115 V Blanketrol II, Model 222S Hyper-Hypothermia System . Used to lower or to raise a patients temperature and /or maintain a desired patient temperature through conductive heat transfer. Part number 86165.
Class I - DangerousWhat Should You Do?
- Check if you have this product: Model Number 222S; Serial #s: 133-2-00697, 133-2-00698; 133-2-00699; 133-2-00700; 133-2-00701; 133-2-00702; and 133-2-00703:.Part # 86165
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
β οΈ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Cincinnati Sub-Zero Products Inc
- Reason for Recall:
- The incorrect voltage component was placed into the device during manufacturing. This may make the device unable to maintain patient temperature during cooling mode. The device will continue to indicate that the unit is cooling because the compressor will continue to run, however water in the reservoir will begin to return to ambient temperature. The actual water temperature will be displayed but
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
115 V Blanketrol II, Model 222S Hyper-Hypothermia System . Used to lower or to raise a patients temperature and /or maintain a desired patient temperature through conductive heat transfer. Part number 86165.
Product Codes/Lot Numbers:
Model Number 222S; Serial #s: 133-2-00697, 133-2-00698; 133-2-00699; 133-2-00700; 133-2-00701; 133-2-00702; and 133-2-00703:.Part # 86165
Distribution:
Distributed in: MI
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1679-2014
Related Recalls
Due to FDA's safety communication relating to nontuberculous mycobacteria (NTM) infections related to heater/cooler devices, an updated Instructions For Use includes a more robust cleaning & disinfection procedure.
Hemotherm CE, model 400CE, Heater-Cooling Device
Cincinnati Sub-Zero Products
Improperly rated fuses were used in the manufacture of the device.
The Hemotherm 400CE 115v, containing power board p/n 39945
Cincinnati Sub-Zero Products
The device may not maintain water flow or temperature control.