R3(TM) 28 MM I.D., 62 MM O.D. CONSTRAINED LINER, REF 71339162, QTY: (1), STERILE EO. Product Usage: The R3 Constrained liner Acetabular System is a cemented or uncemented prosthesis intended to replace a hip joint. The Constrained liner is intended for primary or revision patients at high risk for hip dislocation due to a history of prior dislocation, bone loss, soft tissue laxity, neuromuscular disease, or intra-operative instability and for whom all other options to constrained acetabular components have been considered. The R3 Constrained liner is intended for single use only.

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Lot numbers: 14HM08148, 14HM08176, and 14HM08206
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Smith & Nephew, Inc.
Reason for Recall:
The R3 Constrained outer liner locking groove diameter is 0.030 inches oversized and may not allow the locking ring to engage into the R3 shell.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

R3(TM) 28 MM I.D., 62 MM O.D. CONSTRAINED LINER, REF 71339162, QTY: (1), STERILE EO. Product Usage: The R3 Constrained liner Acetabular System is a cemented or uncemented prosthesis intended to replace a hip joint. The Constrained liner is intended for primary or revision patients at high risk for hip dislocation due to a history of prior dislocation, bone loss, soft tissue laxity, neuromuscular disease, or intra-operative instability and for whom all other options to constrained acetabular components have been considered. The R3 Constrained liner is intended for single use only.

Product Codes/Lot Numbers:

Lot numbers: 14HM08148, 14HM08176, and 14HM08206

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-1739-2015