BARD¿ BrachySource¿ Iodine125 radioactive seeds consist of a welded titanium capsule containing the I125 absorbed onto a nickel / copper-coated, gold-cored aluminum wire (reference Figure 1 below). The I125 has a half-life of 59.6 days and decays by electron capture with the emission of characteristic photons and Auger electrons. The principal photon emissions are 27.4 and 31 keV x-rays and 35.5 keV gamma. The titanium wall of the BrachySource¿ Seed Implants absorbs the electrons.
Class I - DangerousWhat Should You Do?
- Check if you have this product: Product Code:1251CSR, BrachySource¿ Seed Implants in Re-Usable Mick¿ Applicators (60 seeds at 0.28mCi) with Lot BBCP0079 Product Code: 1251CSRA, BrachySource¿ Seed Implants in Shielded Mick¿ Applicators (10 seeds at 0.28mCi) with Lot BBCQ0026 Product Code: PS1251LS, Loose BrachySource¿ Seed Implants (with Lot BBCQ0050) Spacers (with Lot 23498276) 100 Seeds at 0.36 mCi.
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Bard Brachytherapy, Inc.
- Reason for Recall:
- Bard Medical Division is recalling the BARD¿ BrachySource¿ Iodine125 Radioactive Seeds in Mick¿ Applicators and BARD¿ BrachySource¿ Iodine125 Radioactive Seeds Loose in Vial with Sterile Spacers. Because the incorrect dosage level radioactive seeds were mixed up and sent to two hospitals. One Hospital ordered BrachySource¿ Iodine125 0.36 mCi radioactive seeds but instead received BrachySource¿ Iodine125 0.28 mCi radioactive seeds and vice versa for the other hospital affected.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
BARD¿ BrachySource¿ Iodine125 radioactive seeds consist of a welded titanium capsule containing the I125 absorbed onto a nickel / copper-coated, gold-cored aluminum wire (reference Figure 1 below). The I125 has a half-life of 59.6 days and decays by electron capture with the emission of characteristic photons and Auger electrons. The principal photon emissions are 27.4 and 31 keV x-rays and 35.5 keV gamma. The titanium wall of the BrachySource¿ Seed Implants absorbs the electrons.
Product Codes/Lot Numbers:
Product Code:1251CSR, BrachySource¿ Seed Implants in Re-Usable Mick¿ Applicators (60 seeds at 0.28mCi) with Lot BBCP0079 Product Code: 1251CSRA, BrachySource¿ Seed Implants in Shielded Mick¿ Applicators (10 seeds at 0.28mCi) with Lot BBCQ0026 Product Code: PS1251LS, Loose BrachySource¿ Seed Implants (with Lot BBCQ0050) Spacers (with Lot 23498276) 100 Seeds at 0.36 mCi.
Distribution:
Distributed in: US
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1702-2018