AccuChek Spirit Combo insulin infusion pump sold as part of Accu-Chek Combo System.

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    All Serial Numbers Models: 1. 05458544001, Accu-Chek Combo Kit mg US/English version. 2. 05504686001, Accu-Chek Combo Kit mg US/Spanish version. 3. 05391512001, Accu-Chek Spirit Combo Replacement Pump US/EN. 4. 05391750001, Accu-Chek Spirit Combo Demo Pump  US. 5. 04759702001, Accu-Chek Spirit Pump Kit US/CA Back Up. 6. 04540492001, Accu-Chek Spirit Pump Kit US/CA. 7. 04759729001, Accu-Chek Spirit Replacement Pump.
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

โš ๏ธ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Roche Diabetes Care, Inc.
Reason for Recall:
Roche has become aware of an increase of mechanical errors experienced by customers using the Accu-Chek Spirit Combo insulin infusion pump and Accu-Chek Spirit insulin infusion pump when the insulin pumps display E6 and El0 error messages. If the user does not act upon the E6 and E10 error messages appropriately, insulin delivery will be stopped and if unnoticed may lead to severe hyperglycemia u
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

AccuChek Spirit Combo insulin infusion pump sold as part of Accu-Chek Combo System.

Product Codes/Lot Numbers:

All Serial Numbers Models: 1. 05458544001, Accu-Chek Combo Kit mg US/English version. 2. 05504686001, Accu-Chek Combo Kit mg US/Spanish version. 3. 05391512001, Accu-Chek Spirit Combo Replacement Pump US/EN. 4. 05391750001, Accu-Chek Spirit Combo Demo Pump  US. 5. 04759702001, Accu-Chek Spirit Pump Kit US/CA Back Up. 6. 04540492001, Accu-Chek Spirit Pump Kit US/CA. 7. 04759729001, Accu-Chek Spirit Replacement Pump.

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-1733-2015

Related Recalls

The reason for recall is the meters may show an incorrect measurement unit (mmol/L rather than mg/dL) and may result in the wrong unit of measure. Which may subsequently cause an error in interpreting a result as approximately 18 times lower that the actual blood glucose. Potential immediate health consequences of the glucose meter providing a reading in the wrong units of measurement (UOM) may result in inappropriate rescue therapy for presumed hypoglycemia including urgent and repeat administration of carbohydrates. This in turn could lead to frequent episodes of hyperglycemia which is unrecognized and undertreated.Long range consequences of continued use of the affected meters could include mismanagement of the patient s diabetes for longer periods of time potentially resulting in a major health event (in general persistent and/or severe hyperglycemia) that may lead them to seek medical attention, and depending on the duration of hyperglycemia, could contribute to the development of microvascular (i.e., retinopathy, neuropathy, nephropathy) and/or macrovascular (i.e., myocardial infarction, ischemic stroke) complications of diabetes and could lead to fetal damage (abnormal fetal growth and altered organ development and maturation, fetal hypoxemia, spontaneous abortion, and congenital anomalies) in pregnant women. This issue was identified by three initial consumer complaints, with reports that the meters were presenting the incorrect unit of measure.

May 13, 2024 Diagnostic Equipment Nationwide View Details โ†’