Biomet Oxford Partial Knee Phase 3 / Domed Lateral Femoral Drill Guide: Item Number: 32-420321 Phase 3 Femoral Drill Guide Small

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Lot Numbers: ZB160201 ZB160401 ZB161101 ZB170401 ZB170601 ZB170801 ZB170901
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Zimmer Biomet, Inc.
Reason for Recall:
Incorrect raw material used by the supplier in the manufacturing of the screw component, which could potentially lead to corrosion.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

Biomet Oxford Partial Knee Phase 3 / Domed Lateral Femoral Drill Guide: Item Number: 32-420321 Phase 3 Femoral Drill Guide Small

Product Codes/Lot Numbers:

Lot Numbers: ZB160201 ZB160401 ZB161101 ZB170401 ZB170601 ZB170801 ZB170901

Distribution:

Distributed in: CA, IN, LA, NM, NY, TX, WI

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-1736-2019

Related Recalls

Zimmer Biomet is conducting a medical device correction to update the Instructions for Use (IFU) and the Surgical Techniques for the Polaris Spinal System Translation Screw and Cypher MIS Screw System. Users should be aware that changes were made to the contraindications section indicating that the Translation Screws should not be used with a direct current stimulation, as a high probability of construct failure exists when using Translation Screws containing CoCr with direct current simulation system.