, BLG625624S, BLG625651S, BLG625653S, BLG625655S, BLG625656S, BLG625489S, BLG625612S, BLG625910S, BLG624312S, BLG625331S, BLG624790S, BLG624843S, BLG624880S, BLG624881S, BLG624878S, BLG625644S, BLG624773S, BLG625148S, BLG625224S, BLG624958S, BLG625000S, BLG624751S, BLG624884S, BLG624890S, BLG624895S, BLG624896S, BLG624901S, BLG625171S, BLG625173S, BLG625175S, BLG625197S, BLG625571S, BLG625597S, BLG625605S, BLG625626S, BLG625630S, BLG625705S, BLG625876S, BLG626149S, BLG625650S, BLG625631S, BLG625228S, BLG625243S, BLG624759S, BLG625953S, BLG624816S, BLG624819S, BLG624827S, BLG624863S, BLG624889S, BLG624792S, BLG624795S, BLG625013S, BLG625023S, BLG625025S, BLG625304S, BLG625308S, BLG625311S, BLG625312S, BLG625317S, BLG625318S, BLG625319S, BLG625320S, BLG625321S, BLG625323S, BLG625241S, BLG624874S, BLG625239S, BLG625225S, BLG624879S, BLG625220S, BLG624947S, BLG624952S, BLG624954S, BLG625001S, BLG625652S, BLG624750S, BLG624754S, BLG624761S, BLG624762S, BLG624763S, BLG624765S, BLG625189S, BLG625303S, BLG625751S, BLG625746S, BLG625756S, BLG625861S, BLG625730S, BLG625749S, BLG624968S, BLG624969S, BLG624979S, BLG624984S, BLG625215S, BLG625216S, BLG625218S, BLG625219S, BLG625238S, BLG625245S, BLG624988S, BLG625842S, BLG625868S, BLG625945S, BLG626037S, BLG624976S, BLG624977S, BLG625213S, BLG625240S, BLG626737S, BLG625172S, BLG625195S, BLG624885S, BLG624886S, BLG624899S, BLG625092S, BLG625185S, BLG625192S, BLG624999S, BLG625231S, BLG624778S, BLG624848S, BLG625196S, BLG625301S, BLG624927S, BLG624956S, BLG625040S, BLG626471S, BLG626710S, BLG626712S, BLG626732S, BLG626736S, BLG626739S, BLG624897S, BLG625326S, BLG625637S, BLG625654S, BLG625657S, BLG625002S, BLG625103S, BLG625139S, BLG625146S, BLG625194S, BLG625211S, BLG625226S, BLG625236S, BLG625237S, BLG625250S, BLG625305S, BLG625419S, BLG625420S, BLG625615S, BLG625617S, BLG625234S, BLG625205S, BLG625266S, BLG624990S, BLG625004S, BLG625014S, BLG625249S, BLG625273S, BLG625274S, BLG625281S, BLG625284S, BLG625350S, BLG625366S, BLG625435S, BLG625440S, BLG625444S, BLG625589S, BLG625590S, BLG625591S, BLG625592S, BLG625593S, BLG625594S, BLG625600S, BLG625602S, BLG625603S, BLG625604S, BLG625742S, BLG625744S, BLG625747S, BLG625748S, BLG625754S, BLG626025S, BLG626053S, BLG625235S, BLG625275S, BLG625276S, BLG625278S, BLG625285S, BLG625288S, BLG625289S, BLG625291S, BLG625293S, BLG625294S, BLG625297S, BLG625300S, BLG625360S, BLG625368S, BLG625369S, BLG625373S, BLG625377S, BLG625382S, BLG625383S, BLG625384S, BLG624998S, BLG626720S, BLG624883S, BLG624891S, BLG625176S, BLG625599S, BLG625174S, BLG625186S, BLG624752S, BLG624756S, BLG624757S, BLG625179S, BLG625180S, BLG624950S, BLG624783S, BLG625010S, BLG625610S, BLG624804S, BLG624808S, BLG624993S, BLG625455S, BLG625642S, BLG625872S, BLG624900S, BLG624796S, BLG625272S, BLG624963S, BLG625280S, BLG625363S, BLG625158S, BLG625262S, BLG625267S, BLG625270S, BLG625299S, BLG625333S, BLG625334S, BLG625381S, BLG625534S, BLG625690S, BLG625699S, BLG625263S, BLG625327S, BLG625328S, BLG625330S, BLG625361S, BLG625370S, BLG625372S, BLG625375S, BLG625379S, BLG625396S, BLG625398S, BLG625399S, BLG625401S, BLG625429S, BLG625442S, BLG625446S, BLG625460S, BLG625514S, BLG625525S, BLG625533S, BLG625535S, BLG625541S, BLG625689S, BLG625698S, BLG625704S, BLG625715S, BLG625783S, BLG625374S, BLG625454S, BLG625457S, BLG625459S, BLG625695S, BLG625761S, BLG625925S, BLG625929S, BLG626003S, BLG625347S, BLG625351S, BLG625352S, BLG625353S, BLG625436S, BLG625438S, BLG625439S, BLG625441S, BLG625998S, BLG626046S, BLG625091S, BLG625314S, BLG625316S, BLG625344S, BLG625461S, BLG625465S, BLG625468S, BLG625520S, BLG625529S, BLG625530S, BLG625531S, BLG625532S, BLG625536S, BLG625537S, BLG625696S, BLG625697S, BLG625707S, BLG625712S, BLG625713S, BLG625762S, BLG625765S, BLG625766S, BLG625769S, BLG625770S, BLG625771S, BLG625777S, BLG625778S, BLG625784S, BLG625785S, BLG625786S, BLG625789S, BLG625984S, BLG625988S, BLG626010S, BLG626011S, BLG626028S, BLG626038S, BLG626096S, BLG626100S, BLG626108S, BLG625283S, BLG625286S, BLG625292S, BLG625338S, BLG625341S, BLG625342S, BLG625343S, BLG625873S, BLG625874S, BLG625759S, BLG625775S, BLG625776S, BLG625157S, BLG625210S, BLG625296S, BLG625395S, BLG625409S, BLG625469S, BLG625688S, BLG625767S, BLG625773S, BLG625790S, BLG625886S, BLG625916S, BLG625924S, BLG625978S, BLG625979S, BLG626013S, BLG625449S, BLG625521S, BLG625526S, BLG625935S, BLG626015S, BLG626106S, BLG625515S, BLG625517S, BLG625518S, BLG625519S, BLG625760S, BLG626047S, BLG625133S, BLG625138S, BLG625279S, BLG625306S, BLG625307S, BLG625310S, BLG625322S, BLG625371S, BLG625380S, BLG625407S, BLG625425S, BLG625451S, BLG625524S, BLG625774S, BLG625788S, BLG625999S, BLG626009S, BLG625132S, BLG625135S, BLG625156S, BLG625298S, BLG625329S, BLG625376S, BLG625378S, BLG625408S, BLG625463S, BLG625522S, BLG625540S, BLG625694S, BLG625763S, BLG625768S, BLG625772S, BLG625779S, BLG625782S, BLG625791S, BLG625875S, BLG625880S, BLG625882S, BLG625922S, BLG625926S, BLG625983S, BLG626000S, BLG626002S, BLG626005S, BLG626026S, BLG626029S, BLG626031S, BLG626034S, BLG626040S, BLG626048S, BLG626049S, BLG626050S, BLG626102S, BLG626107S, BLG626109S, BLG625365S, BLG625412S, BLG625413S, BLG625416S, BLG625417S, BLG625445S, BLG625692S, BLG625714S, BLG625764S, BLG625792S, BLG625793S, BLG625927S, BLG625977S, BLG625990S, BLG625993S, BLG626021S, BLG626042S, BLG626044S, BLG626052S, BLG626054S, BLG626055S, BLG625313S, BLG625795S, BLG625796S, BLG625981S, BLG625781S, BLG625153S, BLG625359S, BLG625392S, BLG625394S, BLG625424S, BLG625443S, BLG625512S, BLG625538S, BLG625539S, BLG625711S, BLG625780S, BLG625794S, BLG625755S, BLG625456S, BLG624784S, BLG625230S, BLG625658S, BLG624894S, BLG625187S, BLG625703S, BLG625108S, BLG625207S, BLG625362S, BLG625287S, BLG625295S, BLG625315S, BLG625467S, BLG625494S, BLG625508S, BLG625523S, BLG625559S, BLG625560S, BLG625564S, BLG625706S, BLG625716S, BLG625845S, BLG625846S, BLG626036S, BLG626714S, BLG625975S, BLG625976S, BLG625595S, BLG626027S, BLG624794S, BLG624873S, BLG625079S, BLG625081S, BLG625163S, BLG625214S, BLG625421S, BLG625431S, BLG625561S, BLG625583S, BLG625841S, BLG625860S, BLG625866S, BLG625869S, BLG625871S, BLG625878S, BLG626140S, BLG626146S, BLG626147S, BLG626151S, BLG626152S, BLG626458S, BLG626459S, BLG626464S, BLG626465S, BLG624777S, BLG625005S, BLG625008S, BLG624313S, BLG624767S, BLG625190S, BLG626113S, BLG624805S, BLG625100S, BLG625110S, BLG625198S, BLG625199S, BLG625389S, BLG625390S, BLG625426S, BLG625432S, BLG625437S, BLG625969S, BLG625971S, BLG626019S, BLG626022S, BLG625607S, BLG624776S, BLG624749S, BLG626103S, BLG624791S, BLG625120S, BLG625121S, BLG625928S, BLG625938S, BLG625965S, BLG625972S, BLG625140S, BLG625125S, BLG625130S, BLG625919S, BLG625920S, BLG625939S, BLG625940S, BLG625941S, BLG624844S, BLG624924S, BLG625037S, BLG625038S, BLG624846S, BLG624920S, BLG625119S, BLG625122S, BLG625930S, BLG625936S, BLG626024S, BLG624921S, BLG625128S, BLG624788S, BLG624835S, BLG624838S, BLG624839S, BLG624845S, BLG624866S, BLG625039S, BLG625047S, BLG625050S, BLG625124S, BLG625127S, BLG625129S, BLG625908S, BLG624829S, BLG624830S, BLG624925S, BLG625036S, BLG625909S, BLG625937S, BLG624824S, BLG624850S, BLG624853S, BLG624825S, BLG624828S, BLG624865S, BLG624877S, BLG625042S, BLG625049S, BLG625055S, BLG624849S, BLG624841S, BLG624869S, BLG624875S, BLG624915S, BLG625106S, BLG624771S, BLG624772S, BLG624774S, BLG624775S, BLG624858S, BLG624870S, BLG624871S, BLG624872S, BLG625016S, BLG625017S, BLG625018S, BLG625019S, BLG625020S, BLG625072S, BLG625145S, BLG625149S, BLG625150S, BLG625151S, BLG625165S, BLG625166S, BLG625169S, BLG625170S, BLG625257S, BLG625258S, BLG625259S, BLG625260S, BLG625261S, BLG625264S, BLG625393S, BLG625450S, BLG625452S, BLG625464S, BLG625466S, BLG625470S, BLG625471S, BLG625473S, BLG625474S, BLG625476S, BLG625611S, BLG625614S, BLG625629S, BLG625954S, BLG625955S, BLG625956S, BLG626007S, BLG624810S, BLG624811S, BLG624812S, BLG624813S, BLG624814S, BLG624815S, BLG624817S, BLG624818S, BLG624820S, BLG624821S, BLG624822S, BLG624807S, BLG624809S, BLG625021S, BLG625022S, BLG625105S, BLG625155S, BLG625159S, BLG625160S, BLG625271S, BLG626018S, BLG625028S, BLG625510S, BLG624766S, BLG624768S, BLG624770S, BLG624823S, BLG624854S, BLG624855S, BLG624856S, BLG624857S, BLG625076S, BLG625077S, BLG625078S, BLG625080S, BLG625082S, BLG625083S, BLG625085S, BLG625086S, BLG625087S, BLG625088S, BLG625208S, BLG625209S, BLG625253S, BLG625254S, BLG625255S, BLG625256S, BLG625339S, BLG625345S, BLG625346S, BLG625491S, BLG625633S, BLG625634S, BLG625635S, BLG625639S, BLG625643S, BLG625645S, BLG625646S, BLG625648S, BLG625649S, BLG625957S, BLG625958S, BLG625959S, BLG625960S, BLG625961S, BLG624887S, BLG624888S, BLG625447S, BLG625844S, BLG625883S, BLG626098S, BLG624836S, BLG624837S, BLG624867S, BLG624842S, BLG624780S, BLG624781S, BLG624802S, BLG624803S, BLG624859S, BLG624860S, BLG624861S, BLG625009S, BLG625074S, BLG625075S, BLG625094S, BLG625095S, BLG625098S, BLG625099S, BLG625101S, BLG625102S, BLG625115S, BLG625116S, BLG625117S, BLG625147S, BLG625251S, BLG625252S, BLG625388S, BLG626012S, BLG626014S, BLG624834S, BLG624985S, BLG626110S, BLG624919S, BLG625045S, BLG625046S, BLG625048S, BLG625054S, BLG624941S, BLG625012S, BLG625069S, BLG625206S, BLG625227S, BLG625367S, BLG625404S, BLG625569S, BLG624779S, BLG624797S, BLG624798S, BLG624799S, BLG624909S, BLG624978S, BLG624980S, BLG624981S, BLG624982S, BLG624983S, BLG624986S, BLG624991S, BLG624992S, BLG624997S, BLG625003S, BLG625006S, BLG625026S, BLG625053S, BLG625056S, BLG625057S, BLG625065S, BLG625066S, BLG625068S, BLG625118S, BLG625136S, BLG625141S, BLG625143S, BLG625144S, BLG625349S, BLG625354S, BLG625356S, BLG625358S, BLG625385S, BLG625397S, BLG625402S, BLG625738S, BLG625739S, BLG625752S, BLG625753S, BLG625912S, BLG625914S, BLG625915S, BLG625917S, BLG625942S, BLG625944S, BLG625946S, BLG625948S, BLG625950S, BLG625973S, BLG625982S, BLG625986S, BLG625987S, BLG625989S, BLG625991S, BLG625994S, BLG625995S, BLG626004S, BLG626016S, BLG626017S, BLG625638S, BLG625640S, BLG625572S, BLG625575S, BLG625641S, BLG625732S, BLG625736S, BLG624833S, BLG625007S, BLG624786S, BLG624840S, BLG624876S, BLG624831S, BLG624847S, BLG624882S, BLG625178S, BLG625183S, BLG625184S, BLG625109S, BLG626731S, BLG625406S, BLG625411S, BLG625418S, BLG625576S, BLG625578S, BLG625579S, BLG625580S, BLG625582S, BLG625584S, BLG625585S, BLG625586S, BLG625587S, BLG625588S, BLG625847S, BLG625881S, BLG625217S, BLG625222S, BLG625483S, BLG625516S, BLG625562S, BLG625710S, BLG625838S, BLG625839S, BLG625865S, BLG625870S, BLG625877S, BLG626136S, BLG626138S, BLG625029S, BLG625843S, BLG625947S, BLG626032S, BLG625011S, BLG625015S, BLG625403S, BLG625405S, BLG625632S, BLG625717S, BLG625728S, BLG625731S, BLG625733S, BLG625734S, BLG625735S, BLG625737S, BLG624787S, BLG624864S, BLG625131S, BLG624785S, BLG625089S, BLG625090S, BLG625911S, BLG625980S, BLG625137S, BLG625934S, BLG624902S, BLG625290S, BLG625478S, BLG625485S, BLG625486S, BLG625495S, BLG625496S, BLG625497S, BLG625498S, BLG625499S, BLG625527S, BLG625528S, BLG625567S, BLG625570S, BLG625613S, BLG625691S, BLG625693S, BLG625700S, BLG625701S, BLG625708S, BLG625709S, BLG624826S, BLG624851S, BLG624913S, BLG624989S, BLG625126S, BLG625974S, BLG626023S, BLG625142S, BLG625154S, BLG625277S, BLG625355S, BLG625879S, BLG625885S, BLG625472S, BLG625487S, BLG625859S, BLG626142S, BLG626155S, BLG626452S, BLG626453S, BLG626466S, BLG626729S, BLG626748S, BLG625024S, BLG625067S, BLG625484S, BLG625574S, BLG625577S, BLG625840S, BLG625858S, BLG625862S, BLG625863S, BLG625864S, BLG625884S, BLG625887S, BLG625943S, BLG625951S, BLG626030S, BLG626033S, BLG626039S, BLG626043S, BLG626099S, BLG626101S, BLG626104S, BLG626111S, BLG626115S, BLG626137S, BLG626139S, BLG626143S, BLG626144S, BLG626145S, BLG626148S, BLG626150S, BLG626454S, BLG626456S, BLG626457S, BLG626460S, BLG626461S, BLG626462S, BLG626463S, BLG626468S, BLG626469S, BLG626470S, BLG626717S, BLG626718S, BLG626719S, BLG626726S, BLG626727S, BLG626735S, BLG626738S, BLG626740S, BLG626741S, BLG626742S, BLG626743S, BLG626744S, BLG626745S, BLG626746S, BLG626747S, BLG626749S, BLG625391S, BLG625070S, BLG625743S, BLG624801S, BLG625027S, BLG625627S, BLG625758S, BLG625787S, BLG625073S, BLG624914S, BLG624917S, BLG624922S, BLG624923S, BLG625030S, BLG625041S, BLG625123S, BLG625134S, BLG625918S, BLG625932S, BLG625933S, BLG624789S, BLG624994S, BLG625043S, BLG625921S, BLG625475S, BLG625477S, BLG625480S, BLG625482S, BLG625488S, BLG625490S, BLG625492S, BLG625500S, BLG625501S, BLG625504S, BLG625506S, BLG624862S, BLG625044S, BLG625107S, BLG625618S, BLG626006S, BLG625563S, BLG626455S, BLG625268S, BLG626114S, BLG625161S, BLG625162S, BLG625348S, BLG625410S, BLG625415S, BLG625071S, BLG625096S, BLG625112S, BLG625113S, BLG625332S, BLG625335S, BLG625336S, BLG625337S, BLG625622S, BLG625952S, BLG625963S, BLG625964S, BLG625913S, BLG624832S, BLG625931S, BLG626724S, BLG626728S, BLG626467S, BLG626711S, BLG626713S, BLG626716S, BLG626721S, BLG626722S, BLG626723S, BLG626725S, BLG626730S, BLG626734S; GTIN 00643169980143, Lot Serial Numbers: BLG626995S, BLG628179S, BLG628075S, BLG628182S, BLG627589S, BLG625906S, BLG626056S, BLG626163S, BLG626164S, BLG626489S, BLG626492S, BLG626501S, BLG626649S, BLG626651S, BLG626957S, BLG626958S, BLG626961S, BLG627146S, BLG627148S, BLG627150S, BLG627192S, BLG627224S, BLG627704S, BLG628016S, BLG626173S, BLG626174S, BLG626175S, BLG626176S, BLG626179S, BLG626180S, BLG626182S, BLG626183S, BLG626185S, BLG626186S, BLG626188S, BLG626194S, BLG626195S, BLG626858S, BLG626859S, BLG626868S, BLG626892S, BLG626908S, BLG626933S, BLG627199S, BLG627286S, BLG626407S, BLG626542S, BLG626608S, BLG626592S, BLG626593S, BLG626594S, BLG626595S, BLG626596S, BLG626597S, BLG626094S, BLG626122S, BLG627285S, BLG625724S, BLG626370S, BLG625721S, BLG625825S, BLG625826S, BLG625893S, BLG626062S, BLG625681S, BLG625809S, BLG625819S, BLG625850S, BLG625895S, BLG625904S, BLG626058S, BLG626092S, BLG626119S, BLG626808S, BLG626902S, BLG627371S, BLG627486S, BLG628037S, BLG627370S, BLG626229S, BLG626339S, BLG627206S, BLG627459S, BLG627690S, BLG628078S, BLG625816S, BLG625897S, BLG626181S, BLG626313S, BLG626354S, BLG626424S, BLG626886S, BLG626814S, BLG626888S, BLG627336S, BLG627343S, BLG626648S, BLG627323S, BLG627404S, BLG627296S, BLG627554S, BLG627960S, BLG628080S, BLG628110S, BLG625817S, BLG627291S, BLG627764S, BLG627469S, BLG627489S, BLG625899S, BLG625683S, BLG625849S, BLG626067S, BLG626073S, BLG626130S, BLG626131S, BLG626197S, BLG626416S, BLG626425S, BLG626427S, BLG626429S, BLG626444S, BLG625851S, BLG625660S, BLG625804S, BLG626087S, BLG626423S, BLG626431S, BLG626578S, BLG626599S, BLG626602S, BLG626840S, BLG626900S, BLG626913S, BLG626926S, BLG626949S, BLG627234S, BLG627376S, BLG627426S, BLG627464S, BLG627692S, BLG627694S, BLG627697S, BLG627769S, BLG628000S, BLG628009S, BLG628091S, BLG628093S, BLG628113S, BLG626187S, BLG626352S, BLG626866S, BLG625813S, BLG625823S, BLG626080S, BLG626082S, BLG625723S, BLG625803S, BLG626893S, BLG626572S, BLG626573S, BLG626574S, BLG626645S, BLG626647S, BLG626666S, BLG626667S, BLG626797S, BLG626800S, BLG626818S, BLG626847S, BLG626852S, BLG626863S, BLG626869S, BLG626970S, BLG627181S, BLG627182S, BLG627183S, BLG627184S, BLG627185S, BLG627186S, BLG627187S, BLG627188S, BLG627189S, BLG627190S, BLG627191S, BLG626059S, BLG626910S, BLG626942S, BLG627211S, BLG627274S, BLG627300S, BLG627375S, BLG627684S, BLG627685S, BLG627687S, BLG627696S, BLG627727S, BLG628082S, BLG626956S, BLG625543S, BLG625801S, BLG625856S, BLG626071S, BLG626078S, BLG626085S, BLG626200S, BLG626668S, BLG626932S, BLG626936S, BLG626941S, BLG626426S, BLG625810S, BLG625821S, BLG625828S, BLG625829S, BLG625830S, BLG626064S, BLG626065S, BLG626412S, BLG626701S, BLG628014S, BLG626091S, BLG626093S, BLG626118S, BLG626212S, BLG626231S, BLG626211S, BLG626251S, BLG626801S, BLG626802S, BLG626804S, BLG626807S, BLG626816S, BLG626867S, BLG626899S, BLG626912S, BLG626945S, BLG626950S, BLG626951S, BLG626953S, BLG626971S, BLG626974S, BLG627314S, BLG627318S, BLG627321S, BLG627407S, BLG627412S, BLG626256S, BLG626333S, BLG626487S, BLG626965S, BLG627222S, BLG627349S, BLG627425S, BLG627632S, BLG628102S, BLG626332S, BLG627667S, BLG627851S, BLG628088S, BLG626700S, BLG625661S, BLG625800S, BLG626063S, BLG626074S, BLG626075S, BLG626076S, BLG626090S, BLG626120S, BLG626123S, BLG626124S, BLG626228S, BLG626233S, BLG626270S, BLG626362S, BLG626368S, BLG626422S, BLG626565S, BLG626566S, BLG626569S, BLG626591S, BLG626611S, BLG626614S, BLG626791S, BLG626792S, BLG627421S, BLG625898S, BLG626133S, BLG626292S, BLG625674S, BLG625685S, BLG625806S, BLG625808S, BLG625891S, BLG625892S, BLG625900S, BLG626060S, BLG626061S, BLG626083S, BLG626084S, BLG626126S, BLG626156S, BLG626157S, BLG626189S, BLG626190S, BLG626192S, BLG626193S, BLG626198S, BLG626199S, BLG626207S, BLG626209S, BLG626226S, BLG626382S, BLG626434S, BLG626503S, BLG626990S, BLG626383S, BLG626769S, BLG626948S, BLG627422S, BLG625833S, BLG625664S, BLG625665S, BLG625667S, BLG625671S, BLG625672S, BLG625673S, BLG625678S, BLG625679S, BLG625680S, BLG625682S, BLG625684S, BLG625686S, BLG625687S, BLG625718S, BLG625719S, BLG625720S, BLG625722S, BLG625725S, BLG625726S, BLG625805S, BLG625807S, BLG626158S, BLG626159S, BLG626160S, BLG626161S, BLG626265S, BLG626266S, BLG626267S, BLG626268S, BLG626269S, BLG626271S, BLG626272S, BLG626273S, BLG626274S, BLG626275S, BLG626276S, BLG626278S, BLG626279S, BLG626280S, BLG626281S, BLG626282S, BLG626283S, BLG626284S, BLG626285S, BLG626286S, BLG626287S, BLG626288S, BLG626289S, BLG626290S, BLG626375S, BLG626473S, BLG626477S, BLG626478S, BLG626483S, BLG626486S, BLG626495S, BLG626496S, BLG626497S, BLG626498S, BLG626502S, BLG626512S, BLG626516S, BLG626527S, BLG626529S, BLG626536S, BLG626537S, BLG626538S, BLG626539S, BLG626540S, BLG626772S, BLG626773S, BLG626774S, BLG626775S, BLG626778S, BLG627449S, BLG626777S, BLG627423S, BLG625815S, BLG625827S, BLG626069S, BLG626165S, BLG626543S, BLG626603S, BLG625659S, BLG625812S, BLG626072S, BLG627312S, BLG627333S, BLG627344S, BLG625544S, BLG626125S, BLG626191S, BLG626196S, BLG626963S, BLG625802S, BLG627508S, BLG627628S, BLG627699S, BLG627757S, BLG627758S, BLG627852S, BLG628067S, BLG628069S, BLG628154S, BLG628156S, BLG628181S, BLG628207S, BLG626436S, BLG626508S, BLG626889S, BLG626521S, BLG626864S, BLG626920S, BLG626988S, BLG626989S, BLG626997S, BLG627009S, BLG627017S, BLG627028S, BLG627196S, BLG627245S, BLG627393S, BLG627434S, BLG627495S, BLG627497S, BLG627523S, BLG627559S, BLG627577S, BLG627698S, BLG627714S, BLG627879S, BLG628007S, BLG628011S, BLG628137S, BLG626442S, BLG626488S, BLG626507S, BLG626513S, BLG626514S, BLG626515S, BLG626517S, BLG626523S, BLG626861S, BLG626883S, BLG626885S, BLG626897S, BLG626929S, BLG626986S, BLG627095S, BLG627108S, BLG627214S, BLG627231S, BLG627241S, BLG627260S, BLG627290S, BLG627292S, BLG627316S, BLG627330S, BLG627347S, BLG627373S, BLG627429S, BLG627468S, BLG627472S, BLG627473S, BLG627481S, BLG627544S, BLG627622S, BLG627695S, BLG627702S, BLG627720S, BLG627728S, BLG627763S, BLG627798S, BLG627853S, BLG627875S, BLG627882S, BLG627889S, BLG627890S, BLG628004S, BLG628006S, BLG628015S, BLG628023S, BLG628057S, BLG628061S, BLG628140S, BLG626944S, BLG627003S, BLG627004S, BLG626474S, BLG626898S, BLG627011S, BLG627013S, BLG627014S, BLG627034S, BLG627991S, BLG628145S, BLG627057S, BLG627074S, BLG627079S, BLG627238S, BLG627247S, BLG627255S, BLG627380S, BLG627545S, BLG627551S, BLG627553S, BLG627558S, BLG627620S, BLG627664S, BLG627676S, BLG627688S, BLG627715S, BLG627722S, BLG627855S, BLG628126S, BLG628136S, BLG628141S, BLG628188S, BLG626946S, BLG626964S, BLG627015S, BLG627043S, BLG627050S, BLG627086S, BLG627335S, BLG628138S, BLG628146S, BLG626998S, BLG626437S, BLG626509S, BLG626524S, BLG626528S, BLG626903S, BLG626924S, BLG627374S, BLG627501S, BLG627555S, BLG627707S, BLG626938S, BLG626939S, BLG626433S, BLG626441S, BLG626472S, BLG626475S, BLG626485S, BLG626499S, BLG626500S, BLG626504S, BLG626510S, BLG626518S, BLG626519S, BLG626520S, BLG626530S, BLG626531S, BLG626896S, BLG626909S, BLG626962S, BLG626969S, BLG626972S, BLG626996S, BLG627007S, BLG627023S, BLG627254S, BLG627282S, BLG626479S, BLG626482S, BLG626490S, BLG626494S, BLG626511S, BLG626906S, BLG626937S, BLG626955S, BLG626978S, BLG626983S, BLG627019S, BLG627976S, BLG627995S, BLG628049S, BLG628050S, BLG628144S, BLG627953S, BLG626940S, BLG626128S, BLG626413S, BLG626414S, BLG626491S, BLG626613S, BLG626684S, BLG626790S, BLG626904S, BLG627337S, BLG627348S, BLG627420S, BLG627453S, BLG627461S, BLG627474S, BLG628028S, BLG628099S, BLG625811S, BLG626081S, BLG626129S, BLG626230S, BLG626439S, BLG626570S, BLG626901S, BLG627484S, BLG627693S, BLG628017S, BLG628092S, BLG628112S, BLG626356S, BLG626421S, BLG626811S, BLG627470S, BLG625727S, BLG625837S, BLG627896S, BLG627952S, BLG628003S, BLG628077S, BLG628125S, BLG626350S, BLG627424S, BLG627319S, BLG627106S, BLG627401S, BLG627729S, BLG627471S, BLG627341S, BLG626428S, BLG627054S, BLG626553S, BLG626575S, BLG626875S, BLG627734S, BLG626135S, BLG626178S, BLG627419S, BLG626663S, BLG627414S, BLG627416S, BLG627334S, BLG626369S, BLG626134S, BLG626239S, BLG626505S, BLG627415S, BLG627320S, BLG627351S, BLG627313S, BLG627391S, BLG627394S, BLG626637S, BLG626650S, BLG626086S, BLG626132S, BLG626171S, BLG626235S, BLG626236S, BLG626365S, BLG626367S, BLG626415S, BLG626992S, BLG626564S, BLG626162S, BLG626315S, BLG626344S, BLG626358S, BLG626534S, BLG626556S, BLG626604S, BLG626606S, BLG626607S, BLG626609S, BLG626621S, BLG626627S, BLG626633S, BLG626634S, BLG626638S, BLG626639S, BLG626646S, BLG626674S, BLG626675S, BLG626676S, BLG626679S, BLG626683S, BLG626688S, BLG626689S, BLG626690S, BLG626693S, BLG626764S, BLG626766S, BLG626821S, BLG626822S, BLG626829S, BLG626831S, BLG626854S, BLG627121S, BLG627193S, BLG627208S, BLG627209S, BLG627223S, BLG627226S, BLG627229S, BLG627230S, BLG627258S, BLG627272S, BLG627309S, BLG626172S, BLG626184S, BLG626253S, BLG626298S, BLG626314S, BLG626325S, BLG626373S, BLG626376S, BLG626535S, BLG626541S, BLG626562S, BLG626610S, BLG626622S, BLG626624S, BLG626626S, BLG626630S, BLG626640S, BLG626670S, BLG626671S, BLG626672S, BLG626694S, BLG626698S, BLG626756S, BLG626770S, BLG626776S, BLG626779S, BLG626784S, BLG626786S, BLG626820S, BLG626832S, BLG626855S, BLG627113S, BLG627114S, BLG627123S, BLG627125S, BLG627200S, BLG627202S, BLG627205S, BLG627217S, BLG627220S, BLG627228S, BLG627232S, BLG627252S, BLG627269S, BLG627363S, BLG627364S, BLG626787S, BLG627207S, BLG627266S, BLG626345S, BLG626347S, BLG626322S, BLG626341S, BLG626351S, BLG626551S, BLG626771S, BLG626329S, BLG626753S, BLG626830S, BLG626169S, BLG626170S, BLG626295S, BLG626319S, BLG626321S, BLG626331S, BLG626357S, BLG626631S, BLG626706S, BLG626707S, BLG626544S, BLG626550S, BLG626561S, BLG626620S, BLG626765S, BLG626780S, BLG626857S, BLG626959S, BLG627195S, BLG627237S, BLG627239S, BLG627242S, BLG627243S, BLG627244S, BLG627365S, BLG626335S, BLG627256S, BLG627287S, BLG627368S, BLG626116S, BLG627511S, BLG626232S, BLG626598S, BLG626805S, BLG626809S, BLG626810S, BLG626881S, BLG626966S, BLG626973S, BLG627322S, BLG627288S, BLG625668S, BLG626860S, BLG627369S, BLG626612S, BLG627709S, BLG627726S, BLG628030S, BLG628213S, BLG626635S, BLG626877S, BLG626884S, BLG626227S, BLG626346S, BLG626555S, BLG626755S, BLG626215S, BLG626293S, BLG626294S, BLG626297S, BLG626302S, BLG626317S, BLG626318S, BLG626320S, BLG626323S, BLG626588S, BLG626589S, BLG626590S, BLG626703S, BLG626785S, BLG626837S, BLG627020S, BLG627021S, BLG627066S, BLG627154S, BLG627157S, BLG627158S, BLG627159S, BLG627160S, BLG627161S, BLG627162S, BLG627163S, BLG627166S, BLG627227S, BLG627233S, BLG627235S, BLG627549S, BLG627550S, BLG625663S, BLG625848S, BLG626088S, BLG626880S, BLG626914S, BLG627405S, BLG626930S, BLG626931S, BLG627000S, BLG627216S, BLG627675S, BLG627677S, BLG625662S, BLG625814S, BLG625822S, BLG625835S, BLG626089S, BLG626237S, BLG626338S, BLG626340S, BLG626342S, BLG626386S, BLG626554S, BLG626567S, BLG626813S, BLG626815S, BLG626882S, BLG626890S, BLG626895S, BLG626916S, BLG626917S, BLG626952S, BLG626954S, BLG626967S, BLG627215S, BLG627225S, BLG627410S, BLG627413S, BLG627518S, BLG627582S, BLG627584S, BLG627585S, BLG627586S, BLG627598S, BLG627708S, BLG628072S, BLG625831S, BLG625832S, BLG625834S, BLG625836S, BLG625888S, BLG625889S, BLG625901S, BLG625903S, BLG625905S, BLG626241S, BLG626242S, BLG626243S, BLG626245S, BLG626246S, BLG626260S, BLG626261S, BLG626299S, BLG626493S, BLG626506S, BLG626525S, BLG626557S, BLG626558S, BLG626559S, BLG626560S, BLG626563S, BLG626584S, BLG626585S, BLG626705S, BLG626819S, BLG626833S, BLG626834S, BLG626835S, BLG626836S, BLG627048S, BLG627051S, BLG627052S, BLG627053S, BLG627065S, BLG627067S, BLG627068S, BLG627069S, BLG627070S, BLG627071S, BLG627078S, BLG627080S, BLG627109S, BLG627111S, BLG627145S, BLG627147S, BLG627149S, BLG627151S, BLG627240S, BLG627253S, BLG627259S, BLG627262S, BLG627263S, BLG627265S, BLG627280S, BLG627281S, BLG627284S, BLG627324S, BLG627325S, BLG627326S, BLG627432S, BLG627435S, BLG627436S, BLG627437S, BLG627475S, BLG627476S, BLG627477S, BLG627483S, BLG627487S, BLG627492S, BLG627579S, BLG627580S, BLG627633S, BLG627634S, BLG627681S, BLG625676S, BLG625677S, BLG625907S, BLG626095S, BLG626247S, BLG626248S, BLG626249S, BLG626250S, BLG626252S, BLG626254S, BLG626258S, BLG626259S, BLG626291S, BLG626324S, BLG626326S, BLG626327S, BLG626328S, BLG626330S, BLG626359S, BLG626364S, BLG626377S, BLG626378S, BLG626389S, BLG626390S, BLG626391S, BLG626581S, BLG626262S, BLG626264S, BLG625666S, BLG625669S, BLG625670S, BLG625675S, BLG625818S, BLG625820S, BLG625824S, BLG626393S, BLG626448S, BLG626580S, BLG627083S, BLG627084S, BLG627170S, BLG627171S, BLG627172S, BLG627203S, BLG627212S, BLG627564S, BLG627574S, BLG627576S, BLG627578S, BLG627635S, BLG627657S, BLG627659S, BLG627660S, BLG627662S, BLG627663S, BLG628070S, BLG628071S, BLG626121S, BLG626891S, BLG626915S, BLG626921S, BLG627219S, BLG627317S, BLG627350S, BLG627706S, BLG628100S, BLG628184S, BLG626296S, BLG626316S, BLG626355S, BLG626547S, BLG626548S, BLG626549S, BLG626632S, BLG626695S, BLG626708S, BLG627204S, BLG627221S, BLG627257S, BLG627267S, BLG627289S, BLG626334S, BLG626337S, BLG626353S, BLG626374S, BLG626616S, BLG626618S, BLG626628S, BLG626629S, BLG626677S, BLG626678S, BLG626686S, BLG626696S, BLG626697S, BLG626699S, BLG626751S, BLG626752S, BLG626754S, BLG626789S, BLG626794S, BLG626856S, BLG627118S, BLG627130S, BLG627210S, BLG627213S, BLG627307S, BLG626763S, BLG627038S, BLG627041S, BLG627088S, BLG627094S, BLG627167S, BLG627367S, BLG627381S, BLG627411S, BLG627444S, BLG627479S, BLG627493S, BLG627506S, BLG627513S, BLG627624S, BLG626750S, BLG625896S, BLG627395S, BLG627396S, BLG626451S, BLG626476S, BLG626480S, BLG626481S, BLG626484S, BLG626838S, BLG626844S, BLG626845S, BLG626871S, BLG626872S, BLG626873S, BLG626874S, BLG626878S, BLG626879S, BLG626919S, BLG626922S, BLG626923S, BLG626991S, BLG627010S, BLG627018S, BLG627327S, BLG627442S, BLG627443S, BLG627445S, BLG627446S, BLG627447S, BLG627700S, BLG627711S, BLG627712S, BLG627713S, BLG627716S, BLG627717S, BLG627718S, BLG627719S, BLG627725S, BLG626935S, BLG626361S, BLG626336S, BLG627033S, BLG627044S, BLG627045S, BLG627049S, BLG627055S, BLG627063S, BLG627073S, BLG627076S, BLG627077S, BLG627081S, BLG627082S, BLG627085S, BLG627087S, BLG627089S, BLG627090S, BLG627091S, BLG627102S, BLG627156S, BLG627306S, BLG627366S, BLG627406S, BLG627408S, BLG627409S, BLG627427S, BLG627428S, BLG627433S, BLG627438S, BLG627452S, BLG627454S, BLG627455S, BLG627456S, BLG627478S, BLG627480S, BLG627488S, BLG627491S, BLG627516S, BLG627543S, BLG627560S, BLG627596S, BLG627623S, BLG627631S, BLG627636S, BLG627638S, BLG627639S, BLG627646S, BLG627650S, BLG627651S, BLG627656S, BLG627658S, BLG627686S, BLG627733S, BLG627737S, BLG627738S, BLG627740S, BLG627741S, BLG627742S, BLG627748S, BLG627761S, BLG627766S, BLG627778S, BLG627779S, BLG627781S, BLG627786S, BLG627789S, BLG627790S, BLG627791S, BLG627792S, BLG627794S, BLG627795S, BLG627796S, BLG627797S, BLG627799S, BLG627800S, BLG627802S, BLG627804S, BLG627807S, BLG627816S, BLG627818S, BLG627819S, BLG627820S, BLG627821S, BLG627822S, BLG627824S, BLG627825S, BLG627826S, BLG627829S, BLG627830S, BLG627831S, BLG627887S, BLG627888S, BLG627891S, BLG627892S, BLG627893S, BLG627894S, BLG627895S, BLG627897S, BLG627898S, BLG627899S, BLG627900S, BLG627901S, BLG627902S, BLG627904S, BLG627905S, BLG627906S, BLG627907S, BLG627908S, BLG627909S, BLG627924S, BLG627925S, BLG627955S, BLG627963S, BLG627970S, BLG627992S, BLG627993S, BLG627997S, BLG628044S, BLG628045S, BLG628046S, BLG628047S, BLG628048S, BLG628051S, BLG628052S, BLG628053S, BLG628055S, BLG628059S, BLG628062S, BLG628064S, BLG628074S, BLG628076S, BLG628081S, BLG626576S, BLG626577S, BLG626709S, BLG626839S, BLG626841S, BLG626842S, BLG626843S, BLG627022S, BLG627026S, BLG627027S, BLG627030S, BLG627031S, BLG627032S, BLG627035S, BLG627036S, BLG627037S, BLG627039S, BLG627046S, BLG627047S, BLG627059S, BLG627060S, BLG627062S, BLG627064S, BLG627173S, BLG627175S, BLG627198S, BLG627268S, BLG627328S, BLG627329S, BLG627340S, BLG627399S, BLG627448S, BLG627457S, BLG627500S, BLG627502S, BLG627548S, BLG627557S, BLG627565S, BLG627566S, BLG627567S, BLG627599S, BLG627600S, BLG627602S, BLG627661S, BLG627673S, BLG627682S, BLG627780S, BLG627782S, BLG627784S, BLG627793S, BLG627806S, BLG627811S, BLG627812S, BLG627815S, BLG628063S, BLG628066S, BLG628114S, BLG626204S, BLG626244S, BLG625894S, BLG626664S, BLG626862S, BLG627339S, BLG626925S, BLG626928S, BLG626960S, BLG627218S, BLG627311S, BLG626870S, BLG627346S, BLG626371S, BLG626372S, BLG626545S, BLG626546S, BLG626641S, BLG626643S, BLG626704S, BLG626167S, BLG626213S, BLG626625S, BLG626680S, BLG626681S, BLG626691S, BLG626781S, BLG626782S, BLG627499S, BLG626692S, BLG627271S, BLG627273S, BLG627275S, BLG627278S, BLG627283S, BLG627040S, BLG625890S, BLG626360S, BLG626757S, BLG626759S, BLG626760S, BLG627854S, BLG626405S, BLG626533S, BLG627856S, BLG626388S, BLG627841S, BLG626823S, BLG626348S, BLG626761S, BLG626826S, BLG628013S, BLG625853S, BLG625902S, BLG626381S, BLG626343S, BLG627846S, BLG626977S, BLG627776S, BLG626394S, BLG626349S, BLG627120S, BLG627849S, BLG627858S, BLG627270S, BLG627771S, BLG627840S, BLG627843S, BLG627845S, BLG627850S, BLG625857S, BLG626206S, BLG626214S, BLG626300S, BLG626301S, BLG626396S, BLG626398S, BLG626399S, BLG626400S, BLG626401S, BLG626402S, BLG626403S, BLG626404S, BLG626406S, BLG626408S, BLG626409S, BLG626417S, BLG626418S, BLG626419S, BLG626420S, BLG626435S, BLG626438S, BLG626450S, BLG626579S, BLG626582S, BLG626583S, BLG626586S, BLG626587S, BLG627092S, BLG627096S, BLG627097S, BLG627099S, BLG627100S, BLG627101S, BLG627103S, BLG627104S, BLG627115S, BLG627117S, BLG627119S, BLG627122S, BLG627124S, BLG627126S, BLG627127S, BLG627129S, BLG627131S, BLG627132S, BLG627152S, BLG627153S, BLG627164S, BLG627168S, BLG627169S, BLG627302S, BLG627303S, BLG627304S, BLG627308S, BLG627352S, BLG627377S, BLG627379S, BLG627392S, BLG627397S, BLG627398S, BLG627402S, BLG627430S, BLG627431S, BLG627439S, BLG627441S, BLG627458S, BLG627466S, BLG627494S, BLG627498S, BLG627504S, BLG627505S, BLG627507S, BLG627510S, BLG627512S, BLG627514S, BLG627515S, BLG627517S, BLG627519S, BLG627520S, BLG627521S, BLG627522S, BLG627524S, BLG627525S, BLG627527S, BLG627528S, BLG627530S, BLG627531S, BLG627532S, BLG627533S, BLG627534S, BLG627535S, BLG627536S, BLG627537S, BLG627538S, BLG627539S, BLG627540S, BLG627541S, BLG627546S, BLG627552S, BLG627569S, BLG627571S, BLG627572S, BLG627573S, BLG627581S, BLG627587S, BLG627588S, BLG627590S, BLG627591S, BLG627593S, BLG627594S, BLG627595S, BLG627601S, BLG627665S, BLG627666S, BLG627668S, BLG627669S, BLG627670S, BLG627678S, BLG627679S, BLG626642S, BLG626410S, BLG626411S, BLG626655S, BLG626166S, BLG626392S, BLG626397S, BLG626617S, BLG625854S, BLG627842S, BLG627847S, BLG625855S, BLG627042S, BLG627061S, BLG627093S, BLG627098S, BLG627440S, BLG627450S, BLG627509S, BLG627547S, BLG627568S, BLG627575S, BLG627592S, BLG627597S, BLG627603S, BLG627625S, BLG627626S, BLG627627S, BLG627630S, BLG627640S, BLG627643S, BLG627644S, BLG627645S, BLG627647S, BLG627648S, BLG627652S, BLG627653S, BLG627654S, BLG627655S, BLG627701S, BLG627705S, BLG627724S, BLG627730S, BLG627736S, BLG627739S, BLG627743S, BLG627744S, BLG627746S, BLG627747S, BLG627752S, BLG627753S, BLG627754S, BLG6277
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
- Reason for Recall:
- There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
CRT-D DTBA2QQ VIVA QUAD XT IS4/DF4 INTL, Model Number DTBA2QQ; Implantable Cardioverter Defibrillators
Product Codes/Lot Numbers:
GTIN 00643169720169, Lot Serial Numbers: BLG624311S, BLG624755S, BLG625177S, BLG625181S, BLG625182S, BLG625188S, BLG625193S, BLG625324S, BLG625325S, BLG625434S, BLG625596S, BLG625598S, BLG625606S, BLG625608S, BLG625636S, BLG625647S, BLG625702S, BLG625740S, BLG625741S, BLG625745S, BLG625962S, BLG626105S, BLG626141S, BLG624936S, BLG624987S, BLG624995S, BLG625164S, BLG625167S, BLG625168S, BLG625242S, BLG625247S, BLG625565S, BLG625566S, BLG625568S, BLG625573S, BLG625616S, BLG625619S, BLG625620S, BLG625624S, BLG625651S, BLG625653S, BLG625655S, BLG625656S, BLG625489S, BLG625612S, BLG625910S, BLG624312S, BLG625331S, BLG624790S, BLG624843S, BLG624880S, BLG624881S, BLG624878S, BLG625644S, BLG624773S, BLG625148S, BLG625224S, BLG624958S, BLG625000S, BLG624751S, BLG624884S, BLG624890S, BLG624895S, BLG624896S, BLG624901S, BLG625171S, BLG625173S, BLG625175S, BLG625197S, BLG625571S, BLG625597S, BLG625605S, BLG625626S, BLG625630S, BLG625705S, BLG625876S, BLG626149S, BLG625650S, BLG625631S, BLG625228S, BLG625243S, BLG624759S, BLG625953S, BLG624816S, BLG624819S, BLG624827S, BLG624863S, BLG624889S, BLG624792S, BLG624795S, BLG625013S, BLG625023S, BLG625025S, BLG625304S, BLG625308S, BLG625311S, BLG625312S, BLG625317S, BLG625318S, BLG625319S, BLG625320S, BLG625321S, BLG625323S, BLG625241S, BLG624874S, BLG625239S, BLG625225S, BLG624879S, BLG625220S, BLG624947S, BLG624952S, BLG624954S, BLG625001S, BLG625652S, BLG624750S, BLG624754S, BLG624761S, BLG624762S, BLG624763S, BLG624765S, BLG625189S, BLG625303S, BLG625751S, BLG625746S, BLG625756S, BLG625861S, BLG625730S, BLG625749S, BLG624968S, BLG624969S, BLG624979S, BLG624984S, BLG625215S, BLG625216S, BLG625218S, BLG625219S, BLG625238S, BLG625245S, BLG624988S, BLG625842S, BLG625868S, BLG625945S, BLG626037S, BLG624976S, BLG624977S, BLG625213S, BLG625240S, BLG626737S, BLG625172S, BLG625195S, BLG624885S, BLG624886S, BLG624899S, BLG625092S, BLG625185S, BLG625192S, BLG624999S, BLG625231S, BLG624778S, BLG624848S, BLG625196S, BLG625301S, BLG624927S, BLG624956S, BLG625040S, BLG626471S, BLG626710S, BLG626712S, BLG626732S, BLG626736S, BLG626739S, BLG624897S, BLG625326S, BLG625637S, BLG625654S, BLG625657S, BLG625002S, BLG625103S, BLG625139S, BLG625146S, BLG625194S, BLG625211S, BLG625226S, BLG625236S, BLG625237S, BLG625250S, BLG625305S, BLG625419S, BLG625420S, BLG625615S, BLG625617S, BLG625234S, BLG625205S, BLG625266S, BLG624990S, BLG625004S, BLG625014S, BLG625249S, BLG625273S, BLG625274S, BLG625281S, BLG625284S, BLG625350S, BLG625366S, BLG625435S, BLG625440S, BLG625444S, BLG625589S, BLG625590S, BLG625591S, BLG625592S, BLG625593S, BLG625594S, BLG625600S, BLG625602S, BLG625603S, BLG625604S, BLG625742S, BLG625744S, BLG625747S, BLG625748S, BLG625754S, BLG626025S, BLG626053S, BLG625235S, BLG625275S, BLG625276S, BLG625278S, BLG625285S, BLG625288S, BLG625289S, BLG625291S, BLG625293S, BLG625294S, BLG625297S, BLG625300S, BLG625360S, BLG625368S, BLG625369S, BLG625373S, BLG625377S, BLG625382S, BLG625383S, BLG625384S, BLG624998S, BLG626720S, BLG624883S, BLG624891S, BLG625176S, BLG625599S, BLG625174S, BLG625186S, BLG624752S, BLG624756S, BLG624757S, BLG625179S, BLG625180S, BLG624950S, BLG624783S, BLG625010S, BLG625610S, BLG624804S, BLG624808S, BLG624993S, BLG625455S, BLG625642S, BLG625872S, BLG624900S, BLG624796S, BLG625272S, BLG624963S, BLG625280S, BLG625363S, BLG625158S, BLG625262S, BLG625267S, BLG625270S, BLG625299S, BLG625333S, BLG625334S, BLG625381S, BLG625534S, BLG625690S, BLG625699S, BLG625263S, BLG625327S, BLG625328S, BLG625330S, BLG625361S, BLG625370S, BLG625372S, BLG625375S, BLG625379S, BLG625396S, BLG625398S, BLG625399S, BLG625401S, BLG625429S, BLG625442S, BLG625446S, BLG625460S, BLG625514S, BLG625525S, BLG625533S, BLG625535S, BLG625541S, BLG625689S, BLG625698S, BLG625704S, BLG625715S, BLG625783S, BLG625374S, BLG625454S, BLG625457S, BLG625459S, BLG625695S, BLG625761S, BLG625925S, BLG625929S, BLG626003S, BLG625347S, BLG625351S, BLG625352S, BLG625353S, BLG625436S, BLG625438S, BLG625439S, BLG625441S, BLG625998S, BLG626046S, BLG625091S, BLG625314S, BLG625316S, BLG625344S, BLG625461S, BLG625465S, BLG625468S, BLG625520S, BLG625529S, BLG625530S, BLG625531S, BLG625532S, BLG625536S, BLG625537S, BLG625696S, BLG625697S, BLG625707S, BLG625712S, BLG625713S, BLG625762S, BLG625765S, BLG625766S, BLG625769S, BLG625770S, BLG625771S, BLG625777S, BLG625778S, BLG625784S, BLG625785S, BLG625786S, BLG625789S, BLG625984S, BLG625988S, BLG626010S, BLG626011S, BLG626028S, BLG626038S, BLG626096S, BLG626100S, BLG626108S, BLG625283S, BLG625286S, BLG625292S, BLG625338S, BLG625341S, BLG625342S, BLG625343S, BLG625873S, BLG625874S, BLG625759S, BLG625775S, BLG625776S, BLG625157S, BLG625210S, BLG625296S, BLG625395S, BLG625409S, BLG625469S, BLG625688S, BLG625767S, BLG625773S, BLG625790S, BLG625886S, BLG625916S, BLG625924S, BLG625978S, BLG625979S, BLG626013S, BLG625449S, BLG625521S, BLG625526S, BLG625935S, BLG626015S, BLG626106S, BLG625515S, BLG625517S, BLG625518S, BLG625519S, BLG625760S, BLG626047S, BLG625133S, BLG625138S, BLG625279S, BLG625306S, BLG625307S, BLG625310S, BLG625322S, BLG625371S, BLG625380S, BLG625407S, BLG625425S, BLG625451S, BLG625524S, BLG625774S, BLG625788S, BLG625999S, BLG626009S, BLG625132S, BLG625135S, BLG625156S, BLG625298S, BLG625329S, BLG625376S, BLG625378S, BLG625408S, BLG625463S, BLG625522S, BLG625540S, BLG625694S, BLG625763S, BLG625768S, BLG625772S, BLG625779S, BLG625782S, BLG625791S, BLG625875S, BLG625880S, BLG625882S, BLG625922S, BLG625926S, BLG625983S, BLG626000S, BLG626002S, BLG626005S, BLG626026S, BLG626029S, BLG626031S, BLG626034S, BLG626040S, BLG626048S, BLG626049S, BLG626050S, BLG626102S, BLG626107S, BLG626109S, BLG625365S, BLG625412S, BLG625413S, BLG625416S, BLG625417S, BLG625445S, BLG625692S, BLG625714S, BLG625764S, BLG625792S, BLG625793S, BLG625927S, BLG625977S, BLG625990S, BLG625993S, BLG626021S, BLG626042S, BLG626044S, BLG626052S, BLG626054S, BLG626055S, BLG625313S, BLG625795S, BLG625796S, BLG625981S, BLG625781S, BLG625153S, BLG625359S, BLG625392S, BLG625394S, BLG625424S, BLG625443S, BLG625512S, BLG625538S, BLG625539S, BLG625711S, BLG625780S, BLG625794S, BLG625755S, BLG625456S, BLG624784S, BLG625230S, BLG625658S, BLG624894S, BLG625187S, BLG625703S, BLG625108S, BLG625207S, BLG625362S, BLG625287S, BLG625295S, BLG625315S, BLG625467S, BLG625494S, BLG625508S, BLG625523S, BLG625559S, BLG625560S, BLG625564S, BLG625706S, BLG625716S, BLG625845S, BLG625846S, BLG626036S, BLG626714S, BLG625975S, BLG625976S, BLG625595S, BLG626027S, BLG624794S, BLG624873S, BLG625079S, BLG625081S, BLG625163S, BLG625214S, BLG625421S, BLG625431S, BLG625561S, BLG625583S, BLG625841S, BLG625860S, BLG625866S, BLG625869S, BLG625871S, BLG625878S, BLG626140S, BLG626146S, BLG626147S, BLG626151S, BLG626152S, BLG626458S, BLG626459S, BLG626464S, BLG626465S, BLG624777S, BLG625005S, BLG625008S, BLG624313S, BLG624767S, BLG625190S, BLG626113S, BLG624805S, BLG625100S, BLG625110S, BLG625198S, BLG625199S, BLG625389S, BLG625390S, BLG625426S, BLG625432S, BLG625437S, BLG625969S, BLG625971S, BLG626019S, BLG626022S, BLG625607S, BLG624776S, BLG624749S, BLG626103S, BLG624791S, BLG625120S, BLG625121S, BLG625928S, BLG625938S, BLG625965S, BLG625972S, BLG625140S, BLG625125S, BLG625130S, BLG625919S, BLG625920S, BLG625939S, BLG625940S, BLG625941S, BLG624844S, BLG624924S, BLG625037S, BLG625038S, BLG624846S, BLG624920S, BLG625119S, BLG625122S, BLG625930S, BLG625936S, BLG626024S, BLG624921S, BLG625128S, BLG624788S, BLG624835S, BLG624838S, BLG624839S, BLG624845S, BLG624866S, BLG625039S, BLG625047S, BLG625050S, BLG625124S, BLG625127S, BLG625129S, BLG625908S, BLG624829S, BLG624830S, BLG624925S, BLG625036S, BLG625909S, BLG625937S, BLG624824S, BLG624850S, BLG624853S, BLG624825S, BLG624828S, BLG624865S, BLG624877S, BLG625042S, BLG625049S, BLG625055S, BLG624849S, BLG624841S, BLG624869S, BLG624875S, BLG624915S, BLG625106S, BLG624771S, BLG624772S, BLG624774S, BLG624775S, BLG624858S, BLG624870S, BLG624871S, BLG624872S, BLG625016S, BLG625017S, BLG625018S, BLG625019S, BLG625020S, BLG625072S, BLG625145S, BLG625149S, BLG625150S, BLG625151S, BLG625165S, BLG625166S, BLG625169S, BLG625170S, BLG625257S, BLG625258S, BLG625259S, BLG625260S, BLG625261S, BLG625264S, BLG625393S, BLG625450S, BLG625452S, BLG625464S, BLG625466S, BLG625470S, BLG625471S, BLG625473S, BLG625474S, BLG625476S, BLG625611S, BLG625614S, BLG625629S, BLG625954S, BLG625955S, BLG625956S, BLG626007S, BLG624810S, BLG624811S, BLG624812S, BLG624813S, BLG624814S, BLG624815S, BLG624817S, BLG624818S, BLG624820S, BLG624821S, BLG624822S, BLG624807S, BLG624809S, BLG625021S, BLG625022S, BLG625105S, BLG625155S, BLG625159S, BLG625160S, BLG625271S, BLG626018S, BLG625028S, BLG625510S, BLG624766S, BLG624768S, BLG624770S, BLG624823S, BLG624854S, BLG624855S, BLG624856S, BLG624857S, BLG625076S, BLG625077S, BLG625078S, BLG625080S, BLG625082S, BLG625083S, BLG625085S, BLG625086S, BLG625087S, BLG625088S, BLG625208S, BLG625209S, BLG625253S, BLG625254S, BLG625255S, BLG625256S, BLG625339S, BLG625345S, BLG625346S, BLG625491S, BLG625633S, BLG625634S, BLG625635S, BLG625639S, BLG625643S, BLG625645S, BLG625646S, BLG625648S, BLG625649S, BLG625957S, BLG625958S, BLG625959S, BLG625960S, BLG625961S, BLG624887S, BLG624888S, BLG625447S, BLG625844S, BLG625883S, BLG626098S, BLG624836S, BLG624837S, BLG624867S, BLG624842S, BLG624780S, BLG624781S, BLG624802S, BLG624803S, BLG624859S, BLG624860S, BLG624861S, BLG625009S, BLG625074S, BLG625075S, BLG625094S, BLG625095S, BLG625098S, BLG625099S, BLG625101S, BLG625102S, BLG625115S, BLG625116S, BLG625117S, BLG625147S, BLG625251S, BLG625252S, BLG625388S, BLG626012S, BLG626014S, BLG624834S, BLG624985S, BLG626110S, BLG624919S, BLG625045S, BLG625046S, BLG625048S, BLG625054S, BLG624941S, BLG625012S, BLG625069S, BLG625206S, BLG625227S, BLG625367S, BLG625404S, BLG625569S, BLG624779S, BLG624797S, BLG624798S, BLG624799S, BLG624909S, BLG624978S, BLG624980S, BLG624981S, BLG624982S, BLG624983S, BLG624986S, BLG624991S, BLG624992S, BLG624997S, BLG625003S, BLG625006S, BLG625026S, BLG625053S, BLG625056S, BLG625057S, BLG625065S, BLG625066S, BLG625068S, BLG625118S, BLG625136S, BLG625141S, BLG625143S, BLG625144S, BLG625349S, BLG625354S, BLG625356S, BLG625358S, BLG625385S, BLG625397S, BLG625402S, BLG625738S, BLG625739S, BLG625752S, BLG625753S, BLG625912S, BLG625914S, BLG625915S, BLG625917S, BLG625942S, BLG625944S, BLG625946S, BLG625948S, BLG625950S, BLG625973S, BLG625982S, BLG625986S, BLG625987S, BLG625989S, BLG625991S, BLG625994S, BLG625995S, BLG626004S, BLG626016S, BLG626017S, BLG625638S, BLG625640S, BLG625572S, BLG625575S, BLG625641S, BLG625732S, BLG625736S, BLG624833S, BLG625007S, BLG624786S, BLG624840S, BLG624876S, BLG624831S, BLG624847S, BLG624882S, BLG625178S, BLG625183S, BLG625184S, BLG625109S, BLG626731S, BLG625406S, BLG625411S, BLG625418S, BLG625576S, BLG625578S, BLG625579S, BLG625580S, BLG625582S, BLG625584S, BLG625585S, BLG625586S, BLG625587S, BLG625588S, BLG625847S, BLG625881S, BLG625217S, BLG625222S, BLG625483S, BLG625516S, BLG625562S, BLG625710S, BLG625838S, BLG625839S, BLG625865S, BLG625870S, BLG625877S, BLG626136S, BLG626138S, BLG625029S, BLG625843S, BLG625947S, BLG626032S, BLG625011S, BLG625015S, BLG625403S, BLG625405S, BLG625632S, BLG625717S, BLG625728S, BLG625731S, BLG625733S, BLG625734S, BLG625735S, BLG625737S, BLG624787S, BLG624864S, BLG625131S, BLG624785S, BLG625089S, BLG625090S, BLG625911S, BLG625980S, BLG625137S, BLG625934S, BLG624902S, BLG625290S, BLG625478S, BLG625485S, BLG625486S, BLG625495S, BLG625496S, BLG625497S, BLG625498S, BLG625499S, BLG625527S, BLG625528S, BLG625567S, BLG625570S, BLG625613S, BLG625691S, BLG625693S, BLG625700S, BLG625701S, BLG625708S, BLG625709S, BLG624826S, BLG624851S, BLG624913S, BLG624989S, BLG625126S, BLG625974S, BLG626023S, BLG625142S, BLG625154S, BLG625277S, BLG625355S, BLG625879S, BLG625885S, BLG625472S, BLG625487S, BLG625859S, BLG626142S, BLG626155S, BLG626452S, BLG626453S, BLG626466S, BLG626729S, BLG626748S, BLG625024S, BLG625067S, BLG625484S, BLG625574S, BLG625577S, BLG625840S, BLG625858S, BLG625862S, BLG625863S, BLG625864S, BLG625884S, BLG625887S, BLG625943S, BLG625951S, BLG626030S, BLG626033S, BLG626039S, BLG626043S, BLG626099S, BLG626101S, BLG626104S, BLG626111S, BLG626115S, BLG626137S, BLG626139S, BLG626143S, BLG626144S, BLG626145S, BLG626148S, BLG626150S, BLG626454S, BLG626456S, BLG626457S, BLG626460S, BLG626461S, BLG626462S, BLG626463S, BLG626468S, BLG626469S, BLG626470S, BLG626717S, BLG626718S, BLG626719S, BLG626726S, BLG626727S, BLG626735S, BLG626738S, BLG626740S, BLG626741S, BLG626742S, BLG626743S, BLG626744S, BLG626745S, BLG626746S, BLG626747S, BLG626749S, BLG625391S, BLG625070S, BLG625743S, BLG624801S, BLG625027S, BLG625627S, BLG625758S, BLG625787S, BLG625073S, BLG624914S, BLG624917S, BLG624922S, BLG624923S, BLG625030S, BLG625041S, BLG625123S, BLG625134S, BLG625918S, BLG625932S, BLG625933S, BLG624789S, BLG624994S, BLG625043S, BLG625921S, BLG625475S, BLG625477S, BLG625480S, BLG625482S, BLG625488S, BLG625490S, BLG625492S, BLG625500S, BLG625501S, BLG625504S, BLG625506S, BLG624862S, BLG625044S, BLG625107S, BLG625618S, BLG626006S, BLG625563S, BLG626455S, BLG625268S, BLG626114S, BLG625161S, BLG625162S, BLG625348S, BLG625410S, BLG625415S, BLG625071S, BLG625096S, BLG625112S, BLG625113S, BLG625332S, BLG625335S, BLG625336S, BLG625337S, BLG625622S, BLG625952S, BLG625963S, BLG625964S, BLG625913S, BLG624832S, BLG625931S, BLG626724S, BLG626728S, BLG626467S, BLG626711S, BLG626713S, BLG626716S, BLG626721S, BLG626722S, BLG626723S, BLG626725S, BLG626730S, BLG626734S; GTIN 00643169980143, Lot Serial Numbers: BLG626995S, BLG628179S, BLG628075S, BLG628182S, BLG627589S, BLG625906S, BLG626056S, BLG626163S, BLG626164S, BLG626489S, BLG626492S, BLG626501S, BLG626649S, BLG626651S, BLG626957S, BLG626958S, BLG626961S, BLG627146S, BLG627148S, BLG627150S, BLG627192S, BLG627224S, BLG627704S, BLG628016S, BLG626173S, BLG626174S, BLG626175S, BLG626176S, BLG626179S, BLG626180S, BLG626182S, BLG626183S, BLG626185S, BLG626186S, BLG626188S, BLG626194S, BLG626195S, BLG626858S, BLG626859S, BLG626868S, BLG626892S, BLG626908S, BLG626933S, BLG627199S, BLG627286S, BLG626407S, BLG626542S, BLG626608S, BLG626592S, BLG626593S, BLG626594S, BLG626595S, BLG626596S, BLG626597S, BLG626094S, BLG626122S, BLG627285S, BLG625724S, BLG626370S, BLG625721S, BLG625825S, BLG625826S, BLG625893S, BLG626062S, BLG625681S, BLG625809S, BLG625819S, BLG625850S, BLG625895S, BLG625904S, BLG626058S, BLG626092S, BLG626119S, BLG626808S, BLG626902S, BLG627371S, BLG627486S, BLG628037S, BLG627370S, BLG626229S, BLG626339S, BLG627206S, BLG627459S, BLG627690S, BLG628078S, BLG625816S, BLG625897S, BLG626181S, BLG626313S, BLG626354S, BLG626424S, BLG626886S, BLG626814S, BLG626888S, BLG627336S, BLG627343S, BLG626648S, BLG627323S, BLG627404S, BLG627296S, BLG627554S, BLG627960S, BLG628080S, BLG628110S, BLG625817S, BLG627291S, BLG627764S, BLG627469S, BLG627489S, BLG625899S, BLG625683S, BLG625849S, BLG626067S, BLG626073S, BLG626130S, BLG626131S, BLG626197S, BLG626416S, BLG626425S, BLG626427S, BLG626429S, BLG626444S, BLG625851S, BLG625660S, BLG625804S, BLG626087S, BLG626423S, BLG626431S, BLG626578S, BLG626599S, BLG626602S, BLG626840S, BLG626900S, BLG626913S, BLG626926S, BLG626949S, BLG627234S, BLG627376S, BLG627426S, BLG627464S, BLG627692S, BLG627694S, BLG627697S, BLG627769S, BLG628000S, BLG628009S, BLG628091S, BLG628093S, BLG628113S, BLG626187S, BLG626352S, BLG626866S, BLG625813S, BLG625823S, BLG626080S, BLG626082S, BLG625723S, BLG625803S, BLG626893S, BLG626572S, BLG626573S, BLG626574S, BLG626645S, BLG626647S, BLG626666S, BLG626667S, BLG626797S, BLG626800S, BLG626818S, BLG626847S, BLG626852S, BLG626863S, BLG626869S, BLG626970S, BLG627181S, BLG627182S, BLG627183S, BLG627184S, BLG627185S, BLG627186S, BLG627187S, BLG627188S, BLG627189S, BLG627190S, BLG627191S, BLG626059S, BLG626910S, BLG626942S, BLG627211S, BLG627274S, BLG627300S, BLG627375S, BLG627684S, BLG627685S, BLG627687S, BLG627696S, BLG627727S, BLG628082S, BLG626956S, BLG625543S, BLG625801S, BLG625856S, BLG626071S, BLG626078S, BLG626085S, BLG626200S, BLG626668S, BLG626932S, BLG626936S, BLG626941S, BLG626426S, BLG625810S, BLG625821S, BLG625828S, BLG625829S, BLG625830S, BLG626064S, BLG626065S, BLG626412S, BLG626701S, BLG628014S, BLG626091S, BLG626093S, BLG626118S, BLG626212S, BLG626231S, BLG626211S, BLG626251S, BLG626801S, BLG626802S, BLG626804S, BLG626807S, BLG626816S, BLG626867S, BLG626899S, BLG626912S, BLG626945S, BLG626950S, BLG626951S, BLG626953S, BLG626971S, BLG626974S, BLG627314S, BLG627318S, BLG627321S, BLG627407S, BLG627412S, BLG626256S, BLG626333S, BLG626487S, BLG626965S, BLG627222S, BLG627349S, BLG627425S, BLG627632S, BLG628102S, BLG626332S, BLG627667S, BLG627851S, BLG628088S, BLG626700S, BLG625661S, BLG625800S, BLG626063S, BLG626074S, BLG626075S, BLG626076S, BLG626090S, BLG626120S, BLG626123S, BLG626124S, BLG626228S, BLG626233S, BLG626270S, BLG626362S, BLG626368S, BLG626422S, BLG626565S, BLG626566S, BLG626569S, BLG626591S, BLG626611S, BLG626614S, BLG626791S, BLG626792S, BLG627421S, BLG625898S, BLG626133S, BLG626292S, BLG625674S, BLG625685S, BLG625806S, BLG625808S, BLG625891S, BLG625892S, BLG625900S, BLG626060S, BLG626061S, BLG626083S, BLG626084S, BLG626126S, BLG626156S, BLG626157S, BLG626189S, BLG626190S, BLG626192S, BLG626193S, BLG626198S, BLG626199S, BLG626207S, BLG626209S, BLG626226S, BLG626382S, BLG626434S, BLG626503S, BLG626990S, BLG626383S, BLG626769S, BLG626948S, BLG627422S, BLG625833S, BLG625664S, BLG625665S, BLG625667S, BLG625671S, BLG625672S, BLG625673S, BLG625678S, BLG625679S, BLG625680S, BLG625682S, BLG625684S, BLG625686S, BLG625687S, BLG625718S, BLG625719S, BLG625720S, BLG625722S, BLG625725S, BLG625726S, BLG625805S, BLG625807S, BLG626158S, BLG626159S, BLG626160S, BLG626161S, BLG626265S, BLG626266S, BLG626267S, BLG626268S, BLG626269S, BLG626271S, BLG626272S, BLG626273S, BLG626274S, BLG626275S, BLG626276S, BLG626278S, BLG626279S, BLG626280S, BLG626281S, BLG626282S, BLG626283S, BLG626284S, BLG626285S, BLG626286S, BLG626287S, BLG626288S, BLG626289S, BLG626290S, BLG626375S, BLG626473S, BLG626477S, BLG626478S, BLG626483S, BLG626486S, BLG626495S, BLG626496S, BLG626497S, BLG626498S, BLG626502S, BLG626512S, BLG626516S, BLG626527S, BLG626529S, BLG626536S, BLG626537S, BLG626538S, BLG626539S, BLG626540S, BLG626772S, BLG626773S, BLG626774S, BLG626775S, BLG626778S, BLG627449S, BLG626777S, BLG627423S, BLG625815S, BLG625827S, BLG626069S, BLG626165S, BLG626543S, BLG626603S, BLG625659S, BLG625812S, BLG626072S, BLG627312S, BLG627333S, BLG627344S, BLG625544S, BLG626125S, BLG626191S, BLG626196S, BLG626963S, BLG625802S, BLG627508S, BLG627628S, BLG627699S, BLG627757S, BLG627758S, BLG627852S, BLG628067S, BLG628069S, BLG628154S, BLG628156S, BLG628181S, BLG628207S, BLG626436S, BLG626508S, BLG626889S, BLG626521S, BLG626864S, BLG626920S, BLG626988S, BLG626989S, BLG626997S, BLG627009S, BLG627017S, BLG627028S, BLG627196S, BLG627245S, BLG627393S, BLG627434S, BLG627495S, BLG627497S, BLG627523S, BLG627559S, BLG627577S, BLG627698S, BLG627714S, BLG627879S, BLG628007S, BLG628011S, BLG628137S, BLG626442S, BLG626488S, BLG626507S, BLG626513S, BLG626514S, BLG626515S, BLG626517S, BLG626523S, BLG626861S, BLG626883S, BLG626885S, BLG626897S, BLG626929S, BLG626986S, BLG627095S, BLG627108S, BLG627214S, BLG627231S, BLG627241S, BLG627260S, BLG627290S, BLG627292S, BLG627316S, BLG627330S, BLG627347S, BLG627373S, BLG627429S, BLG627468S, BLG627472S, BLG627473S, BLG627481S, BLG627544S, BLG627622S, BLG627695S, BLG627702S, BLG627720S, BLG627728S, BLG627763S, BLG627798S, BLG627853S, BLG627875S, BLG627882S, BLG627889S, BLG627890S, BLG628004S, BLG628006S, BLG628015S, BLG628023S, BLG628057S, BLG628061S, BLG628140S, BLG626944S, BLG627003S, BLG627004S, BLG626474S, BLG626898S, BLG627011S, BLG627013S, BLG627014S, BLG627034S, BLG627991S, BLG628145S, BLG627057S, BLG627074S, BLG627079S, BLG627238S, BLG627247S, BLG627255S, BLG627380S, BLG627545S, BLG627551S, BLG627553S, BLG627558S, BLG627620S, BLG627664S, BLG627676S, BLG627688S, BLG627715S, BLG627722S, BLG627855S, BLG628126S, BLG628136S, BLG628141S, BLG628188S, BLG626946S, BLG626964S, BLG627015S, BLG627043S, BLG627050S, BLG627086S, BLG627335S, BLG628138S, BLG628146S, BLG626998S, BLG626437S, BLG626509S, BLG626524S, BLG626528S, BLG626903S, BLG626924S, BLG627374S, BLG627501S, BLG627555S, BLG627707S, BLG626938S, BLG626939S, BLG626433S, BLG626441S, BLG626472S, BLG626475S, BLG626485S, BLG626499S, BLG626500S, BLG626504S, BLG626510S, BLG626518S, BLG626519S, BLG626520S, BLG626530S, BLG626531S, BLG626896S, BLG626909S, BLG626962S, BLG626969S, BLG626972S, BLG626996S, BLG627007S, BLG627023S, BLG627254S, BLG627282S, BLG626479S, BLG626482S, BLG626490S, BLG626494S, BLG626511S, BLG626906S, BLG626937S, BLG626955S, BLG626978S, BLG626983S, BLG627019S, BLG627976S, BLG627995S, BLG628049S, BLG628050S, BLG628144S, BLG627953S, BLG626940S, BLG626128S, BLG626413S, BLG626414S, BLG626491S, BLG626613S, BLG626684S, BLG626790S, BLG626904S, BLG627337S, BLG627348S, BLG627420S, BLG627453S, BLG627461S, BLG627474S, BLG628028S, BLG628099S, BLG625811S, BLG626081S, BLG626129S, BLG626230S, BLG626439S, BLG626570S, BLG626901S, BLG627484S, BLG627693S, BLG628017S, BLG628092S, BLG628112S, BLG626356S, BLG626421S, BLG626811S, BLG627470S, BLG625727S, BLG625837S, BLG627896S, BLG627952S, BLG628003S, BLG628077S, BLG628125S, BLG626350S, BLG627424S, BLG627319S, BLG627106S, BLG627401S, BLG627729S, BLG627471S, BLG627341S, BLG626428S, BLG627054S, BLG626553S, BLG626575S, BLG626875S, BLG627734S, BLG626135S, BLG626178S, BLG627419S, BLG626663S, BLG627414S, BLG627416S, BLG627334S, BLG626369S, BLG626134S, BLG626239S, BLG626505S, BLG627415S, BLG627320S, BLG627351S, BLG627313S, BLG627391S, BLG627394S, BLG626637S, BLG626650S, BLG626086S, BLG626132S, BLG626171S, BLG626235S, BLG626236S, BLG626365S, BLG626367S, BLG626415S, BLG626992S, BLG626564S, BLG626162S, BLG626315S, BLG626344S, BLG626358S, BLG626534S, BLG626556S, BLG626604S, BLG626606S, BLG626607S, BLG626609S, BLG626621S, BLG626627S, BLG626633S, BLG626634S, BLG626638S, BLG626639S, BLG626646S, BLG626674S, BLG626675S, BLG626676S, BLG626679S, BLG626683S, BLG626688S, BLG626689S, BLG626690S, BLG626693S, BLG626764S, BLG626766S, BLG626821S, BLG626822S, BLG626829S, BLG626831S, BLG626854S, BLG627121S, BLG627193S, BLG627208S, BLG627209S, BLG627223S, BLG627226S, BLG627229S, BLG627230S, BLG627258S, BLG627272S, BLG627309S, BLG626172S, BLG626184S, BLG626253S, BLG626298S, BLG626314S, BLG626325S, BLG626373S, BLG626376S, BLG626535S, BLG626541S, BLG626562S, BLG626610S, BLG626622S, BLG626624S, BLG626626S, BLG626630S, BLG626640S, BLG626670S, BLG626671S, BLG626672S, BLG626694S, BLG626698S, BLG626756S, BLG626770S, BLG626776S, BLG626779S, BLG626784S, BLG626786S, BLG626820S, BLG626832S, BLG626855S, BLG627113S, BLG627114S, BLG627123S, BLG627125S, BLG627200S, BLG627202S, BLG627205S, BLG627217S, BLG627220S, BLG627228S, BLG627232S, BLG627252S, BLG627269S, BLG627363S, BLG627364S, BLG626787S, BLG627207S, BLG627266S, BLG626345S, BLG626347S, BLG626322S, BLG626341S, BLG626351S, BLG626551S, BLG626771S, BLG626329S, BLG626753S, BLG626830S, BLG626169S, BLG626170S, BLG626295S, BLG626319S, BLG626321S, BLG626331S, BLG626357S, BLG626631S, BLG626706S, BLG626707S, BLG626544S, BLG626550S, BLG626561S, BLG626620S, BLG626765S, BLG626780S, BLG626857S, BLG626959S, BLG627195S, BLG627237S, BLG627239S, BLG627242S, BLG627243S, BLG627244S, BLG627365S, BLG626335S, BLG627256S, BLG627287S, BLG627368S, BLG626116S, BLG627511S, BLG626232S, BLG626598S, BLG626805S, BLG626809S, BLG626810S, BLG626881S, BLG626966S, BLG626973S, BLG627322S, BLG627288S, BLG625668S, BLG626860S, BLG627369S, BLG626612S, BLG627709S, BLG627726S, BLG628030S, BLG628213S, BLG626635S, BLG626877S, BLG626884S, BLG626227S, BLG626346S, BLG626555S, BLG626755S, BLG626215S, BLG626293S, BLG626294S, BLG626297S, BLG626302S, BLG626317S, BLG626318S, BLG626320S, BLG626323S, BLG626588S, BLG626589S, BLG626590S, BLG626703S, BLG626785S, BLG626837S, BLG627020S, BLG627021S, BLG627066S, BLG627154S, BLG627157S, BLG627158S, BLG627159S, BLG627160S, BLG627161S, BLG627162S, BLG627163S, BLG627166S, BLG627227S, BLG627233S, BLG627235S, BLG627549S, BLG627550S, BLG625663S, BLG625848S, BLG626088S, BLG626880S, BLG626914S, BLG627405S, BLG626930S, BLG626931S, BLG627000S, BLG627216S, BLG627675S, BLG627677S, BLG625662S, BLG625814S, BLG625822S, BLG625835S, BLG626089S, BLG626237S, BLG626338S, BLG626340S, BLG626342S, BLG626386S, BLG626554S, BLG626567S, BLG626813S, BLG626815S, BLG626882S, BLG626890S, BLG626895S, BLG626916S, BLG626917S, BLG626952S, BLG626954S, BLG626967S, BLG627215S, BLG627225S, BLG627410S, BLG627413S, BLG627518S, BLG627582S, BLG627584S, BLG627585S, BLG627586S, BLG627598S, BLG627708S, BLG628072S, BLG625831S, BLG625832S, BLG625834S, BLG625836S, BLG625888S, BLG625889S, BLG625901S, BLG625903S, BLG625905S, BLG626241S, BLG626242S, BLG626243S, BLG626245S, BLG626246S, BLG626260S, BLG626261S, BLG626299S, BLG626493S, BLG626506S, BLG626525S, BLG626557S, BLG626558S, BLG626559S, BLG626560S, BLG626563S, BLG626584S, BLG626585S, BLG626705S, BLG626819S, BLG626833S, BLG626834S, BLG626835S, BLG626836S, BLG627048S, BLG627051S, BLG627052S, BLG627053S, BLG627065S, BLG627067S, BLG627068S, BLG627069S, BLG627070S, BLG627071S, BLG627078S, BLG627080S, BLG627109S, BLG627111S, BLG627145S, BLG627147S, BLG627149S, BLG627151S, BLG627240S, BLG627253S, BLG627259S, BLG627262S, BLG627263S, BLG627265S, BLG627280S, BLG627281S, BLG627284S, BLG627324S, BLG627325S, BLG627326S, BLG627432S, BLG627435S, BLG627436S, BLG627437S, BLG627475S, BLG627476S, BLG627477S, BLG627483S, BLG627487S, BLG627492S, BLG627579S, BLG627580S, BLG627633S, BLG627634S, BLG627681S, BLG625676S, BLG625677S, BLG625907S, BLG626095S, BLG626247S, BLG626248S, BLG626249S, BLG626250S, BLG626252S, BLG626254S, BLG626258S, BLG626259S, BLG626291S, BLG626324S, BLG626326S, BLG626327S, BLG626328S, BLG626330S, BLG626359S, BLG626364S, BLG626377S, BLG626378S, BLG626389S, BLG626390S, BLG626391S, BLG626581S, BLG626262S, BLG626264S, BLG625666S, BLG625669S, BLG625670S, BLG625675S, BLG625818S, BLG625820S, BLG625824S, BLG626393S, BLG626448S, BLG626580S, BLG627083S, BLG627084S, BLG627170S, BLG627171S, BLG627172S, BLG627203S, BLG627212S, BLG627564S, BLG627574S, BLG627576S, BLG627578S, BLG627635S, BLG627657S, BLG627659S, BLG627660S, BLG627662S, BLG627663S, BLG628070S, BLG628071S, BLG626121S, BLG626891S, BLG626915S, BLG626921S, BLG627219S, BLG627317S, BLG627350S, BLG627706S, BLG628100S, BLG628184S, BLG626296S, BLG626316S, BLG626355S, BLG626547S, BLG626548S, BLG626549S, BLG626632S, BLG626695S, BLG626708S, BLG627204S, BLG627221S, BLG627257S, BLG627267S, BLG627289S, BLG626334S, BLG626337S, BLG626353S, BLG626374S, BLG626616S, BLG626618S, BLG626628S, BLG626629S, BLG626677S, BLG626678S, BLG626686S, BLG626696S, BLG626697S, BLG626699S, BLG626751S, BLG626752S, BLG626754S, BLG626789S, BLG626794S, BLG626856S, BLG627118S, BLG627130S, BLG627210S, BLG627213S, BLG627307S, BLG626763S, BLG627038S, BLG627041S, BLG627088S, BLG627094S, BLG627167S, BLG627367S, BLG627381S, BLG627411S, BLG627444S, BLG627479S, BLG627493S, BLG627506S, BLG627513S, BLG627624S, BLG626750S, BLG625896S, BLG627395S, BLG627396S, BLG626451S, BLG626476S, BLG626480S, BLG626481S, BLG626484S, BLG626838S, BLG626844S, BLG626845S, BLG626871S, BLG626872S, BLG626873S, BLG626874S, BLG626878S, BLG626879S, BLG626919S, BLG626922S, BLG626923S, BLG626991S, BLG627010S, BLG627018S, BLG627327S, BLG627442S, BLG627443S, BLG627445S, BLG627446S, BLG627447S, BLG627700S, BLG627711S, BLG627712S, BLG627713S, BLG627716S, BLG627717S, BLG627718S, BLG627719S, BLG627725S, BLG626935S, BLG626361S, BLG626336S, BLG627033S, BLG627044S, BLG627045S, BLG627049S, BLG627055S, BLG627063S, BLG627073S, BLG627076S, BLG627077S, BLG627081S, BLG627082S, BLG627085S, BLG627087S, BLG627089S, BLG627090S, BLG627091S, BLG627102S, BLG627156S, BLG627306S, BLG627366S, BLG627406S, BLG627408S, BLG627409S, BLG627427S, BLG627428S, BLG627433S, BLG627438S, BLG627452S, BLG627454S, BLG627455S, BLG627456S, BLG627478S, BLG627480S, BLG627488S, BLG627491S, BLG627516S, BLG627543S, BLG627560S, BLG627596S, BLG627623S, BLG627631S, BLG627636S, BLG627638S, BLG627639S, BLG627646S, BLG627650S, BLG627651S, BLG627656S, BLG627658S, BLG627686S, BLG627733S, BLG627737S, BLG627738S, BLG627740S, BLG627741S, BLG627742S, BLG627748S, BLG627761S, BLG627766S, BLG627778S, BLG627779S, BLG627781S, BLG627786S, BLG627789S, BLG627790S, BLG627791S, BLG627792S, BLG627794S, BLG627795S, BLG627796S, BLG627797S, BLG627799S, BLG627800S, BLG627802S, BLG627804S, BLG627807S, BLG627816S, BLG627818S, BLG627819S, BLG627820S, BLG627821S, BLG627822S, BLG627824S, BLG627825S, BLG627826S, BLG627829S, BLG627830S, BLG627831S, BLG627887S, BLG627888S, BLG627891S, BLG627892S, BLG627893S, BLG627894S, BLG627895S, BLG627897S, BLG627898S, BLG627899S, BLG627900S, BLG627901S, BLG627902S, BLG627904S, BLG627905S, BLG627906S, BLG627907S, BLG627908S, BLG627909S, BLG627924S, BLG627925S, BLG627955S, BLG627963S, BLG627970S, BLG627992S, BLG627993S, BLG627997S, BLG628044S, BLG628045S, BLG628046S, BLG628047S, BLG628048S, BLG628051S, BLG628052S, BLG628053S, BLG628055S, BLG628059S, BLG628062S, BLG628064S, BLG628074S, BLG628076S, BLG628081S, BLG626576S, BLG626577S, BLG626709S, BLG626839S, BLG626841S, BLG626842S, BLG626843S, BLG627022S, BLG627026S, BLG627027S, BLG627030S, BLG627031S, BLG627032S, BLG627035S, BLG627036S, BLG627037S, BLG627039S, BLG627046S, BLG627047S, BLG627059S, BLG627060S, BLG627062S, BLG627064S, BLG627173S, BLG627175S, BLG627198S, BLG627268S, BLG627328S, BLG627329S, BLG627340S, BLG627399S, BLG627448S, BLG627457S, BLG627500S, BLG627502S, BLG627548S, BLG627557S, BLG627565S, BLG627566S, BLG627567S, BLG627599S, BLG627600S, BLG627602S, BLG627661S, BLG627673S, BLG627682S, BLG627780S, BLG627782S, BLG627784S, BLG627793S, BLG627806S, BLG627811S, BLG627812S, BLG627815S, BLG628063S, BLG628066S, BLG628114S, BLG626204S, BLG626244S, BLG625894S, BLG626664S, BLG626862S, BLG627339S, BLG626925S, BLG626928S, BLG626960S, BLG627218S, BLG627311S, BLG626870S, BLG627346S, BLG626371S, BLG626372S, BLG626545S, BLG626546S, BLG626641S, BLG626643S, BLG626704S, BLG626167S, BLG626213S, BLG626625S, BLG626680S, BLG626681S, BLG626691S, BLG626781S, BLG626782S, BLG627499S, BLG626692S, BLG627271S, BLG627273S, BLG627275S, BLG627278S, BLG627283S, BLG627040S, BLG625890S, BLG626360S, BLG626757S, BLG626759S, BLG626760S, BLG627854S, BLG626405S, BLG626533S, BLG627856S, BLG626388S, BLG627841S, BLG626823S, BLG626348S, BLG626761S, BLG626826S, BLG628013S, BLG625853S, BLG625902S, BLG626381S, BLG626343S, BLG627846S, BLG626977S, BLG627776S, BLG626394S, BLG626349S, BLG627120S, BLG627849S, BLG627858S, BLG627270S, BLG627771S, BLG627840S, BLG627843S, BLG627845S, BLG627850S, BLG625857S, BLG626206S, BLG626214S, BLG626300S, BLG626301S, BLG626396S, BLG626398S, BLG626399S, BLG626400S, BLG626401S, BLG626402S, BLG626403S, BLG626404S, BLG626406S, BLG626408S, BLG626409S, BLG626417S, BLG626418S, BLG626419S, BLG626420S, BLG626435S, BLG626438S, BLG626450S, BLG626579S, BLG626582S, BLG626583S, BLG626586S, BLG626587S, BLG627092S, BLG627096S, BLG627097S, BLG627099S, BLG627100S, BLG627101S, BLG627103S, BLG627104S, BLG627115S, BLG627117S, BLG627119S, BLG627122S, BLG627124S, BLG627126S, BLG627127S, BLG627129S, BLG627131S, BLG627132S, BLG627152S, BLG627153S, BLG627164S, BLG627168S, BLG627169S, BLG627302S, BLG627303S, BLG627304S, BLG627308S, BLG627352S, BLG627377S, BLG627379S, BLG627392S, BLG627397S, BLG627398S, BLG627402S, BLG627430S, BLG627431S, BLG627439S, BLG627441S, BLG627458S, BLG627466S, BLG627494S, BLG627498S, BLG627504S, BLG627505S, BLG627507S, BLG627510S, BLG627512S, BLG627514S, BLG627515S, BLG627517S, BLG627519S, BLG627520S, BLG627521S, BLG627522S, BLG627524S, BLG627525S, BLG627527S, BLG627528S, BLG627530S, BLG627531S, BLG627532S, BLG627533S, BLG627534S, BLG627535S, BLG627536S, BLG627537S, BLG627538S, BLG627539S, BLG627540S, BLG627541S, BLG627546S, BLG627552S, BLG627569S, BLG627571S, BLG627572S, BLG627573S, BLG627581S, BLG627587S, BLG627588S, BLG627590S, BLG627591S, BLG627593S, BLG627594S, BLG627595S, BLG627601S, BLG627665S, BLG627666S, BLG627668S, BLG627669S, BLG627670S, BLG627678S, BLG627679S, BLG626642S, BLG626410S, BLG626411S, BLG626655S, BLG626166S, BLG626392S, BLG626397S, BLG626617S, BLG625854S, BLG627842S, BLG627847S, BLG625855S, BLG627042S, BLG627061S, BLG627093S, BLG627098S, BLG627440S, BLG627450S, BLG627509S, BLG627547S, BLG627568S, BLG627575S, BLG627592S, BLG627597S, BLG627603S, BLG627625S, BLG627626S, BLG627627S, BLG627630S, BLG627640S, BLG627643S, BLG627644S, BLG627645S, BLG627647S, BLG627648S, BLG627652S, BLG627653S, BLG627654S, BLG627655S, BLG627701S, BLG627705S, BLG627724S, BLG627730S, BLG627736S, BLG627739S, BLG627743S, BLG627744S, BLG627746S, BLG627747S, BLG627752S, BLG627753S, BLG627754S, BLG6277
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1739-2023
Related Recalls
ICD-DR DDBB2D1 EVERA XT IS1/DF1 INTL, Model Number DDBB2D1; Implantable Cardioverter Defibrillators
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.
ICD-DR DDMC3D1 EVERA MRI S OUS/US DF1, Model Number DDMC3D1; Implantable Cardioverter Defibrillators
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.
ICD DDPA2D1G COBALT XT DR MRI DF1 GOLD, Model Number DDPA2D1G; Implantable Cardioverter Defibrillators
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.