Neusoft NeuViz 64 Multi-slice CT Scanner System
Class I - DangerousWhat Should You Do?
- Check if you have this product: NeuViz 16/64 Series
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Neusoft Medical System Co
- Reason for Recall:
- The A-plane compensator is made of Teflon which will become fragile due to X-ray exposure passing through it. The thinnest part of the compensator is subject to radiation degradation and as a result has been found to fail prematurely. Failure is indicated by pin or ring artifact in the image
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Neusoft NeuViz 64 Multi-slice CT Scanner System
Product Codes/Lot Numbers:
NeuViz 16/64 Series
Distribution:
Distributed in: US, IL, MA, TX, SC
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1693-2015