JResultNet drivers using ASTM coding language. Calculator/data processing module, for clinical use - Stand-alone software package that provides connectivity between in-vitro diagnostic instruments and laboratory information's systems, aiding the flow of laboratory information, including test orders and results, specimen information, and patient information
Class I - DangerousWhat Should You Do?
- Check if you have this product: All drive versions were affected.
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Data Innovations, LLC
- Reason for Recall:
- Software issue resulting in the results of one patient potentially being associated with another.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
JResultNet drivers using ASTM coding language. Calculator/data processing module, for clinical use - Stand-alone software package that provides connectivity between in-vitro diagnostic instruments and laboratory information's systems, aiding the flow of laboratory information, including test orders and results, specimen information, and patient information
Product Codes/Lot Numbers:
All drive versions were affected.
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1680-2017
Related Recalls
Quality control results are not processed for Abbott Diagnostics Alinity ci series (abdicqii) instrument. This has been determined to be a driver malfunction. When an instrument transmits results from a quality control (QC) sample ID ("Q^Control" in SPM segment), and the option Hold Results Until Complete is enabled, the results do not parse into IM. The communication Trace shows the sample being transmitted, but no information is logged in the Specimen Event Log (SEL). There is a potential for harm if QC results are not processed, downstream auto verification rules may misfire based on missing QC results leading to a delay in, or incorrect treatment. QC results will not be available for evaluation by the lab. There have been no reported instances of patient harm from this malfunction.
Potential issues with the ProVue driver have been identified which may cause data to be misassociated with patient results under a given set of circumstances.
When connected to the Sysmex UF-100, the Instrument Manager driver sysu100i driver software versions 7.00.0001, 7.00.0002, 7.00.0003, 7.00.0004, 7.00.0005, 8.00.0001, 8.00.0002, 8.00.0003, and 8.00.0004, and is configured with option "Strip Zeros from Result" set to "Yes" may result in sample error