STA - Deficient VIII, Product Code: 00725. STA - Deficient VIII is an immuno-depleted human plasma intended for use in tests for the determination of factor VIII activity in plasma on STA-R and STA Compact analyzer platforms as follows: (A) STA-R Max, Product Codes: 58989, 58990 (B) STA-R Evolution, Product Codes: 58986, 58987, 58988 (C) STA Compact Max, Product Codes: 59026, 59027 (D) STA Compact, Product Codes: 58978, 58979
Class I - DangerousWhat Should You Do?
- Check if you have this product: All lots This cleaning sequence has been implemented in the following available software versions. Customers not possessing these software levels will be upgraded by Stago Field Support personnel: STA-R¿ Max (all serial numbers): Software version 4.05.04 or higher, and Methodology Update (MU) to enhance database version MS215. STA-R¿ Evolution (serial numbers higher than 8012756): Software version 3.04.09 or higher STA¿ Compact Max (all serial numbers): Software version 110.05 or higher.
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Diagnostica Stago, Inc.
- Reason for Recall:
- A risk of sample-to-sample cross-contamination of Emicizumab for factor VIII assays on Stago instruments, clotting times will be shortened that may cause factor VIII levels reported to be falsely increased
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
STA - Deficient VIII, Product Code: 00725. STA - Deficient VIII is an immuno-depleted human plasma intended for use in tests for the determination of factor VIII activity in plasma on STA-R and STA Compact analyzer platforms as follows: (A) STA-R Max, Product Codes: 58989, 58990 (B) STA-R Evolution, Product Codes: 58986, 58987, 58988 (C) STA Compact Max, Product Codes: 59026, 59027 (D) STA Compact, Product Codes: 58978, 58979
Product Codes/Lot Numbers:
All lots This cleaning sequence has been implemented in the following available software versions. Customers not possessing these software levels will be upgraded by Stago Field Support personnel: STA-R¿ Max (all serial numbers): Software version 4.05.04 or higher, and Methodology Update (MU) to enhance database version MS215. STA-R¿ Evolution (serial numbers higher than 8012756): Software version 3.04.09 or higher STA¿ Compact Max (all serial numbers): Software version 110.05 or higher.
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1699-2020
Related Recalls
Product: STA Liatest D-Di; REF: 00515;
Diagnostica Stago
A potential risk of underestimation of D-Dimer (D-Di) levels with the two listed lots.
Potential for contamination of STA-LIQUID ANTI-Xa by Stago Fibrinogen reagents in UFH and LMWH applications.
Potential for contamination of STA-LIQUID ANTI-Xa by Stago Fibrinogen reagents in UFH and LMWH applications.