Various eye packs, cataract packs

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Tray Number, Sterilization Lot, Lot Number, Bag Serial Number 1) BLSE37K, 2106292, 820211, 15017475-004 2) MECA21P, 2108022, 786211, 44234851-005 3) CMEY17AD, 2108301, 760211, 15101886-004 4) CXEY80X, 2109083, 749211, 15109749-004
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
American Contract Systems, Inc.
Reason for Recall:
Product was exposed to multiple sterilization cycles without validation for multiple exposures.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
ongoing

Product Information

Full Description:

Various eye packs, cataract packs

Product Codes/Lot Numbers:

Tray Number, Sterilization Lot, Lot Number, Bag Serial Number 1) BLSE37K, 2106292, 820211, 15017475-004 2) MECA21P, 2108022, 786211, 44234851-005 3) CMEY17AD, 2108301, 760211, 15101886-004 4) CXEY80X, 2109083, 749211, 15109749-004

Distribution:

Distributed in: AZ, FL, IA, IL, MA, MN, MO, OH, NE, PA, RI, SD, TX

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-1705-2022

Related Recalls

Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.

Sep 4, 2025 Other Medical Devices Nationwide View Details →

Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.

Sep 4, 2025 Infusion Pumps Nationwide View Details →

Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.

Sep 4, 2025 Other Medical Devices Nationwide View Details →