Company:
Datascope Corporation
Reason for Recall:
The Better Bladder device provided as part of some custom tubing kits has been recalled by its manufacturer. Seal separation from the housing may allow air to enter the housing and collapse the balloon. This may increase resistance to flow in the venous line and cause a drop in ECMO circuit blood flow.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

BEQ-TOP 49101 Childrens Large ECC Pack, Custom Tubing Kit, Catalog No. 701067069R01

Product Codes/Lot Numbers:

Lot 3000059724

Distribution:

Distributed in: AL, AR, CA, DC, FL, IL, LA, MO, NC, PA, SC, TX, WA

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-1717-2018

Related Recalls

The Battery Charging Station is an optional dual-bay charger used to charge the Cardiosave IABP Lithium- Ion batteries when they are not being used to operate the IABP. Getinge has identified that a protruding screw in the left battery bay limits full insertion of certain batteries and prevents proper charging. The right bay of the Battery Charging Station is not affected by this issue.