ICD-DR DDMB1D4 EVERA MRI XT IS-1/DF4 US, Model Number DDMB1D4; Implantable Cardioverter Defibrillators
Class I - DangerousWhat Should You Do?
- Check if you have this product: GTIN 00643169720497, Lot Serial Numbers: PFZ236436H, PFZ242339H, PFZ247708H, PFZ247891H, PFZ250783H, PFZ252117H, PFZ259560H, PFZ265053H, PFZ271895H, PFZ248688H, PFZ254957H, PFZ260373H, PFZ273832H, PFZ244136H, PFZ245315H, PFZ264212H, PFZ271087H, PFZ273388H, PFZ238454H, PFZ242886H, PFZ245032H, PFZ245482H, PFZ251539H, PFZ251620H, PFZ254361H, PFZ256071H, PFZ258481H, PFZ259582H, PFZ261497H, PFZ263430H, PFZ264999H, PFZ268418H, PFZ269828H, PFZ274151H, PFZ236422H, PFZ239293H, PFZ240140H, PFZ242722H, PFZ244519H, PFZ244850H, PFZ246467H, PFZ247093H, PFZ248638H, PFZ250072H, PFZ250429H, PFZ251666H, PFZ263854H, PFZ268573H, PFZ268973H, PFZ240263H, PFZ242174H, PFZ247965H, PFZ263650H, PFZ268685H, PFZ270530H, PFZ240280H, PFZ240282H, PFZ240283H, PFZ240284H, PFZ240285H, PFZ240286H, PFZ240287H, PFZ240288H, PFZ240292H, PFZ240294H, PFZ240295H, PFZ240296H, PFZ245429H, PFZ245430H, PFZ245435H, PFZ245437H, PFZ253305H, PFZ253306H, PFZ253312H, PFZ253315H, PFZ253317H, PFZ253318H, PFZ256364H, PFZ256365H, PFZ256366H, PFZ256367H, PFZ256368H, PFZ256371H, PFZ257391H, PFZ261794H, PFZ270374H, PFZ271142H, PFZ271143H, PFZ236326H, PFZ236764H, PFZ237716H, PFZ238433H, PFZ243679H, PFZ244764H, PFZ246310H, PFZ246566H, PFZ246737H, PFZ247439H, PFZ249902H, PFZ249948H, PFZ261645H, PFZ263628H, PFZ263982H, PFZ264671H, PFZ266539H, PFZ273368H, PFZ274260H, PFZ242175H, PFZ243885H, PFZ243969H, PFZ243986H, PFZ244929H, PFZ251483H, PFZ256616H, PFZ260183H, PFZ260192H, PFZ261455H, PFZ266251H, PFZ273240H, PFZ274766H, PFZ269396H, PFZ271560H, PFZ274773H, PFZ236231H, PFZ237619H, PFZ237945H, PFZ238006H, PFZ238412H, PFZ239804H, PFZ240431H, PFZ240466H, PFZ241697H, PFZ244314H, PFZ245164H, PFZ245800H, PFZ246090H, PFZ246094H, PFZ246236H, PFZ247740H, PFZ247933H, PFZ248109H, PFZ248283H, PFZ248339H, PFZ248508H, PFZ248626H, PFZ250107H, PFZ250741H, PFZ252021H, PFZ252022H, PFZ252024H, PFZ252025H, PFZ252026H, PFZ252039H, PFZ252071H, PFZ253127H, PFZ256234H, PFZ257883H, PFZ257885H, PFZ257888H, PFZ258670H, PFZ258724H, PFZ259092H, PFZ259128H, PFZ259407H, PFZ259634H, PFZ259768H, PFZ260228H, PFZ263134H, PFZ263135H, PFZ263137H, PFZ263138H, PFZ263139H, PFZ263140H, PFZ263143H, PFZ265232H, PFZ267609H, PFZ267724H, PFZ267731H, PFZ267737H, PFZ267739H, PFZ269088H, PFZ269495H, PFZ270011H, PFZ270368H, PFZ273934H, PFZ274111H, PFZ274511H, PFZ274556H, PFZ274558H, PFZ274559H, PFZ274560H, PFZ274567H, PFZ243356H, PFZ256224H, PFZ270631H, PFZ243177H, PFZ245235H, PFZ245238H, PFZ245239H, PFZ248236H, PFZ253010H, PFZ253013H, PFZ253014H, PFZ253018H, PFZ255999H, PFZ256274H, PFZ256276H, PFZ256450H, PFZ256544H, PFZ256548H, PFZ261926H, PFZ261958H, PFZ261973H, PFZ271928H, PFZ271929H, PFZ271932H, PFZ271933H, PFZ271976H, PFZ272004H, PFZ272005H, PFZ274401H, PFZ239090H, PFZ237479H, PFZ240615H, PFZ242632H, PFZ243623H, PFZ248360H, PFZ263391H, PFZ264792H, PFZ265032H, PFZ272288H, PFZ273988H, PFZ238134H, PFZ238770H, PFZ238948H, PFZ239048H, PFZ239924H, PFZ243193H, PFZ243471H, PFZ244459H, PFZ245096H, PFZ245511H, PFZ247396H, PFZ247402H, PFZ247423H, PFZ247705H, PFZ248090H, PFZ248306H, PFZ248432H, PFZ249082H, PFZ249612H, PFZ249670H, PFZ250503H, PFZ250730H, PFZ250824H, PFZ251229H, PFZ251391H, PFZ254208H, PFZ254210H, PFZ255320H, PFZ255542H, PFZ260053H, PFZ261405H, PFZ263362H, PFZ263815H, PFZ264223H, PFZ264866H, PFZ269425H, PFZ269692H, PFZ270244H, PFZ270447H, PFZ270664H, PFZ272164H, PFZ272184H, PFZ272588H, PFZ274269H, PFZ274795H, PFZ274923H, PFZ246344H, PFZ238808H, PFZ236349H, PFZ236357H, PFZ236510H, PFZ237250H, PFZ238749H, PFZ238772H, PFZ239182H, PFZ239228H, PFZ240223H, PFZ241137H, PFZ245192H, PFZ245291H, PFZ246070H, PFZ246128H, PFZ247049H, PFZ247190H, PFZ247302H, PFZ247658H, PFZ248280H, PFZ249488H, PFZ250211H, PFZ250293H, PFZ251357H, PFZ253683H, PFZ253903H, PFZ255550H, PFZ256881H, PFZ258761H, PFZ260058H, PFZ273435H, PFZ240150H, PFZ240152H, PFZ242461H, PFZ244741H, PFZ260613H, PFZ261716H, PFZ270494H, PFZ243399H, PFZ245243H, PFZ247184H, PFZ247429H, PFZ247438H, PFZ253424H, PFZ253437H, PFZ255273H, PFZ258736H, PFZ258815H, PFZ258823H, PFZ260077H, PFZ261251H, PFZ261265H, PFZ263062H, PFZ263063H, PFZ263709H, PFZ264832H, PFZ269116H, PFZ272410H, PFZ273386H, PFZ274108H, PFZ274296H, PFZ238491H, PFZ238496H, PFZ247570H, PFZ247584H, PFZ248358H, PFZ250648H, PFZ252528H, PFZ252530H, PFZ252531H, PFZ252532H, PFZ258278H, PFZ260083H, PFZ260085H, PFZ260346H, PFZ262933H, PFZ262937H, PFZ262939H, PFZ262940H, PFZ262951H, PFZ272259H, PFZ273823H, PFZ236765H, PFZ261148H, PFZ236500H, PFZ243102H, PFZ243266H, PFZ244846H, PFZ244953H, PFZ245185H, PFZ249732H, PFZ255095H, PFZ256608H, PFZ258947H, PFZ259174H, PFZ260577H, PFZ261128H, PFZ261474H, PFZ268923H, PFZ270060H, PFZ270685H, PFZ272446H, PFZ274261H, PFZ243968H, PFZ246920H, PFZ262748H, PFZ272598H, PFZ253150H, PFZ238813H, PFZ240249H, PFZ240250H, PFZ240475H, PFZ240480H, PFZ242413H, PFZ243706H, PFZ245132H, PFZ245134H, PFZ248245H, PFZ248461H, PFZ248464H, PFZ248467H, PFZ252362H, PFZ252368H, PFZ252419H, PFZ252421H, PFZ252422H, PFZ252425H, PFZ259462H, PFZ260752H, PFZ260791H, PFZ260938H, PFZ260939H, PFZ264256H, PFZ264634H, PFZ264635H, PFZ264748H, PFZ267747H, PFZ267748H, PFZ267749H, PFZ269096H, PFZ269619H, PFZ271555H, PFZ271652H, PFZ273263H, PFZ273399H, PFZ273540H, PFZ268502H, PFZ236530H, PFZ243945H, PFZ248167H, PFZ256447H, PFZ268505H, PFZ236106H, PFZ241267H, PFZ256587H, PFZ236202H, PFZ236415H, PFZ237482H, PFZ238551H, PFZ238655H, PFZ238944H, PFZ238956H, PFZ239964H, PFZ240269H, PFZ240322H, PFZ240363H, PFZ241925H, PFZ241930H, PFZ241931H, PFZ241934H, PFZ241937H, PFZ241938H, PFZ242873H, PFZ245977H, PFZ245978H, PFZ246448H, PFZ248376H, PFZ248677H, PFZ250118H, PFZ250729H, PFZ251079H, PFZ253394H, PFZ254619H, PFZ254893H, PFZ255508H, PFZ256105H, PFZ256620H, PFZ259066H, PFZ259656H, PFZ260045H, PFZ261329H, PFZ261493H, PFZ261524H, PFZ263426H, PFZ264254H, PFZ265152H, PFZ267368H, PFZ268316H, PFZ268317H, PFZ268620H, PFZ269200H, PFZ271137H, PFZ271185H, PFZ273105H, PFZ273909H, PFZ273960H, PFZ274546H, PFZ274977H, PFZ249479H, PFZ252162H, PFZ254043H, PFZ255981H, PFZ257069H, PFZ258636H, PFZ258644H, PFZ259670H, PFZ259706H, PFZ259857H, PFZ259925H, PFZ261052H, PFZ261526H, PFZ262485H, PFZ262603H, PFZ263642H, PFZ263906H, PFZ263915H, PFZ264412H, PFZ264667H, PFZ264685H, PFZ264769H, PFZ266154H, PFZ267980H, PFZ268046H, PFZ268062H, PFZ268119H, PFZ268501H, PFZ269147H, PFZ269611H, PFZ269782H, PFZ269987H, PFZ270687H, PFZ271688H, PFZ272192H, PFZ272663H, PFZ273530H, PFZ273653H, PFZ273741H, PFZ274310H, PFZ274350H, PFZ274799H, PFZ274814H, PFZ274938H, PFZ231761H, PFZ236629H, PFZ236836H, PFZ236837H, PFZ236839H, PFZ238982H, PFZ241226H, PFZ241945H, PFZ242870H, PFZ246829H, PFZ247533H, PFZ247765H, PFZ249300H, PFZ249435H, PFZ249438H, PFZ250170H, PFZ250837H, PFZ254562H, PFZ255461H, PFZ256160H, PFZ257618H, PFZ258540H, PFZ259426H, PFZ259984H, PFZ260402H, PFZ264204H, PFZ265101H, PFZ265291H, PFZ269097H, PFZ269545H, PFZ270331H, PFZ271404H, PFZ272722H, PFZ272723H, PFZ274076H, PFZ241334H, PFZ251305H, PFZ251741H, PFZ260600H, PFZ239511H, PFZ242715H, PFZ249463H, PFZ249476H, PFZ254241H, PFZ256177H, PFZ258200H, PFZ258469H, PFZ263400H, PFZ266644H, PFZ267787H, PFZ268775H, PFZ270617H, PFZ271773H, PFZ271780H, PFZ236166H, PFZ236167H, PFZ236348H, PFZ236476H, PFZ238709H, PFZ238725H, PFZ238826H, PFZ238901H, PFZ238957H, PFZ238958H, PFZ238978H, PFZ239064H, PFZ239091H, PFZ239093H, PFZ239289H, PFZ239309H, PFZ239315H, PFZ239339H, PFZ239985H, PFZ240024H, PFZ240370H, PFZ241072H, PFZ242346H, PFZ242630H, PFZ242757H, PFZ243195H, PFZ243270H, PFZ243281H, PFZ243402H, PFZ243644H, PFZ243861H, PFZ243895H, PFZ244002H, PFZ244272H, PFZ244554H, PFZ244586H, PFZ244712H, PFZ244887H, PFZ244961H, PFZ244962H, PFZ245075H, PFZ245938H, PFZ245943H, PFZ246132H, PFZ246762H, PFZ246885H, PFZ247045H, PFZ247266H, PFZ247300H, PFZ247374H, PFZ247520H, PFZ247883H, PFZ247992H, PFZ248012H, PFZ248119H, PFZ248683H, PFZ249570H, PFZ249591H, PFZ249607H, PFZ249692H, PFZ249703H, PFZ249929H, PFZ250140H, PFZ250198H, PFZ250210H, PFZ250294H, PFZ250296H, PFZ251266H, PFZ251269H, PFZ251322H, PFZ251429H, PFZ251499H, PFZ251554H, PFZ251700H, PFZ251942H, PFZ252258H, PFZ252455H, PFZ252970H, PFZ253638H, PFZ253731H, PFZ253896H, PFZ253928H, PFZ253951H, PFZ254017H, PFZ254708H, PFZ254829H, PFZ255083H, PFZ255247H, PFZ255362H, PFZ255467H, PFZ255627H, PFZ255803H, PFZ255807H, PFZ255861H, PFZ256012H, PFZ257060H, PFZ257284H, PFZ257308H, PFZ257858H, PFZ258385H, PFZ258801H, PFZ258854H, PFZ258856H, PFZ258857H, PFZ259141H, PFZ259206H, PFZ259405H, PFZ259762H, PFZ259944H, PFZ260031H, PFZ260062H, PFZ260073H, PFZ260170H, PFZ260231H, PFZ260401H, PFZ260793H, PFZ260863H, PFZ260982H, PFZ261496H, PFZ261832H, PFZ261972H, PFZ262353H, PFZ262362H, PFZ263053H, PFZ263111H, PFZ263512H, PFZ263633H, PFZ263705H, PFZ263744H, PFZ263829H, PFZ263850H, PFZ264137H, PFZ264150H, PFZ264291H, PFZ264380H, PFZ264383H, PFZ264522H, PFZ264532H, PFZ265027H, PFZ265082H, PFZ265084H, PFZ265538H, PFZ265547H, PFZ265550H, PFZ265595H, PFZ265611H, PFZ265681H, PFZ265734H, PFZ265925H, PFZ265954H, PFZ266073H, PFZ266146H, PFZ266174H, PFZ266777H, PFZ267970H, PFZ267974H, PFZ268132H, PFZ268283H, PFZ268809H, PFZ269169H, PFZ269527H, PFZ269562H, PFZ269660H, PFZ269787H, PFZ269947H, PFZ270104H, PFZ270245H, PFZ270320H, PFZ270326H, PFZ270993H, PFZ271581H, PFZ271725H, PFZ272148H, PFZ272167H, PFZ272248H, PFZ272285H, PFZ272336H, PFZ272449H, PFZ272458H, PFZ272475H, PFZ272623H, PFZ272624H, PFZ272703H, PFZ272779H, PFZ273601H, PFZ273796H, PFZ273940H, PFZ273970H, PFZ274158H, PFZ274357H, PFZ274370H, PFZ274371H, PFZ274620H, PFZ274665H, PFZ274745H, PFZ274757H, PFZ274768H, PFZ274800H, PFZ274837H, PFZ274850H, PFZ274858H, PFZ274903H, PFZ274950H, PFZ274968H, PFZ240336H, PFZ243293H, PFZ246170H, PFZ251468H, PFZ255056H, PFZ260914H, PFZ260918H, PFZ266142H, PFZ266148H, PFZ266149H, PFZ271428H, PFZ271452H, PFZ236492H, PFZ237757H, PFZ238156H, PFZ239184H, PFZ239236H, PFZ244137H, PFZ244551H, PFZ244860H, PFZ244866H, PFZ246561H, PFZ247975H, PFZ248268H, PFZ248657H, PFZ256036H, PFZ256518H, PFZ256948H, PFZ261072H, PFZ262805H, PFZ268220H, PFZ270601H, PFZ270606H, PFZ247259H, PFZ247884H, PFZ258478H, PFZ260676H, PFZ265434H, PFZ265450H, PFZ267876H, PFZ267891H, PFZ240225H, PFZ263311H, PFZ271066H, PFZ241355H, PFZ241360H, PFZ245638H, PFZ245639H, PFZ245640H, PFZ252581H, PFZ262182H, PFZ262183H, PFZ262185H, PFZ262187H, PFZ265574H, PFZ238328H, PFZ242428H, PFZ256451H, PFZ262146H, PFZ264920H, PFZ236021H, PFZ237623H, PFZ239300H, PFZ240083H, PFZ240090H, PFZ240389H, PFZ244701H, PFZ246184H, PFZ247333H, PFZ247892H, PFZ248214H, PFZ248382H, PFZ250820H, PFZ251511H, PFZ251820H, PFZ251881H, PFZ252876H, PFZ252993H, PFZ254858H, PFZ255014H, PFZ255017H, PFZ255023H, PFZ255081H, PFZ255268H, PFZ255483H, PFZ256027H, PFZ256082H, PFZ256271H, PFZ256826H, PFZ257080H, PFZ257543H, PFZ258741H, PFZ258969H, PFZ258970H, PFZ260754H, PFZ261087H, PFZ262771H, PFZ262860H, PFZ262910H, PFZ263833H, PFZ263838H, PFZ263980H, PFZ266888H, PFZ268634H, PFZ269528H, PFZ270417H, PFZ271012H, PFZ273679H, PFZ269434H, PFZ272191H, PFZ272374H, PFZ272395H, PFZ240611H, PFZ244039H, PFZ245378H, PFZ246782H, PFZ252614H, PFZ254658H, PFZ254660H, PFZ257466H, PFZ263508H, PFZ263867H, PFZ264139H, PFZ264494H, PFZ266057H, PFZ267566H, PFZ268766H, PFZ269725H, PFZ273367H, PFZ274828H, PFZ271922H, PFZ274189H, PFZ245249H, PFZ245253H, PFZ245254H, PFZ245256H, PFZ245258H, PFZ245261H, PFZ245262H, PFZ245263H, PFZ245264H, PFZ245265H, PFZ245266H, PFZ245268H, PFZ245273H, PFZ252480H, PFZ252481H, PFZ252484H, PFZ252486H, PFZ256844H, PFZ256845H, PFZ256846H, PFZ256847H, PFZ256848H, PFZ256849H, PFZ256850H, PFZ256853H, PFZ256854H, PFZ256855H, PFZ256856H, PFZ256857H, PFZ256859H, PFZ256862H, PFZ256903H, PFZ256905H, PFZ256906H, PFZ256907H, PFZ256908H, PFZ256909H, PFZ261297H, PFZ261305H, PFZ261308H, PFZ261316H, PFZ261330H, PFZ261351H, PFZ261355H, PFZ261358H, PFZ261361H, PFZ261380H, PFZ261384H, PFZ261390H, PFZ261392H, PFZ261398H, PFZ261408H, PFZ261409H, PFZ261415H, PFZ261434H, PFZ261439H, PFZ261440H, PFZ261886H, PFZ261889H, PFZ261892H, PFZ261893H, PFZ261894H, PFZ261895H, PFZ261897H, PFZ261906H, PFZ261910H, PFZ261911H, PFZ261918H, PFZ261975H, PFZ262092H, PFZ262118H, PFZ262125H, PFZ262126H, PFZ262134H, PFZ262136H, PFZ262174H, PFZ262179H, PFZ239241H, PFZ254253H, PFZ258322H, PFZ268917H, PFZ237630H, PFZ239197H, PFZ239270H, PFZ239271H, PFZ242204H, PFZ243395H, PFZ244895H, PFZ246735H, PFZ249517H, PFZ249546H, PFZ251447H, PFZ252034H, PFZ252129H, PFZ252155H, PFZ252640H, PFZ252650H, PFZ252651H, PFZ257059H, PFZ258638H, PFZ258643H, PFZ258652H, PFZ258653H, PFZ258654H, PFZ259208H, PFZ259943H, PFZ261268H, PFZ262895H, PFZ262897H, PFZ262903H, PFZ262904H, PFZ262911H, PFZ269049H, PFZ272433H, PFZ272571H, PFZ270177H, PFZ272933H, PFZ236755H, PFZ242014H, PFZ242158H, PFZ245149H, PFZ252999H, PFZ254551H, PFZ262535H, PFZ242699H, PFZ242742H, PFZ246428H, PFZ252976H, PFZ260673H, PFZ265364H, PFZ237709H, PFZ244291H, PFZ247489H, PFZ247904H, PFZ248691H, PFZ250205H, PFZ255280H, PFZ256043H, PFZ258564H, PFZ272668H, PFZ239509H, PFZ249622H, PFZ250462H, PFZ258317H, PFZ259868H, PFZ262894H, PFZ263726H, PFZ268876H, PFZ272232H, PFZ274464H, PFZ239520H, PFZ260091H, PFZ262181H, PFZ262322H, PFZ262988H, PFZ236403H, PFZ237408H, PFZ237674H, PFZ242683H, PFZ244052H, PFZ244902H, PFZ254446H, PFZ254557H, PFZ254666H, PFZ247615H, PFZ248703H, PFZ253298H, PFZ253299H, PFZ242492H, PFZ242494H, PFZ242496H, PFZ245659H, PFZ245660H, PFZ245662H, PFZ253274H, PFZ253278H, PFZ253280H, PFZ264749H, PFZ265139H, PFZ268061H, PFZ274377H, PFZ254662H, PFZ238862H, PFZ240461H, PFZ244752H, PFZ244779H, PFZ250337H, PFZ251563H, PFZ253681H, PFZ254505H, PFZ259672H, PFZ260546H, PFZ262688H, PFZ263835H, PFZ263927H, PFZ265349H, PFZ265351H, PFZ273267H, PFZ274455H, PFZ274538H, PFZ245398H, PFZ245399H, PFZ245400H, PFZ245401H, PFZ245403H, PFZ245406H, PFZ245408H, PFZ249893H, PFZ251925H, PFZ253510H, PFZ253514H, PFZ257916H, PFZ257925H, PFZ257928H, PFZ258025H, PFZ258029H, PFZ258034H, PFZ258057H, PFZ258060H, PFZ258061H, PFZ258063H, PFZ262844H, PFZ262846H, PFZ262847H, PFZ262848H, PFZ262849H, PFZ263124H, PFZ263125H, PFZ267677H, PFZ267679H, PFZ267682H, PFZ267683H, PFZ267684H, PFZ268653H, PFZ272129H, PFZ244107H, PFZ244146H, PFZ246884H, PFZ237998H, PFZ240099H, PFZ253663H, PFZ254447H, PFZ254862H, PFZ255765H, PFZ258480H, PFZ259187H, PFZ260308H, PFZ265630H, PFZ266887H, PFZ269060H, PFZ271884H, PFZ273183H, PFZ274053H, PFZ274686H, PFZ240224H, PFZ241351H, PFZ247106H, PFZ256015H, PFZ260385H, PFZ260695H, PFZ260856H, PFZ260969H, PFZ261476H, PFZ262968H, PFZ265868H, PFZ271095H, PFZ274533H, PFZ269893H, PFZ271097H, PFZ238117H, PFZ247567H, PFZ260398H, PFZ243061H, PFZ250146H, PFZ262016H, PFZ268122H, PFZ272106H, PFZ238213H, PFZ238214H, PFZ238410H, PFZ247575H, PFZ254100H, PFZ254106H, PFZ254406H, PFZ254432H, PFZ260069H, PFZ262385H, PFZ262386H, PFZ262387H, PFZ263852H, PFZ263986H, PFZ269264H, PFZ270771H, PFZ271580H, PFZ271582H, PFZ271583H, PFZ271584H, PFZ272528H, PFZ273335H, PFZ274336H, PFZ236841H, PFZ237284H, PFZ237797H, PFZ238686H, PFZ239369H, PFZ239875H, PFZ240371H, PFZ241644H, PFZ242627H, PFZ242798H, PFZ242969H, PFZ243074H, PFZ245382H, PFZ245448H, PFZ246808H, PFZ247411H, PFZ247695H, PFZ248979H, PFZ249426H, PFZ249457H, PFZ249958H, PFZ250113H, PFZ250580H, PFZ250775H, PFZ252802H, PFZ252806H, PFZ252810H, PFZ252813H, PFZ252815H, PFZ252816H, PFZ254005H, PFZ254807H, PFZ254892H, PFZ255882H, PFZ257923H, PFZ257942H, PFZ259311H, PFZ259723H, PFZ260451H, PFZ260608H, PFZ260837H, PFZ261417H, PFZ262714H, PFZ262967H, PFZ263507H, PFZ264565H, PFZ266956H, PFZ268527H, PFZ268814H, PFZ268815H, PFZ268816H, PFZ269424H, PFZ272621H, PFZ272838H, PFZ273547H, PFZ274122H, PFZ274202H, PFZ274418H, PFZ238246H, PFZ240136H, PFZ242611H, PFZ242672H, PFZ247248H, PFZ247281H, PFZ248254H, PFZ248499H, PFZ248738H, PFZ248785H, PFZ248982H, PFZ249041H, PFZ250273H, PFZ250749H, PFZ251633H, PFZ253771H, PFZ253904H, PFZ253988H, PFZ254248H, PFZ258189H, PFZ258298H, PFZ259068H, PFZ260232H, PFZ260356H, PFZ260593H, PFZ261279H, PFZ268547H, PFZ270302H, PFZ270731H, PFZ272775H, PFZ240253H, PFZ256483H, PFZ256536H, PFZ271112H, PFZ238794H, PFZ236108H, PFZ236163H, PFZ236442H, PFZ238239H, PFZ239151H, PFZ239749H, PFZ240266H, PFZ240393H, PFZ241689H, PFZ243066H, PFZ243276H, PFZ243417H, PFZ243526H, PFZ243563H, PFZ243604H, PFZ243723H, PFZ244151H, PFZ244315H, PFZ245044H, PFZ245240H, PFZ245310H, PFZ245446H, PFZ246505H, PFZ246548H, PFZ246768H, PFZ247191H, PFZ247442H, PFZ248021H, PFZ248200H, PFZ249778H, PFZ249859H, PFZ250022H, PFZ250808H, PFZ251121H, PFZ252659H, PFZ252885H, PFZ253682H, PFZ254144H, PFZ254488H, PFZ254961H, PFZ255165H, PFZ255660H, PFZ257723H, PFZ260129H, PFZ261057H, PFZ261168H, PFZ261359H, PFZ261745H, PFZ262969H, PFZ263501H, PFZ263716H, PFZ264004H, PFZ264066H, PFZ264228H, PFZ265206H, PFZ265697H, PFZ265806H, PFZ265807H, PFZ266018H, PFZ266196H, PFZ268837H, PFZ270287H, PFZ270835H, PFZ271210H, PFZ271967H, PFZ272626H, PFZ272949H, PFZ273041H, PFZ274440H, PFZ274639H, PFZ274738H, PFZ240314H, PFZ241712H, PFZ243032H, PFZ243141H, PFZ243627H, PFZ250793H, PFZ253609H, PFZ254074H, PFZ259012H, PFZ260866H, PFZ261784H, PFZ261878H, PFZ262773H, PFZ272576H, PFZ274326H, PFZ240865H, PFZ242375H, PFZ244009H, PFZ246337H, PFZ246996H, PFZ253506H, PFZ253507H, PFZ253509H, PFZ253516H, PFZ253518H, PFZ253520H, PFZ253523H, PFZ253525H, PFZ253526H, PFZ255928H, PFZ262246H, PFZ262251H, PFZ262347H, PFZ262379H, PFZ262399H, PFZ267383H, PFZ267384H, PFZ267386H, PFZ267387H, PFZ267388H, PFZ267391H, PFZ270646H, PFZ270649H, PFZ270650H, PFZ270651H, PFZ270652H, PFZ270653H, PFZ270655H, PFZ270656H, PFZ270659H, PFZ270660H, PFZ272583H, PFZ274782H, PFZ255523H, PFZ259702H, PFZ261921H, PFZ263981H, PFZ268385H, PFZ268386H, PFZ268470H, PFZ273163H, PFZ274051H, PFZ250830H, PFZ259125H, PFZ266041H, PFZ237950H, PFZ237951H, PFZ237954H, PFZ239694H, PFZ241653H, PFZ241654H, PFZ242452H, PFZ244507H, PFZ246111H, PFZ246114H, PFZ246126H, PFZ249464H, PFZ249921H, PFZ250241H, PFZ250747H, PFZ251263H, PFZ252443H, PFZ252456H, PFZ252673H, PFZ252674H, PFZ252675H, PFZ252677H, PFZ252679H, PFZ252680H, PFZ253055H, PFZ256819H, PFZ256823H, PFZ256824H, PFZ260044H, PFZ260354H, PFZ260460H, PFZ260807H, PFZ261032H, PFZ261991H, PFZ262005H, PFZ262009H, PFZ262011H, PFZ262078H, PFZ266842H, PFZ269031H, PFZ269447H, PFZ270654H, PFZ270982H, PFZ273966H, PFZ236550H, PFZ236551H, PFZ244279H, PFZ258380H, PFZ260636H, PFZ246144H, PFZ246547H, PFZ247306H, PFZ250485H, PFZ255157H, PFZ259312H, PFZ259578H, PFZ264367H, PFZ268849H, PFZ240153H, PFZ258132H, PFZ261668H, PFZ236061H, PFZ236413H, PFZ237343H, PFZ237435H, PFZ238194H, PFZ238503H, PFZ238547H, PFZ238908H, PFZ240356H, PFZ241084H, PFZ243531H, PFZ243556H, PFZ243954H, PFZ245027H, PFZ245686H, PFZ246895H, PFZ249680H, PFZ249797H, PFZ250058H, PFZ250610H, PFZ251202H, PFZ251721H, PFZ252145H, PFZ253649H, PFZ254849H, PFZ254878H, PFZ254925H, PFZ254951H, PFZ255501H, PFZ255522H, PFZ255880H, PFZ256179H, PFZ257651H, PFZ258955H, PFZ258962H, PFZ261313H, PFZ263297H, PFZ265074H, PFZ265906H, PFZ266704H, PFZ268290H, PFZ268805H, PFZ268910H, PFZ269054H, PFZ269422H, PFZ269496H, PFZ269515H, PFZ270431H, PFZ270474H, PFZ270721H, PFZ271881H, PFZ272467H, PFZ273670H, PFZ273787H, PFZ274004H, PFZ274934H, PFZ243511H, PFZ247519H, PFZ251033H, PFZ269123H, PFZ247177H, PFZ249635H, PFZ255143H, PFZ259837H, PFZ268295H, PFZ270822H, PFZ273673H, PFZ237542H, PFZ239828H, PFZ243950H, PFZ245418H, PFZ251151H, PFZ264517H, PFZ265005H, PFZ266862H, PFZ267162H, PFZ271113H, PFZ273057H, PFZ273714H, PFZ236423H, PFZ238806H, PFZ242829H, PFZ248717H, PFZ249949H, PFZ263836H, PFZ264402H, PFZ269575H, PFZ273588H, PFZ240085H, PFZ251895H, PFZ251992H, PFZ256029H, PFZ260532H, PFZ262464H, PFZ262465H, PFZ270042H, PFZ268969H, PFZ268981H, PFZ270487H, PFZ271873H, PFZ236679H, PFZ237180H, PFZ237715H, PFZ237973H, PFZ238688H, PFZ239007H, PFZ239420H, PFZ239838H, PFZ241227H, PFZ242988H, PFZ243065H, PFZ243473H, PFZ244010H, PFZ244366H, PFZ244441H, PFZ246253H, PFZ246928H, PFZ247299H, PFZ247868H, PFZ248260H, PFZ248569H, PFZ249035H, PFZ249746H, PFZ249938H, PFZ249939H, PFZ251369H, PFZ251412H, PFZ253443H, PFZ253665H, PFZ254539H, PFZ254731H, PFZ254799H, PFZ254819H, PFZ255235H, PFZ255813H, PFZ255938H, PFZ256174H, PFZ256569H, PFZ257494H, PFZ258577H, PFZ258879H, PFZ259176H, PFZ259668H, PFZ259719H, PFZ259839H, PFZ260211H, PFZ260766H, PFZ261217H, PFZ261323H, PFZ261416H, PFZ262500H, PFZ263418H, PFZ263558H, PFZ263608H, PFZ263610H, PFZ264105H, PFZ264158H, PFZ264490H, PFZ265323H, PFZ266835H, PFZ268247H, PFZ268351H, PFZ269300H, PFZ269744H, PFZ269795H, PFZ269999H, PFZ270528H, PFZ270896H, PFZ271171H, PFZ272656H, PFZ273019H, PFZ273454H, PFZ274737H, PFZ236268H, PFZ236273H, PFZ237203H, PFZ238104H, PFZ238483H, PFZ238873H, PFZ239179H, PFZ239207H, PFZ239822H, PFZ242490H, PFZ242574H, PFZ243158H, PFZ243702H, PFZ243888H, PFZ244297H, PFZ244861H, PFZ246267H, PFZ246791H, PFZ247056H, PFZ247141H, PFZ247455H, PFZ247637H, PFZ249452H, PFZ250638H, PFZ250640H, PFZ250653H, PFZ250784H, PFZ251234H, PFZ251689H, PFZ251716H, PFZ251794H, PFZ252132H, PFZ253850H, PFZ253884H, PFZ253981H, PFZ254310H, PFZ254480H, PFZ254623H, PFZ254803H, PFZ255592H, PFZ255965H, PFZ259344H, PFZ260995H, PFZ261375H, PFZ261382H, PFZ261468H, PFZ263013H, PFZ263583H, PFZ263602H, PFZ264030H, PFZ264458H, PFZ264495H, PFZ265326H, PFZ265384H, PFZ265575H, PFZ266958H, PFZ268042H, PFZ268667H, PFZ268846H, PFZ269466H, PFZ269739H, PFZ270364H, PFZ270490H, PFZ272843H, PFZ273084H, PFZ273775H, PFZ273916H, PFZ274116H, PFZ274523H, PFZ274902H, PFZ236405H, PFZ236547H, PFZ236705H, PFZ245488H, PFZ245489H, PFZ245490H, PFZ245491H, PFZ245495H, PFZ245496H, PFZ245500H, PFZ245504H, PFZ245761H, PFZ251223H, PFZ251490H, PFZ251523H, PFZ251546H, PFZ252066H, PFZ252634H, PFZ253362H, PFZ253365H, PFZ253368H, PFZ253370H, PFZ253376H, PFZ253377H, PFZ253382H, PFZ253383H, PFZ253386H, PFZ253390H, PFZ253391H, PFZ262964H, PFZ262966H, PFZ262973H, PFZ263020H, PFZ263025H, PFZ263027H, PFZ263030H, PFZ263031H, PFZ263032H, PFZ263038H, PFZ263041H, PFZ263042H, PFZ263047H, PFZ245671H, PFZ246982H, PFZ247174H, PFZ247175H, PFZ255264H, PFZ256511H, PFZ260417H, PFZ238648H, PFZ238791H, PFZ249423H, PFZ249722H, PFZ260911H, PFZ236265H, PFZ237803H, PFZ241213H, PFZ241215H, PFZ241216H, PFZ241217H, PFZ241219H, PFZ241220H, PFZ241221H, PFZ241223H, PFZ243901H, PFZ244550H, PFZ244581H, PFZ245563H, PFZ245564H, PFZ245566H, PFZ245567H, PFZ245568H, PFZ245569H, PFZ245572H, PFZ245573H, PFZ245574H, PFZ245577H, PFZ245578H, PFZ245579H, PFZ245580H, PFZ245581H, PFZ245583H, PFZ245584H, PFZ249384H, PFZ249977H, PFZ253135H, PFZ253136H, PFZ253139H, PFZ253144H, PFZ253145H, PFZ253146H, PFZ254613H, PFZ255791H, PFZ257649H, PFZ258146H, PFZ258716H, PFZ258717H, PFZ258718H, PFZ258720H, PFZ258806H, PFZ258809H, PFZ258810H, PFZ259696H, PFZ259697H, PFZ261043H, PFZ261342H, PFZ261343H, PFZ261399H, PFZ261419H, PFZ261750H, PFZ262992H, PFZ262996H, PFZ262997H, PFZ262998H, PFZ263689H, PFZ263690H, PFZ263691H, PFZ265011H, PFZ265012H, PFZ265013H, PFZ266557H, PFZ268821H, PFZ268822H, PFZ268823H, PFZ271018H, PFZ271019H, PFZ271021H, PFZ271022H, PFZ272751H, PFZ272752H, PFZ272820H, PFZ272824H, PFZ273051H, PFZ273052H, PFZ273224H, PFZ273225H, PFZ273562H, PFZ273572H, PFZ274651H, PFZ274719H, PFZ274720H, PFZ274722H, PFZ242844H, PFZ243233H, PFZ243575H, PFZ243654H, PFZ243789H, PFZ244370H, PFZ244620H, PFZ246799H, PFZ248667H, PFZ250286H, PFZ250718H, PFZ251218H, PFZ255088H, PFZ257323H, PFZ258778H, PFZ259113H, PFZ259912H, PFZ260624H, PFZ260915H, PFZ261242H, PFZ261429H, PFZ261656H, PFZ261657H, PFZ261658H, PFZ264419H, PFZ267590H, PFZ267591H, PFZ267592H, PFZ267593H, PFZ267594H, PFZ267596H, PFZ267598H, PFZ271005H, PFZ271006H, PFZ271007H, PFZ273503H, PFZ274974H, PFZ236130H, PFZ238921H, PFZ239147H, PFZ243073H, PFZ243930H, PFZ244145H, PFZ246641H, PFZ249960H, PFZ250180H, PFZ251232H, PFZ251890H, PFZ251954H, PFZ252214H, PFZ253886H, PFZ254547H, PFZ255434H, PFZ259006H, PFZ259071H, PFZ259291H, PFZ259533H, PFZ259707H, PFZ260135H, PFZ260296H, PFZ260505H, PFZ260557H, PFZ261203H, PFZ262657H, PFZ263213H, PFZ263289H, PFZ263331H, PFZ263595H, PFZ263821H, PFZ264101H, PFZ264185H, PFZ264589H, PFZ265041H, PFZ265618H, PFZ265703H, PFZ265792H, PFZ266185H, PFZ268618H, PFZ269397H, PFZ269726H, PFZ270264H, PFZ271971H, PFZ271987H, PFZ272898H, PFZ273081H, PFZ273122H, PFZ273626H, PFZ273790H, PFZ274833H, PFZ246901H, PFZ247962H, PFZ248407H, PFZ250244H, PFZ250571H, PFZ255343H, PFZ259189H, PFZ259219H, PFZ259232H, PFZ259235H, PFZ262493H, PFZ262503H, PFZ262508H, PFZ267464H, PFZ236830H, PFZ236832H, PFZ236833H, PFZ236834H, PFZ237477H, PFZ237487H, PFZ238280H, PFZ238284H, PFZ238289H, PFZ238291H, PFZ239035H, PFZ239036H, PFZ239497H, PFZ239498H, PFZ239701H, PFZ239776H, PFZ240432H, PFZ240433H, PFZ240434H, PFZ241179H, PFZ241180H, PFZ241181H, PFZ241182H, PFZ241184H, PFZ241185H, PFZ241186H, PFZ241188H, PFZ241190H, PFZ241193H, PFZ241195H, PFZ241197H, PFZ241198H, PFZ241202H, PFZ241204H, PFZ246307H, PFZ246309H, PFZ246480H, PFZ246481H, PFZ246689H, PFZ246691H, PFZ247221H, PFZ247222H, PFZ247223H, PFZ247859H, PFZ255438H, PFZ255440H, PFZ256550H, PFZ257298H, PFZ257301H, PFZ257302H, PFZ257304H, PFZ257305H, PFZ257316H, PFZ257322H, PFZ257325H, PFZ257329H, PFZ257330H, PFZ257331H, PFZ257334H, PFZ257335H, PFZ257336H, PFZ257337H, PFZ257339H, PFZ257826H, PFZ259274H, PFZ261385H, PFZ261386H, PFZ261698H, PFZ261699H, PFZ261700H, PFZ261701H, PFZ261702H, PFZ261703H, PFZ261704H, PFZ261705H, PFZ261706H, PFZ261707H, PFZ261708H, PFZ261710H, PFZ261714H, PFZ261715H, PFZ261719H, PFZ261722H, PFZ261723H, PFZ261724H, PFZ261727H, PFZ265762H, PFZ265764H, PFZ265864H, PFZ266419H, PFZ266423H, PFZ266428H, PFZ266442H, PFZ266448H, PFZ266449H, PFZ266451H, PFZ266454H, PFZ266457H, PFZ266463H, PFZ266464H, PFZ266466H, PFZ266467H, PFZ266468H, PFZ266474H, PFZ266477H, PFZ266481H, PFZ266483H, PFZ266484H, PFZ272810H, PFZ274724H, PFZ237517H, PFZ237520H, PFZ237521H, PFZ237535H, PFZ237550H, PFZ237551H, PFZ237556H, PFZ237557H, PFZ237569H, PFZ237578H, PFZ237580H, PFZ237581H, PFZ237588H, PFZ237590H, PFZ237595H, PFZ237597H, PFZ239958H, PFZ239969H, PFZ241192H, PFZ241196H, PFZ241665H, PFZ241667H, PFZ241670H, PFZ241672H, PFZ241675H, PFZ241676H, PFZ241681H, PFZ241705H, PFZ241715H, PFZ241731H, PFZ241733H, PFZ241734H, PFZ241736H, PFZ241739H, PFZ245080H, PFZ245083H, PFZ245531H, PFZ245836H, PFZ245837H, PFZ245841H, PFZ245847H, PFZ245849H, PFZ245869H, PFZ245877H, PFZ245882H, PFZ245884H, PFZ245886H, PFZ245889H, PFZ245893H, PFZ245895H, PFZ245902H, PFZ245905H, PFZ245906H, PFZ245909H, PFZ245910H, PFZ245911H, PFZ245912H, PFZ246106H, PFZ246549H, PFZ246680H, PFZ246681H, PFZ253241H, PFZ253242H, PFZ253243H, PFZ253245H, PFZ253249H, PFZ253250H, PFZ253252H, PFZ253255H, PFZ253257H, PFZ253258H, PFZ253260H, PFZ253266H, PFZ253267H, PFZ253269H, PFZ253271H, PFZ253272H, PFZ254967H, PFZ255974H, PFZ256333H, PFZ256525H, PFZ256541H, PFZ256611H, PFZ256749H, PFZ257482H, PFZ258035H, PFZ258036H, PFZ258037H, PFZ258038H, PFZ258042H, PFZ258043H, PFZ262228H, PFZ262232H, PFZ262234H, PFZ262235H, PFZ262236H, PFZ262237H, PFZ262240H, PFZ262241H, PFZ262242H, PFZ262244H, PFZ262248H, PFZ262249H, PFZ262252H, PFZ265019H, PFZ265924H, PFZ265959H, PFZ265962H, PFZ265966H, PFZ266420H, PFZ267242H, PFZ267246H, PFZ267250H, PFZ267251H, PFZ267252H, PFZ267253H, PFZ267256H, PFZ267259H, PFZ267260H, PFZ267261H, PFZ267263H, PFZ267266H, PFZ267267H, PFZ267268H, PFZ267269H, PFZ267270H, PFZ267271H, PFZ267272H, PFZ267273H, PFZ267275H, PFZ267277H, PFZ267278H, PFZ267280H, PFZ267281H, PFZ267282H, PFZ267283H, PFZ267298H, PFZ271610H, PFZ271611H, PFZ271612H, PFZ271615H, PFZ271616H, PFZ271617H, PFZ271618H, PFZ271619H, PFZ271620H, PFZ271622H, PFZ271623H, PFZ271624H, PFZ271627H, PFZ271630H, PFZ271631H, PFZ271632H, PFZ271633H, PFZ271634H, PFZ263197H, PFZ238998H, PFZ246118H, PFZ247057H, PFZ252590H, PFZ258154H, PFZ263953H, PFZ265669H, PFZ270644H, PFZ251404H, PFZ250645H, PFZ263144H, PFZ267810H, PFZ240194H, PFZ266694H, PFZ266737H, PFZ244311H, PFZ247667H, PFZ249076H, PFZ262886H, PFZ238399H, PFZ242782H, PFZ242846H, PFZ245692H, PFZ246181H, PFZ247110H, PFZ256111H, PFZ262308H, PFZ262309H, PFZ262356H, PFZ262359H, PFZ267506H, PFZ267567H, PFZ267568H, PFZ267572H, PFZ267573H, PFZ270363H, PFZ271198H, PFZ271200H, PFZ271201H, PFZ271203H, PFZ271204H, PFZ271205H, PFZ271206H, PFZ271207H, PFZ272001H, PFZ272033H, PFZ239959H, PFZ240013H, PFZ246658H, PFZ249301H, PFZ256020H, PFZ264797H, PFZ273424H, PFZ238101H, PFZ238179H, PFZ240418H, PFZ242849H, PFZ243264H, PFZ244482H, PFZ246827H, PFZ249193H, PFZ250088H, PFZ250754H, PFZ251188H, PFZ251899H, PFZ252320H, PFZ253392H, PFZ254685H, PFZ259595H, PFZ270272H, PFZ236322H, PFZ238058H, PFZ239799H, PFZ250678H, PFZ253885H, PFZ255561H, PFZ260391H, PFZ260523H, PFZ270260H, PFZ273729H, PFZ274381H, PFZ245117H, PFZ237093H, PFZ237106H, PFZ237110H, PFZ237111H, PFZ238136H, PFZ238199H, PFZ238202H, PFZ238423H, PFZ238424H, PFZ245684H, PFZ245733H, PFZ245736H, PFZ245739H, PFZ245740H, PFZ246025H, PFZ246033H, PFZ246278H, PFZ250042H, PFZ252327H, PFZ252328H, PFZ252330H, PFZ252332H, PFZ252335H, PFZ252336H, PFZ258136H, PFZ258137H, PFZ258140H, PFZ258141H, PFZ258145H, PFZ258148H, PFZ258150H, PFZ258152H, PFZ261695H, PFZ261697H, PFZ265907H, PFZ266432H, PFZ266433H, PFZ266437H, PFZ267758H, PFZ267759H, PFZ270572H, PFZ273244H, PFZ274168H, PFZ237680H, PFZ237683H, PFZ242802H, PFZ247712H, PFZ251550H, PFZ254312H, PFZ254767H, PFZ258709H, PFZ260372H, PFZ260379H, PFZ263563H, PFZ267552H, PFZ268100H, PFZ272819H, PFZ249007H, PFZ249009H, PFZ253322H, PFZ260029H, PFZ265453H, PFZ271904H, PFZ272893H, PFZ272894H, PFZ252553H, PFZ259569H, PFZ263132H, PFZ270062H, PFZ274476H, PFZ274573H, PFZ274668H, PFZ271882H, PFZ243711H, PFZ258570H, PFZ259793H, PFZ260340H, PFZ266919H, PFZ268609H, PFZ270113H, PFZ270541H, PFZ274693H, PFZ237771H, PFZ241160H, PFZ241161H, PFZ241164H, PFZ241166H, PFZ244658H, PFZ245462H, PFZ245463H, PFZ245464H, PFZ245468H, PFZ246825H, PFZ247166H, PFZ247843H, PFZ249170H, PFZ250566H, PFZ250752H, PFZ251976H, PFZ252837H, PFZ252838H, PFZ252913H, PFZ253338H, PFZ253340H, PFZ253344H, PFZ255734H, PFZ256533H, PFZ256554H, PFZ258124H, PFZ258125H, PFZ258128H, PFZ258133H, PFZ258134H, PFZ260746H, PFZ260760H, PFZ261231H, PFZ261374H, PFZ262195H, PFZ262197H, PFZ262200H, PFZ262201H, PFZ262204H, PFZ262509H, PFZ262542H, PFZ263509H, PFZ264753H, PFZ264754H, PFZ265103H, PFZ265119H, PFZ265551H, PFZ265587H, PFZ265588H, PFZ266062H, PFZ267330H, PFZ267331H, PFZ267332H, PFZ267334H, PFZ267335H, PFZ267336H, PFZ268871H, PFZ270832H, PFZ272358H, PFZ272855H, PFZ272856H, PFZ272857H, PFZ272858H, PFZ272859H, PFZ272860H, PFZ272861H, PFZ272863H, PFZ272864H, PFZ274030H, PFZ274034H, PFZ274598H, PFZ274670H, PFZ236215H, PFZ240603H, PFZ240604H, PFZ240605H, PFZ240613H, PFZ240616H, PFZ240619H, PFZ240624H, PFZ240626H, PFZ240627H, PFZ240628H, PFZ240630H, PFZ245281H, PFZ245283H, PFZ245284H, PFZ245288H, PFZ245290H, PFZ245292H, PFZ245294H, PFZ245300H, PFZ245302H, PFZ245303H, PFZ256911H, PFZ256913H, PFZ256915H, PFZ256916H, PFZ256919H, PFZ256920H, PFZ256921H, PFZ256925H, PFZ256926H, PFZ264193H, PFZ266340H, PFZ266341H, PFZ266342H, PFZ266344H, PFZ266345H, PFZ266347H, PFZ270171H, PFZ237714H, PFZ238418H, PFZ238421H, PFZ238890H, PFZ239571H, PFZ240614H, PFZ241975H, PFZ241977H, PFZ241982H, PFZ241983H, PFZ241984H, PFZ242012H, PFZ242016H, PFZ242021H, PFZ242026H, PFZ242028H, PFZ242031H, PFZ242032H, PFZ242034H, PFZ242036H, PFZ242037H, PFZ255948H, PFZ256112H, PFZ256116H, PFZ256119H, PFZ256120H, PFZ256121H, PFZ256130H, PFZ256132H, PFZ256140H, PFZ256149H, PFZ256153H, PFZ256154H, PFZ256156H, PFZ256157H, PFZ256361H, PFZ264759H, PFZ264945H, PFZ264953H, PFZ265014H, PFZ265394H, PFZ265623H, PFZ265945H, PFZ265984H, PFZ266159H, PFZ266989H, PFZ266991H, PFZ266992H, PFZ266993H, PFZ266995H, PFZ266996H, PFZ266998H, PFZ266999H, PFZ267000H, PFZ267001H, PFZ267002H, PFZ267003H, PFZ267004H, PFZ267005H, PFZ267009H, PFZ267012H, PFZ267013H, PFZ267016H, PFZ267017H, PFZ267018H, PFZ267019H, PFZ267020H, PFZ267022H, PFZ267023H, PFZ267024H, PFZ267025H, PFZ267026H, PFZ268906H, PFZ269266H, PFZ238021H, PFZ239396H, PFZ243886H, PFZ248326H, PFZ270961H, PFZ241691H, PFZ241692H, PFZ241698H, PFZ242440H, PFZ252842H, PFZ252909H, PFZ252911H, PFZ258830H, PFZ258841H, PFZ258842H, PFZ259879H, PFZ259880H, PFZ259881H, PFZ259882H, PFZ270766H, PFZ271275H, PFZ271305H, PFZ242896H, PFZ246435H, PFZ246817H, PFZ246882H, PFZ246904H, PFZ247631H, PFZ248226H, PFZ250546H, PFZ250925H, PFZ251074H, PFZ251114H, PFZ252642H, PFZ253057H, PFZ253735H, PFZ254571H, PFZ255684H, PFZ256331H, PFZ259535H, PFZ260116H, PFZ260800H, PFZ260927H, PFZ260972H, PFZ261761H, PFZ268186H, PFZ269856H, PFZ270643H, PFZ270804H, PFZ270940H, PFZ272832H, PFZ273036H, PFZ273277H, PFZ273661H, PFZ274467H, PFZ236205H, PFZ239493H, PFZ239494H, PFZ243516H, PFZ246078H, PFZ247105H, PFZ248034H, PFZ255469H, PFZ270285H, PFZ271117H, PFZ238170H, PFZ243519H, PFZ259613H, PFZ260556H, PFZ264662H, PFZ268584H, PFZ268586H, PFZ268702H, PFZ270756H, PFZ273807H, PFZ274918H, PFZ238385H, PFZ239136H, PFZ240301H, PFZ242914H, PFZ243610H, PFZ244936H, PFZ245168H, PFZ246441H, PFZ246648H, PFZ246961H, PFZ247152H, PFZ247634H, PFZ248102H, PFZ248564H, PFZ248881H, PFZ249180H, PFZ249593H, PFZ249888H, PFZ250723H, PFZ250753H, PFZ250942H, PFZ250964H, PF
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
- Reason for Recall:
- There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
ICD-DR DDMB1D4 EVERA MRI XT IS-1/DF4 US, Model Number DDMB1D4; Implantable Cardioverter Defibrillators
Product Codes/Lot Numbers:
GTIN 00643169720497, Lot Serial Numbers: PFZ236436H, PFZ242339H, PFZ247708H, PFZ247891H, PFZ250783H, PFZ252117H, PFZ259560H, PFZ265053H, PFZ271895H, PFZ248688H, PFZ254957H, PFZ260373H, PFZ273832H, PFZ244136H, PFZ245315H, PFZ264212H, PFZ271087H, PFZ273388H, PFZ238454H, PFZ242886H, PFZ245032H, PFZ245482H, PFZ251539H, PFZ251620H, PFZ254361H, PFZ256071H, PFZ258481H, PFZ259582H, PFZ261497H, PFZ263430H, PFZ264999H, PFZ268418H, PFZ269828H, PFZ274151H, PFZ236422H, PFZ239293H, PFZ240140H, PFZ242722H, PFZ244519H, PFZ244850H, PFZ246467H, PFZ247093H, PFZ248638H, PFZ250072H, PFZ250429H, PFZ251666H, PFZ263854H, PFZ268573H, PFZ268973H, PFZ240263H, PFZ242174H, PFZ247965H, PFZ263650H, PFZ268685H, PFZ270530H, PFZ240280H, PFZ240282H, PFZ240283H, PFZ240284H, PFZ240285H, PFZ240286H, PFZ240287H, PFZ240288H, PFZ240292H, PFZ240294H, PFZ240295H, PFZ240296H, PFZ245429H, PFZ245430H, PFZ245435H, PFZ245437H, PFZ253305H, PFZ253306H, PFZ253312H, PFZ253315H, PFZ253317H, PFZ253318H, PFZ256364H, PFZ256365H, PFZ256366H, PFZ256367H, PFZ256368H, PFZ256371H, PFZ257391H, PFZ261794H, PFZ270374H, PFZ271142H, PFZ271143H, PFZ236326H, PFZ236764H, PFZ237716H, PFZ238433H, PFZ243679H, PFZ244764H, PFZ246310H, PFZ246566H, PFZ246737H, PFZ247439H, PFZ249902H, PFZ249948H, PFZ261645H, PFZ263628H, PFZ263982H, PFZ264671H, PFZ266539H, PFZ273368H, PFZ274260H, PFZ242175H, PFZ243885H, PFZ243969H, PFZ243986H, PFZ244929H, PFZ251483H, PFZ256616H, PFZ260183H, PFZ260192H, PFZ261455H, PFZ266251H, PFZ273240H, PFZ274766H, PFZ269396H, PFZ271560H, PFZ274773H, PFZ236231H, PFZ237619H, PFZ237945H, PFZ238006H, PFZ238412H, PFZ239804H, PFZ240431H, PFZ240466H, PFZ241697H, PFZ244314H, PFZ245164H, PFZ245800H, PFZ246090H, PFZ246094H, PFZ246236H, PFZ247740H, PFZ247933H, PFZ248109H, PFZ248283H, PFZ248339H, PFZ248508H, PFZ248626H, PFZ250107H, PFZ250741H, PFZ252021H, PFZ252022H, PFZ252024H, PFZ252025H, PFZ252026H, PFZ252039H, PFZ252071H, PFZ253127H, PFZ256234H, PFZ257883H, PFZ257885H, PFZ257888H, PFZ258670H, PFZ258724H, PFZ259092H, PFZ259128H, PFZ259407H, PFZ259634H, PFZ259768H, PFZ260228H, PFZ263134H, PFZ263135H, PFZ263137H, PFZ263138H, PFZ263139H, PFZ263140H, PFZ263143H, PFZ265232H, PFZ267609H, PFZ267724H, PFZ267731H, PFZ267737H, PFZ267739H, PFZ269088H, PFZ269495H, PFZ270011H, PFZ270368H, PFZ273934H, PFZ274111H, PFZ274511H, PFZ274556H, PFZ274558H, PFZ274559H, PFZ274560H, PFZ274567H, PFZ243356H, PFZ256224H, PFZ270631H, PFZ243177H, PFZ245235H, PFZ245238H, PFZ245239H, PFZ248236H, PFZ253010H, PFZ253013H, PFZ253014H, PFZ253018H, PFZ255999H, PFZ256274H, PFZ256276H, PFZ256450H, PFZ256544H, PFZ256548H, PFZ261926H, PFZ261958H, PFZ261973H, PFZ271928H, PFZ271929H, PFZ271932H, PFZ271933H, PFZ271976H, PFZ272004H, PFZ272005H, PFZ274401H, PFZ239090H, PFZ237479H, PFZ240615H, PFZ242632H, PFZ243623H, PFZ248360H, PFZ263391H, PFZ264792H, PFZ265032H, PFZ272288H, PFZ273988H, PFZ238134H, PFZ238770H, PFZ238948H, PFZ239048H, PFZ239924H, PFZ243193H, PFZ243471H, PFZ244459H, PFZ245096H, PFZ245511H, PFZ247396H, PFZ247402H, PFZ247423H, PFZ247705H, PFZ248090H, PFZ248306H, PFZ248432H, PFZ249082H, PFZ249612H, PFZ249670H, PFZ250503H, PFZ250730H, PFZ250824H, PFZ251229H, PFZ251391H, PFZ254208H, PFZ254210H, PFZ255320H, PFZ255542H, PFZ260053H, PFZ261405H, PFZ263362H, PFZ263815H, PFZ264223H, PFZ264866H, PFZ269425H, PFZ269692H, PFZ270244H, PFZ270447H, PFZ270664H, PFZ272164H, PFZ272184H, PFZ272588H, PFZ274269H, PFZ274795H, PFZ274923H, PFZ246344H, PFZ238808H, PFZ236349H, PFZ236357H, PFZ236510H, PFZ237250H, PFZ238749H, PFZ238772H, PFZ239182H, PFZ239228H, PFZ240223H, PFZ241137H, PFZ245192H, PFZ245291H, PFZ246070H, PFZ246128H, PFZ247049H, PFZ247190H, PFZ247302H, PFZ247658H, PFZ248280H, PFZ249488H, PFZ250211H, PFZ250293H, PFZ251357H, PFZ253683H, PFZ253903H, PFZ255550H, PFZ256881H, PFZ258761H, PFZ260058H, PFZ273435H, PFZ240150H, PFZ240152H, PFZ242461H, PFZ244741H, PFZ260613H, PFZ261716H, PFZ270494H, PFZ243399H, PFZ245243H, PFZ247184H, PFZ247429H, PFZ247438H, PFZ253424H, PFZ253437H, PFZ255273H, PFZ258736H, PFZ258815H, PFZ258823H, PFZ260077H, PFZ261251H, PFZ261265H, PFZ263062H, PFZ263063H, PFZ263709H, PFZ264832H, PFZ269116H, PFZ272410H, PFZ273386H, PFZ274108H, PFZ274296H, PFZ238491H, PFZ238496H, PFZ247570H, PFZ247584H, PFZ248358H, PFZ250648H, PFZ252528H, PFZ252530H, PFZ252531H, PFZ252532H, PFZ258278H, PFZ260083H, PFZ260085H, PFZ260346H, PFZ262933H, PFZ262937H, PFZ262939H, PFZ262940H, PFZ262951H, PFZ272259H, PFZ273823H, PFZ236765H, PFZ261148H, PFZ236500H, PFZ243102H, PFZ243266H, PFZ244846H, PFZ244953H, PFZ245185H, PFZ249732H, PFZ255095H, PFZ256608H, PFZ258947H, PFZ259174H, PFZ260577H, PFZ261128H, PFZ261474H, PFZ268923H, PFZ270060H, PFZ270685H, PFZ272446H, PFZ274261H, PFZ243968H, PFZ246920H, PFZ262748H, PFZ272598H, PFZ253150H, PFZ238813H, PFZ240249H, PFZ240250H, PFZ240475H, PFZ240480H, PFZ242413H, PFZ243706H, PFZ245132H, PFZ245134H, PFZ248245H, PFZ248461H, PFZ248464H, PFZ248467H, PFZ252362H, PFZ252368H, PFZ252419H, PFZ252421H, PFZ252422H, PFZ252425H, PFZ259462H, PFZ260752H, PFZ260791H, PFZ260938H, PFZ260939H, PFZ264256H, PFZ264634H, PFZ264635H, PFZ264748H, PFZ267747H, PFZ267748H, PFZ267749H, PFZ269096H, PFZ269619H, PFZ271555H, PFZ271652H, PFZ273263H, PFZ273399H, PFZ273540H, PFZ268502H, PFZ236530H, PFZ243945H, PFZ248167H, PFZ256447H, PFZ268505H, PFZ236106H, PFZ241267H, PFZ256587H, PFZ236202H, PFZ236415H, PFZ237482H, PFZ238551H, PFZ238655H, PFZ238944H, PFZ238956H, PFZ239964H, PFZ240269H, PFZ240322H, PFZ240363H, PFZ241925H, PFZ241930H, PFZ241931H, PFZ241934H, PFZ241937H, PFZ241938H, PFZ242873H, PFZ245977H, PFZ245978H, PFZ246448H, PFZ248376H, PFZ248677H, PFZ250118H, PFZ250729H, PFZ251079H, PFZ253394H, PFZ254619H, PFZ254893H, PFZ255508H, PFZ256105H, PFZ256620H, PFZ259066H, PFZ259656H, PFZ260045H, PFZ261329H, PFZ261493H, PFZ261524H, PFZ263426H, PFZ264254H, PFZ265152H, PFZ267368H, PFZ268316H, PFZ268317H, PFZ268620H, PFZ269200H, PFZ271137H, PFZ271185H, PFZ273105H, PFZ273909H, PFZ273960H, PFZ274546H, PFZ274977H, PFZ249479H, PFZ252162H, PFZ254043H, PFZ255981H, PFZ257069H, PFZ258636H, PFZ258644H, PFZ259670H, PFZ259706H, PFZ259857H, PFZ259925H, PFZ261052H, PFZ261526H, PFZ262485H, PFZ262603H, PFZ263642H, PFZ263906H, PFZ263915H, PFZ264412H, PFZ264667H, PFZ264685H, PFZ264769H, PFZ266154H, PFZ267980H, PFZ268046H, PFZ268062H, PFZ268119H, PFZ268501H, PFZ269147H, PFZ269611H, PFZ269782H, PFZ269987H, PFZ270687H, PFZ271688H, PFZ272192H, PFZ272663H, PFZ273530H, PFZ273653H, PFZ273741H, PFZ274310H, PFZ274350H, PFZ274799H, PFZ274814H, PFZ274938H, PFZ231761H, PFZ236629H, PFZ236836H, PFZ236837H, PFZ236839H, PFZ238982H, PFZ241226H, PFZ241945H, PFZ242870H, PFZ246829H, PFZ247533H, PFZ247765H, PFZ249300H, PFZ249435H, PFZ249438H, PFZ250170H, PFZ250837H, PFZ254562H, PFZ255461H, PFZ256160H, PFZ257618H, PFZ258540H, PFZ259426H, PFZ259984H, PFZ260402H, PFZ264204H, PFZ265101H, PFZ265291H, PFZ269097H, PFZ269545H, PFZ270331H, PFZ271404H, PFZ272722H, PFZ272723H, PFZ274076H, PFZ241334H, PFZ251305H, PFZ251741H, PFZ260600H, PFZ239511H, PFZ242715H, PFZ249463H, PFZ249476H, PFZ254241H, PFZ256177H, PFZ258200H, PFZ258469H, PFZ263400H, PFZ266644H, PFZ267787H, PFZ268775H, PFZ270617H, PFZ271773H, PFZ271780H, PFZ236166H, PFZ236167H, PFZ236348H, PFZ236476H, PFZ238709H, PFZ238725H, PFZ238826H, PFZ238901H, PFZ238957H, PFZ238958H, PFZ238978H, PFZ239064H, PFZ239091H, PFZ239093H, PFZ239289H, PFZ239309H, PFZ239315H, PFZ239339H, PFZ239985H, PFZ240024H, PFZ240370H, PFZ241072H, PFZ242346H, PFZ242630H, PFZ242757H, PFZ243195H, PFZ243270H, PFZ243281H, PFZ243402H, PFZ243644H, PFZ243861H, PFZ243895H, PFZ244002H, PFZ244272H, PFZ244554H, PFZ244586H, PFZ244712H, PFZ244887H, PFZ244961H, PFZ244962H, PFZ245075H, PFZ245938H, PFZ245943H, PFZ246132H, PFZ246762H, PFZ246885H, PFZ247045H, PFZ247266H, PFZ247300H, PFZ247374H, PFZ247520H, PFZ247883H, PFZ247992H, PFZ248012H, PFZ248119H, PFZ248683H, PFZ249570H, PFZ249591H, PFZ249607H, PFZ249692H, PFZ249703H, PFZ249929H, PFZ250140H, PFZ250198H, PFZ250210H, PFZ250294H, PFZ250296H, PFZ251266H, PFZ251269H, PFZ251322H, PFZ251429H, PFZ251499H, PFZ251554H, PFZ251700H, PFZ251942H, PFZ252258H, PFZ252455H, PFZ252970H, PFZ253638H, PFZ253731H, PFZ253896H, PFZ253928H, PFZ253951H, PFZ254017H, PFZ254708H, PFZ254829H, PFZ255083H, PFZ255247H, PFZ255362H, PFZ255467H, PFZ255627H, PFZ255803H, PFZ255807H, PFZ255861H, PFZ256012H, PFZ257060H, PFZ257284H, PFZ257308H, PFZ257858H, PFZ258385H, PFZ258801H, PFZ258854H, PFZ258856H, PFZ258857H, PFZ259141H, PFZ259206H, PFZ259405H, PFZ259762H, PFZ259944H, PFZ260031H, PFZ260062H, PFZ260073H, PFZ260170H, PFZ260231H, PFZ260401H, PFZ260793H, PFZ260863H, PFZ260982H, PFZ261496H, PFZ261832H, PFZ261972H, PFZ262353H, PFZ262362H, PFZ263053H, PFZ263111H, PFZ263512H, PFZ263633H, PFZ263705H, PFZ263744H, PFZ263829H, PFZ263850H, PFZ264137H, PFZ264150H, PFZ264291H, PFZ264380H, PFZ264383H, PFZ264522H, PFZ264532H, PFZ265027H, PFZ265082H, PFZ265084H, PFZ265538H, PFZ265547H, PFZ265550H, PFZ265595H, PFZ265611H, PFZ265681H, PFZ265734H, PFZ265925H, PFZ265954H, PFZ266073H, PFZ266146H, PFZ266174H, PFZ266777H, PFZ267970H, PFZ267974H, PFZ268132H, PFZ268283H, PFZ268809H, PFZ269169H, PFZ269527H, PFZ269562H, PFZ269660H, PFZ269787H, PFZ269947H, PFZ270104H, PFZ270245H, PFZ270320H, PFZ270326H, PFZ270993H, PFZ271581H, PFZ271725H, PFZ272148H, PFZ272167H, PFZ272248H, PFZ272285H, PFZ272336H, PFZ272449H, PFZ272458H, PFZ272475H, PFZ272623H, PFZ272624H, PFZ272703H, PFZ272779H, PFZ273601H, PFZ273796H, PFZ273940H, PFZ273970H, PFZ274158H, PFZ274357H, PFZ274370H, PFZ274371H, PFZ274620H, PFZ274665H, PFZ274745H, PFZ274757H, PFZ274768H, PFZ274800H, PFZ274837H, PFZ274850H, PFZ274858H, PFZ274903H, PFZ274950H, PFZ274968H, PFZ240336H, PFZ243293H, PFZ246170H, PFZ251468H, PFZ255056H, PFZ260914H, PFZ260918H, PFZ266142H, PFZ266148H, PFZ266149H, PFZ271428H, PFZ271452H, PFZ236492H, PFZ237757H, PFZ238156H, PFZ239184H, PFZ239236H, PFZ244137H, PFZ244551H, PFZ244860H, PFZ244866H, PFZ246561H, PFZ247975H, PFZ248268H, PFZ248657H, PFZ256036H, PFZ256518H, PFZ256948H, PFZ261072H, PFZ262805H, PFZ268220H, PFZ270601H, PFZ270606H, PFZ247259H, PFZ247884H, PFZ258478H, PFZ260676H, PFZ265434H, PFZ265450H, PFZ267876H, PFZ267891H, PFZ240225H, PFZ263311H, PFZ271066H, PFZ241355H, PFZ241360H, PFZ245638H, PFZ245639H, PFZ245640H, PFZ252581H, PFZ262182H, PFZ262183H, PFZ262185H, PFZ262187H, PFZ265574H, PFZ238328H, PFZ242428H, PFZ256451H, PFZ262146H, PFZ264920H, PFZ236021H, PFZ237623H, PFZ239300H, PFZ240083H, PFZ240090H, PFZ240389H, PFZ244701H, PFZ246184H, PFZ247333H, PFZ247892H, PFZ248214H, PFZ248382H, PFZ250820H, PFZ251511H, PFZ251820H, PFZ251881H, PFZ252876H, PFZ252993H, PFZ254858H, PFZ255014H, PFZ255017H, PFZ255023H, PFZ255081H, PFZ255268H, PFZ255483H, PFZ256027H, PFZ256082H, PFZ256271H, PFZ256826H, PFZ257080H, PFZ257543H, PFZ258741H, PFZ258969H, PFZ258970H, PFZ260754H, PFZ261087H, PFZ262771H, PFZ262860H, PFZ262910H, PFZ263833H, PFZ263838H, PFZ263980H, PFZ266888H, PFZ268634H, PFZ269528H, PFZ270417H, PFZ271012H, PFZ273679H, PFZ269434H, PFZ272191H, PFZ272374H, PFZ272395H, PFZ240611H, PFZ244039H, PFZ245378H, PFZ246782H, PFZ252614H, PFZ254658H, PFZ254660H, PFZ257466H, PFZ263508H, PFZ263867H, PFZ264139H, PFZ264494H, PFZ266057H, PFZ267566H, PFZ268766H, PFZ269725H, PFZ273367H, PFZ274828H, PFZ271922H, PFZ274189H, PFZ245249H, PFZ245253H, PFZ245254H, PFZ245256H, PFZ245258H, PFZ245261H, PFZ245262H, PFZ245263H, PFZ245264H, PFZ245265H, PFZ245266H, PFZ245268H, PFZ245273H, PFZ252480H, PFZ252481H, PFZ252484H, PFZ252486H, PFZ256844H, PFZ256845H, PFZ256846H, PFZ256847H, PFZ256848H, PFZ256849H, PFZ256850H, PFZ256853H, PFZ256854H, PFZ256855H, PFZ256856H, PFZ256857H, PFZ256859H, PFZ256862H, PFZ256903H, PFZ256905H, PFZ256906H, PFZ256907H, PFZ256908H, PFZ256909H, PFZ261297H, PFZ261305H, PFZ261308H, PFZ261316H, PFZ261330H, PFZ261351H, PFZ261355H, PFZ261358H, PFZ261361H, PFZ261380H, PFZ261384H, PFZ261390H, PFZ261392H, PFZ261398H, PFZ261408H, PFZ261409H, PFZ261415H, PFZ261434H, PFZ261439H, PFZ261440H, PFZ261886H, PFZ261889H, PFZ261892H, PFZ261893H, PFZ261894H, PFZ261895H, PFZ261897H, PFZ261906H, PFZ261910H, PFZ261911H, PFZ261918H, PFZ261975H, PFZ262092H, PFZ262118H, PFZ262125H, PFZ262126H, PFZ262134H, PFZ262136H, PFZ262174H, PFZ262179H, PFZ239241H, PFZ254253H, PFZ258322H, PFZ268917H, PFZ237630H, PFZ239197H, PFZ239270H, PFZ239271H, PFZ242204H, PFZ243395H, PFZ244895H, PFZ246735H, PFZ249517H, PFZ249546H, PFZ251447H, PFZ252034H, PFZ252129H, PFZ252155H, PFZ252640H, PFZ252650H, PFZ252651H, PFZ257059H, PFZ258638H, PFZ258643H, PFZ258652H, PFZ258653H, PFZ258654H, PFZ259208H, PFZ259943H, PFZ261268H, PFZ262895H, PFZ262897H, PFZ262903H, PFZ262904H, PFZ262911H, PFZ269049H, PFZ272433H, PFZ272571H, PFZ270177H, PFZ272933H, PFZ236755H, PFZ242014H, PFZ242158H, PFZ245149H, PFZ252999H, PFZ254551H, PFZ262535H, PFZ242699H, PFZ242742H, PFZ246428H, PFZ252976H, PFZ260673H, PFZ265364H, PFZ237709H, PFZ244291H, PFZ247489H, PFZ247904H, PFZ248691H, PFZ250205H, PFZ255280H, PFZ256043H, PFZ258564H, PFZ272668H, PFZ239509H, PFZ249622H, PFZ250462H, PFZ258317H, PFZ259868H, PFZ262894H, PFZ263726H, PFZ268876H, PFZ272232H, PFZ274464H, PFZ239520H, PFZ260091H, PFZ262181H, PFZ262322H, PFZ262988H, PFZ236403H, PFZ237408H, PFZ237674H, PFZ242683H, PFZ244052H, PFZ244902H, PFZ254446H, PFZ254557H, PFZ254666H, PFZ247615H, PFZ248703H, PFZ253298H, PFZ253299H, PFZ242492H, PFZ242494H, PFZ242496H, PFZ245659H, PFZ245660H, PFZ245662H, PFZ253274H, PFZ253278H, PFZ253280H, PFZ264749H, PFZ265139H, PFZ268061H, PFZ274377H, PFZ254662H, PFZ238862H, PFZ240461H, PFZ244752H, PFZ244779H, PFZ250337H, PFZ251563H, PFZ253681H, PFZ254505H, PFZ259672H, PFZ260546H, PFZ262688H, PFZ263835H, PFZ263927H, PFZ265349H, PFZ265351H, PFZ273267H, PFZ274455H, PFZ274538H, PFZ245398H, PFZ245399H, PFZ245400H, PFZ245401H, PFZ245403H, PFZ245406H, PFZ245408H, PFZ249893H, PFZ251925H, PFZ253510H, PFZ253514H, PFZ257916H, PFZ257925H, PFZ257928H, PFZ258025H, PFZ258029H, PFZ258034H, PFZ258057H, PFZ258060H, PFZ258061H, PFZ258063H, PFZ262844H, PFZ262846H, PFZ262847H, PFZ262848H, PFZ262849H, PFZ263124H, PFZ263125H, PFZ267677H, PFZ267679H, PFZ267682H, PFZ267683H, PFZ267684H, PFZ268653H, PFZ272129H, PFZ244107H, PFZ244146H, PFZ246884H, PFZ237998H, PFZ240099H, PFZ253663H, PFZ254447H, PFZ254862H, PFZ255765H, PFZ258480H, PFZ259187H, PFZ260308H, PFZ265630H, PFZ266887H, PFZ269060H, PFZ271884H, PFZ273183H, PFZ274053H, PFZ274686H, PFZ240224H, PFZ241351H, PFZ247106H, PFZ256015H, PFZ260385H, PFZ260695H, PFZ260856H, PFZ260969H, PFZ261476H, PFZ262968H, PFZ265868H, PFZ271095H, PFZ274533H, PFZ269893H, PFZ271097H, PFZ238117H, PFZ247567H, PFZ260398H, PFZ243061H, PFZ250146H, PFZ262016H, PFZ268122H, PFZ272106H, PFZ238213H, PFZ238214H, PFZ238410H, PFZ247575H, PFZ254100H, PFZ254106H, PFZ254406H, PFZ254432H, PFZ260069H, PFZ262385H, PFZ262386H, PFZ262387H, PFZ263852H, PFZ263986H, PFZ269264H, PFZ270771H, PFZ271580H, PFZ271582H, PFZ271583H, PFZ271584H, PFZ272528H, PFZ273335H, PFZ274336H, PFZ236841H, PFZ237284H, PFZ237797H, PFZ238686H, PFZ239369H, PFZ239875H, PFZ240371H, PFZ241644H, PFZ242627H, PFZ242798H, PFZ242969H, PFZ243074H, PFZ245382H, PFZ245448H, PFZ246808H, PFZ247411H, PFZ247695H, PFZ248979H, PFZ249426H, PFZ249457H, PFZ249958H, PFZ250113H, PFZ250580H, PFZ250775H, PFZ252802H, PFZ252806H, PFZ252810H, PFZ252813H, PFZ252815H, PFZ252816H, PFZ254005H, PFZ254807H, PFZ254892H, PFZ255882H, PFZ257923H, PFZ257942H, PFZ259311H, PFZ259723H, PFZ260451H, PFZ260608H, PFZ260837H, PFZ261417H, PFZ262714H, PFZ262967H, PFZ263507H, PFZ264565H, PFZ266956H, PFZ268527H, PFZ268814H, PFZ268815H, PFZ268816H, PFZ269424H, PFZ272621H, PFZ272838H, PFZ273547H, PFZ274122H, PFZ274202H, PFZ274418H, PFZ238246H, PFZ240136H, PFZ242611H, PFZ242672H, PFZ247248H, PFZ247281H, PFZ248254H, PFZ248499H, PFZ248738H, PFZ248785H, PFZ248982H, PFZ249041H, PFZ250273H, PFZ250749H, PFZ251633H, PFZ253771H, PFZ253904H, PFZ253988H, PFZ254248H, PFZ258189H, PFZ258298H, PFZ259068H, PFZ260232H, PFZ260356H, PFZ260593H, PFZ261279H, PFZ268547H, PFZ270302H, PFZ270731H, PFZ272775H, PFZ240253H, PFZ256483H, PFZ256536H, PFZ271112H, PFZ238794H, PFZ236108H, PFZ236163H, PFZ236442H, PFZ238239H, PFZ239151H, PFZ239749H, PFZ240266H, PFZ240393H, PFZ241689H, PFZ243066H, PFZ243276H, PFZ243417H, PFZ243526H, PFZ243563H, PFZ243604H, PFZ243723H, PFZ244151H, PFZ244315H, PFZ245044H, PFZ245240H, PFZ245310H, PFZ245446H, PFZ246505H, PFZ246548H, PFZ246768H, PFZ247191H, PFZ247442H, PFZ248021H, PFZ248200H, PFZ249778H, PFZ249859H, PFZ250022H, PFZ250808H, PFZ251121H, PFZ252659H, PFZ252885H, PFZ253682H, PFZ254144H, PFZ254488H, PFZ254961H, PFZ255165H, PFZ255660H, PFZ257723H, PFZ260129H, PFZ261057H, PFZ261168H, PFZ261359H, PFZ261745H, PFZ262969H, PFZ263501H, PFZ263716H, PFZ264004H, PFZ264066H, PFZ264228H, PFZ265206H, PFZ265697H, PFZ265806H, PFZ265807H, PFZ266018H, PFZ266196H, PFZ268837H, PFZ270287H, PFZ270835H, PFZ271210H, PFZ271967H, PFZ272626H, PFZ272949H, PFZ273041H, PFZ274440H, PFZ274639H, PFZ274738H, PFZ240314H, PFZ241712H, PFZ243032H, PFZ243141H, PFZ243627H, PFZ250793H, PFZ253609H, PFZ254074H, PFZ259012H, PFZ260866H, PFZ261784H, PFZ261878H, PFZ262773H, PFZ272576H, PFZ274326H, PFZ240865H, PFZ242375H, PFZ244009H, PFZ246337H, PFZ246996H, PFZ253506H, PFZ253507H, PFZ253509H, PFZ253516H, PFZ253518H, PFZ253520H, PFZ253523H, PFZ253525H, PFZ253526H, PFZ255928H, PFZ262246H, PFZ262251H, PFZ262347H, PFZ262379H, PFZ262399H, PFZ267383H, PFZ267384H, PFZ267386H, PFZ267387H, PFZ267388H, PFZ267391H, PFZ270646H, PFZ270649H, PFZ270650H, PFZ270651H, PFZ270652H, PFZ270653H, PFZ270655H, PFZ270656H, PFZ270659H, PFZ270660H, PFZ272583H, PFZ274782H, PFZ255523H, PFZ259702H, PFZ261921H, PFZ263981H, PFZ268385H, PFZ268386H, PFZ268470H, PFZ273163H, PFZ274051H, PFZ250830H, PFZ259125H, PFZ266041H, PFZ237950H, PFZ237951H, PFZ237954H, PFZ239694H, PFZ241653H, PFZ241654H, PFZ242452H, PFZ244507H, PFZ246111H, PFZ246114H, PFZ246126H, PFZ249464H, PFZ249921H, PFZ250241H, PFZ250747H, PFZ251263H, PFZ252443H, PFZ252456H, PFZ252673H, PFZ252674H, PFZ252675H, PFZ252677H, PFZ252679H, PFZ252680H, PFZ253055H, PFZ256819H, PFZ256823H, PFZ256824H, PFZ260044H, PFZ260354H, PFZ260460H, PFZ260807H, PFZ261032H, PFZ261991H, PFZ262005H, PFZ262009H, PFZ262011H, PFZ262078H, PFZ266842H, PFZ269031H, PFZ269447H, PFZ270654H, PFZ270982H, PFZ273966H, PFZ236550H, PFZ236551H, PFZ244279H, PFZ258380H, PFZ260636H, PFZ246144H, PFZ246547H, PFZ247306H, PFZ250485H, PFZ255157H, PFZ259312H, PFZ259578H, PFZ264367H, PFZ268849H, PFZ240153H, PFZ258132H, PFZ261668H, PFZ236061H, PFZ236413H, PFZ237343H, PFZ237435H, PFZ238194H, PFZ238503H, PFZ238547H, PFZ238908H, PFZ240356H, PFZ241084H, PFZ243531H, PFZ243556H, PFZ243954H, PFZ245027H, PFZ245686H, PFZ246895H, PFZ249680H, PFZ249797H, PFZ250058H, PFZ250610H, PFZ251202H, PFZ251721H, PFZ252145H, PFZ253649H, PFZ254849H, PFZ254878H, PFZ254925H, PFZ254951H, PFZ255501H, PFZ255522H, PFZ255880H, PFZ256179H, PFZ257651H, PFZ258955H, PFZ258962H, PFZ261313H, PFZ263297H, PFZ265074H, PFZ265906H, PFZ266704H, PFZ268290H, PFZ268805H, PFZ268910H, PFZ269054H, PFZ269422H, PFZ269496H, PFZ269515H, PFZ270431H, PFZ270474H, PFZ270721H, PFZ271881H, PFZ272467H, PFZ273670H, PFZ273787H, PFZ274004H, PFZ274934H, PFZ243511H, PFZ247519H, PFZ251033H, PFZ269123H, PFZ247177H, PFZ249635H, PFZ255143H, PFZ259837H, PFZ268295H, PFZ270822H, PFZ273673H, PFZ237542H, PFZ239828H, PFZ243950H, PFZ245418H, PFZ251151H, PFZ264517H, PFZ265005H, PFZ266862H, PFZ267162H, PFZ271113H, PFZ273057H, PFZ273714H, PFZ236423H, PFZ238806H, PFZ242829H, PFZ248717H, PFZ249949H, PFZ263836H, PFZ264402H, PFZ269575H, PFZ273588H, PFZ240085H, PFZ251895H, PFZ251992H, PFZ256029H, PFZ260532H, PFZ262464H, PFZ262465H, PFZ270042H, PFZ268969H, PFZ268981H, PFZ270487H, PFZ271873H, PFZ236679H, PFZ237180H, PFZ237715H, PFZ237973H, PFZ238688H, PFZ239007H, PFZ239420H, PFZ239838H, PFZ241227H, PFZ242988H, PFZ243065H, PFZ243473H, PFZ244010H, PFZ244366H, PFZ244441H, PFZ246253H, PFZ246928H, PFZ247299H, PFZ247868H, PFZ248260H, PFZ248569H, PFZ249035H, PFZ249746H, PFZ249938H, PFZ249939H, PFZ251369H, PFZ251412H, PFZ253443H, PFZ253665H, PFZ254539H, PFZ254731H, PFZ254799H, PFZ254819H, PFZ255235H, PFZ255813H, PFZ255938H, PFZ256174H, PFZ256569H, PFZ257494H, PFZ258577H, PFZ258879H, PFZ259176H, PFZ259668H, PFZ259719H, PFZ259839H, PFZ260211H, PFZ260766H, PFZ261217H, PFZ261323H, PFZ261416H, PFZ262500H, PFZ263418H, PFZ263558H, PFZ263608H, PFZ263610H, PFZ264105H, PFZ264158H, PFZ264490H, PFZ265323H, PFZ266835H, PFZ268247H, PFZ268351H, PFZ269300H, PFZ269744H, PFZ269795H, PFZ269999H, PFZ270528H, PFZ270896H, PFZ271171H, PFZ272656H, PFZ273019H, PFZ273454H, PFZ274737H, PFZ236268H, PFZ236273H, PFZ237203H, PFZ238104H, PFZ238483H, PFZ238873H, PFZ239179H, PFZ239207H, PFZ239822H, PFZ242490H, PFZ242574H, PFZ243158H, PFZ243702H, PFZ243888H, PFZ244297H, PFZ244861H, PFZ246267H, PFZ246791H, PFZ247056H, PFZ247141H, PFZ247455H, PFZ247637H, PFZ249452H, PFZ250638H, PFZ250640H, PFZ250653H, PFZ250784H, PFZ251234H, PFZ251689H, PFZ251716H, PFZ251794H, PFZ252132H, PFZ253850H, PFZ253884H, PFZ253981H, PFZ254310H, PFZ254480H, PFZ254623H, PFZ254803H, PFZ255592H, PFZ255965H, PFZ259344H, PFZ260995H, PFZ261375H, PFZ261382H, PFZ261468H, PFZ263013H, PFZ263583H, PFZ263602H, PFZ264030H, PFZ264458H, PFZ264495H, PFZ265326H, PFZ265384H, PFZ265575H, PFZ266958H, PFZ268042H, PFZ268667H, PFZ268846H, PFZ269466H, PFZ269739H, PFZ270364H, PFZ270490H, PFZ272843H, PFZ273084H, PFZ273775H, PFZ273916H, PFZ274116H, PFZ274523H, PFZ274902H, PFZ236405H, PFZ236547H, PFZ236705H, PFZ245488H, PFZ245489H, PFZ245490H, PFZ245491H, PFZ245495H, PFZ245496H, PFZ245500H, PFZ245504H, PFZ245761H, PFZ251223H, PFZ251490H, PFZ251523H, PFZ251546H, PFZ252066H, PFZ252634H, PFZ253362H, PFZ253365H, PFZ253368H, PFZ253370H, PFZ253376H, PFZ253377H, PFZ253382H, PFZ253383H, PFZ253386H, PFZ253390H, PFZ253391H, PFZ262964H, PFZ262966H, PFZ262973H, PFZ263020H, PFZ263025H, PFZ263027H, PFZ263030H, PFZ263031H, PFZ263032H, PFZ263038H, PFZ263041H, PFZ263042H, PFZ263047H, PFZ245671H, PFZ246982H, PFZ247174H, PFZ247175H, PFZ255264H, PFZ256511H, PFZ260417H, PFZ238648H, PFZ238791H, PFZ249423H, PFZ249722H, PFZ260911H, PFZ236265H, PFZ237803H, PFZ241213H, PFZ241215H, PFZ241216H, PFZ241217H, PFZ241219H, PFZ241220H, PFZ241221H, PFZ241223H, PFZ243901H, PFZ244550H, PFZ244581H, PFZ245563H, PFZ245564H, PFZ245566H, PFZ245567H, PFZ245568H, PFZ245569H, PFZ245572H, PFZ245573H, PFZ245574H, PFZ245577H, PFZ245578H, PFZ245579H, PFZ245580H, PFZ245581H, PFZ245583H, PFZ245584H, PFZ249384H, PFZ249977H, PFZ253135H, PFZ253136H, PFZ253139H, PFZ253144H, PFZ253145H, PFZ253146H, PFZ254613H, PFZ255791H, PFZ257649H, PFZ258146H, PFZ258716H, PFZ258717H, PFZ258718H, PFZ258720H, PFZ258806H, PFZ258809H, PFZ258810H, PFZ259696H, PFZ259697H, PFZ261043H, PFZ261342H, PFZ261343H, PFZ261399H, PFZ261419H, PFZ261750H, PFZ262992H, PFZ262996H, PFZ262997H, PFZ262998H, PFZ263689H, PFZ263690H, PFZ263691H, PFZ265011H, PFZ265012H, PFZ265013H, PFZ266557H, PFZ268821H, PFZ268822H, PFZ268823H, PFZ271018H, PFZ271019H, PFZ271021H, PFZ271022H, PFZ272751H, PFZ272752H, PFZ272820H, PFZ272824H, PFZ273051H, PFZ273052H, PFZ273224H, PFZ273225H, PFZ273562H, PFZ273572H, PFZ274651H, PFZ274719H, PFZ274720H, PFZ274722H, PFZ242844H, PFZ243233H, PFZ243575H, PFZ243654H, PFZ243789H, PFZ244370H, PFZ244620H, PFZ246799H, PFZ248667H, PFZ250286H, PFZ250718H, PFZ251218H, PFZ255088H, PFZ257323H, PFZ258778H, PFZ259113H, PFZ259912H, PFZ260624H, PFZ260915H, PFZ261242H, PFZ261429H, PFZ261656H, PFZ261657H, PFZ261658H, PFZ264419H, PFZ267590H, PFZ267591H, PFZ267592H, PFZ267593H, PFZ267594H, PFZ267596H, PFZ267598H, PFZ271005H, PFZ271006H, PFZ271007H, PFZ273503H, PFZ274974H, PFZ236130H, PFZ238921H, PFZ239147H, PFZ243073H, PFZ243930H, PFZ244145H, PFZ246641H, PFZ249960H, PFZ250180H, PFZ251232H, PFZ251890H, PFZ251954H, PFZ252214H, PFZ253886H, PFZ254547H, PFZ255434H, PFZ259006H, PFZ259071H, PFZ259291H, PFZ259533H, PFZ259707H, PFZ260135H, PFZ260296H, PFZ260505H, PFZ260557H, PFZ261203H, PFZ262657H, PFZ263213H, PFZ263289H, PFZ263331H, PFZ263595H, PFZ263821H, PFZ264101H, PFZ264185H, PFZ264589H, PFZ265041H, PFZ265618H, PFZ265703H, PFZ265792H, PFZ266185H, PFZ268618H, PFZ269397H, PFZ269726H, PFZ270264H, PFZ271971H, PFZ271987H, PFZ272898H, PFZ273081H, PFZ273122H, PFZ273626H, PFZ273790H, PFZ274833H, PFZ246901H, PFZ247962H, PFZ248407H, PFZ250244H, PFZ250571H, PFZ255343H, PFZ259189H, PFZ259219H, PFZ259232H, PFZ259235H, PFZ262493H, PFZ262503H, PFZ262508H, PFZ267464H, PFZ236830H, PFZ236832H, PFZ236833H, PFZ236834H, PFZ237477H, PFZ237487H, PFZ238280H, PFZ238284H, PFZ238289H, PFZ238291H, PFZ239035H, PFZ239036H, PFZ239497H, PFZ239498H, PFZ239701H, PFZ239776H, PFZ240432H, PFZ240433H, PFZ240434H, PFZ241179H, PFZ241180H, PFZ241181H, PFZ241182H, PFZ241184H, PFZ241185H, PFZ241186H, PFZ241188H, PFZ241190H, PFZ241193H, PFZ241195H, PFZ241197H, PFZ241198H, PFZ241202H, PFZ241204H, PFZ246307H, PFZ246309H, PFZ246480H, PFZ246481H, PFZ246689H, PFZ246691H, PFZ247221H, PFZ247222H, PFZ247223H, PFZ247859H, PFZ255438H, PFZ255440H, PFZ256550H, PFZ257298H, PFZ257301H, PFZ257302H, PFZ257304H, PFZ257305H, PFZ257316H, PFZ257322H, PFZ257325H, PFZ257329H, PFZ257330H, PFZ257331H, PFZ257334H, PFZ257335H, PFZ257336H, PFZ257337H, PFZ257339H, PFZ257826H, PFZ259274H, PFZ261385H, PFZ261386H, PFZ261698H, PFZ261699H, PFZ261700H, PFZ261701H, PFZ261702H, PFZ261703H, PFZ261704H, PFZ261705H, PFZ261706H, PFZ261707H, PFZ261708H, PFZ261710H, PFZ261714H, PFZ261715H, PFZ261719H, PFZ261722H, PFZ261723H, PFZ261724H, PFZ261727H, PFZ265762H, PFZ265764H, PFZ265864H, PFZ266419H, PFZ266423H, PFZ266428H, PFZ266442H, PFZ266448H, PFZ266449H, PFZ266451H, PFZ266454H, PFZ266457H, PFZ266463H, PFZ266464H, PFZ266466H, PFZ266467H, PFZ266468H, PFZ266474H, PFZ266477H, PFZ266481H, PFZ266483H, PFZ266484H, PFZ272810H, PFZ274724H, PFZ237517H, PFZ237520H, PFZ237521H, PFZ237535H, PFZ237550H, PFZ237551H, PFZ237556H, PFZ237557H, PFZ237569H, PFZ237578H, PFZ237580H, PFZ237581H, PFZ237588H, PFZ237590H, PFZ237595H, PFZ237597H, PFZ239958H, PFZ239969H, PFZ241192H, PFZ241196H, PFZ241665H, PFZ241667H, PFZ241670H, PFZ241672H, PFZ241675H, PFZ241676H, PFZ241681H, PFZ241705H, PFZ241715H, PFZ241731H, PFZ241733H, PFZ241734H, PFZ241736H, PFZ241739H, PFZ245080H, PFZ245083H, PFZ245531H, PFZ245836H, PFZ245837H, PFZ245841H, PFZ245847H, PFZ245849H, PFZ245869H, PFZ245877H, PFZ245882H, PFZ245884H, PFZ245886H, PFZ245889H, PFZ245893H, PFZ245895H, PFZ245902H, PFZ245905H, PFZ245906H, PFZ245909H, PFZ245910H, PFZ245911H, PFZ245912H, PFZ246106H, PFZ246549H, PFZ246680H, PFZ246681H, PFZ253241H, PFZ253242H, PFZ253243H, PFZ253245H, PFZ253249H, PFZ253250H, PFZ253252H, PFZ253255H, PFZ253257H, PFZ253258H, PFZ253260H, PFZ253266H, PFZ253267H, PFZ253269H, PFZ253271H, PFZ253272H, PFZ254967H, PFZ255974H, PFZ256333H, PFZ256525H, PFZ256541H, PFZ256611H, PFZ256749H, PFZ257482H, PFZ258035H, PFZ258036H, PFZ258037H, PFZ258038H, PFZ258042H, PFZ258043H, PFZ262228H, PFZ262232H, PFZ262234H, PFZ262235H, PFZ262236H, PFZ262237H, PFZ262240H, PFZ262241H, PFZ262242H, PFZ262244H, PFZ262248H, PFZ262249H, PFZ262252H, PFZ265019H, PFZ265924H, PFZ265959H, PFZ265962H, PFZ265966H, PFZ266420H, PFZ267242H, PFZ267246H, PFZ267250H, PFZ267251H, PFZ267252H, PFZ267253H, PFZ267256H, PFZ267259H, PFZ267260H, PFZ267261H, PFZ267263H, PFZ267266H, PFZ267267H, PFZ267268H, PFZ267269H, PFZ267270H, PFZ267271H, PFZ267272H, PFZ267273H, PFZ267275H, PFZ267277H, PFZ267278H, PFZ267280H, PFZ267281H, PFZ267282H, PFZ267283H, PFZ267298H, PFZ271610H, PFZ271611H, PFZ271612H, PFZ271615H, PFZ271616H, PFZ271617H, PFZ271618H, PFZ271619H, PFZ271620H, PFZ271622H, PFZ271623H, PFZ271624H, PFZ271627H, PFZ271630H, PFZ271631H, PFZ271632H, PFZ271633H, PFZ271634H, PFZ263197H, PFZ238998H, PFZ246118H, PFZ247057H, PFZ252590H, PFZ258154H, PFZ263953H, PFZ265669H, PFZ270644H, PFZ251404H, PFZ250645H, PFZ263144H, PFZ267810H, PFZ240194H, PFZ266694H, PFZ266737H, PFZ244311H, PFZ247667H, PFZ249076H, PFZ262886H, PFZ238399H, PFZ242782H, PFZ242846H, PFZ245692H, PFZ246181H, PFZ247110H, PFZ256111H, PFZ262308H, PFZ262309H, PFZ262356H, PFZ262359H, PFZ267506H, PFZ267567H, PFZ267568H, PFZ267572H, PFZ267573H, PFZ270363H, PFZ271198H, PFZ271200H, PFZ271201H, PFZ271203H, PFZ271204H, PFZ271205H, PFZ271206H, PFZ271207H, PFZ272001H, PFZ272033H, PFZ239959H, PFZ240013H, PFZ246658H, PFZ249301H, PFZ256020H, PFZ264797H, PFZ273424H, PFZ238101H, PFZ238179H, PFZ240418H, PFZ242849H, PFZ243264H, PFZ244482H, PFZ246827H, PFZ249193H, PFZ250088H, PFZ250754H, PFZ251188H, PFZ251899H, PFZ252320H, PFZ253392H, PFZ254685H, PFZ259595H, PFZ270272H, PFZ236322H, PFZ238058H, PFZ239799H, PFZ250678H, PFZ253885H, PFZ255561H, PFZ260391H, PFZ260523H, PFZ270260H, PFZ273729H, PFZ274381H, PFZ245117H, PFZ237093H, PFZ237106H, PFZ237110H, PFZ237111H, PFZ238136H, PFZ238199H, PFZ238202H, PFZ238423H, PFZ238424H, PFZ245684H, PFZ245733H, PFZ245736H, PFZ245739H, PFZ245740H, PFZ246025H, PFZ246033H, PFZ246278H, PFZ250042H, PFZ252327H, PFZ252328H, PFZ252330H, PFZ252332H, PFZ252335H, PFZ252336H, PFZ258136H, PFZ258137H, PFZ258140H, PFZ258141H, PFZ258145H, PFZ258148H, PFZ258150H, PFZ258152H, PFZ261695H, PFZ261697H, PFZ265907H, PFZ266432H, PFZ266433H, PFZ266437H, PFZ267758H, PFZ267759H, PFZ270572H, PFZ273244H, PFZ274168H, PFZ237680H, PFZ237683H, PFZ242802H, PFZ247712H, PFZ251550H, PFZ254312H, PFZ254767H, PFZ258709H, PFZ260372H, PFZ260379H, PFZ263563H, PFZ267552H, PFZ268100H, PFZ272819H, PFZ249007H, PFZ249009H, PFZ253322H, PFZ260029H, PFZ265453H, PFZ271904H, PFZ272893H, PFZ272894H, PFZ252553H, PFZ259569H, PFZ263132H, PFZ270062H, PFZ274476H, PFZ274573H, PFZ274668H, PFZ271882H, PFZ243711H, PFZ258570H, PFZ259793H, PFZ260340H, PFZ266919H, PFZ268609H, PFZ270113H, PFZ270541H, PFZ274693H, PFZ237771H, PFZ241160H, PFZ241161H, PFZ241164H, PFZ241166H, PFZ244658H, PFZ245462H, PFZ245463H, PFZ245464H, PFZ245468H, PFZ246825H, PFZ247166H, PFZ247843H, PFZ249170H, PFZ250566H, PFZ250752H, PFZ251976H, PFZ252837H, PFZ252838H, PFZ252913H, PFZ253338H, PFZ253340H, PFZ253344H, PFZ255734H, PFZ256533H, PFZ256554H, PFZ258124H, PFZ258125H, PFZ258128H, PFZ258133H, PFZ258134H, PFZ260746H, PFZ260760H, PFZ261231H, PFZ261374H, PFZ262195H, PFZ262197H, PFZ262200H, PFZ262201H, PFZ262204H, PFZ262509H, PFZ262542H, PFZ263509H, PFZ264753H, PFZ264754H, PFZ265103H, PFZ265119H, PFZ265551H, PFZ265587H, PFZ265588H, PFZ266062H, PFZ267330H, PFZ267331H, PFZ267332H, PFZ267334H, PFZ267335H, PFZ267336H, PFZ268871H, PFZ270832H, PFZ272358H, PFZ272855H, PFZ272856H, PFZ272857H, PFZ272858H, PFZ272859H, PFZ272860H, PFZ272861H, PFZ272863H, PFZ272864H, PFZ274030H, PFZ274034H, PFZ274598H, PFZ274670H, PFZ236215H, PFZ240603H, PFZ240604H, PFZ240605H, PFZ240613H, PFZ240616H, PFZ240619H, PFZ240624H, PFZ240626H, PFZ240627H, PFZ240628H, PFZ240630H, PFZ245281H, PFZ245283H, PFZ245284H, PFZ245288H, PFZ245290H, PFZ245292H, PFZ245294H, PFZ245300H, PFZ245302H, PFZ245303H, PFZ256911H, PFZ256913H, PFZ256915H, PFZ256916H, PFZ256919H, PFZ256920H, PFZ256921H, PFZ256925H, PFZ256926H, PFZ264193H, PFZ266340H, PFZ266341H, PFZ266342H, PFZ266344H, PFZ266345H, PFZ266347H, PFZ270171H, PFZ237714H, PFZ238418H, PFZ238421H, PFZ238890H, PFZ239571H, PFZ240614H, PFZ241975H, PFZ241977H, PFZ241982H, PFZ241983H, PFZ241984H, PFZ242012H, PFZ242016H, PFZ242021H, PFZ242026H, PFZ242028H, PFZ242031H, PFZ242032H, PFZ242034H, PFZ242036H, PFZ242037H, PFZ255948H, PFZ256112H, PFZ256116H, PFZ256119H, PFZ256120H, PFZ256121H, PFZ256130H, PFZ256132H, PFZ256140H, PFZ256149H, PFZ256153H, PFZ256154H, PFZ256156H, PFZ256157H, PFZ256361H, PFZ264759H, PFZ264945H, PFZ264953H, PFZ265014H, PFZ265394H, PFZ265623H, PFZ265945H, PFZ265984H, PFZ266159H, PFZ266989H, PFZ266991H, PFZ266992H, PFZ266993H, PFZ266995H, PFZ266996H, PFZ266998H, PFZ266999H, PFZ267000H, PFZ267001H, PFZ267002H, PFZ267003H, PFZ267004H, PFZ267005H, PFZ267009H, PFZ267012H, PFZ267013H, PFZ267016H, PFZ267017H, PFZ267018H, PFZ267019H, PFZ267020H, PFZ267022H, PFZ267023H, PFZ267024H, PFZ267025H, PFZ267026H, PFZ268906H, PFZ269266H, PFZ238021H, PFZ239396H, PFZ243886H, PFZ248326H, PFZ270961H, PFZ241691H, PFZ241692H, PFZ241698H, PFZ242440H, PFZ252842H, PFZ252909H, PFZ252911H, PFZ258830H, PFZ258841H, PFZ258842H, PFZ259879H, PFZ259880H, PFZ259881H, PFZ259882H, PFZ270766H, PFZ271275H, PFZ271305H, PFZ242896H, PFZ246435H, PFZ246817H, PFZ246882H, PFZ246904H, PFZ247631H, PFZ248226H, PFZ250546H, PFZ250925H, PFZ251074H, PFZ251114H, PFZ252642H, PFZ253057H, PFZ253735H, PFZ254571H, PFZ255684H, PFZ256331H, PFZ259535H, PFZ260116H, PFZ260800H, PFZ260927H, PFZ260972H, PFZ261761H, PFZ268186H, PFZ269856H, PFZ270643H, PFZ270804H, PFZ270940H, PFZ272832H, PFZ273036H, PFZ273277H, PFZ273661H, PFZ274467H, PFZ236205H, PFZ239493H, PFZ239494H, PFZ243516H, PFZ246078H, PFZ247105H, PFZ248034H, PFZ255469H, PFZ270285H, PFZ271117H, PFZ238170H, PFZ243519H, PFZ259613H, PFZ260556H, PFZ264662H, PFZ268584H, PFZ268586H, PFZ268702H, PFZ270756H, PFZ273807H, PFZ274918H, PFZ238385H, PFZ239136H, PFZ240301H, PFZ242914H, PFZ243610H, PFZ244936H, PFZ245168H, PFZ246441H, PFZ246648H, PFZ246961H, PFZ247152H, PFZ247634H, PFZ248102H, PFZ248564H, PFZ248881H, PFZ249180H, PFZ249593H, PFZ249888H, PFZ250723H, PFZ250753H, PFZ250942H, PFZ250964H, PF
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1712-2023
Related Recalls
ICD-DR DDBB2D1 EVERA XT IS1/DF1 INTL, Model Number DDBB2D1; Implantable Cardioverter Defibrillators
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.
ICD-DR DDMC3D1 EVERA MRI S OUS/US DF1, Model Number DDMC3D1; Implantable Cardioverter Defibrillators
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.
ICD DDPA2D1G COBALT XT DR MRI DF1 GOLD, Model Number DDPA2D1G; Implantable Cardioverter Defibrillators
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.